Is it possible to install a usb-6009 in Dasylab?

I need to install a usb-6009 in Dasylab, but Dasylab does not recogise daqmx-devices, or so it seems.
In MAX, I can see that I have traditional NI DAQ installed (7.4.2f3), but the usb devices are listed as daqmx devices.
Is there any way to trick Dasylab into seeing my device?
I usually work in Labview, but now I need to replicate a measurement from  a colleague from Ethiopia and Dasylab is confusing in the beginning.
Would it be easier to use a pci card (we have pci 6220 available)?
max version4.3.0f0
daqmx 8.6.1f0
traditional NI daq 7.4.2f3
dasylab 8.00.00 from 2004, old, I know)
labview 8.2
Solved!
Go to Solution.

I want to apologize in advance for incomplete technical competence
in the subject, because my amateur level of training.
I also want to apologize to experienced professionals
and especially Mr. CJ_Butler,
for my English machine translation.
I hope you will indulgent  to the Russian citizen,
who is actively interested in operation
of the device NI USB-9006 program DASYLab.
In Russia, unfortunately, very few know
about object-oriented programming in the program DASYLab
(compared with a developed system of training LabVIEV).
Given the popularity of the brand as the NI USB- 6009 ,
I would like to understand in detail the stages of installation,
work in the program DASYLab. ( I use DASYLab 11 , in Windows XP)
My first question is as follows :
why I can not directly see and work with the NI USB- 6009 program DASYLab,
as it comes with a sound card or other devices e.g. , Measurement Computing?
When you open the tab «Hardware Setup» error appears: «No hardware driver loaded»?
Can we do without NI MAX ( task creation and other unnecessary actions (!),
and produce all presets in the program DASYLab (Hardware Settings)?
Indeed, in the program directory has the files as : nddi.dll, niezio.dll?
Why they can not determine the direct control of NI USB- 6009 in DASYLab?
2) Is there a " prime-driver » DASYLab for NI USB- 6009 in DASYLab 11 ? ? ?
Or how else can modify DASYLab driver  (DASYLab NI-DAQmx (?))  for NI USB- 6009 ?
http://dasylab.com/content/driver.php?action=filelist&sid=33 (?)
http://dasylab.com/content/driver.php?action=filelist&sid=575 (?)
Thank You in advance for the opinions expressed.
Igor Viktorovich, from Russia
https://fbcdn-sphotos-c-a.akamaihd.net/hphotos-ak-prn2/t1.0-9/10302230_395793633896399_2521902383856967762_n.jpg

Similar Messages

  • Is it possible to install a USB 2.0 card in an old eMac?

    Hi,
    This is probably an embarrassingly ignorant question, but I'd be grateful to know whether I can install a USB 2.0 card in my old eMac 1GHz PowerPC G4 (1GB RAM) running 10.5.6, which only has the old USB 1 ports buitl in. Basically, I'd like to be able to sync my iPhone and attach other newer peripheral devices, such as external hard drives with USB 2.0 connections.
    Thanks!

    This is probably an embarrassingly ignorant question, but I'd be grateful to know whether I can install a USB 2.0 card in my old eMac 1GHz PowerPC G4 (1GB RAM) running 10.5.6, which only has the old USB 1 ports buitl in.
    You can't.
    (40568)

  • How to install drivers for Ni USb 6009

    Hi
    I am trying to use an NI-USB 6009 and was wondering if its possible to install just the drivers for this hardware. I know that I can install the NI-Daq base but I would really like to install the minimum amount of software onto my computer that has linux. However, I did try to install the NI daq base and the installation is failing because of an input/output error, (I am not really sure why).
    I would really like to be able to just install the drivers for the USB-6009. I was just wondering if that was possible and where I could find them.
    Thanks

    Hi Cannonball,
    You'll know pretty quickly after installing DAQmx Base and running the lsdaq utility as to if you need different firmware. The 6008 and 6009 must have a different firmware loaded to be able to run DAQmx Base 3.x. This is not a problem though as there is a firmware updating utility that comes with DAQmx Base and it's fairly straigh forward. For more information, see the link below.
    KB 491FPHB9: Firmware Incompatibility With Different Versions of DAQmx Base.
    http://digital.ni.com/public.nsf/allkb/91CBCFE9E171C845862572CF0077D8F9?OpenDocument
    PBear
    NI RF

  • Is it possible to install 3.0 USB port on HP 2000 412NR Notebook?

    Is it possible to install 3.0 USB port on HP 2000 412NR Notebook? Is there a usb board upgrade available?

    Hi,
    The following one has 2 USB 3.0 ports:
       http://www.pcmag.com/article2/0,2817,2422860,00.as​p
    Regards.
    BH
    **Click the KUDOS thumb up on the left to say 'Thanks'**
    Make it easier for other people to find solutions by marking a Reply 'Accept as Solution' if it solves your problem.

  • Is it possible to install Oracle linux from usb?

    Hi Guys,
    Is it possible to install Oracle linux from usb? If yes, may I know the exact procedures to do so.
    I am using linux ubuntu and I tried to make a bootable usb with the dvd iso of oracle linux 6 using Unetbootin software. The installation started well.
    but after formatting the harddisk, I got an error as shown below
    Missing ISO 9660 image
    The installer has tried to mount image #1, but cannot find it on the hard drive.
    I think ISO 9660 image is something connected with the CD image or something. But not sure...
    Can anyone help me out with this situation.... Thank you..
    Edited by: 839414 on Feb 24, 2011 12:47 AM

    HI
    I know this is an old post but has anyone resolved this issue? I am trying to install OEL 6U4 via usb with no luck. Keep getting Missing ISO 9660 Image. I have tried all the tools under the sun to write it to USB, also tried copying the iso to the Boot disk.. as stated above... no luck ....
    I have checked the installation files using md5 and they are fine... just in case downloaded it again as well as 6U3 and 6U0.
    I have installed almost every version of linux there is... but OEL is definitely the most difficult and after 3 days, I am ready to give up. Any idea?
    BTW. I am using utilities on windows 8 to write the iso to USB.
    Thanks
    P

  • USB 6009 and DAQmx Base Concerns

    Hello,
    I just received a USB-6009 multifunction DAQ device along with DAQmx Base for use on Windows XP with LabWindows/CVI ver 7.1.  Both the 6009 and DAQmxBase are successfully installed however I have several questions, I hope I am missing something obvious:
    1. The Task Configuration Utility is not needed for use with LabWindows/CVI?  I believe I read this in the Getting Started Guide.
    2. There is no function panel for the DAQmxBase functions?  I even searched the installation CD and found no .fp files.  So the only way to know what functions are available with DAQmxBase is to open the NIDAQmxBase.h file and search through the function prototypes?  Then, to know what a function does it is necessary to open the C Function Reference Manual via the Windows Start button and search for the function name of interest?
    3. When creating a digital input or output channel with DAQmxBaseCreateDIChan or DOChan is it only possible to setup a "channel" as an entire port?  Is it not possible to setup a channel as a single input or output line?  Similarly, once the channel is configured is it only possible to read or write an entire port with DAQmxBaseReadDigitalU8 or WriteDigitalU8?  I was expecting something similar to the DIG_out_line function used in Traditional NI-DAQ where a board number, port number, line number and state are passed so that a signal output line can be set on or off.
    Thanks in advance for the assistance,
    - Aaron

    Aaron T. wrote:
    Hello,   I just received a USB-6009 multifunction DAQ device along with DAQmx Base for use on Windows XP with LabWindows/CVI ver 7.1.  Both the 6009 and DAQmxBase are successfully installed however I have several questions, I hope I am missing something obvious:   1. The Task Configuration Utility is not needed for use with LabWindows/CVI?  I believe I read this in the Getting Started Guide.  
    **The Task Configuration Utility is not necessary, except that there is significant functionality missing from DAQmx Base but present in the utility. That includes a lot of what DAQmx calls "proerties", and includes things like the convert rate for AI scanning.
    2. There is no function panel for the DAQmxBase functions?  I even searched the installation CD and found no .fp files.  So the only way to know what functions are available with DAQmxBase is to open the NIDAQmxBase.h file and search through the function prototypes?  Then, to know what a function does it is necessary to open the C Function Reference Manual via the Windows Start button and search for the function name of interest?  
    **I don't use CVI so I don't know anything about this. It strikes me that even if you have such a panel reading the documentation is a good idea
    3. When creating a digital input or output channel with DAQmxBaseCreateDIChan or DOChan is it only possible to setup a "channel" as an entire port?  Is it not possible to setup a channel as a single input or output line?  Similarly, once the channel is configured is it only possible to read or write an entire port with DAQmxBaseReadDigitalU8 or WriteDigitalU8?  I was expecting something similar to the DIG_out_line function used in Traditional NI-DAQ where a board number, port number, line number and state are passed so that a signal output line can be set on or off.  
    **Well, the ReadMe file says that USB-6008/9 is supported with both Line I/O and Port I/O. On the other hand, the documentation for DAQmxBaseCreateDOChan and DAQmxBaseCreateDIChan say for the lineGrouping parameter, "Always pass DAQmx_Val_ChanForAllLines." So your question is a good one! Perhaps one of the NI engineers can comment on this.
    - Aaron
    Thanks in advance for the assistance,
    John Weeks
    WaveMetrics, Inc.
    Phone (503) 620-3001
    Fax (503) 620-6754
    www.wavemetrics.com

  • Usb 6009 multiple analog inputs

    I am currently attempting to sample two different analog inputs at different sampling rates using a USB 6009.  I keep getting the 'resource reserved' error and am wondering if this is not possible using this DAQ.  Questions:
    1.  Does creating two analog input channels on the device cause this error?
    2.  Is it possible to sample at different rates on channels created in the same task?  (i am trying to 'slow down' the second analog input to display switch points to a customer)
    3.  Running multiple analog inputs using independent timing would be better achieved by switching to a higher end DAQ?  If so which would you recommend?
    I have attached my vi.  Thank you in advance for your help. 
    I surf therefore I am....
    Attachments:
    demo_nolvl.vi ‏27 KB

    The DAQ boards only have 1 timing clock for the analog inputs/outputs.  So you can only have 1 sample rate on a given card.  I would recommend just sampling at the highest of the desired rates in a single task.
    There are only two ways to tell somebody thanks: Kudos and Marked Solutions
    Unofficial Forum Rules and Guidelines

  • USB 6009 DAQmx Deployment

    Hi.
       I have created a 4 channel differential data logging scope with a NI USB-6009. I would like to deploy this 'easily' to many of my colegues. I have app builder and the ability to build apps, but I would like to include the correct drivers for the 6009 in with the installer (although I have selected NI_DAQmx Core Runtime and NI_DAQmx MAX config support).
       For the running of the app, I have created a Task that consists of 4 channels. I would like it for my end user not to have to configure this. Does anyone know how I might acheive this? My experience of deployment is limited.
    Many thanks, Alec

    Hi Alec,
    It's also worth installing DAQmx MAX Configuration Support from the install builder if you still want to be able to modify the tasks from MAX as well.
    Best Regards,
    Chris
    National Instruments - Tech Support

  • USB 6009 thermocouple noise

    I'm using NI-DAQmx 7.5, Labview 7.0, and a USB-6009.
    I'm trying to get a more accurate reading from my j-type thermocouple.  It is now distorted with noise and jumps rapidly +/- 20 degrees around room temp without any heat being applied.  When heat is applied, it climbs as it should but with the same variation.  I haven't moved to the labview side of it yet, I'm still trying to get a usable reading in MAX.  I have it wired as a differential input with a floating signal source.  I've installed 10k ohm resistors for bias current return as instructed.  Installing the resistors helped but still leaves me with the readings shown in the attached pic.  I've also tried grounding my USB 6009 with no results.  Please help. john
    Attachments:
    thermocouple reading.jpg ‏41 KB

    Hi John,
    What you are running into is a problem with the actual voltage resolution of your USB-6009.  This device has 14-bits of resolution over a minimum range of +/- 1V.  That means that the voltage resolution is (1 - (-1))V / (2^14) = 2V / (16384) = 122uV.  While this is an excellent resolution for such a low cost USB device, you will still notice problems when trying to make a thermocouple measurement.  The reason for that is that Thermocouples make a very small output voltage (see image below).
    Common thermocouple types have the following response:
    K type: 41 microVolts per degree Celsius
    J type: 56 microVolts per degree Celsius
    T type: 52 microVolts per degree Celsius
    There are a few options for ways to get around this limitation.
    Good: You could read the signal in as a voltage and then manually convert that voltage into a temperature.  This would be the least accurate method, but it should allow you to average out your signal and then use the Convert Thermocouple Reading.vi to show the voltage as a temperature.
    Better: If you want to keep using the USB-6009 to measure thermocouple, what you really need to do is first amplify your signal before you connect it with your device.  This should increase your resolution of the thermocouple allowing you to get more accurate temperature specs.  This will require some more complex programming to properly scale your measurements, but it will give you more accurate readings for relatively little cost.
    Best: Opt for getting a USB device that is designed for taking temperature measurements such as the USB-9211A.  This device would give you a voltage resolution of 160mV/2^24 = 9.54 nV!  That's over 12500x the resolution of the USB-6009 that you currently have.
    Hopefully the above information helps you better understand what is going on and why.  If you have any other questions feel free to post them here.
    Message Edited by Otis [DE] on 01-13-2006 11:37 AM
    Message Edited by Otis [DE] on 01-13-2006 11:38 AM
    Otis
    Training and Certification
    Product Support Engineer
    National Instruments
    Attachments:
    761029-Thermocouple_Voltages.GIF ‏10 KB

  • USB-6009 Digital Output Current

    You need to have all those grounds connected together - USB-6009 Digital Ground, ULN2003 GND, and the power supply ground. In this case it will not matter how they are connected.  If they are not connected, then one of the devices could be damaged.
    Lynn

    Hi, I'm trying to increase the voltage of the digital outputs of USB-6009 device. I read a couple topics about using a relay but I could not manage it.
    I was thinking to use 5V supply on the card since it has 200mA current on it, but when I use it with open-collector output, it can not switch the relay. When I measure the current between
    5V and ground: 200mA,
    5V and output, I read a value around 30-40 mA.
    Why can't I use that 200mA with an output? This is the relay I'm using.
    If this is not possible, can I use an external 5V supply (with more current) and a digital output to switch the relay without damage the 6009?

  • USB 6009 Up/Down Counters

    Hello to everybody,
    I'm going to buy a USB 6009 for a low cost application and it fits my requirements except the counter.
    In fact, I would need an Up/Down counter while USB 6009 has an event counter type only.
    Do you know whether is possible to create the count direction by using one of the DIO bit?
    I have been thinking to design an external circuit for creating the right signal but I have no idea for a while.
    Please, let me know any suggestion.
    Thanks in advance.
    Regards 

    Hi,
    These device can be used only for simple event counting using PFI 0 as an event counter input. The counter gate is not accessible on the USB 6008/6009 devices.
    No Quadrature Encoder Capability is available: USB-6009 has counter. However, in DAQmx Base, externally controlled count direction is not supported. Notice that for a CI input task in the NI-DAQmx Base configuration utility, you can only specify count up. You cannot select "externally controlled." or "Count Down".
    I hope it helps,
    Marco

  • Usb - 6009 Device Recongition Problems

    Hi,
    I just got a USB-6009 DAQ. I have labview 7.0, drivers software and the supplied software installed with no problems. In the measurment nad automation explorer (MAX) the device is recongnised and declared problem free. I tried to test the device in labview using help tutorials which suggests the DAQ assitant block. After selecting Analogue input and then voltage I am told that no device is detected i.e. Where I should be selecting my terminal(s) I am not giving any options.
    If anyone has had the same problems or know how to fix them I would be thankful of your help.
    Kyle

    Hi Kyle,
    The USB-6009 device works on DAQmx base. This is a restricted version of NI-DAQmx which is used on some of our products. Unfortunetly it is incompatible with DAQmx and the DAQ Assistant uses DAQmx. Therefore you can't use the UDB-6009 with the DAQ Assistant.
    To use the device, load up the USB-600x Interactive Control which should be located in the following directory.
    C:\Program Files\National Instruments\LabVIEW 7.0\examples\daqmxbase
    Take a look at this and figure out how the code works. If you want to configure it to do your own specific functions then I would suggest that you use this program as a base and customise it for your purposes.
    Best regards,
    Peter H.
    Applications Engineer.
    NIUK.

  • USB-6009 with Labview 2014 on Windows 8.1

    Hi,
    I have trouble connecting USB-6009 OEM with my new laptop with windows 8.1. My USB-6009 OEM device is not shown up in NI Max. So when I check Device Manager, there is a question mark on "NI USB-6009 OEM FW". I tried to install Labview 2011 but still the same. When I tried to choose the driver in Device Manager manually, I can only find NI-VISA driver as the closest one. It is still not working when I open DAQmx to check my card. It is still not detected in NI Max and anywhere within labview. When I used NI-VISA driver, and try to connect the device in NI Max, I get timeout error. I think I need to see NI-USB driver instead of NI-VISA driver. Any help would be appreciated.
    Cheers,

    Hi Thoo,
    You may also want to try reinstalling the DAQmx 14.5 driver first. DAQmx 14.5 is supported on Windows 8.1. Some of the driver files may have been corrupted and reinstalling the driver may solve the issue. If not, using the DAQmx Base driver could be a good option as Dennis suggested.
    Thanks,
    Frank
    Application Engineer
    National Instruments

  • USB-6009 AI and AO

    Hello,
    I have an USB-6009 DAQmx device. I would like to measure analogue input with the possible highest sampling rate (48kHz) . At the same time, I need to ramp a voltage analogue output from 0 to 5 Volts, lets say during 2 minutes. These two tasks do not require close syncing. Regarding to the spec, the highest rate of the AO is 150 Hz. It is ok for my application.
    So the AI measurement will use device internal clock, and the AO can only work with software timing. Is it possible to independently ramping the voltage output and at the same time, reading the analogue input? If not, is there a workaround for this device?
    Thanks for advice!
    Regards,
    Solved!
    Go to Solution.

    I have made a test version (the final solution will do measurement for 2 minutes, so I will need to ramp up the analogue signal from 0 to 5 Volts in 2 minutes, and I think 10 msec will give a sufficiently smooth ramp curve --> 24000 steps over the 5V range).
    In this test VI I do measurement on two analogue inputs using internal timing, and in parallel, I do the ramp signal with software timing (I guess it is not advised to go below 10 msec because Windows can slow down too much?).
    Additional question: when I use DAQ on the two input channels, I understand that, the available highest rate is half of the maximum. But when I plot the two channels, the points are at the same time on the XY graph. How does  LabView interleave data from the two channels during measurement?
    Thanks for help and advice,
    Regards,
    Attachments:
    Test1.vi ‏213 KB

  • USB 6009 Signals Readings

    Hello,
    I'm using USB 6009 to read signals from PMT and other equipment for experiments and I have a couple of questions:
    1. Is it possible to read a signal with out DAQ assist? I've looked around the forums and haven't seen any relevant material.
    2. This is probably true, but I haven't checked it yet. Is it possible to read two signals in the same loop (eg, AI0 & AI2 in a while loop)?
    I'd appreciate if you'll point me to the right directions.
    Thank you,
    A
    P.S
    I should probably mention that I want to read the signals with out using DAQ assist because of the  assumption that DAQ assist  is slower than a direct measurement.
    Solved!
    Go to Solution.

    Hi Ravens Fan,
    Thank you for the answers and link. This is what I'm trying to do, and I've read the post on this link (the error code page): http://digital.ni.com/public.nsf/allkb/485201B647950BF886257537006CEB89 I think my error is mostly case #4 on the page (mostly because I dont use DAQ assist).
    But, I can't figure out how to terminate/clear channel to release the device for the next one. Any ideas?
    Thank you.
    Attachments:
    error.JPG ‏55 KB

Maybe you are looking for

  • How to create internal table in Adobe form - SFP transaction?

    How shd i create an internal table under Global data ??? Regards, Gurmukh singh

  • 40gb Ipod says "do not disconnect" all the time... please help!

    My 40gb ipod just says do not disconnect all the time when it is connected to my pc. itunes says the update is complete and my computer is recognising the ipod. please help!!!

  • Direct payer 'Incoming payment method E'

    Dear Friends, 1) What is Direct payer..? 2)In transaction FD03, for a  Customer under company code tab, payment transactions what does 'Incoming payment method E' mean..?.        Can you kindly explain me ..? Thanks in advance.

  • Udf for context change

    Hi All,             I write a UDF to change context but it is not working properly             can any please guide me. code is ,   for (int i = 0 ; i<Plant.length ; i ++) int b = 0;       for (int j = 0  ; j<Plant.length ; j++)               if (i!=

  • Technical Usages for EhP4 upgrade for CP/WSD BP activation

    Dear All, I have installed EhP4 ready system. Now in the process of upgrading it to EhP4 system. This system is mainly for activation V1.604 Best Practices of Consumer Products/Wholesale Distribution. Please let me know which technical usages I can u