Cannot configure SCXI 1102 in parrallel mode

Hi all...
I have a little problem, i have a system made of 3 Daqmx cards : 2 PXI-6254 and 1 PXI-6229. these cards are linked to SCXI modules. The acquisition of the data are done in parrell i.e. the first PXI 6254 is linked to 2 SCXI-1125, the 2nde PXI-6254 is linked to one SCXI 1125 and one SCXI 1102B and the third one is linked to one SCXI 1180 and one SCXI 1102...
So i defined 3 different Daqmx Task in MAX (one for each DAQmx card) and i configured the SCXI rack.. but when i try to configure the SCXI 1102 B or the SCXI 1102  i CAN'T choose the parrallel mode i can only choose multiplexed mode... and therefore, when i run my application, i have the following error:
Error -200619 occurred at DAQmx Start Task.vi:6
Possible reason(s):
Measurements: Chassis cannot be used for more than one scanning operation at the same time.
Do only one scanning operation, or combine multiple scanning operations into a single operation.
Can't anybody explain me why ???
Thanks a lot for your futures answers ...

Not necssarily, You can have parallel mode for other SCXI modules while acquiring from SCXI 1102 in multiplexed mode
See this Link

Similar Messages

  • INST-07408: Unable to install or configure the product on a 32-bit JVM on a 64-bit machine. Make sure to install and configure the product in supported modes.

    Hi there, i have an issue installing oracle forms and reports. its throwing this error"INST-07408: Unable to install or configure the product on a 32-bit JVM on a 64-bit machine. Make sure to install and configure the product in supported modes." on step 5. I have jdk 64bit installed, weblogic server was installed successfully but iam getting stuck somehow.
    Below are details of my pc.
    OS Name    Microsoft Windows 7 Professional
    Version    6.1.7601 Service Pack 1 Build 7601
    Other OS Description     Not Available
    OS Manufacturer    Microsoft Corporation
    System Name   ...........-HP
    System Manufacturer    Hewlett-Packard
    System Model    HP ProBook 4530s
    System Type    x64-based PC
    Processor    Intel(R) Core(TM) i5-2450M CPU @ 2.50GHz, 2501 Mhz, 2 Core(s), 4 Logical Processor(s)
    BIOS Version/Date    Hewlett-Packard 68SRR Ver. F.23, 09/03/2012
    SMBIOS Version    2.6
    Windows Directory    C:\windows
    System Directory    C:\windows\system32
    Boot Device    \Device\HarddiskVolume1
    Locale    United Kingdom
    Hardware Abstraction Layer    Version = "6.1.7601.17514"
    User Name    ...........-HP\................
    Time Zone    South Africa Standard Time
    Installed Physical Memory (RAM)    4.00 GB
    Total Physical Memory    3.94 GB
    Available Physical Memory    1.00 GB
    Total Virtual Memory    7.87 GB
    Available Virtual Memory    1.25 GB
    Page File Space    3.94 GB
    Page File    C:\pagefile.sys

    To clarify further....  It is important to note that Forms/Repors 11.1.2.x software (including WLS and JDK) must be of the same archetecture as the machine.  In other words, you cannot use the 32bit Forms/Reports 11.1.2 software on 64bit Windows.  If you are using a 64bit OS, you must install the 64bit software.  Again, this would include Forms/Reports, WLS, and JDK

  • SCXI 1001, SCXI-1102 and SCXI-1303 not working correctly with DAQmx

    I have recently upgraded from Labview 6.5 to Labview 8.2 and now my SCXI 1001 with a SCXI 1102 and SCXI 1303 terminal block no longer reads the correct voltage.  Using the old version of labview thermocouple measurements are correct at 0.5mV; however using the new version of labview and DAQmx the thermocouples are reading 5V.  Everything is installed correctly as it still works with the older version of Labview, is there something that needs to be reset or change so that teh SCXI will function properly with DAQmx?

    Hello ilabtech,
    From my understanding you are using an SCXI 1001 chassis with an SCXI 1102 and 1303 terminal block to perform thermocouple measurements.  You are upgrading to LabVIEW 8.2 and DAQmx.  I am assuming you were using LabVIEW 6.1 since there isn’t a LabVIEW 6.5.  You mentioned that are able to get a correct reading of 0.5 mV with LabVIEW 6.1 but you see a rail of 5V with LabVIEW 8.2 and DAQmx. 
    My first suggestion is that you ensure that the SCXI is set at the appropriate range and for the correct thermocouple type.  You can also configure these settings in MAX as detailed on page 27 of the SCXI 1102 User Manual and Specifications Guide. 
    My second suggestion is for you to start off using an example program to see if you see the same error.  When you are in a LabVIEW 8.2 VI you can go to Help>>Find Examples to open the NI Example Finder.  You then choose Hardware Input and Output>>DAQmx>>Analog Measurements>>Temperature>> Cont. Acq Thermocouple Samples-Int CLK.vi.   This example program is supported by your hardware and is a good resource to use to troubleshoot.
    If these suggestions aren't helpful please answer the following questions in your reply.
    1.        I would like to know what range you have set the SCXI 1102 to read. 
    2.       What version of DAQmx are you using?  You can find the version my opening Measurement and Automation explorer (MAX), expanding the “Software “and clicking on NI-DAQmx.
    3.       What operating system are you using?
    4.  Does your Traditional DAQ program work in LabVIEW 8.2?
    Regards,
    Ima
    Applications Engineer
    National Instruments
    LabVIEW Introduction Course - Six Hours
    Getting Started with NI-DAQmx

  • Multiple tc types on one scxi 1102

    Does anybody know if you can use two different thermocouple types on one SCXI-1102 module? 
    I need to measure both Type K and Type T thermocouples.  I am using a SCXI 1303 connector block and plan to use the internal CJC.
    Thanks in advance
    doloop

    Hello doloop,
    You can indeed configure a task to have to two different types of
    thermocouples, and acquire from multiple thermocouple types in one
    DAQmx Task.  For each channel, the NI-DAQmx driver will apply
    different conversion coefficients to convert voltage measurements from
    different thermocouple types into temperature measurements.  You
    can configure this easily in Measurement & Automation Explorer
    (MAX), by creating a DAQmx Task of the thermocouple type, and adding
    multiple physical channels from the SCXI 1102 module to the channel
    list.  For each channel in the scan list, under the settings tab,
    you will have the option to select what type of thermocouple to
    associate with each channel, along with CJC settings.  You can
    also do this programmatically in LabVIEW using the DAQmx Create Channel
    VI.  I've attached a shot of a block diagram that shows how to
    program this.
    I hope this helps!
    Travis G.
    Applications Engineering
    Message Edited by Travis G. on 04-18-2006 05:09 PM
    Attachments:
    MultipleThermocoupleTask.Jpg ‏68 KB

  • I don't Configuring Access Manager in SSL Mode

    i only install am7.1 and ws7.0 in windows2003 pack 1.
    then, i read "Sun Java SystemAccess Manager 7.1 Postinstallation Guide" .
    it said that "Login to theWeb Server console.The default port is 8888." but i can't find the default port .
    i think my web server console's default port is 8989.

    Hi,
    As a part of my requirement, I need to Configure Access Manager in SSL Mode. For that, I followed all the steps(Change http to https in web server instance in Access Manager, Install Certificate, Modify AMConfig.properties) mentioned in the PostInstallation Guide of Sun Access Manager to configure the SSL using Selfsigned certificate. so, after doing all these steps, as soon as the hit the Access Manager URL httsp://machinename:portno/amserver/UI/Login, it shows "page cannot be displayed" error. I have checked the web server with SSL enabled in it and its running fine.
    On one of forum post, I read that you need to set this property to true "com.sun.am.jssproxy.trustAllServerCerts" if you are not installing the ROOTCA certificate however this is not listed in the AM documentation.
    Any help on this would be highly appreciated. Let me know if am missing any steps.

  • Using SCXI-1520 in Parallel Mode.

    Hi all,
    I am using SCXI-1000 chassis and PCI-1024E DAQ card.
    I want to use my SCXI-1520 in Parallel mode.
    Is there any other way to configure my 1520 in parallel mode other than directly connecting to DAQ board using SCXI-1180 feed through panel and SCXi-1302?
    Some of my friends are telling that there are some Patch files, through which we can use the scxi-1520 in parallel mode. Is it true? If it is true please give me how use the Patch files( or alternative solutions).
    Thanq
    Regards
    Preetam

    Preetam,
    As explained in the SCXI-1520 user manual, Traditional NI-DAQ does not support true parallel mode operation of the SCXI-1520. However, NI-DAQmx does. I would recommend installing NI-DAQ 7.2, which includes the latest version of both Traditional NI-DAQ and NI-DAQmx. Then you can use the NI-DAQmx driver to control your SCXI-1520 in parallel mode. To configure the SCXI-1520 in parallel mode with NI-DAQmx, simply select the Parallel Mode check box when setting up your chassis and modules in MAX. I hope this information helps.
    SCXI-1520 User Manual
    NI-DAQ 7.2 Download

  • API0259: The object cannot be edited in Read-Only mode!

    All,
    I'm getting the following error when trying to import an MDL file into my repository:
    API0259: The object cannot be edited in Read-Only mode!
    Does anyone know what this relates to?

    Hmmmm, sounds like it is hanging up on a conflict with the configuration name from your source repository and the configuration names in your target. Not sure that I have time to try and replicate that on my own system today.
    What I might suggest you try is to, on your source, set the project configuration back to the standard "DEFAULT_CONFIGURATION", get your module configured to use it, and re-export the mapping. Then you can fix your configuration back to where it should be.
    Do the same re-configuration on the target prior to import and see if that resolves the problem.
    Now, clearly this is not optimal and it would be really nice to find out WHY the conflict is occurring, but it would at least get this mapping imported for now.
    Mike
    And Vader, it is not that I don't have full faith that Oracle always raises the correct error message for a problem..... but, well, I've USED Oracle for too many years not to consider the possibility.
    Case in point in OWB, try putting an invalid value into the OWB-supplied WB_Date_From_Julian function
    e.g. select WB_Date_From_Julian(-1) from dual;
    10:55:12 Error: ORA-21000: error number argument to raise_application_error of -1854 is out of range
    ORA-06512: at "OWB_MGR.WB_DATE_FROM_JULIAN", line 7
    because some bright puppy at Oracle decided to try and raise an Oracle Built-in Error via raise_application_error instead of just using "raise", and - of course - the Oracle error code isn't within the allowable range.
    Error messages are nice, but you can't always trust 'em - which is why I look to expanded possibilities when neccessary.

  • Connection of the pressure transmitter to the scxi-1102 module

    Dear sir,
    we have a differential pressure transmitter whose specifications are as follows:
    Excitation:24 Vdc, output:4-20mA.(make: Cole-parmer instrumentation company)
    now, we have a NI scxi-1102 and scxi-1520 module with corresponding terminal blocks.
    Our question is how to connect this transmitter to the module and to which module? can we connect it to the scxi-1102? And for the excitation what we should do? please explain the detailed procedure for that.
    As the excitation is 24 Vdc, i think we have to use 1102 and has to be excited externally.
    Since, i m a mechanical engineer and don't have much knowledge of electronics, please explain all the details as simply as possible.
    regards...
    Lalit.

    I am assuming a couple of things here. They are:
    The input you are looking at is a slow responding signal ie less than 2 hz. The 1102 has a built in 2 hz filter.
    and or the transducer you are connecting to has a slow update rate.
    You are using scxi 1300, 1303, or 1308 terminal block to connect to your chassis.
    Channel 0 isn't being used.
    That being said, in Max, set the gain to 1
    input mode to differential
    You can do the scaling in Max if you want to.That's what I typically do.
    Linear scAle, 1 volt = 4 ma and 5 volt = 20 ma
    On your terminal block, connect a 250 ohm precision resistor across Ch 0 in ( pin 96 & 64)
    In series with your current loop ( cut the wire of the existing loop)
    Connect one end to pin 96 and the other to pin 64.
    Put a voltmeter across the input. red lead to 96, black lead to 64
    If the reading is negative flip flop the wires. If positive, good guess.
    you should be ready to rock and roll.
    GB
    Greg Bush CLAD

  • I cannot configure yahoo small business after upgrading to Mavericks Help needed

    Hi Guys,
    it is as simple as that:
    I cannot configure yahoo small business after upgrading to Mavericks Help needed
    Thanks

    Please read this whole message before doing anything.
    This procedure is a diagnostic test. It’s unlikely to solve your problem. Don’t be disappointed when you find that nothing has changed after you complete it.
    The purpose of the test is to determine whether the problem is caused by third-party software that loads automatically at startup or login, by a peripheral device, or by corruption of certain system caches. 
    Disconnect all wired peripherals except those needed for the test, and remove all aftermarket expansion cards. Boot in safe mode and log in to the account with the problem. Note: If FileVault is enabled on some models, or if a firmware password is set, or if the boot volume is a software RAID, you can’t do this. Ask for further instructions.
    Safe mode is much slower to boot and run than normal, and some things won’t work at all, including sound output and  Wi-Fi on certain models. The next normal boot may also be somewhat slow.
    The login screen appears even if you usually log in automatically. You must know your login password in order to log in. If you’ve forgotten the password, you will need to reset it before you begin. Test while in safe mode. Same problem? After testing, reboot as usual (i.e., not in safe mode) and verify that you still have the problem. Post the results of the test.

  • Cannot configure my iPhone 3gs after i have updated from 5.1.1 to os 6.0

    hi i need some help, im using iphone 3gs, when i connect it to my mac it says theres an update to my iphone which i did install, but few minutes after its telling me that i need to restore my iPhone. i did restore it but i cannot configure it anymore, in itunes it says there is no sim card installed in the iphone that your attempting to activate.

    in itunes its telling me theres no sim card installed in the iphone you are attemptng to activate.

  • How can I set the gain on a SCXI-1102 module without having my computer lock-up for 3 seconds?

    Forgive my insolence, but did it never occur to anyone that spinning in a hard loop while polling a status register to set a gain for 3 seconds isn�t a great idea?
    I have a number of SCXI-1102 modules to set the gains on, and it takes several seconds for each. Reading the documentation on the subject indicates that the SCXI_Set_Gain call polls a status register waiting for the gain to settle before returning � and it doesn�t even yield while doing so.
    Are there any other ways I can set the gain? Is there a way that doesn't poll the register that returns in a reasonable amount of time?
    AFAICT, there's no good reason to only allow one module to be s
    etting its gain at a time. An asynchronous call it really needed - one that makes it the API programmer's responsibility to check the status of the register, or one that spawns a thread and sleeps between polls, and signals an event or invokes a callback or some-such-thing. Or you could even just sleep for three seconds after setting all the gains, this way the delay would be 3 seconds total � not 3 seconds for each gain set.

    This NI-DAQ function call directly accesses the nidaq32.dll. While the nidaq32.dll is multithread safe, it is not a multithreaded DLL. This is also described in the KnowledgeBase entry 1V9DSHDM: "Can NI-DAQ Handle Different Operations at the Same Time? Is It a Multi-threaded DLL?"
    You are correct that the SCXI_Set_Gain function is not asynchronous. Thus, you do have to wait till that function returns before calling the next one. LabVIEW has found a way around this. So, you can actually use the LabVIEW functions to set the gain on these modules without having to wait as you do with SCXI_Set_Gain.
    Regards,
    Geneva L.
    Applications Engineer
    National Instruments
    http://www.ni.com/ask

  • Configuring Weblogic Domain in Offline Mode

    Hi,
    I wrote a java code which invoke wlst command required for configuring weblogic domain in offline mode.
    please give any idea even i do not know exactly what are the required jar file but i have used weblogic.jar,wlfullclient.jar,jython.jar,and some jar files from modules folder which has given in someone post
    i am getting the following error
    java.lang.reflect.InvocationTargetException
    at sun.reflect.NativeMethodAccessorImpl.invoke0(Native Method)
    at sun.reflect.NativeMethodAccessorImpl.invoke(NativeMethodAccessorImpl.java:39)
    at sun.reflect.DelegatingMethodAccessorImpl.invoke(DelegatingMethodAccessorImpl.java:25)
    at java.lang.reflect.Method.invoke(Method.java:597)
    at weblogic.management.scripting.utils.WLSTUtil.initOfflineContext(WLSTUtil.java:291)
    at weblogic.management.scripting.utils.WLSTUtil.setupOfflineInternal(WLSTUtil.java:267)
    at weblogic.management.scripting.utils.WLSTUtil.setupOffline(WLSTUtil.java:234)
    at weblogic.management.scripting.utils.WLSTInterpreter.<init>(WLSTInterpreter.java:134)
    at weblogic.management.scripting.utils.WLSTInterpreter.<init>(WLSTInterpreter.java:76)
    at DomainConfiguration.<init>(DomainConfiguration.java:15)
    at DomainConfiguration.main(DomainConfiguration.java:61)
    Caused by: java.lang.NoClassDefFoundError: com/oracle/cie/wizard/ControllerProxy
    at java.lang.Class.getDeclaredMethods0(Native Method)
    at java.lang.Class.privateGetDeclaredMethods(Class.java:2427)
    at java.lang.Class.getMethod0(Class.java:2670)
    at java.lang.Class.getMethod(Class.java:1603)
    at com.oracle.cie.domain.AbstractManager.getInstanceObject(AbstractManager.java:88)
    at com.oracle.cie.domain.AbstractManager.load(AbstractManager.java:46)
    at com.oracle.cie.domain.ConfigManager.<clinit>(ConfigManager.java:37)
    at com.oracle.cie.domain.DomainConstants.<clinit>(DomainConstants.java:20)
    at com.oracle.cie.domain.script.jython.WLScriptContext.init(WLScriptContext.java:220)
    at com.oracle.cie.domain.script.jython.WLScriptContext.setup(WLScriptContext.java:162)
    at com.oracle.cie.domain.script.jython.WLST_offline.setupContext(WLST_offline.java:105)
    ... 11 more
    thanks in advance

    Hi,
    Make sure that you have weblogic.jar and jython-modules.jar
    You will find weblogic.jar under %BEA_Home% / server / lib
    and you will find jython-modules.jar under %BEA_Home% / common / wlst/modules
    Regards,
    Kal

  • My iPhone 3GS 16gb, is stuck in restore mode, downloaded and tried to do a restore multiple times. Keep getting an error The iPhone "iPhone" could not be restored. Unknown error occured (-1). Cannot get it out of restore mode, keeps rebooting!

    My iPhone 3GS 16gb, is stuck in restore mode, downloaded and tried to do a restore multiple times. Keep getting an error The iPhone "iPhone" could not be restored. Unknown error occured (-1). Cannot get it out of restore mode, keeps rebooting!
    I have tried to hold the "home" and "power" button to reboot
    I have tried powering it off
    I have tried, to unplug the cable, hold the "home" button and then plug it in.
    Can't check the SN or warranty, cause it is stuck in restore mode.
    Is the hard drive gone? Should there be something replaced.

    http://support.apple.com/kb/TS3694#error1
    Error 1 or -1
    This may indicate a hardware issue with your device. Follow Troubleshooting security software issues, and restore your device on a different known-good computer. If the errors persist on another computer, the device may need service.
    Usually this is caused by your antivirus or firewall blocking iTunes access to the Internet. It can also occur if the phone is jailbroken or if the computer you are using was ever used to jailbreak an iDevice.

  • I seem to have lost some of my settings and now I cannot configure my airport connection. How do I get my settings back?

    Hello. First time using this. Still trying to figure out how to post my questions. I seem to have lost some of my settings and now I cannot configure my airport connection. How do I get my settings back?

    If you are using cable and a Nethear router sounds like from the information you provide before your Airport setting are setup improperly. PPoE is not something a cable service would need to work. Fallow the steps I will provide bellow, I will add some screen shots also to assist if necessary.
    1 - Open System Preferences from  (Menu)
    2 - Click "Network"
    3 - From the Network window you are going to wanna make sure Location is : Automatic and Show is : Airport
    4 - Once the Show: Airport is select you should see a window like this, my window is selected for Ethernet simply because I am no longer running Tiger but the window will look the same.
    5 - You will want to click on the PPPoE tab and you should see this
    6 - You're gonna wanna make sure " connection using PPPoE " is not select, if it is make sure to uncheck this option
    7 - Once this is uncheck make sure to click "Apply Now"
    8 - You will want to click back on to the "TCP/IP" tab and make sure IPv4 is set to " Using DHCP "
    9 - If the option to " Apply Now" is available again please click this
    10 - Once this is done go to your Aiport Menu and see if your network is listed and select your network and eneter the password if necessary.
    11 - If after these steps you still have issues I would recommend Power Cylcing your Router and Modem for about 5 minutes
    12 - Make sure all lights are off when unplugging the Router and the Modem
    13 - Some Modems have backup batterys so you may have to push in a tiny button on the back with a pin

  • I cannot configure pop account in mail, says it is offline.  What to do?

    I cannot configure pop account in mail, says it is offline.  What to do?

    Welcome to Apple Communities
    Download it here and install. Sometimes the download fails with Software Update

Maybe you are looking for

  • Error creating a file in filestream folder

    So, we have a filestream table that we have been using and copying a significant number of image file into over the past month (about 6 million).  So far, the copying has been going well but have run into a problem which I cannot find an explanation

  • How do I set up an indicator to display for only 1 second?

    How do I set up an indicator to display for only 1 second? I am reading keys from a serial device and I turn on an indicator that corresponds to the key pressed on the serial device but how do I set this up so that it only stays lit (enabled) for a s

  • How to change tax code while processing sales order dynamically

    Hi      we need to change the tax condition value whenever Value inclusive tax more than 5250  then it has to capture different tax code ( 12.5 percent ) else it should capture existing tax code ( 5 percent ), is there any user exits or can we do thr

  • How can i connect to a SQL Server 2000 database usgin JSP?

    I need (URGENT) to connect to SQL Server 2000 databse using JSP. I do not know how to program using JSP, so if anyone has any code snippet please let me see it. What i need is to give my site some login/password security. Anyone? Thx.

  • Create huge Waited-Session in a particular moment in Oracle 10g

    Friends, I m new in Oracle 10g database . I m working in a bank and Our production server runs in Oracle 10g in AIX server . We also running an application server with Oracle 10g database . Recently I got a strange problem in my production server .I