Mac G4 crashes using continuous generation with PCI-1200 DAQ card

I've been trying to use a PCI-1200 DAQ card with a Mac G4 to create a continuous analog output signal. The problem is that the computer locks up after a while (a few minutes). I have been using the Generate Continuous Sine Wave vi, but the problem remains.
I've tried slow signals (10Hz) and fast signals (2kHz) but the computer still seems to crash. I have no clue what is going on!

bissdp,
The first thing I would recommend is making sure you have the latest version of the driver:
www.ni.com > Support > Drivers and Updates > Current Versions > Multifunction DAQ
Another you could try is to determine what component is causing the computer to lock up. Use a process explorer utility to try to see what processes are locked up. A search in google might turn up utilities like this one: (http://www.clarkwoodsoftware.com/peekaboo/)
Keep in mind heat or speed issues. It may sound funny, but you may want to try opening the computer case and making sure the board has enough air flow to keep it cool.
These are just some ideas for now, I hope they help.
Alejandro Asenjo
Applications Engineer
National Instruments

Similar Messages

  • How to avoid dependance between output digital port with a PCI 1200 daq card?

    Hello,
    I don't manage to fix the state of a line on one port (PA0 for ex) without change the state of another line of another digital port (PB0 for ex)
    I use a PCI 1200 Daq card with Labview 5.0.1 (under Windows 98)
    How to configure a robust state (high logic level) on a line of a port??
    Does anybody have an idea on this problem?
    Thanks a lot in advance and happy new year to the community
    JF LAVEZARD

    jflab a écrit:
    Hello,
    I don't manage to fix the state of a line on one port (PA0 for ex) without change the state of another line of another digital port (PB0 for ex)
    I use a PCI 1200 Daq card with Labview 5.0.1 (under Windows 98)
    How to configure a robust state (high logic level) on a line of a port??
    Does anybody have an idea on this problem?
    Thanks a lot in advance and happy new year to the community
    JF LAVEZARD
    Hello,
    I've found the problem
    The PCI1200 has a chip 8255 for Digital I/O
    If I use the bad VI, it reconfigures the port and all the lines go in a low logic state
    so I'd to write only on the port or the line without reconfigure the port
    So the good VI in the palett is: Port Write and not Digital Port write!
    See attached file..
    Thanks
    Best Regards
    Attachments:
    port_dio.jpg ‏771 KB

  • I am trying to sample 2khz signal at 8khz,with an accuracy upto 6place of decimal ,using pci 1200 DAQ card.

    how can i externally trigger this operation.i need to sample my signal at 4 points and store it in a excel sheet and the values should be same,as the the 2khz signal and the 8 khz signal both are comming from the same source.

    could u tell me what more details would be required....i will be more than glad to let u know about it.

  • Using a counter on PCI-6023E DAQ card

    Hi there,
    I'm about to use a 6023 card to measure a periodic pulse signal. I need to count the number of the pulses. But...I've heard that the counters on my card are unusable in that case. Is that correct? Do I have to chage my card? Or can I do it with 6023E?
    Thanks in advance;
    Radek

    Hi DAC,
    The counter on your 6023E board is the DAQ-STC counter/timer. This counter fully supports counting the number of pulses. Depending on your application development environment, if you search the shipping examples you will find examples that perform this function.
    This is so long as your digital pulses are 0-5 V or TTL compatible.
    Other applications supported by your counter are:
    Simple event counting
    Gated-event counting
    Single-period measurement
    Single pulse-width measurement
    Single pulse generation
    Single triggered pulse generation
    Retriggerable single pulse generation
    Continuous pulse-train generation
    Buffered event counting
    Buffered period measurement
    Buffered semi-period measurement
    Buffered pulse-width measurement
    Fre
    quency measurement
    Buffered frequency measurement
    Finite pulse-train generation
    Quadrature encoders
    Hope that helps. Have a good day.
    Ron
    Application Engineering
    National Instruments

  • What is the maximum number of samples that can be acquired with a single trigger using PCI-6036E DAQ card?

    1. My PCI-6036E card has a FIFO of 1024 bytes in AI. Maximum how many samples can be acquired with single trigger for digitizing a waveform?
    2. PCI-6036E DAQ card has got a AO with 10ks/s update rate? Can I generate a pulsed output with the AO of this card? I need a pulse of, say, 5 V and 1 kHz frequency.

    Giri,
    1.The trigger is usually for enabling the acquisition. I mean, when it is logic zero, you acquire nothing; however, when it is logic one, you start acquiring your waveform,and you can acquire continuously without having to worry about how many samples, unless you want to set your program for doing so. You can acquire continuously by implementing buffer acquisition. Finally, the FIFO that is in the board will not affect the buffer acquisition.
    2.Yes, you can do that! You can generate your 5V signal with that specific frequency, and you should not have any problems, because the freq fits in the update rate.
    I hope this information is helpful.
    L Aguila
    Applications Engineer
    National Instruments

  • My old MacBook pro died, got a new Mac can I use my cs4 with my new computer?

    MMy old MacBook pro died, got a new Mac, can I use my cs4 with my new computer? Do I need to buy a new license?

    Using Time Machine, or any other such program, to transfer Adobe software from one machine to another works probably less than five percent of the time. Uninstall from your new machine. Run the cleaner tool (do a search for Cleaner Tool on this site if you don't know what it is). Then reinstall properly from original media or downloads.
    --OB

  • Noise in thermocouple measurement using PCI-6024E DAQ card

    I am using a PCI-6024E DAQ card for thermocouple(TC) measurement. The
    TC is installed in the mold cavity of an injection molding (IM)
    machine. The leads from the TC are connected to the DAQ card using a
    CB-68LP board. I made a  temperature mesurment task in LabView 7.0
    with the two leads from TC as differential inputs and used CJC value of
    25 deg C.
    The problem is that when the IM machine is OFF, I get a reasonable temp
    reading with some slight noise but as soon as I turn ON the IM machine,
    the temp readings go wild with a lot of variations around the mean
    value. The temp values also don't go to higher values when the TC is in
    contact with the molten plastic in the IM machine.
    Do I need to ground something here or put a resistor in the I/O board?
    BTW the IM machine is grounded and the TC give correct temp values when
    connected to a hand-held temp reader.
    Any help will be highly appreciated.

    Hello  Aftab,
    Looks like you need to use bias resistors to reduce the noise. Check out the following tutorial for more information about how to use bias resistors.

  • Implementing a Lock In Amplifier using a PCI-6034E DAQ card

    I am looking for a lock-in amplifier and have read about the Lock-in Amplifier kit provided by NI. I am wonder in is it possible to implement a solution with a PCI-6034E DAQ card instead of the 4472? And if so what would be the limitions of using the 6034 for this application?

    The requirements for the Lock In Amplifier are contained within the document linked bellow. The S series can be used with the Lock In Amplifier but not the E-series.
    http://ae.natinst.com/operations/ae/public.nsf/fca7838c4500dc10862567a100753500/ebaadbd271adfec586256c86007b63f9?OpenDocument
    Regards
    Jon Bowers
    Applications Engineer
    NIUK

  • 32 bit I/O using PCI-6533 DAQ card

    I am thinking of purchasing the PCI-6533 DAQ card and need to know if the the 32 bits can be configured individually as input or output. I know that they can for unstrobed and static applications but I need to be using strobed or non-static. I also know that the you can set the 8 bit ports as input or output ports for highspeed. So my question is... Can I configure each individual line as input or output for highspeed applications?

    Hi,
    Unfortunately, the 653x boards can only be configured on a line-by-line basis for asynchronous digital I/O.
    When in synchronous pattern acquisition/generation mode, the 653x is divided into 2 groups (2 ports in each group). These two groups can be operated independently as input or output and at different clock frequencies. However, all lines on both ports of a group are configured in the same way.
    If you are looking for completely individually controlled lines in a pattern generation/acquisition application, you might consider the new high-speed digital board (NI 655x).
    Hope that helps. Have a good day.
    Ron

  • How to switch multiple relays using PCI 6601 DAQ card

    What is the specification of the digital output value given from the PCI 6601 DAQ card?How to develop LAB VIEW code to switch a relay?Can anybody please explain?

    Hi hasi,
    did you read the manual and the specs of the PCI6601? It says: "5V TTL/CMOS compatible"...
    You shouldn't expect to much current from a simple DO pin - probably it's not enough to drive your relay!
    Best regards,
    GerdW
    CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
    Kudos are welcome

  • Im using a pci-6221 daq-card.I want to know the difference(use)of a line output and a port output.

    Im using a pci-6221 daq-card.I want to know the difference(use)of a line output and a port output.

    Hello Hellraiser24,
    This question has actually been discussed in another thread here:
    http://forums.ni.com/ni/board/message?board.id=170&message.id=95895&requireLogin=False
    I hope this helps,
    E.Lee
    Eric
    DE For Life!

  • I need to interface SCC-TC02 with help of SC 2345 and PCI-6024E DAQ Card.please tell me the procedure and if possible send me the program

    Dear Friends,
               I need to interface SCC-TC02 with help of SC 2345 and PCI-6024E DAQ Card.But i can not understand example program.please send me the procedure to interface these three componnets and if possible send me the program.
    With Regards,
    Eswaramoorthy K V

    nce of nth triggering pulse. i need to know when the when the ist trigger occurs and when the nth trigger occurs . please tell me how to to . what i know is that event triggering has to be done with start and stop trigger. please tell me how it has to doneSuresh;
    What you will need to do is to set up a digital start and stop triggered Analog Input acquisition. Then you will need to have a counter set as event counter, having the specific number of pulses you need configured as the preset value, configured to count down, and generating a pulse after the terminal count has been reached. That counter output will be the stop trigger of the Analog Input operation. In summary, you will have the external pulse being both the digital trigger of the Analog Input operation and the source pin of the counter, and the counter output being the digital stop signal for the analog input.
    I'm attaching a Labview VI that does the start and stop analog input acquisition. You will need to include the counter part and set the stop s
    ignal to be the counter output.
    Hope this helps.
    Filipe
    Attachments:
    AI_Start-Stop_D-Trig.zip ‏25 KB

  • How do i generate the TTL pulse from PCI 6024E DAQ card

    Hi, I'm trying to write a program which generates a single TTL pulse when it is run. My goal is to generate the pulse, send it to the output channel of my PCI 6024E DAQ card, and then use a BNC cable, to display the signal on the oscilloscope. I have written the code and I think it appears fine but when I try to view the signal on the scope I am unable to see the pulse. I have tried slowing down the sweep on the scope and have adjusted the voltage and time scales but have not seen any response. After examing my DAQ card, I noticed that none of the PFI/TRIG terminals on the card are wired; the output channel connects to terminals for DAC0OUT and Ground. I'm wondering if this pin needs to be wired in orde
    r to generate the pulse since PFI0/TRIG1 is the start trigger terminal. Any help would be appreciated. Thanks.
    Attachments:
    ttl.vi ‏26 KB

    Hi,
    You are looking for a retriggerable pulse generation. Please check the Generate Retriggerable Pulse (DAQ-STC).vi example that ships with LabVIEW.
    Regards,
    RamziH.

  • How can I connect the output of a force transducer unit(BNC cable) to the PCI-6036E DAQ card via the SCB-68 block connector?

    I m using the 404A force transducer unit by Aurora Scientific and i want to connect the (converted to Volt) output, which is a BNC cable, to the PCI-6036E DAQ card via the SCB-68 connector block, facing difficulties in choosing the correct input mode for my signal(couldn t clear this using the manual), setting the switches and which channel shall i use in the block connector in order to configure ot with LABVIEW

    Hi MSc Student,
    Connecting your transducer to the 6036E card via the SCB-68 connector block should be pretty straight forward. You have 2 choices of input mode, Referenced Single Ended (RSE) this is where you will connect the screen of the BNC cable to a ground connection for the DAQ card in the SCB-68 and the other to any analogue input channel . These are listed in the device pinouts in the help section of MAX, under the 68 pin devices section. The terminal block numbers correspond to the DAQ card's pin numbers. Another method, which avoids ground loop problems would be to use the Differential connection and connect the signal and ground connections of the BNC cable to a differential analogue input, these are listed in the 6036E manual and in th
    e MAX help section again. the disadvantage to using this method is 2 analogue input channels are used and you must use the lowest 8 analogue input channels for differential inputs.
    The switch settings that you need can be found in the manual for the SCB-68 in table 2-1, the manual can be found here : http://www.ni.com/pdf/manuals/320745b.pdf
    The switches are for power to extra signal conditioning and the cold junction compensation for thermocouples which in your case you do not need so set the switches to the positions in the first box ie Direct feedthrough, with temperature sensor disabled and accessory power disabled. The channel you use should not matter the only ting to remember will be that if you use differential mode measurement that you will be connecting the signals to AI(n) and AI(n+8) so differential input 0 is across the connectors for AI0 and AI(8). The channel numbers corespond to the analogue input number for the card.
    I hope this information helps,
    Chris M
    App
    lications Engineer, National Instruments

  • Low sample-rat​e measuremen​ts on the PCI-6115 DAQ card

    I need to measure an analog signal at a sampling rate of a few tens to hundreds of Hz in sync with the rising edge of an external clock. I have a PCI-6115 DAQ card w/ Labview 6.1 and NI-DAQ 6.9.2. The PCI-6115 is a high speed card and has a minimum sample rate of 10 KS/s. Is there any way of implementing a low sampling rate measurement using the PCI-6115 in sync with an external clock?
    Thanks in advance.

    Kuldeep,
    It is possible to do what you are describing above (in fact I don't think an external clock is required to do this), however, bear in mind that the reason for this minimum sampling frequency is due to the ADCs used on this high speed board. The ADCs used are pipelined ADCs, meaning that when a signal is digitized, it is digitized in distinct stages within the ADC (in the case of the 6115, I think there are 3 stages involved). Data is moved from one stage of the ADC to the next each time a sample clock pulse is recieved. If too much time elapses between these clock edges, the signal to be digitized can actually 'leak' off of the ADC. This can result in improper digitization, which can lead to less accurate measurements. So, while it is possible to mak
    e the device sample below it's minimum rate, it may be advisable to sample faster than the rates required by your application, and either average multiple data points per measurement, or throw away extra points taken.
    I hope this helps,
    Dan

Maybe you are looking for

  • Old 144G Apple TV - External Hard Drive

    My old Apple TV has a 144G hard drive and I want to know if I can use the USB connection in the back to connect to my iMac for an external hard drive? Would hate to waste the memory simply because I upgraded to the new Apple TV.

  • Payments made to vendor against the PO should appear in PO

    Hi Thanks for the reply is there is any possibliy so that i can see the vendor down payment and the other progreesive payments made to vendor. Also i want to link the payments made to vendor i.e what all payments is made to vendor aganist the PO shou

  • Profit center document is not created for invoice cancellation document.

    Hi all, when i create the invoice, Profit center document is getting created, where when we try to create a cancellation document for the same invoice, Profit center document is not getting created, can anybody let me know where can i check this prob

  • Moving a set number of frames

    I'm ashamed I have to ask this, but I can't find the code anywhere Instead of prevFrame(); which only goes back to the previous frame, I would like to go back THREE frames. This seems like it should be so easy, but I just can't find the right command

  • Distributed Multinode  Multi-OD Server Instances

    Trying to set up a HAA scaling configuration in network of replicated nodes (HP Keystone, HP-UX)in multimaster configuration with many Oracle Directory Server Instances on different front-end machines (HP Bl Server, Linux). The doc is in someway cont