How work the Visa session?

I'm working with the Visa session. When I receive from an external processor the hex value 0X0A then suddenly aboard the reading instruction. When I want to receive the value 0XAA0A55 then I receive only 0XAA0A. Why?

Hi
defuflo is right. The read ends due to the termination character. If you do binary serial transfer you have to do the following:
You have to set the "Termination Character Enable" property to false. In LV 6.0.2 use a property node which input comes from the "duplicate VISA resource name" output of the "VISA configure serial port", in LV 6.1 it is a parameter to that function.
Second use a property node to set the "Serial Settingserial End Mode For Reads" to "None".
Waldemar
Waldemar
Using 7.1.1, 8.5.1, 8.6.1, 2009 on XP and RT
Don't forget to give Kudos to good answers and/or questions

Similar Messages

  • How work the decompiler flash?

    i wnat to know how work the decompiler flash..
    thank nissan.

    Hello & welcome Iotu!
    http://www.tech-recipes.com/rx/4836/os-x-how-to-burn-a-multisession-cd-or-dvd/
    http://myrightbrain.wordpress.com/2006/08/12/how-to-do-multisession-burns-with-m ac-os-x/

  • FCS and image sequences (TGA...) : how works the proxy creation?

    Hello,
    I would like to know how work FCS with image sequences (TGA, TIFF sequences...) for proxy creation.
    For example, the image sequences are created by 3D workstation at the speed of one frame per second.
    Does FCS create a proxy each time the folder is checked? The old proxy is delated?
    Thank you for your answer

    Jim is right. You can't replace a clip with an image sequence.
    If you are rendering your comps from one system, follow Jim's sage advice.
    But, in the cases when I do a multi-system network 3d render, there is
    no way for all systems to write to the same Uncompressed AVI file...
    so rendering to an image sequence is the only option.
    You say:
    ...for speed reasons -  converting them to lower res quicktime files
    and using those to edit in Premiere Pro.
    You might consider converting your sequences to both a low-res proxy for editing,
    and also convert to full-res lossless codec AVI files at the same time for swapping out later.

  • How works the analogo filter of the 4472?

    I need understand how filter the signals the 4472 board.
    Thanks

    Hello!
    That's a pretty broad question, the best thing to do would be to take a look at the 4472 manual on pages 3-4 through 3-9 at the following URL:
    http://www.ni.com/pdf/manuals/322940c.pdf
    This is an extensive explanation of how the antialiasing filters work, including several graphs of outputs.
    Hope this helps!
    Brian Spears
    Applications Engineer
    National Instruments

  • How works the assignation of a group to a role visitor in WLP

    Hi everyone,
    I have some problems with Siteminder integration with security in Weblogic Portal.
    Basically when I try to show the groups in my SM security provider for assign to a Role Visitor it produces a siteminder API error:
    weblogic.management.utils.NotFoundException:
         at com.netegrity.siteminder.weblogic.sspi.auth.ci.h(DashoA10*..)
         at com.netegrity.siteminder.weblogic.sspi.auth.SiteMinderAuthenticationProviderImpl.listMemberGroups(DashoA10*..)
         at com.netegrity.siteminder.weblogic.sspi.auth.SiteMinderAuthenticationProviderMBeanImpl.listMemberGroups(DashoA10*..)
         at sun.reflect.GeneratedMethodAccessor4239.invoke(Unknown Source)
         at sun.reflect.DelegatingMethodAccessorImpl.invoke(DelegatingMethodAccessorImpl.java:25)
         at java.lang.reflect.Method.invoke(Method.java:585)
         at com.bea.p13n.usermgmt.AtnProviderProxy$Runner.run(AtnProviderProxy.java:156)
         at weblogic.security.acl.internal.AuthenticatedSubject.doAs(AuthenticatedSubject.java:363)
         at weblogic.security.service.SecurityManager.runAs(Unknown Source)
         at com.bea.p13n.usermgmt.AtnProviderProxy.invoke(AtnProviderProxy.java:113)
         at $Proxy369.listMemberGroups(Unknown Source)
         at com.bea.p13n.security.management.authentication.internal.GroupProvider.getParentGroupNames(GroupProvider.java:363)
         ... 112 more
    Caused by: com.netegrity.siteminder.weblogic.sspi.auth.dw: Could not obtain groups for user: fnp_sane_plasan_plapro_actpet_bpm_grupActuacio
         at com.netegrity.siteminder.weblogic.sspi.auth.b9.b(DashoA10*..)
    This error is reported to the provider but we haven't solution yet and we need a workaround and I would like to know how is done the relationship between the role visitor and the assigned groups, is this information in LDAP? who I can introduce this relationship manually?
    Thank you.

    I have the role visitor created. I am trying to assign a group to the role visitor in PAT, but when I selected the Siteminder Provider to view all the groups this error happens.
    The Siteminder error is recognized by Siteminder as a bug and is making a patch (5 months ago) and we haven't answer yet, and then I would like to find a workaround to this problem.
    I try it with the API but Oracle recognize me another bug when I try to create a visitor role for Enterprise Scoped Applicaction, only works in Web Application scope. (1 month ago)
    And then the last solution is doing this assign manually... and I would like to know how I can do it in LDAP (if LDAP is the repository of this information)

  • Counter function : How works the Set attribute function

    Hi,
    I use Labview and a 6713 DAQ
    I want to use the "set attribute vi" but I don't exactly what means the "attribute value number" and the attribute value type" in that vi
    >Where can I find some data about that?
    >What is the relation between the internal clock frequency and the attribute value number if I generate a pulse train ?
    >How can I configure the frequency of my DAQ ? ( I can use 20MHz or 100KHz )
    Any help is greatly appreciated
    Renaud

    Renaud,
    You can read the on-line help for the Set Attribute VI by right-clicking over the VI and selecting "Help." The help will describe each of those parameters. In addition, you can search through counter examples for your 6713 by selecting from the menu bar Help>>Search Examples>>I/O Interfaces>>Counters. Any example that demonstrates the functionality of the DAQ-STC counter will work for your board. In addition, you will find examples on pulse train generation.
    Hope this helps.
    Regards,
    Erin

  • How is the 'Restore Session' feature disabled entirely?

    Ive come to the end of my days using FF if this continues.
    Im sick and tired of the session restore feature trying to RESTORE a session that I closed out, sometimes by using the power button quickly...say because I crossed a site that I thought might be infested or something.
    Can it be DISABLED?
    If so, please tell me how and I appreciate the help.
    If not, Mozilla, this has cost you a user..and Im betting not just one.
    GET RID of this feature if it cant be disabled. I HATE it. It serves NO purpose other than to irritate users like me who have to shut down the brower or even the pc because some crack head put a virus on a site and we're trying to NOT get it...then when we reopen FF it just RESTORES the SESSION thereby OPENING the very site we DIDNT WANT to OPEN!
    Im fed up.
    If I cant disable this feature, Im done.
    Any help is appreciated.

    Close all tabs before closing the browser. If you do not clear the history then others can check that as well, so you may want to use [[Clear Recent History]].

  • How works the plc

    how can i get the info from several serial scanners thru plc on my visual basic plataform, without use lookout or labview?

    how can i get the info from several serial scanners thru plc on my visual basic plataform, without use lookout or labview?

  • VISA session to an instrument handle

    I tried to aquire data from TEKTDS 220. I use "set condition" to "aqu
    status",
    then "read waveform" subVIs. So there is a problem: I need to connect the
    "VISA
    session" to "instrument handle", but it doesn't work. Could you please tell
    me
    how to connect these two type of wires?
    Thanks a lot!

    It looks like you're trying to mix the tktds2x0 IVI driver with the tktds2xx LabVIEW native driver. The IVI driver has an "instrument handle" which is an IVI session, and the LabVIEW driver has a "VISA session". The two wires are not compatible. You should either use just the IVI driver or just the LabVIEW driver.

  • Remotely closing VISA sessions??​?

    I have a network with three computers running LabView 6.1 The master computer collects data via fieldpoint ethernet devices and serial devices. If the master should go down (power loss, software freeze ect.) one of the slave computers takes over. The problem I am having is closing the VISA session origially opened be the master (which is now down). The new master can never talk to the serial devices because of VISA session errors. How can I remotely close a VISA session????????????

    I haven't worked with the fieldpoint ethernet devices,
    but other ethernet devices either:
    a) accepted a reboot command (with a password protection) - the reboot would clear VISA sessions.
    or
    b) could set a timeout on ethernet connections - ie, detect that the modem disconnected, close the VISA sessions.
    Check to see if those options are available.
    Les Hammer

  • How maintaining the connection table for sticky/persistent/non-persistent?

    Question about how to maintain the connection table for the source(client) and destination(server) in the CSM(or CSS).
    I know the sticky has the table and max size such as 128K(css11501)as per CCO but not clear how works the persistant and non-persistant case.
    Q1) persistant. does it maintain the conntion table to tracking the session? then, any information the table size?
    Q2) non-persistant. is this also have connection table? then, how it works?
    why I'm asking is want to understand how the session keep tracking. for example, the router based on the routing table(stateless) versus PIX firewall has stateful table. As analogue, is the non-persistnet stateless and statefull for the persistent and sticky?
    Thnaks in advance,

    The CSS uses FCB to maintain information about active connections.
    Each connections requires 2 FCB - one for client to vip and one for server to client.
    When you boot the CSS it will immediately reserver a good amount of memory to create a list of FCB.
    Each connection will then take 2 FCB from the list.
    You can do a 'flow stat' from llama mode to verify how much free/used FCB you have.
    When running low on FCB, the CSS will try to allocate more memory.
    Gilles.

  • Why Open VISA Session?

    In the driver here, the VISA Open sub-VI states that it "must be called first" when using the device. There is a similar sub-VI for closing the session.
    How true is this?
    I can open and run a different sub-VI (such as Write Single) with no immediately obvious problems, just by specifying the correct device in the VISA Session field. Are errors or performance hits building up in the background?
    My problem is that the KS 3988 will be called on in sub VI's within sub VI's within sub VI's (relative to the main program). I'm not clear on where, exactly, I should be putting that VISA Open panel. This is especially confusing if the "dup VISA session" needs to be wired into every use of the Write Single sub-VI.

    Here is a picture of the VI which actually calls the KS 3988.
    My problem is that this thing is just the sub VI of a VI which is itself the sub VI of a VI that is running in a frame. The top-level VI that contains the frames has four such frames.
    Now, those top three VI's all read out of the black Admin sub VI, and the bottom can both read from it and write to it. Currently, all programs use the same KS 3988 via VISA, but that needs to be something we can change on the fly.
    So, where does the Open sub VI go? Do I need to include, in the frame-containing VI, something like this?
    Message Edited by DJDDA on 07-31-2008 04:10 PM
    Attachments:
    convertfix.jpg ‏17 KB
    convertvi1.jpg ‏26 KB
    convertvicontext.jpg ‏66 KB

  • How works Get Property Value / Type Automatic?

    Hi,
    I need to get in Labview some variables from Teststand. The VI I wrote use "Get Property Value" from Teststand and works fine... when I specify the type of variable I whant from Teststand (Boolean, I32,...).
    But I need a VI which retrieve the value of my variable independently of the variable type.
    I tried to set the type option "Automatic" but the "Get Property Value" always give back an empty array of boolean....
    How works the "Automatic" option by the "Get Property Value"?
    Is there a possibility in Labview to get the type of a Teststand variable?
    Cheers,
    Risotto

    Risotto,
    the "automatic" feature is no feature but an entry made because of the
    polymorphic design of the "get property value" VI. if you would change
    the order of the VIs contained in the polymorphic VI, you would
    retrieve other kind of values.
    therefore never use the automatic setting of this VI. the
    simplest way to solve such thing is to pass the data in variant and an
    additional numeric value for identifying the type of the value.
    Norbert B.
    - NI Germany
    CEO: What exactly is stopping us from doing this?
    Expert: Geometry
    Marketing Manager: Just ignore it.

  • How work a buffer?

    Hi,
    I work on a buffer. I want to acquire an .avi in an buffer of ten seconds. For example, all images acquired by the camera come in a buffer
    9[1][X][X][X][X][X][X][X][X][X] -> [2][1][X][X][X][X][X][X][X][X]  etc...  [1][2][3][4][5][6][7][8][9][10] and after [11][1][2][3][4][5][6][7][8][9]    and when I push on capture it copy only the ten last seconds. I think that is a ring buffer.I do two .vi but I don't understand how works the buffers.           
    Attachments:
    Essaione.vi ‏22 KB
    Buffer.vi ‏52 KB

    wow .. too many questions in one post
    you need to learn BASIC ABAP to be precise..
    1. Internal table are like array of structure...(hope u can relate it to other coding language now) and workarea is just one structure.. so a work area is used to hold single line of a internal table of same format..
    2. text - integer.. integer to text... => YES.. just define a variable as type STRING. and do what ever operation you want on this.
    3. use ROUND for rounding off..
    4. Please NO MORE TOO MANY QUESTION on ONE thread... Thanks
    -Somu.

  • Driver visa session looping issue

    Hi,
    This is the first time I've ever setup a loop with a visa driver session in it. I am attempting to collect an array of flux data. The array indexes correspond to physical (x,y)  locations on the item i'm measureing the flux of. This is why i need the data to be read in as a 5x7 array of flux. The 5x 7 matrix is collected 3 times (at 3 different z heights). However, I can't figure out why I am getting three 6x7 matrices instead of a 5x7. I can't tell if it is my shift register setup, where my visa session is implemented, etc. I just can't figure it out.
    Could someone look at my attached code ( it's pretty simple, just 3 nested for loops with the visa session inside, and that data is fed out to 3 matrices, which is then stripped down further which can be ignored) 
    Thank you very much,
    chelsea
    Solved!
    Go to Solution.
    Attachments:
    visa loop test1.vi ‏31 KB

    First of all, I don't see a point in any of those sequence structures.  All they are doing is creating clutter.
    Secondly, why are you appending the second to last results to the whole results?  That is where your extra index is coming from.  Get rid of that shift register and the Build Array.
    There are only two ways to tell somebody thanks: Kudos and Marked Solutions
    Unofficial Forum Rules and Guidelines

Maybe you are looking for

  • How to show a compressed image in the report?

    Hi I have created a new report.what i do in application is i  compress the image and save it in database.now i need to retrieve the compressed image and display in the report. I have used the following code to decompress the binary data save in the i

  • Music gone from iphone after ios6 update

    Ever since the update all my music is gone from my iPhone 3Gs, and I can't seem to get it back.

  • Share Point 2013 - User Settings not initialised

    Hi there,  We are using the SharePoint 2013 server on top of Windows 2012. We've recently created  user profile application service and activated the Site Feed feature.  After that one of our users experiences an annoying thing, she is being redirect

  • Switching from PSA 2 to PSA 3 Starter Edition

    I previously used PSA 2 on my older Windows XP computer. I recently purchased a new computer with Windows Vista. I am using PSA 3 Starter Edition because PSA 2 would always have to shut down about a minuted after opening it ( a compatibility problem

  • Email program giving Dump

    I'm using the function module SO_DOCUMENT_SEND_API1 and after that I'm using this command.   SUBMIT rsconn01 WITH mode = 'INT'   WITH output = 'X'   AND RETURN. But it is giving Dump. Says Error when calling a remote function call. RFC_GET4 I'm getti