Floating voltage in BNC 2090, PCI 6024E

Hi experts, I have a single ended connection on a BNC 2090 connected to PCI 6024E board. However, when I leave it connected(without any input) I see that the voltage gradually increases to +10V. Even when I input a signal, I see that the 10V forms a baseline, which hinders any measurement. Is this some sort of grounding problem? Kindly post any helpful thoughts on this matter. thanks. Roshan.

Roshan,
I believe that the 2090 has a switch for each channel which configures the inputs for either Single Ended configuration (AI channel is read with respect to your 6024's Analog Input Ground) or Differnetial configuration (AI channel reads signal on BNC connector differentially). Ensure that the terminal configuration you have programmed for you channel matches the terminal configuration selected by the switch on the 2090. If you do not make any explicit setting, the 6024 will default to differential mode.
I hope this helps,
Dan

Similar Messages

  • BNC-2090 & PCI-6036E: MIO power or LPM power?

    Which power mode do I choose?
    There's a choice of "MIO-16" and "LPM-16" on the back of the BNC-2090.
    Neither the string "MIO" nor the string "LPM" appears in the owner's manual for the PCI-6036E [322339c.pdf].

    The PCI-6036 is an MIO board so you'll want to use the "MIO-16" selection. The MIO boards are our newer series of boards whereas LPM series is our older legacy boards.
    Erick D.
    NI Applications Engineer

  • Strange voltage observed in NI BNC 2090

    Hi all, I have a BNC 2090 and AT-MIO-16O-2 DAQ board. I am recently
    (starting this morning) experiencing strange voltage readings from the
    ACH1, 2,3,4and 5. The voltage input from ACH0 is about -.2V; and range
    from -1 to -3 in other channels. These happens even when nothing is
    plugged into the BNC other than the its connection to the DAQ in the
    computer.
    any suggestion on what might be happening?
    Roshan.

    This is actually very common. When nothing is connected to the pins the voltage will charge up on the leads. Sense the lines are not connected to anything and the voltage has no where to go the lines will normally completely saturate the capacitors on the board.
    This is not a bad thing and it should not affect your measurements, if you do not scan the channels that are not connected. If you are worried about the unused floating signals, you can wire the unused inputs to ground.
    If you are performing measurements and notice that the inputs are floating you should check your grounding methods. In general use RSE for floating sources, NRSE and DIFF for grounded signal sources. If you want to use NRSE or DIFF on a floating signal you will need to add a 10-
    100kOhm bias resistor from the negative lead to AIGND to reference the signal. Refer to this KB for more help.
    http://digital.ni.com/public.nsf/websearch/DF70FCE​03D397BFE86256A3700780FFA?OpenDocument
    I hope this helps,
    Joshua

  • Using PFI1-PFI9 on BNC-2090 to trigger recording on a PCI-MIO-16-XE-50

    Hi there,
    I've got another question about triggering using the BNC-2090. Now I can get an output of a particular digital line from the DIO-96 I'm using using a ribbon connector connected to CB-50 Connector block. This connector block has individual line outputs, so is it possible to take one of these individual lines, plug it into PFI1-PFI9 of the digital connection area of the BNC-2090 and have the PCI-MIO-16-XE-50 use the appropriate digital PFI line as a trigger? I ask this because I assume the default trigger line whey I click "Use external trigger" in NIDAQ Tools for Igor Pro is PFI0, so I was wondering if there's a way to change that. Thanks a lot!
    Arul Thangavel

    Arul,
    Yes, you can connect the dio line of the dio-96 to any of the PFI lines that are located in the pinout of the bnc-2090. The default line to be used for triggering an acquisition is PFI0,but you can modify that.
    In LabVIEW, you can select the proper PFI line by configuring it in the "AI Start.VI", or by using the "AI Trigger Config.VI" before using the latter VI.
    In LW/CVI, you can also select the proper PFI line by configuring it with the "DAQ_Config" function call and with the "Select Signal" function call.
    Hope this information is helpful.
    L Aguila
    Applications Engineer
    National Instruments

  • How do i generate the TTL pulse from PCI 6024E DAQ card

    Hi, I'm trying to write a program which generates a single TTL pulse when it is run. My goal is to generate the pulse, send it to the output channel of my PCI 6024E DAQ card, and then use a BNC cable, to display the signal on the oscilloscope. I have written the code and I think it appears fine but when I try to view the signal on the scope I am unable to see the pulse. I have tried slowing down the sweep on the scope and have adjusted the voltage and time scales but have not seen any response. After examing my DAQ card, I noticed that none of the PFI/TRIG terminals on the card are wired; the output channel connects to terminals for DAC0OUT and Ground. I'm wondering if this pin needs to be wired in orde
    r to generate the pulse since PFI0/TRIG1 is the start trigger terminal. Any help would be appreciated. Thanks.
    Attachments:
    ttl.vi ‏26 KB

    Hi,
    You are looking for a retriggerable pulse generation. Please check the Generate Retriggerable Pulse (DAQ-STC).vi example that ships with LabVIEW.
    Regards,
    RamziH.

  • Bnc 2090 offset and other errors

    I have a bnc 2090 plugged into a pci 6070e. The BNC is currently plugged into a DC variable power supply. Right now, when I change the voltage on the power supply, the other channels reflect this change even though it is only plugged into one channel. I have read about ghosting and even when I lower the sampling rate, the problem is not fixed. There is also an offset that is showing up on the MAX test panels and in my VI. This offset steadily decreases and seems to arbitrarily change. Whats going on here?

    About the offset another point, I assume you use a measurment on one coax connector (+ and - in one connector), in that case your switch should be set to RSE.
    In MAX this should be set as well (and in your VI) instead of the default differential measurement.
    Ton
    Free Code Capture Tool! Version 2.1.3 with comments, web-upload, back-save and snippets!
    Nederlandse LabVIEW user groep www.lvug.nl
    My LabVIEW Ideas
    LabVIEW, programming like it should be!

  • Cross talk in BNC 2090

    Thanks for the information provided. I followed your advice by connecting
    DGND1 with a wire to USER1 and DGND2 to USER2. But I still get cross talk.
    It also puzzles with a fact that when I connect a signal source (from a
    functional generator) to ACH1 but in fact the signals show up on the
    Biobench screen as ACH0 as defined. So my hunch is that something wrong
    happens to BNC 2090 or the PCI-MIO-16XE-50 board installed in the computer.
    Would it be possible for me to send in the BNC2090 for a check? If so, what
    are the procedures to follow.

    We handle all repair requests over the phone. The procedure is as follows:
    1. Go to www.ni.com/ask to create your own Service Request number (SR#). Choose the Phone NI option. Fill in the form with the appropriate information.
    2. Call the support number that is given to you and enter in your SR#. You will be directed to an Applications Engineer (AE) that will assist you through the process. The AE will be able to answer any questions you may have about the repair process.
    Have a great day.

  • How does the system Validyne pressure transducer -- Validyne CD280-Dual-- SCB-68-- PCI-6024E-- LABVIEW work and how can we calibrate the pressure transducer?

    Hello there,
    I am completely new to the LABVIEW software.
    I am learning an existing LABVIEW code of my pressure acquisition system (attached) which has the data path as following: Validyne pressure transducer --> Validyne CD280-Dual--> SCB-68--> PCI-6024E--> LABVIEW and I have some question:
    1. How does the complete system from transducer to the LABVIEW work? ie if we apply the pressure at the transducer, it will change the resistance of the transducer then ....?
    2. How can I find the equation which express the relationship between the pressure and the voltage at the oupt of the Validyne CD280-Dual in the LABVIEW?
    3. If I want to do the calibration for the transducer, what are steps?
    Thanks,
    Solved!
    Go to Solution.
    Attachments:
    Experiment.vi ‏143 KB
    Plogger.vi ‏21 KB

    Hello there,
    I do not understand what subVIs you are refering to but I have tried to run the LABVIEW program using the files I attached and it worked fine (pls see the attached pressure plot).
    As I mentined before, I am new to LABVIEW so maybe I do not know.
    Thanks,
    Attachments:
    Pressure plot.jpg ‏111 KB

  • DIO7 of PCI-6024E forgets input/output setting?

    encoder pulses fed to PCI-6024E DIO7 (used as up/down input) to GPCTR1; after a few days DIO7 loads down the encoer signal, dropping the voltage. It seems that the DIO line is not switched to INPUT mode. MAX thinks it is and input but the DIO hardware does not behave like it. Anyone had this problem? Any solutions?

    No, this isn't something that is a known issue. It might not be that the DIO line is loading down the encoder. There might be a problem with the encoder itself. To check this, do a continuity/Ohm check between DIO7 and ground.
    Russell
    Applications Engineer
    National Instruments
    http://www.ni.com/support

  • Can I use a SH68-68EPM cable to connect a PXI-6704 with a BNC-2090?

    Hi,
    I was wondering if I could use a SH68-68-EPM cable to connect a PXI-6704 (sitting in a PXI-1002 Chassis and talking to my (Windows 7) PC via a PCI-8331) with a BNC-2090-box.The manual suggests to use a D1-cable but maybe it is possible to still get the analog outputs from the BNC-box although the cable is different?
    Thanks for your help!

    Hi cars10,
    The BNC-2090 is not compatible with the NI 6704 due to pin mapping.  You can use BNC connections via the CA-1000 or use a card from the analog output series family with the BNC-2090A or BNC-2110 as described in the user manual. 
    Regards,
    h_baker
    National Instruments
    Applications Engineer
    Digital Multimeter Resources

  • Noise in thermocouple measurement using PCI-6024E DAQ card

    I am using a PCI-6024E DAQ card for thermocouple(TC) measurement. The
    TC is installed in the mold cavity of an injection molding (IM)
    machine. The leads from the TC are connected to the DAQ card using a
    CB-68LP board. I made a  temperature mesurment task in LabView 7.0
    with the two leads from TC as differential inputs and used CJC value of
    25 deg C.
    The problem is that when the IM machine is OFF, I get a reasonable temp
    reading with some slight noise but as soon as I turn ON the IM machine,
    the temp readings go wild with a lot of variations around the mean
    value. The temp values also don't go to higher values when the TC is in
    contact with the molten plastic in the IM machine.
    Do I need to ground something here or put a resistor in the I/O board?
    BTW the IM machine is grounded and the TC give correct temp values when
    connected to a hand-held temp reader.
    Any help will be highly appreciated.

    Hello  Aftab,
    Looks like you need to use bias resistors to reduce the noise. Check out the following tutorial for more information about how to use bias resistors.

  • I need to interface SCC-TC02 with help of SC 2345 and PCI-6024E DAQ Card.please tell me the procedure and if possible send me the program

    Dear Friends,
               I need to interface SCC-TC02 with help of SC 2345 and PCI-6024E DAQ Card.But i can not understand example program.please send me the procedure to interface these three componnets and if possible send me the program.
    With Regards,
    Eswaramoorthy K V

    nce of nth triggering pulse. i need to know when the when the ist trigger occurs and when the nth trigger occurs . please tell me how to to . what i know is that event triggering has to be done with start and stop trigger. please tell me how it has to doneSuresh;
    What you will need to do is to set up a digital start and stop triggered Analog Input acquisition. Then you will need to have a counter set as event counter, having the specific number of pulses you need configured as the preset value, configured to count down, and generating a pulse after the terminal count has been reached. That counter output will be the stop trigger of the Analog Input operation. In summary, you will have the external pulse being both the digital trigger of the Analog Input operation and the source pin of the counter, and the counter output being the digital stop signal for the analog input.
    I'm attaching a Labview VI that does the start and stop analog input acquisition. You will need to include the counter part and set the stop s
    ignal to be the counter output.
    Hope this helps.
    Filipe
    Attachments:
    AI_Start-Stop_D-Trig.zip ‏25 KB

  • Can version 7.1 work with a PCI-6024E?

    Hi! I've been trying to run a VI which is being successfully used in my lab to acquire data. However, whenever i try to run it i get stuck when the AI Waveform Scan VI file is being searched. Following your replies to similar questions i've made sure that the traditional NIDAQ is properly installed. is it possible that my data acquisition card (PCI-6024E) cannot support ver. 7.1? what do i do next?
    Thanks,
    Tamar

    from what i understand from your post, your problem has nothing to do with the card.
    your vi cannot run because a subvi is missing: AI Waveform Scan. it is probably in a different library than on the original computer (i would guess the original computer has a different version of LV installed). maybe it is one of the examples in a previous LV version?
    Either way, the easiest solution for you is to go to the original computer, save your vi with all dependencies (as an llb for example), and try again on the new computer.
    ... And here's where I keep assorted lengths of wires...

  • R6850 kit to connect 5B backplane to PCI-6024E DAQ card is it possible?

    I have an old 5B01 backplane that I have connected to an SC-2050. Can I connect the SC-2050 to my PCI-6024E DAQ car using the R6850 kit?

    Lcacho,
    I've reviewed the manuals and specifications for all those devices, and the connection should work.  Please note, however, that given the age of the 5B01 backplane, I cannot confirm this completely.  However, from the information I have, you should be able to connect all these devices.
    Seth B.
    Staff Test Engineer | National Instruments
    Certified LabVIEW Developer
    Certified TestStand Developer
    “Engineers like to solve problems. If there are no problems handily available, they will create their own problems.”- Scott Adams

  • Maximum Frequency on PCI-6024E

    Hello, I'm just wondering what is the maximum frequency can PCI-6024E hold?
    My sensor's range is .016KHz - 16.6665 KHz, if this is too much, can I just use
    a divider to cover what is max limit of the card?
    Thanks!
    KowdTek
    LabVIEW 2009
    One Step At A Time, Maybe Two...

    200Ksamples/sec on a single channel, per the specs:
    http://www.ni.com/pdf/manuals/370719c.pdf
    For optimum accuracy, sample at 5-10 times the highest frequency component of the signal if possible (minimum of 2x per Nyquist), that would be 166.6Ks/sec rate max which is still within range of the card.
    ~~~~~~~~~~~~~~~~~~~~~~~~~~
    "It’s the questions that drive us.”
    ~~~~~~~~~~~~~~~~~~~~~~~~~~

Maybe you are looking for

  • Automatic Reprocess of Trfc messages in PI production server using RSARFCEX

    Hello All, i am getting frequent errors of SYSFAIL in TRFC queue in PI. I am trying to schedule RSARFCEX report as the Background job, please advice me "is this correct solution" and also please tell whether this will effect any other messages other

  • Sharing .mac calenders - gives me a problem

    Im subscribing on some calenders from my girlfriend, we both have .Mac accounts but I get a warning i iCal, when I want to update the subscribed calenders: "the severe idisk.mac.com dismissed your log-in name og the password" (my transation). I know

  • Profit center - actual data and plan data question -?

    Hi Gurus Please clear my doubts in profit cneter accounting. 1) what is PLAN data  -? (What is planning in a bigger picture) why do we need plan data -? 2)ACTAUL data will be posted in the profit center from other SAP moduldes. So with out PLAN data 

  • Duplicate entries in table

    Hi, I am using a standard function module, BBP_GET_BOBUSERS_F4. However, after the execution, im getting duplicate entries in my table. Please suggest cause and solution. Thanks, Prisford Pereira.

  • Please Help : (x)newInstance()

    I want to load class and instantiate object, but i dont i now the type of object : my code java is : Class cls = class.forName(org.com.xxxx); int idx0 = str.lastIndexOf("."); String testObj = str.substring(idx0+1); obj = (testObj)cls.newInstance(); w