Setting up Analog Output in PCI6025E

I have a PCI6025E board that I would like to setup such that it outputs a sawtooth data.
I used a Sawtooth function signal generator (range 0-4V), which I connected to AO Continuous Gen. I want to output this to pin DAC0Out, which I set up in MAX as a channel called test.
For testing, I also set up an analog input channel, using Differential mode (using ACH0 and ACH8). I connected the pin from DAC0Out to ACH0 and AOGND to ACH8.
When I tested the results using MAX, I couldn't get a signal, and MAX shows constant 5 V output. When I use a Probe on the waveform data, it shows the sawtooth signal.
Can anyone give me a suggestion how to fix this problem? I have tried the Labview example Continuous Sine Generation and set it
up to test, but I can't get any results out of it as well.
I'm using this setup because I needed a signal generator capable of outputting a sawtooth signal, as well as trying to measure any possible losses through cables and the DAQ board.
Thanks in advance
elton

Hello elton,
At first glace it looks like you have everything set up correctly. To replicate your situation, I modified the �Simul AIAO Buffer(E-Series).vi� example with a 0-4 v sawtooth signal. I am including copies of this modified example both with and without the input line. They both ran fine, and the copy without the input line tested great with a virtual channel in MAX. If these files don�t address your particular situation, or if you have any other questions, feel free to post a copy of your program and I would be happy to take a look.
Have a nice day!
Robert M
Applications Engineer
National Instruments
Robert Mortensen
Software Engineer
National Instruments
Attachments:
Sawtooth_(with_input).vi ‏105 KB
Sawtooth_(wo_input).vi ‏57 KB

Similar Messages

  • Setting different analog output's parameters in PCI-6229 DAQ card

    Hello,
    We are using DAQ assistant to set the analog output parameters (frequency).
    Is it possible to set different values for each channel using DAQ assistant?
    If not using DAQ assistant, what  is the other way to set it?
    Thank You
    Arunas

    If you are trying to set different frequencies for waveform generation on two channels, the answer is No, it can't be done. If possible, you will have to pick a sample rate high enough to work for both signals, and use an over-sampled waveform.
    Another choice would be to use a second device for the second waveform.
    John Weeks
    WaveMetrics, Inc.
    Phone (503) 620-3001
    Fax (503) 620-6754
    www.wavemetrics.com

  • Set default analog output in MAX

    Hi,
    I would like to set a default value (e.g. 10V) to my analog output of my NI6251 card. The current value is (I assume) 0 V.
    Is there a way to change the default value without using labview?
    Many thanks

    I'm pretty sure you can't set the default analog output of a DAQ card to something other than 0V even using LabVIEW.  When the device powers up it will reset and set all outputs to an inert state which is a good thing for safety conserns.  
    I'd recommend you redesign your system so it too uses this standard of using the inert state of a startup condition.  What kind of system do you have that you want the off state to be 10V?
    Unofficial Forum Rules and Guidelines - Hooovahh - LabVIEW Overlord
    If 10 out of 10 experts in any field say something is bad, you should probably take their opinion seriously.

  • Processing counter input into an analog output

    Greetings to all --
    I've set up a PCI6070E card to do digital input through its "counter 0" channel and output through its "AO 0" channel.
    I'm brand-new to Labview, however, and having trouble figuring out how to design my program.  In short, this is what I want to do:
    1. Count the number of pulses, let's call it P, that arrive at the counter in a given time increment.  (Ideally, this time will be adjustable from 1 microsec to 1 sec.  In reality, I may have to settle for some number of milliseconds or more, as Labview doesn't seem to like to do loops at more than 1khz.  But that's an issue for another day.)
    2. Perfrom a mathematical operation to turn P number of counts into some number N between -10 and 10
    3. Set the analog output to read N volts
    4. Reset counter to zero and start over at (1.) again.
    I can make labview count pulses in, and I can send voltage signals out, and I can use the front panel to get the inputs I need from the user for the mathematical operation,  However, I'm having trouble with counting how mant pulses hit in a given time increment, and then resetting my digital counter to zero.
    Can anyone point me at the right function names or help files, or suggest a rough program flow?  I appreciate it.
    Chad
    [email protected]

    Chad,
    The only way you can perform this type of operation with your device is in software.  See the knowledge base entries below for more details. 
    HW resetting Counter on an STC Device:
    http://digital.ni.com/public.nsf/websearch/55AF6CA​DD87F6F7786256B89005B8AFD?OpenDocument
    How To Reset in Software:
    http://ae.natinst.com/operations/ae/public.nsf/web​/searchinternal/65e3dbc715998c3286256e900075b7f8?O​...
    Regards,
    L. Allen

  • Dock Analog Output Adjustment

    I use an FM modulator to play my iPod Touch in my car. I've used a few and they all seem to have shortcomings... mainly FM co-channel interference.
    The one I use now is a Kensington. I've had the best luck with fade and co-channel interference with it. But the FM pre-emphasis is over modulated with the standard Touch analog output and it only connects via the dock connector.
    I have already lowered the output level of all songs on my iPod with the iTunes Get Info:Options:Volume Setting to eliminate sibilance but would prefer to lower the line level output...
    I KNOW that one day I am going to come up on a track that hasn't been lowered and soil my pants!
    Does anyone know a hack or terminal command that will set the analog output of an iPod Touch?
    Thanks!

    - Ive tried it on different slot (in the other slot the driver installer wont even recognize the card). The 8V was not on PCI connector, but on the card itself. Also, I tried it on both PCs with the same headset. On sons PC works great, on mine doesnt. So connections are probably ok.
    I only have a regular stereo speaker system, but I tried configuring the card to 4 speaker system and plugged the speakers to the rear channel. Theres horrible cracking and the sound breaks up badly. Im 99% sure this is somekind of configuration or driver problem. I really, really dont want to start re-installing OS right now, since Win7 is around the corner and Im probably going to buy it in 2-3 months and doing a clean install then.
    Kinda blows since Ive had this card for more than 6 months and it has never worked. Ive been forced to use the integrated Realtek soundchip and I cry a few tears every time I put some music on.

  • Setting analog outputs to 0

    I would like to start by cautioning you that I am not a programmer.  I have been trained in LabVIew basics, so naturally I am trying to write a much more than basic code. 
    I am writing a test sequence in which I am using a sequence structure to setup and sequence various test functions. For part of the code I am writing I am trying to control a pressure output, I start the task outside a while loop, I write to the task to control control pressure inside the while loop after leaving the while loop and then stiop the task outside the while loop, but the analog output remains at the last commanded value, how do I set it to 0? 
    I have tried writing a constant 0 to the task after I leave the loop and then stoping the task but this does not appear to work. I have a number of analog and digital outputs in this code that I would like to set at 0 before I proceed to the following sequence. 

    My appologies, I have been going through a number of trial and error phases with this code.  Here is an example of when I was trying to write a 0 to the channel, in this version I used a case structure to tell it to go to 0 when the test sequence would be complete.  I arrived at this potential solution after having no success with earlier attempts. 
    As far as state machines go, like I said I have only been to LabView basics I and II, would I be able to convert the sequence structure to a case structure and put that in a while loop of some sort to control the sequencing?
    Attachments:
    Purge Portion 15.vi ‏59 KB

  • Setting sample rate for sinus analog output

    Hello,
    I've been trying to do something very simple : using an analog output of the card PCI 6221 to produce a sinus curve of frequency 50 Hz. For this I used a Vi to create a sinus curve and the different DAQmx VIs. But I have difficulties understanding the principle of virtual channel and I think I'm doing a mistake setting the sample rate and samples number : one time for the sinus vi, second time for "DAQmx - Timing". Should I use the same values for both of these VIs ?
    On my oscilloscope, with frequency=50Hz and sample rate=1kHz, I get a null signal. Then depending on both values, I get differently rated signals. For example with f=1Hz and sr=10kHz, a sinus of frequency 0,7 Hz.
    Solved!
    Go to Solution.
    Attachments:
    Sinus analog output.vi ‏32 KB

    Yes, thanks for your advice. I used the structure given in the example and now it's working fine. I'm still not sure what I did wrong though.
    I would have a second question now (should I create a new topic?):
    I put a continuous sine wave on the analog output. As soon as this is running, (or maybe after a short delay) I want to measure a limited amount of samples on my analog input. How can I be sure, it's not going to start measuring before the output is properly set ?
    I don't think a trigger would solve the problem since I'm going to vary the output Amplitude.

  • How do I set up analog 0-10 VDC output pulses with varying duty cycles?

    I am trying to control a 400 watt laser so that it can pulse on and off within a few milliseconds. The laser controller reads a 0-10 VDC signal, with 0V being 0 watts and 10V being 400 watts. I am using a PCI-6221 DAQ and LabVIEW 8.0.
    It seems that all of the examples have square waves that are symmetric about the 0 volt range. I need a square wave that goes from 0-10VDC with different on/off timespans. I would like to have something as follows:
    1) 0 seconds = laser off
    2) .001 to .002 seconds = laser on at full power
    3) .002 to .009 seconds = laser off
    4) Repeat steps 2-3 500 times
    Any ideas?
    Michael Fidler

    Michael, 
    I ran your code, but was unable to reproduce the problem you are
    seeing. I measured the signal in two fashions’. I brought the signal back into
    the card (LabVIEW_Square Wave.JPG) and into a benchtop scope (Benchtop
    scope_Square Wave.jpg).
    I would like to ask that you run the attached code twice. The
    first time, please run it with the analog input as ‘Dev1/_ao0_vs_aognd’. This
    take the analog output and compares it with the cards AO ground, therefore eliminating
    the need for external connections and simply doing internal routing. The second
    time, please run it with the analog input as “Dev1/ai1”, and wire your AO
    channel to an AI channel. These tests are to ensure that there is not a stray
    capacitance in your scope, the probe, etc. Let me know the results (pictures
    are always great)
    David L.
    Systems Engineering
    National Instruments
    Attachments:
    Cont Gen Voltage Wfm-Int Clk_Test VI (8.2).vi ‏64 KB
    Benchtop scope_Square Wave.jpg ‏38 KB
    LabVIEW_Square Wave.JPG ‏36 KB

  • How to switch analog output between two waveform frequencie​s

    Hello,
    I'm using a fairly simple VI to generate a square wave at our analog output board.  I have the square wave controls in one panel of a sequence structure, and a "release" function in the other panel of the sequence. Currently the VI is programmed to switch from the original square wave to -10V when the Boolean control is switched.
    What I'd like to do now is to modify the VI so that I can swtich between a square wave of a given frequency, say 100 kHz, to a square wave of a different frequency, say 10 kHz, or to DC -10V. But I'd like to be able to vary each frequency and amplitude independently.
    Could anyone give me some programming pointers?
    Thanks,
    Michelle

    Thanks to the three of you for the suggestions. I started trying a few of them, and then realized that what I had already built will work pretty well to update the frequency. I didn't realize that it would update the waveform in "run" mode -- I thought it would only do that in "run continuously" -- so I've found that I can change the frequency of the waveform by doing that. This doesn't allow me to change the frequency and the amplitude simultaneously, but I'm not sure yet whether or not I will need to do that -- I so will probably go ahead with the experiments I need to do with this set-up, then modify it if I find that I need to control both frequency and amplitude simultaneously.
    Thanks for your help!
    M.

  • NI-DAQmx VisualStud​io C++ 6 Single point analog output

    Specs: NI-DAQmx 7, VisualStudio C++ 6.0,  PCI-6722,8channel AO
    We have a very simple application: set a voltage (actually 6 channels) and keep it until we want it changed again, perform the change very quickly in response to an image capturing algorithm. So I don't need any waveforms or buffering.
    In this forum post http://forums.ni.com/ni/board/message?board.id=231​&message.id=3283&query.id=18094 you talk about an AOOnePoint example, but I get an error that the NI-DAQ driver does not support my device.
    I may need to use NI-DAQmx, but how? I would like to use something like AO_VWrite(,,), maybe for 6 channels in one call. But I can't find it in NI-DAQmx. It seems I need to setup buffers and frequencies. I have a working sample, but it seems a slow and certainly overkill of this simple application:
    // Link with \DAQmx ANSI C Dev\lib\msvc\NIDAQmx.lib
    #include "NIDAQmx.h"
    double[2] data;
    int taskHandleAnalog;
    int written;
    void Init()
        DAQmxErrChk (DAQmxCreateTask("",&taskHandleAnalog));
        DAQmxErrChk (DAQmxCreateAOVoltageChan(taskHandleAnalog,"Device and Channel Info","",0,10,DAQmx_Val_Volts,NULL));
        DAQmxErrChk (DAQmxCfgSampClkTiming(taskHandleAnalog,"",1000,DA​Qmx_Val_Rising,DAQmx_Val_ContSamps,NUMBER_OF_AO_SA​AMPLES));
        DAQmxErrChk (DAQmxWriteAnalogF64(taskHandleAnalog,NUMBER_OF_AO​_SAAMPLES,0,1.0,DAQmx_Val_GroupByChannel,data,&wri​tten,NULL));
        DAQmxErrChk (DAQmxStartTask(taskHandleAnalog));
    void SetVoltage( double voltage )
        data[0] = voltage;
        data[1] = voltage;
        DAQmxStopTask(taskHandleAnalog);
        DAQmxErrChk (DAQmxWriteAnalogF64(taskHandleAnalog,NUMBER_OF_AO​_SAAMPLES,true,10.0,DAQmx_Val_GroupByChannel,data,​&written,NULL));

    Hi,
    It looks like you simply wants to output voltages on the analog output channels, but only wants one update at a time with no waveforms or buffering in DAQmx.
    As I'm sure you know there are really just 3 types of measurements.  Single Point, Finite, and Continuous.  Since you want a single value at a time it's just a Single Point operation.
    You can find DAQmx examples for single point operations in this path:
    C:\Program Files\National Instruments\NI-DAQ\Examples\DAQmx ANSI C\Analog Out\Generate Voltage\Volt Update
    Simply place the DAQmx Write Code within a loop and you will be updating one value at a time, but multiple times when "we want it changed again".
    Dennis Morini
    Field Sales Engineer
    National Instruments Denmark
    http://www.ni.com/ask

  • Single shot analog output: How to pulse this. Probably easy, I'm at a loss.

    Hello.
    I have a piece of machinery that my system is connected to.  Every time it reaches a proximity sensor (It's a reciprocating unit), I need to do a single pulse of 5v out of my analog output.  It has to be fast, it has to stop until the sensor is reached again, and it can't stay 'on' for long.
    I've attached my VI.  I'm using a voltage input via DAQ assistant, routed through some logic to produce a 'true' boolean every time that sensor is reached, and that boolean is connected to a case structure with voltage out tasks created in DAQmx.
    My problem is that the machine won't reverse while the case is still true and there's still the 5v being called by the true case on the case structure.  It has to pulse, then stop, then be allowed to come back when the proximity sensor is reached again.  As it is now, the machine reaches the prox and then the whole system stops, as the 'true' condition on the case structure, and subsequent voltage output, remain high.
    I've attached the VI and heirarchy as well.
    I've tried a few different things here, as well as just using a DAQ assistant with N samples, but that would just retrigger when the loop repeats anyway.
    I'd love to sit and bang my head against this til I get it, but I'm under a time constraint.  Help!
    Still confused after 8 years.
    Solved!
    Go to Solution.
    Attachments:
    using create channel.vi ‏94 KB
    logicsub.vi ‏32 KB

    Ralph, without knowing the exact DAQ card that you are using, it is difficult to give you the best solution. If you want to continue with the solution that you currently have, try adding a shift register to the outer while loop to track the previous value of the boolean. In this way, we can keep multiple values that are above the threshold from causing multiple pulses. Next add a time delay after the DAQmx Write, then add another DAQmx Write to set the output back to your low value. Remove all code from your "false" case. See attached VIs. Please let us know what model of DAQ card you have as we may be able to do something clever like a retriggerable pulse train if your card can support it.
    Charles Chickering
    Charles Chickering
    Architecture is art with rules.
    ...and the rules are more like guidelines
    Attachments:
    using create channel.vi ‏98 KB
    logicsub.vi ‏29 KB

  • Synchronizing Analog Output to Counter Input

    Hi!
    I'm really new to LabView. Pls help!
    I'm trying to obtain a set of stimulus signal and response signal from my motor for system identification.
    The stimulus signal will be generated by the Ao0 channel and response signal (quad encoder linear position) will be read into counter0.
    How can I synchronize the stimulus signal and response signal so that the sampling (n samples) for both stimulus and response starts at the same time?
    e.g.
    t=0, stimulus sample 1, response sample 1
    t=1, stimulus sample 2, response sample 2
    t=2000, stimulus sample 2000, response sample 2000 
    I've attached my VI.
    Thanks. 
    Regards,
    Student
    Attachments:
    Acquiring Linear Encoder Samples.vi ‏44 KB

    Hello Htp,
    What do you mean by synchronize? Do you want to start the analog and counter outputs at the same time? If you do, you should have both outputs start with the same digital trigger. For the analog output, you can set up the triggering in the DAQ Assistant in the Task Triggering tab. For the counter output, look at Gen Dig Pulse Train-Continuous-Dig Start.vi found in the Example Finder (Help Menu>>Find Examples). Browse Hardware Input and Output>>DAQmx>>Generating Digital Pulses. Please let me know if you have any questions. Have a great day!
    Sincerely,
    Marni S.

  • Simultaneous analog output

    I am using PCI-6052 with (LabVIEW) .Where two analog outputs are available ,DAC0 and DAC1.I want to simultanouly output the different waveforms on each Channel at the same time.Time is important.
    I tried it with two DAQ assistance vi's setting each channel on each vi ,but at time randomly one vi's gives output not both at the same time.
    Thanks
    Haider

    Hello Haider,
    It sounds like you are trying to get started with NI-DAQmx and want to
    output separate signals on the two analog output lines.  The
    restriction of the hardware is that all of the analog input or output
    channels must be contained with in the same DAQmx Task.  There are
    a couple ways to do this. 
    If you want to stick with the DAQ Assistant, I searched our KnowledgeBases and found the following example program which does analog output for two channels using the DAQ Assistant. 
    If you would like to drill down a little farther and use the lower
    level DAQmx VI's, I would recommend looking at one of the shipping
    examples, the format for multiple channels is Devx/AOy:z where x is the
    device number, y is the starting AO channel, and z is the final AO
    channel. 
    Also, another good reference, especially if you are going to use the lower level VI's is the Learn 10 Functions in NI-DAQmx and Solve 80% of Data Acquisition Applications Tutorial.  Hopefully,
    that'll get you going with your program.  Let us know if you have
    any other questions with either the online example or one of the
    shipping examples. 
    Regards,
    Micaela N
    National Instruments

  • How do I get an analog output signal from my PCI-MIO-16E-4 card?

    I need an analog output from 0-10 V to control the mass range on a mass spectrometer. Do I use the DAC0out channel to do this? The manual only talks about analog out (ao) channels. Is this the same as DAC0out and DAC1out? How do I set up the DAQ assistant for this process? I do not need a waveform, only a constant signal.

    Yes indeeded. The DAC0out and DAC1out are the analog output channels (ao0 and ao1).
    Simply follow the examples that ship with LabVIEW. You can also easily create code with DAQ assistant.

  • Read Data (ASCII) and send to analog output

    Hello
    I have the following MAX configuration:
    Analog input task, continous, 2 samples, rate 100Hz (PCI-6024E)
    Analog output task, continous, 64 samples, rate 100Hz (PCI-6723)
    The cards are synchronized by the RTSI bus.
    With Dasylab I display signals from load cells by the input task. A "Read Data" module sends values to the output task. The "Read Data" module has the following settings: Output in realtime, synchronisation with analog input task, file is an ASCII, block size 4 values, no ascii time channel.
    Additional settings for DAQmx are done by DasyLab:
    Analog input: Driver buffer 30
    Analog output: Driver buffer 30, output start after 2
    When I start DasyLab I get a DAQmx failure (after a couple hours): DAQmxReadAnalogF64, Attempted to read samples that are no longer available. The requested sample was previously available, but has since been overwritten.
    Could anyone please explain, how to configure analog input/output task and the DasyLab settings (Read Data)? - Attached the DasyLab diagram.
    Additional information: The application should read and display signals from a test bench (load cells). The output shall feed external controllers. The input and output signals shall be displayed on the same chart (as realtime as possible).
    Thank you
    Regards Samuel
    Attachments:
    DasyLabDiagram.png ‏67 KB

    Thanks for the prompt reply! - I have changed the settings to the following:
    Analog input task, continous, 16 samples, rate 100Hz, driver buffer 32
    Analog output task, continous, 16 samples, rate 100Hz, driver buffer 32, output start after 32
    In the "Data read" module the block size is now 16 at 100Hz
    With this setting, DasyLab runs, but only for around 12 hours.
    The following error occurs:
    "Read date module name" / The data flow is blocked by one of the following modules.
    Followed by:
    Attempted to read samples that are no longer available. The requested sample was previously available, but has since been overwritten.
    Is there something else I have to care? Could it be problematic to send signals from different modules to the same Chart Recorder module? I use DasyLab version 11.
    Samuel