Error -10007 after self calibration

I am trying to self calibrate an AT-MIO64E3 using 'Calibrate_E_Series'
under CVI5.0. After the calibration using 'SCAN_op' I regularly (not
always however) receive the error message -10007:
"Channel out-of-range for device type or input configuration; either the
combination of channels is not allowed, or you must reverse the scan
order so that channel 0 is last"
Any hints would be appreciated.
Thanks, Matthias
Sent via Deja.com http://www.deja.com/
Before you buy.

Here is the info from the calibration report.  I can email you a PDF of the report and also a screen capture of the error message which states: "Measurement taken during calibration produced an invalid AI gain calibration constant. If performing an external calibration, ensure that the reference voltage passed to the calibration VI or function is correct. Repeat the calibration. If the error persists, conatct National Instruments Technical Support.
CALIBRATION PERFORMANCE TEST DATA
DUT Information
Type: PXI-4461
Tracking Number: 33367
Serial Number: 33367
Notes
Customer Information
Name: Cal Lab
Address:
Purchase Order:
Notes
Environmental Conditions
Temperature: 23.0 C
Humidity: 13.0 %
Operator Information
Operator Name: administrator
Calibration Date: Friday, March 23, 2007
16:27:03
Notes: Error or termination
occurred. This calibration
may not be valid. Error
code: -200718 Error
message: Error -200718
occurred at DAQmx Self
Calibrate.vi at step Self
Calibrate Possible
reason(s): Measurement
taken during calibration
produced an inval
PXI-4461 Serial Number: 33367
Friday, March 23, 2007 16:27:03 Page 1 of 2
Standards used during Calibration
Type Tracking Number Calibration Due Date Notes
Fluke 5500A Multifunction
Calibrator
32261 3/15/2008
DMM 32260 11/21/2007
33250A 29536 10/13/2007
Calibration Results
Test Canceled
Calibration As Found As Left
Test Value Low Limit Reading High Limit PassFail Low Limit Reading High Limit PassFail
N/A N/A N/A N/A N/A N/A N/A N/A N/A
PXI-4461 Serial Number: 33367
Friday, March 23, 2007 16:27:03 Page 2 of 2

Similar Messages

  • PCI-4472 self calibration error under MAX

    Hi,
    Our PCI-4472 appeared to have offsets on the order of several hundred microvolts, so I tried a self calibration under MAX (version 4.1.0.3001) to try to reduce them. The self calibration gave me the error message:
    "Measurement taken during calibration produced an invalid AI gain calibration constant."
    Does anyone know where I can find troubleshooting information for this issue, or have any suggestions on how to fix it?
    Thanks,
    Ron Norton
    Faculty Research Assistant
    Gravitation Experiment Research Group
    Department of Physics
    University of Maryland
    College Park, MD

    I am posting this for an applications engineer, intended for this to post last in a linear fashion ~
    I have attached screen grabs of the self test and calibration panels.
    When running the "self-test" panels within MAX, all tests pass.
    I performed the tests after hitting "reset" in MAX.
    Current versions of software installed: Labview Full Dev 8.2., and Max 8.5.
    All upgrades have been applied.
    Ai1 is setup to measure a voltage differential of 1-5 VDC based upon a "differential" input
    across a precision resistor. I have setup custom scale, 1V = -40`C 5V = +180`C.
    I have tried setting sample rates of: 3 @3Hz, 10@60Hz, 100 @600Hz, etc...
    As I run the self test, and it's the same when I launch Labview and work within a VI, I see
    the wave output with what looks like a square wave. I have put a scope on the wire pair
    and do see steady voltage levels. Sensor has a current loop supplying external excitation.
    Sensor output passes thru a linearization circuit operating at 60Hz.
    In the 3rd image I captured Ai0 which is similarly setup in "differential" 1-5VDC input to scaled output.
    You can see the graph image captures the trailing signal just like on Ai1 analog channel.
    I have verified from Omega that the sensor is wired correctly supplying current loop,
    and also taken the wires and "twisted" the pairs for less EMI/RFI crosstalk/noise.
    The sensor and it's power supply are brand new purchased directly from Omega.
    I am planning to send this to a local vendor for calibration, SE Labs in Santa Clara, but if it's
    a matter of the boards circuitry being damaged than it needs repair. I started here about
    3 months ago and after asking around found the PCI-6250 DAQmx board was available.
    Problem is I don't know the history, besides the 2 year calibration expiring Nov of 2006.
    The 1 thing I have not tried is moving the PCI to another slot, but FYI I have just this week
    moved the PCI card over to another workstation that exceeds minimum hdw requirements.
    This card was installed in a Pentium III @848MHz workstation with 424MB of ram.
    Facts are the graph output of both channels look exactly the same in either pc.
    The previous user was an engineer, whom may have exceeded voltages/parameters
    on the analog and/or digital inputs. Thank you again for your knowledgeable replies!
    Sincerely,
    Phil Johnson
    Hardware Technician
    http://www.digitalpersona.com/

  • Error during self calibration - PXI-4461

    We are running Calibration Executive 3.2.  We are using a PXI chassis and controller and trying to calibrate a PXI-4461 card.
    In running the procedure, we received the following error during Self Calibration:
    Error 200718 occurred at DAQmx Self Calibration.VI at step self calibrate.
    Any guidance in resolving this issue would be appreciated.
    Richard

    Here is the info from the calibration report.  I can email you a PDF of the report and also a screen capture of the error message which states: "Measurement taken during calibration produced an invalid AI gain calibration constant. If performing an external calibration, ensure that the reference voltage passed to the calibration VI or function is correct. Repeat the calibration. If the error persists, conatct National Instruments Technical Support.
    CALIBRATION PERFORMANCE TEST DATA
    DUT Information
    Type: PXI-4461
    Tracking Number: 33367
    Serial Number: 33367
    Notes
    Customer Information
    Name: Cal Lab
    Address:
    Purchase Order:
    Notes
    Environmental Conditions
    Temperature: 23.0 C
    Humidity: 13.0 %
    Operator Information
    Operator Name: administrator
    Calibration Date: Friday, March 23, 2007
    16:27:03
    Notes: Error or termination
    occurred. This calibration
    may not be valid. Error
    code: -200718 Error
    message: Error -200718
    occurred at DAQmx Self
    Calibrate.vi at step Self
    Calibrate Possible
    reason(s): Measurement
    taken during calibration
    produced an inval
    PXI-4461 Serial Number: 33367
    Friday, March 23, 2007 16:27:03 Page 1 of 2
    Standards used during Calibration
    Type Tracking Number Calibration Due Date Notes
    Fluke 5500A Multifunction
    Calibrator
    32261 3/15/2008
    DMM 32260 11/21/2007
    33250A 29536 10/13/2007
    Calibration Results
    Test Canceled
    Calibration As Found As Left
    Test Value Low Limit Reading High Limit PassFail Low Limit Reading High Limit PassFail
    N/A N/A N/A N/A N/A N/A N/A N/A N/A
    PXI-4461 Serial Number: 33367
    Friday, March 23, 2007 16:27:03 Page 2 of 2

  • How many DSA Cards can be simultaneo​usly Self Calibrated​?

    I have two PXI-1045 chassises connected to a single computer via a dual port MXI Express card.
    Each of the two chassises have 12 PXI-4462 4 channel DSA cards.
    I am using the "DAQmx Self Calibrate.vi", which takes about 156 seconds.
    It appears that if place 24 of these DAQmx Self Calibrate VIs on my block diagram,
    then it appears that 4 will finish after 156 seconds, then another 4 will finish after
    another 156 seconds, and so on, until all are finished after a total of 936 seconds.
    I have  found that 1, 2, 3, or 4 can be simultaneously self calibrated, but more than that requires additional time for each group of 4.
    So my question is, is it possible to sell calibrate more than 4 at once?
    My computer is a quad core, but I don't think this is the limiting factor.
    Could it rather be some other resource, such as Triggers on the PXI bus, that are the limiting factor?

    Hey Kevin,
    The problem you are running into is likely caused by the fact that LabVIEW allocates a default of 4 threads per priority per execution system.  DAQmx VIs by default run at the same priority and execution system as their caller (despite the existence of a "DAQ" execution system), so they are probably running at normal/standard:
    While a .dll is being called by LabVIEW the thread is reserved.  So, 4 parallel calls to DAQmx Self Calibrate.vi (which call into nilvaiu.dll inside the subVI) will reserve all 4 threads that LabVIEW has allocated to the normal/standard priority and execution system.
    To raise the number of threads that LabVIEW allocates, add the line "ESys.Normal=24" or whatever number you'd like to LabVIEW.ini.  You might also have to run threadconfig.vi (instructions found here).  The end result should look like this when you run threadconfig.vi (you'll have to restart LabVIEW after editing the .ini file):
    You could technically just use threadconfig.vi to set the number of threads without manually editing the .ini file, but this only lets you allocate up to 8 threads per execution system per priority.  The number shown in < > is two higher than the actual number of threads allocated due to how the enum was defined in the vi.
    Best Regards,
    John Passiak

  • Self-calibration of NI PCI-6036E

    I have a new NI PCI-6036E (low cost) data acquisition board. We set it up using one of the analog input channels in single-ended nonreferenced mode.
    I want to acquire a voltage value which is equal to a pressure reading of the connected pressure controller.
    After connecting all parts together and set up the parameters I still have an offset between the voltage value from the pressure controller (7.85 V) and the card reading (8.05 V). How can I solve this problem?
    I tried to do a self-calibration of the data acquisition board using the provided online E series diagnostic utility on your web page undere: www.ni.com/support/selftest but it says "device detected but not supported". The E-series model I am using sould be supported
    Can I do a quick self-calibration in an other way?
    It would be very nice somebody could give me some advice how to solve this problem.
    Thank you very much for your help!
    beam

    Beam,
    You can programmatically perform a self-calibration. In LabVIEW, you can call the E-Series Calibrate VI within the Data Acquisition>>Calibration and Configuration subpalette. If using a text-based ADE, you can call the Calibrate_E_Series NI-DAQ function.
    You are right in that the PCI-6036E device should be supported by the online diagnostic utility; that should work. Make sure you have NI-DAQ 6.9 or higher and Internet Explorer 5.0 or higher.
    Also, is the voltage you are reading floating or ground-referenced? If the source is floating then you should use bias resistors when configured in NRSE or differential mode. Chapter 4 of the E Series User Manual (linked below) goes over how/when to connect bias resistors. I would take a look at this chapte
    r and use it as a reference for connecting signals. I hope this helps.
    PCI E Series User Manual
    http://digital.ni.com/manuals.nsf/webAdvsearch/06F1C9FB0D0BA5C286256C010057461B?OpenDocument&vid=niwc&node=132100_US
    Regards,
    Todd D.
    NI Applications Engineer

  • Why is Self Calibration of 6711 inconsistent

    Self Calibration of 6711 using E-series calibrate in Labview RT is inconsistent (error issued: -10842) inspite of warming up the board, ensuring that the board has an IRQ assigned and Front Mount connector being removed. Is calibration for 6711 not supported in Labview RT?

    Have you tried using the 6711 as an RDA device and performing the calibration with the configured RDA device? This might get around the error. I hope this helps.
    Regards,
    Todd D.
    NI Applications Engineer

  • We had some accuracy issues with NI 5112 scope in one of the ATE and then I decided to perform self calibration using Labvidew vi to perform "niScope_CalSelfCalibrate(handle, "", 0);" function.

    We had some accuracy issues with NI 5112 scope in one of the ATE and then I decided to perform self calibration using Labvidew vi to perform “niScope_CalSelfCalibrate(handle, "", 0);” function.
    But it made it worse. I tried using option 2 to restore but it did not work.
    Could you pls advice me to resolve this issue.

    Hi Ana10,
    Are you using this digitizer with NI VideoMaster? if not you should probably post this in the Digitizer forums. That said I would suggest using the self calibrate function in MAX for this device rather than the LabVIEW API method just so that you can rule out any errors in correctly configuring the digitizer for self cal in LabVIEW. Also you should ensure that all inputs are disconnected before performing a self cal. If this still results in an error in calibration you could refer to the following document or arrange to return the digitizer to NI for external calibration.
    http://www.ni.com/pdf/manuals/370328e.pdf
    Hope this helps,
    Nick

  • I cant download itunes programme for my samsung rv510 laptop, it says error occurs after I try to instal the 10.7 version? What do I need to do?

    I cant download itunes programme for my samsung rv510 laptop, it says error occurs after I try to instal the 10.7 version? What do I need to do?

    .dmg.part is not a complete file..  It usually indicates that the entire file has not been downloaded yet.  Are you sure you downloaded the entire file?   (i recommend updating itunes through  > Software Update
    Alternatively, you can download it from here: http://support.apple.com/kb/DL1614

  • Error message after installing Office 2013 from DVD

    After i install office 2013 on new Win 8.1 Pro laptop , i can start in control panel mail 32 bit option to configure accounts...
    i found error in event logs:
    Log Name:      Application
    Source:        Application Error
    Date:          2014-06-24 5:20:58 PM
    Event ID:      1000
    Task Category: (100)
    Level:         Error
    Keywords:      Classic
    User:          N/A
    Computer:      PMCHome
    Description:
    Faulting application name: rundll32.exe, version: 6.3.9600.16384, time stamp: 0x52158827
    Faulting module name: KERNELBASE.dll, version: 6.3.9600.16656, time stamp: 0x531812f4
    Exception code: 0xc06d007e
    Fault offset: 0x00012c1a
    Faulting process id: 0x1180
    Faulting application start time: 0x01cf8ff2304d4f6f
    Faulting application path: C:\Windows\SysWOW64\rundll32.exe
    Faulting module path: C:\Windows\SYSTEM32\KERNELBASE.dll
    Report Id: 6ebd5d06-fbe5-11e3-8263-7c7a9101e4d2
    Faulting package full name: 
    Faulting package-relative application ID: 
    Event Xml:
    <Event xmlns="http://schemas.microsoft.com/win/2004/08/events/event">
      <System>
        <Provider Name="Application Error" />
        <EventID Qualifiers="0">1000</EventID>
        <Level>2</Level>
        <Task>100</Task>
        <Keywords>0x80000000000000</Keywords>
        <TimeCreated SystemTime="2014-06-24T21:20:58.000000000Z" />
        <EventRecordID>1851</EventRecordID>
        <Channel>Application</Channel>
        <Computer>PMCHome</Computer>
        <Security />
      </System>
      <EventData>
        <Data>rundll32.exe</Data>
        <Data>6.3.9600.16384</Data>
        <Data>52158827</Data>
        <Data>KERNELBASE.dll</Data>
        <Data>6.3.9600.16656</Data>
        <Data>531812f4</Data>
        <Data>c06d007e</Data>
        <Data>00012c1a</Data>
        <Data>1180</Data>
        <Data>01cf8ff2304d4f6f</Data>
        <Data>C:\Windows\SysWOW64\rundll32.exe</Data>
        <Data>C:\Windows\SYSTEM32\KERNELBASE.dll</Data>
        <Data>6ebd5d06-fbe5-11e3-8263-7c7a9101e4d2</Data>
        <Data>
        </Data>
        <Data>
        </Data>
      </EventData>
    </Event>
    Can some answer why?
    Fresh install, no errors in install at all
    Regards

    Hi,
    Rundll32.exe is responsible for calling Windows DLL (or Dynamic Link Library) files and loading them into your PC's memory.
    Usually the rundll32.exe Application error is not a straight forward error. There are lots of possible reasons for this error, includes a failed or incorrect installation or uninstallation of any softwares, which might have left erroneous Windows registry
    entries.
    Tracking down the exact issue can take lots of investigation. The severity of this error really depends, if everything works well for you, you can just ignore it. However, if you want to correct this, try below steps:
    Since you notice this error right after you installed Office 2013, try to remove and re-install it using default setting.
    Reboot your computer, many programs require a reboot for successful installation.
    Try to update your computer drivers. This is usually the cause of lots of .dll file error.
    Try to run the System File Checker tool (SFC.exe) to correct any incorrect system file. It is a utility in Windows that allows users to scan for corruptions in Windows system files and restore corrupted files:
    http://support.microsoft.com/kb/929833/en-us
    Hope this helps.
    Thanks,
    Ethan Hua CHN
    Forum Support
    Come back and mark the replies as answers if they help and unmark them if they provide no help.
    If you have any feedback on our support, please click
    here

  • I used to update my iPhone 4S applications whenever a new update is available, but at this time, "unknown error" displays after entering my apple id password. Anyone who could help, whom they might have experienced? Thanks

    I used to update my iPhone 4S applications whenever a new update is available, but at this time, "unknown error" displays after entering my apple id password. Anyone who could help, whom they might have experienced? Thanks

    Hi Christianjr,
    When you say you reset your network settings, what do you mean, exactly?
    I saw another post on this issue, and the poster indicated that they resolved the problem by turning off their cellular date option and wifi options in Settings, and then turned them back on.
    Also, resetting your phone may help - press the Home and Power buttons at the same time and continue to hold both down until the Silver Apple appears. Once it is back up, click on the update option and see if it fixes the problem.
    If none of those things work, then need to have details on what you did to reset your network settings....
    Cheers,
    GB

  • I have no trouble viewing apps in the iTunes Store on my iPad. However, every time I try to install a new app, I get an error message after a while - cannot connect to iTunes Store. This issues cropped up two days ago on my new iPad. Please help...

    I have no trouble viewing apps in the iTunes Store on my iPad. However, every time I try to install a new app, I get an error message after a while - cannot connect to iTunes Store. This issues cropped up two days ago on my new iPad. Please help...

    JUst experienced the exact  same problem after changing password.Getting same message. Hope someone has an answer for this.

  • Sorry, something went wrong --- I got an error message after a new webapp was built and a new site collection was created

    Sorry, something went wrong --- I got an error message after a new webapp was built and a new site collection was created.
    This in SharePoint 2013 Farm.
    What should I do?
    1- I did a DisableLoopBack and set the value to "1".
    For hostname do I have to do : 111.111.111.111 (sample) and serversp2013?
    P.S. I fixed this issue once, long time ago and I forgot how it ... be done.

    Hi,
    For troubleshooting your issue, please take steps as below:
    1.Go to IIS, make sure the web application is actived.
    2.Go to IIS -> Applicaiton Pools, make sure the application pool related to your web application is started.
    Reference:
    http://support.microsoft.com/en-us/kb/926642/en-us
    http://www.techrepublic.com/forums/questions/cannot-browse-created-web-application-site-in-sharepoint-services-30/
    Best Regards,
    Eric
    TechNet Community Support
    Please remember to mark the replies as answers if they help, and unmark the answers if they provide no help. If you have feedback for TechNet Support, contact
    [email protected]

  • Error message after upgrading to 7.3

    i upgraded to iTunes 7.3 and i got a message that said something about some files were missing and i needed to uninstall and try again. i uninstalled it and now i can't even download iTunes. i can download but i can't do anything else. anybody else having this problem?

    Yes, i had the same error message after upgrading to 7.3.
    I am using windows XP - i went to the Control Panel -> Add/Remove Programs -> Scrolled to iTunes & clicked on Change (not Remove), then clicked Repair. It all works fine after that.
    Search this forum - i hope you find a solution.

  • I downgrad my iphone from ios5 to ios 4.3.3 and after restoring the ios 4.3.3 i fixed all itunes error but after that my i phone screen turned off and the apple logo didnt shows up any more what ever i did my iphone model 4 gsm

    i downgrad my iphone from ios5 to ios 4.3.3 and after restoring the ios 4.3.3 i fixed all itunes error but after that my i phone screen turned off and the apple logo didnt shows up any more what ever i did my iphone model 4 gsm and fyi  i try all the ways to turn it on but it didnt sucsses

    roaminggnome 
      thanks for helping but i didnt know that is the downgrade is forbidden  and i ask for help and u didnt help me so i just want to thank u for telling me that is the downgrade is forbidden >>> but now i didnt have i phone any more so thank u so much again >>>> 

  • HT1937 I tried to update software on my Iphone 3 from Itunes after i managed to sync phone to laptop but then i had a message (error 1015) after completing saying itunes has detected an iphone in recovery mode, i am now unable to do anything with my phone

    I tried to update software on my Iphone 3 from Itunes after i managed to sync phone to laptop but then i had a message (error 1015) after completing saying itunes has detected an iphone in recovery mode, i am now unable to do anything with my phone, please help!!

    You generally get that error code on a jailbroken phone ?
    If it is Jailbroken then I am afraid you will get no help here.

Maybe you are looking for