Manual for SCC-A107 modules of SC02345 connector box

Hi, I’m a research student and using SC 2345 connecter box. I want to know that the amplification factor of that modules (SCC-A107). Please send me a manual for those e modules.

well easy enough to find
http://sine.ni.com/nips/cds/view/p/lang/en/nid/389​0
lets see a107 oh the seventh item
spec sheet page 9
http://www.ni.com/pdf/products/us/4daqsc253-265_19​4-196.pdf
scc-a107 is 50mv in and gain 200 so 10v out (50 x 200)
i bought a couple  they are two channel single ended
so one channel differential input (i got caught on this too)
hope this helps

Similar Messages

  • Need Test  Scripts & User Training Manuals for AP  &  PO modules

    Hi There,
    f you have any docs.. for AP and PO - test scripts/scenarios and training user manuals.
    Please send me the docs as soon as possible
    I will really appreciate your help

    Go to www.erpschools.com then on the left side menu click user guides and select 11i or R12. It will take you directly to the page where you can find all user guides.
    Thanks
    Prudhvi
    www.erpschools.com
    www.oracleschools.com

  • SAP installation manual for SAP 4.7 for abap module

    SAP installation manual for SAP 4.7 for abap module

    You can find the install guide in the SAP Marketplace:
    service.sap.com/instguides
    (logon is required)
    You have to navigate deeper depends on your platform and exact release.
    ABAP is not a module, it's a stack.
    Peter

  • Why is SCC FT MODULE USED?

    i am about to use SCC FT module for one of my applications?what is this device function and under which situations is this useful?

    Hello,
    The NI SCC-FT01 is a feedthrough module that offers direct connection to analog input or output, digital I/O and Counter channels of the E Series DAQ device. It also includes breadboard area for development of custom signal conditioning circuitry for analog input, analog output, digital I/O, and GPCTR channels of the E Series DAQ device.
    I recommend you to read the User Manual of the FT01. You'll find there a more detailed description.
    Regards,
    Luca
    Regards,
    Luca

  • Converting RTD voltage to temperature using SCC-TC module.

    I'm trying to convert the thermistor voltage from an SCC-TD module to deg C (using VB6). I am using the formulas supplied with the SCC-TC manual. The formula that I'm having problems with is Tk = 1/ [a+b(1nRt)+c(1nRt)^3]. I understand Rt to be the resistance of the thermistor but what is 1n?
    If all else fails is there an API call I can make in VB6 for this conversion. I have seen reference to one in C but not VB6.

    Hello,
    It is not 1n, it is ln i.e., natural logarithm.
    The NI-DAQ Users manual talks about the conversion functions.
    NI-DAQ Users Manual
    Sincerely,
    Sastry V
    Applications Engineer
    National Instruments

  • Search for a function module for deleting document originals (DMS)

    Hi,
    I'm searching for a function module to delete originals in documents (like manually with transaction cv02n). I'm only able to set an delete mark with the module 'BAPI_DOCUMENT_DELETE'.
    The module 'BAPI_DOCUMENT_CHANGE2' can only add new originals to an existing document. I need a possibility to delete all originals in an document without deleting the document itself.
    Has anyone an idea?
    Thanks.
    Jan-Christian Treusch

    Many thanks!
    That's especially the method I've searched for so long....
    best regards.
    Jan-Christian Treusch

  • Can I use SCC-DI01 modules with a PCI-6533 device?

    I am looking into performing buffered acquisition of various digital signals using a PCI-6533 device. Can I use SCC-DI01 modules with this device? If so, would this be via the SC-2345 Carrier? The manual seemed to suggest is was only compatible with E-series devices?
    Basically, I want to monitor 32 digital channels and will have a mixture of 5V and 24V signals.
    Many thanks,
    Douglas

    Hi Douglas
    Could you please clarify where the manual suggests that the module is only compatible with an E series board (is it because only an E series board is referenced in the manual)?
    I can't find any information stating which devices are/aren't supported by the SCC-DI01, but I will continue to investigate and keep you posted with my progress.
    Kind Regards
    Stuart R

  • Manual for building Pocket PC executable

    Hi,
    Where can I find the manual for building Pocket PC executables?    I am using LV 8.0 and is Build for PDA only for 7.0?
    Also, does any one have any suggestions on dumping on my executables onto an Arcom product?
    Thanks.
    hellolv

    You will need the LabVIEW PDA module for LabVIEW 8.0 in order to build PDA executables.  You will also need to license the runtime engine.  All of that information can be found here.  
    Hope that this helps,
    Bob Young
    Bob Young - Test Engineer - Lapsed Certified LabVIEW Developer
    DISTek Integration, Inc. - NI Alliance Member
    mailto:[email protected]

  • Suggest me way for calling Function Module in BSP page

    Dear Friends,
    I want to create sales order using BSP. For that there is an bapi called BAPI_SALES_CREATEFROMDAT2. I have two ways to do that.
    1. I will create all the required structure on the BW system and then i can call the bapi.
    2. I will create another function module which calls the BAPI_SALES_CREATEFROMDAT2. and i will pass all the require parameters to my created Function module.
    In the first case i suppose to create structure for header data, order items and partner data. It is quite lenghty work.
    In the second case i need not to create any kind of structure or types-pool.
    If answer is case1 then should i create exact structure or i can create that many fields contained structure which i require?
    Suggest me that whether should i follow first case or i should follow second case and also which one is more powerful?

    Thanks for giving me answer.
    I have tried to find through BAPI Browser for BAPI_SALESORDER_CREATEFROMDAT2 but unfortunately it is not creating the structure. This is very suprising for me because as per your weblog and as per your answer it should be generated. Now i have two choices :
    1. I create same structure on 6.20 system manually for that i have to create around 100 data elements or may be more than that which are not exits on 6.20 system.
    2. I will create another function module on 4.6c system which calls BAPI_SALESORDER_CREATEFROMDAT2 and i need to pass only selected parameters.
    2nd option is suitable for me but i have doubt that is the proper way for working production?

  • User manual for the PXI-6723

    I've been trying to find a pdf of the user manual for the PXI-6723. I need to know the pinouts of the two connectors. The data sheet I have shows no pinout.
    Arthur Galan
    Senior Technician - Space Science and Engineering
    Southwest Research Institute
    Div 15 / Bldg 178
    6220 Culebra Road
    San Antonio, Texas  78228
    Tel: (210) 522-5713 
    Fax: (210) 520-9935

    Thanks Alan,
    But maybe I'm missing something. The 6723 has two connectors on it. The pinout on the spec sheet shows only one. That's why I thought I was missing the pinout of the 2nd connector. I'll go back and take a look at that again.
     Best Regards,
    Arthur Galan,  CID
    Senior Technician - Space Science and Engineering
    Southwest Research Institute
    Div 15 / Bldg 178
    6220 Culebra Road
    San Antonio, Texas  78228
    Tel: (210) 522-5713 
    Fax: (210) 520-9935

  • Spanish Manual for SAP GRC AC 5.3

    Hello,
    I want to know if someone has an spanish manual for SAP GRC AC 5.3 for the 4 modules (RAR, SPM, CUP & ERM).
    Or if someone knows a link to get them. I need them.
    Thank you in advance.
    Best Regards.
    Pablo mortera.

    Hi Pablo,
    In 5.3 we can use the SPM by two different ways.
    1. FFID based: In this you make a user as a FFID and then you use this FFID to perform all the actions.
    User based roles are for the above.
    3. Role based: In this you assign the roles(maintained in SPM) to the user.
    Role based roles are for above.
    The basic diffeence is in first one you use user (FFID) to perform the activities whereas in second you use role (which are maintained in SPM).
    You can get more details of this in user guide and configuration guide.
    Regards,
    Shweta

  • Manual for MS-7091 Mainboard

    Have searched the MSI website but I cannot find any documentation about the MS-7091 mainboard.
    Does anybody know where I can find the MSI manual for this mainboard?

    don´t know, but according to this google result, it looks like 2 IDE and 4 SATA connectors.
    Anzahl und Typ der vorhandenen Steckplätze   
    3 x PCI,
    1 x PCI Express,
    4 x DDR-RAM,
    2 x EIDE (UDMA-100),
    4 x Serial ATA(S-ATA 150),
    1 x Diskettenlaufwerk
    http://www.medion.de/company/press/?cb_20040024.html

  • How can I use a different search path for FlexRIO IO modules?

    I want to use an IO module that is outside of the LabVIEW tree, but it seems like LabVIEW only searches in shared/flexrio/IO modules, and I can't figure out how to create symlinks in windows.
    Is there a variable I can set somewhere to include other search directories for the IO modules?
    //Olof
    Solved!
    Go to Solution.

    Hello Olaf,
    There are two locations on your Windows computer that are searched for FlexRIO adapter module support files.  The first is the Program Files directory you have indicated in your post; this directory holds the support files for NI manufactured adapter modules.  However, there is another directory that can be used for the location of any non-NI manufactured adapter modules.  For Windows XP this location is: C:\Documents and Settings\All Users\Documents\National Instruments\FlexRIO\IO Modules.  If you are attempting to install support for a non-NI manufactured adapter module, please ensure that you have installed all of the adapter module support files into this folder. 
    If you would like more information on this as it pertains to a custom adapter module development, please email the support email address listed in your Module Development Kit Manual.  This email address will give you direct access to the NI FlexRIO R&D team for support as indicated here.  If you want to discuss adding the ability to search other directories for adapter module support (other than the two listed above), please email the MDK support email address. 
    Regards,
    Browning G
    FlexRIO R&D

  • Need repair manual for CD-rom upgrade on Satellite Pro L20

    Hi
    Is there anywhere I can download a repair manual for the Satellite Pro L20 as I've just purchased a dvdrw drive & want to fit it myself
    TIA

    Hi
    Unfortunately, you will not find any repair manuals on the Toshiba website. Only the Toshiba service partners have such documents and have possibility to use notebook manuals.
    As far as I know the drive is secured with a screw which is placed at the bottom of the unit and under the keyboard.
    After removing of these two screws you can disconnect an optical drive from the connector on the system board.
    But without any experience you should remove any drives. I think you should ask the Toshiba service to help you because he has more experience in such cases

  • Scripting active configuration deployment location for an oracle module

    Hi,
    Does anyone know how to script the setting of the active configuration deployment location for an oracle module?
    I have executed the following which works fine but I can't figure out how to make it active so the module shows up in the control center to allow deployment.
    I can of course do it manually in the design center by right clicking on the module in the design center, selecting 'Configure' and then (strangely) just selecting the location property which then automatically sets the active location and then it shows up in the control center.
    OMBALTER ORACLE_MODULE 'MOD_EXT' ADD REFERENCE LOCATION 'MIG_EXT_LOCATION' SET AS DEFAULT
    Oracle module MOD_EXT altered.
    OMBALTER ORACLE_MODULE 'MOD_EXT' SET REFERENCE METADATA_LOCATION 'MIG_EXT_LOCATION'
    Oracle module MOD_EXT altered.
    Any ideas?
    Regards,
    Brad

    Hi Brad,
    before you set the location you have to set the correct configuration (in case you use more configurations than the default "DEFAULT_CONFIGURATION") as active configuration:
    OMBCAC 'MY_CONFIGURATION'
    Regards,
    Carsten.

Maybe you are looking for

  • OnDisconnect not working as expected on IE7

    I have an application for a poker game which works perfectly in Firefox 1.5 and 2.0, however in IE7 (not too sure about IE6 thou) there is a serious problem. I shall explain.. User logs in. Opens a popup of the game which is just a php script which l

  • Window 8 pro 64 bit wifi hotspot problem

    I am use windows 8 pro 64 bit . I make WiFi hotspot on my laptop ,it successfully started and i enable internet sharing then i connect my android phone to this network, it connect successfully but nothing open in my phone ,it didn't receive any data.

  • Initial Steps in Implementing SAP FICO Module

    Hi Peers I'm a team member in a fresh Implementation for a manufacturing company for FICO. I want to know what are the action steps to be taken before going to implement. Plz help me in how to go about the implementation of FICO module. Regards Rajar

  • Retrieving the destination location of output for a specific instance

    Hi, I have reports that run daily and create output PDFs in Enterprise location. I am now trying to access these output PDFs using a C# program using SDK. I see that when i query for PDF instances, the query does not give me SI_SCHEDULEINFO. How do I

  • Formatting hyperlinks with color background cells

    hello.. i need help figuring out how to fix this error i keep running into. I have a single cell in a table that i want to have a background and the cell has numerous hyperlinks. i created the background color in the single cell just fine, but now al