NI PXI 5610 procedure in Cal Executive

I am currently working with the calibration executive software version 3.3 to calibrate some PXI-5610 cards. For the AmplitudeAccuracy_Super10MHz it has me manually input two measurements. In turn it appears to perform a calculation to determine the As found dB reading on the calibration report. However, I cannot seem to understand why two measurements are made and why the calculations appear inconsistent. For the same test at frequencies from 20MHz to 2.7GHz with a power level of -50dBm I input similar numbers for both values on all frequencies but received vastly different values.
Here is a specific example:
1005MHz @ -50dBm 256average 1st value: -50.54dBm 2nd value: -65.17 As found reading on report: -.5837
1500MHz @ -50dBm 256average 1st value: -50.67dBm 2nd value: -65.12 As found reading on report: -.8231
Does anyone know what the calculations are for this procedure and why/what I might be doing incorrectly to make my measurements inconsistent?

Hi Jerimiah,
I am taking a look into your question and I wanted to clarify a few things. What hardware are you using to make the measurements? Is it the recommended hardware or an unsupported model? Could you provide a screenshot of a couple rows of the AmplitudeAccuracy_Super10MHz section of the calibration report if you have it? It would also be helpful to have a screenshot of the prompts to enter the two measurements.  I am not sure of the calculation that is made based on your inputs, however I suspect the difference may be related to the passband flatness. Let me know what other information you can provide and I will continue looking into an answer.
Steve B

Similar Messages

  • Calibration Executive error during PXI-5610 sequence.

    When running the PXI-5610 procedure in Calibration Executive, I get the prompt "NI-FGEN Create Arbitrary Waveform - Cannot write Real data when Data Processing Mode is Complex. Disable OSP or set Data Processing Mode to Real."  The procedure the errors out with the same message when I click 'Continue'.  I am using the NI PXI-5441 AWG (in place of the NI PXI-5421), but it is a selectable source during the initial configuration of the sequence.
    Am I missing something?  Am I doing something wrong?  Is there a hiccup in the Calibration Executive software when PXI-5441 is selected?
    Any help would be appreciated.

    Hi DaRock,
    We do not recommend making any changes to any of the test code for calibration as these may cause problems in your test sequence. We are still currently looking into the issue, however, in the meantime, if you would like to try and experiment some changes in the test code, please make sure to backup any original versions of the files you edit so you can load them back in they cause any problems.
    Just for your information, in a standard NI-FGEN application, you can manually disable OSP by accessing the OSP Enabled property from the NI-Fgen property node under Arbitrary Waveform Output»Onboard Signal Processing.
    Regards,
    S_Hong
    National Instruments
    Applications Engineer

  • Using Calibration Executive for PXI-5610 using PXI-5124

    Hi guys,
    We've been trying to calibrate PXI-5610 but the Calibration Executive software pop up an error of PXI-5122 requirement but we don't have the module, we have the PXI-5124 instead. We have tried to use the unsupport but it's not working too.
    We have found that their resolutions are different
    PXI 5124 http://sine.ni.com/nips/cds/view/p/lang/en/nid/14231
    PXI 5122 http://sine.ni.com/nips/cds/view/p/lang/en/nid/12615
    Is this the cause of error? Is it possible to use PXI-5124 instead of PXI-5122?
    FYI about our system
    Chassis PXI-1010
    Software: Calibration Executive 3.3
    For any further recommendations, please let us know, thanks in advance.

    Hi, somebody here?
    I'm not sure if we have specified forum for this issue, please help let me know.
    Thanks...

  • Fail to calibrate PXI-5610

    Sory if this isn't the correct Board to post this question but I couldn't find a Calibration Executive Board
    I'm trying to calibrate a PXI-5610 with the Calibration Executive 3.4.1 and I have the latest drivers instaled on my pc, the probem is that the calibration doesn't start and it gives this error:
    "Open VI Reference in cex_Instr_RFSigGen_Init.vi->_NI 5610_RFSigGen Setup.vi->_NI 5610_RFSigGen Setup.vi.ProxyCaller
    VI Path: C:\Program Files\National Instruments\Calibration Executive\InstrumentDrivers\LabVIEW\RFSigGen\rssma\cex_Instr_LabVIEW_RFSigGen_rssma_Init.vi
    LabVIEW:  The VI is not executable. Most likely the VI is broken or one of its subVIs cannot be located.  Open the VI in LabVIEW using File>>Open and verify that it is runnable."
    What can I do to fix this?
    ps: I had to perform the calibration with Calibration Executive 3.2.1 because the client had some urgency...
    Can anybody help
    Thank's in advance
    Bruno Ranhel

    Hi BRanhel, 
    Please take a look at this knowledge base. This is for Calibration Executive 3.3.1 but it could work for the 3.4.1
    Best Regards,
    Vincent.O
    National Instruments France
    #adMrkt{text-align: center;font-size:11px; font-weight: bold;} #adMrkt a {text-decoration: none;} #adMrkt a:hover{font-size: 9px;} #adMrkt a span{display: none;} #adMrkt a:hover span{display: block;}
    Été de LabVIEW 2014
    12 présentations en ligne, du 30 juin au 18 juillet

  • PXI-5610 RF Frequency Response adjustment

    Hello,
    I have noticed that adjustment for 5610 in RF frequency response is taking long time (about 2-3hr) with Cal exec 3.5, and I have tried to get the PXI chassis in a good enviroment and with some extra fans around since the adjustment has the legend "RF Freq. Response accumulated Meas. (Normalized to 45°C)
    I am wondering if the enviroment has some important point here or it needs some update in a particular application inside Cal Executive 3.5?
    I also have this question, Win7 has some advantage over WinXP in this particular or general adjustment?.
    Br,
    Omar

    Hello Omar_Rdz,
    Thanks for using NI forums! Calibration Executive usually takes a considerable amount of time for some of the PXI modules. By reviewing the Cal Exec manual it says that for the 5610 module it could take up to 240 mins (around 4 hrs). But definitely the temperature is a factor that can affect the performance not only of the calibration procedure but the whole system. Have you tried to execute the calibration when the controller and the chassis are cold? Also try to avoid dust accumulation in both the modules and the chassis because these can impact on the general performance of the system.
    Answering your question regarding the OS, Win7 has a better performance compared to WinXP but it will also depend on what type of controller you are using. In order to give you a better answer, could you please tell me what controller and chassis are you using? 
    Regards,

  • PXI 5610 problem

    When running the PXI-5610 adjustment procedure in Calibration Executive , the main power suddenly shut off . then what problem will happen to the PXI 5610?

    Thanks a lot for your reply.
    After the PXI controller 's suddenly power off when performing adjustment ,  we restart the OS and found the PXI 5610's error infomation in the MAX : Error -200437,External calibration constants are invalid,and he card cannot be used in our Auto-Test-System anymore. I guess the calibration data in the EEPROM was missed or uncompleted,but i have no relevant softwre tools to read or write PXI-5610's EEPROM.
    Could you give me some suggestions to solve this BIG problem? Thanks.

  • Cal executive

    We are trying to calibrate a PXIe 5622 for the first time using Cal Executive version 3.4.1. We are using an Agilent E4432B sig gen, Agilent E4418B power meter, E9301A power sensor, 11667B power splitter on the very first measuring test. We have a couple issues, but the main problem is after “characterization” of the power splitter as the procedure has you perform, the first reading it tries to take fails. The unit appears it does not know what it is reading. You manually enter the data on the power meter into the program. We assume it should all be in dbm, but the program does not tell you. We have tried Watts as well with the same results. The results, when it fails, says the reading you entered is “NA dbm” and it is looking for +/- 0.05dbm. We have used the soft front panel and the unit is working fine. We have more units in now and they all do the same thing. We have changed chassis, slots, test equipment and they all fail. What is the first reading the test looking for? 0dbm, .707V, 1W, etc? We have entered every combination we can think of and it always says the result is “NA dbm” and it fails. If we knew what the unit is looking for we might be able to at least see if the program will run.  

    Hello Jeep CJ7,
    Yes, you are correct that this value should be in dBm.
    The next thing we have to do is compare the hardware you are using to the hardware we recommend. You can find more about the hardware we recommenend on page on page 7 of this manual.
    Please find the following information and specifications for your devices:
    ---------------------Sig Gens----------------------------------------
    Agilent E4432B
    Freq Range: 250kHz to 3 GHz
    Power Measurement Level Setting Range: ??
    Frequency accuracy?
    Rohde and Schwarz SMA-B103
    Frequency Range: 900 kHz to 820 MHz
    Power Measurement Level Setting Range: –4 dBm to 10 dBm
    ±0.1 ppm frequency accuracy
    ---------------------Power Meters----------------------------------
    Agilent E4418B
    Frequency Range: 9 kHz to 110 GHz
    Power Measurement Range: –70 dBm to 44 dBm
    Maximum VSWR: ?
    Absolute Power Measurement Uncertainty:?
    Relative Power Measurement Uncertainty: ?
    Rohde and Schwarz NRP-Z91
    Frequency Range: 900 kHz to 820 MHz
    Power Measurement Range: –4 dBm to 3 dBm
    Maximum VSWR: 1.11 (23 ±5 °C)
    Absolute Power Measurement Uncertainty: 0.1 dB (23 ±5 °C)
    Relative Power Measurement Uncertainty: 0.05 dB (23 ±5 °C)
    ---------------------Power Splitter----------------------------------
    Agilent 11667B
    Frequency Range: dc to 26.5 GHz
    Input Power Range: ?
    Maximum VSWR (Output ports): 1.22
    Maximum Amplitude Tracking: ?
    Weinschel WA1507R
    Frequency Range: 900 kHz to 820 MHz
    Input Power Range: –4 dBm to 3 dBm
    Maximum VSWR (Output ports): 1.15
    Maximum Amplitude Tracking: 0.15 dBm
    ---------------------Attenuator----------------------------------
    Mini-Circuits VAT-3+
    Frequency Range: 900 kHz to 820 MHz
    Minimum Power (W): 10 mW
    Maximum VSWR: 1.20
    Nominal Attenuation: 3 dB
    Please find the same for yours.
    ---------------------SMA Termination----------------------------------
    Mini-Circuits ANNE-50+
    50 Ohms
    Please find the same for yours.
    ---------------------SMA Cable----------------------------------
    SMA Cable
    50 Ohms; Maximum VSWR: 1.20
    Frequency Range: 900 kHz to 820 MHz
    Power Measurement Range: –4 dBm to 3 dBm
    Relative Shielding: –100 dB
    Please find the same for yours.
    Also, are you running Cal Exec in automatic or manual mode? Are you doing verify or verify and adjust?
    Please get back to me when you can.
    Thank you,
    Vimal Fernandez || Applications Engineer  || National Instruments || (512)683-1781 || [email protected] 
    Applications Engineer
    National Instruments

  • How do I return two values from a stored procedure into an "Execute SQL Task" within SQL Server 2008 R2

    Hi,
    How do I return two values from a
    stored procedure into an "Execute SQL Task" please? Each of these two values need to be populated into an SSIS variable for later processing, e.g. StartDate and EndDate.
    Thinking about stored procedure output parameters for example. Is there anything special I need to bear in mind to ensure that the SSIS variables are populated with the updated stored procedure output parameter values?
    Something like ?
    CREATE PROCEDURE [etl].[ConvertPeriodToStartAndEndDate]
    @intPeriod INT,
    @strPeriod_Length NVARCHAR(1),
    @dtStart NVARCHAR(8) OUTPUT,
    @dtEnd NVARCHAR(8) OUTPUT
    AS
    then within the SSIS component; -
    Kind Regards,
    Kieran. 
    Kieran Patrick Wood http://www.innovativebusinessintelligence.com http://uk.linkedin.com/in/kieranpatrickwood http://kieranwood.wordpress.com/

    Below execute statement should work along the parameter mapping which you have provided. Also try specifying the parameter size property as default.
    Exec [etl].[ConvertPeriodToStartAndEndDate] ?,?,? output, ? output
    Add a script task to check ssis variables values using,
    Msgbox(Dts.Variables("User::strExtractStartDate").Value)
    Do not forget to add the property "readOnlyVariables" as strExtractStartDate variable to check for only one variable.
    Regards, RSingh

  • Stored procedure by using EXECUTE at the sender JDBC

    We have Scenario from JDBC to RFC (ASYNCH).We are using Oracle DB,at the database side we have a stored Procedure and inside the Procedure we are accessing multiple table.
    My query is how can we excecute update sql statement a stored procedure by using EXECUTE at the sender JDBC?
    EXECUTE SP_STUDENT_UPDATE
    ERROR: 'java.sql.SQLException: ORA-00900: invalid SQL statement '. For details, contact your database server vendor.

    Hi. Peter.
    yes , is possible to update your rows in your sp.
    If you want the data determined from the Query SQL Statement to remain in the database unchanged after being sent successfully, enter <TEST>.This is recommended if the data has not only been read, but also changed by a stored procedure entered under Query SQL Statement.
    http://help.sap.com/saphelp_nw73/helpdata/en/7e/5df96381ec72468a00815dd80f8b63/content.htm?frameset=/en/48/465581929734d9e10000000a42189d/frameset.htm&current_toc=/en/ca/b977f1c7814201954f20bb87ad7aab/plain.htm&node_id=64
    Could you put your sp? Maybe something is  wrong .
    Regards
    Lucho

  • How to print procedure in an executable program

    Hi Everyone
                           I am having one procedure  PR_PRICING_T_1723_2000 using cu02.In this some code is written and i want to print the code lines of this procedure in my executable program output.Can anyone help me in finding where this procedure get stored for example in table or is their any function module or bapi where we can directly pass procedure name and can get the desired lines of code.
    Hitesh

    Are they making the changes under a same CR? If so there is actually no way to track the changes made by individual user, unless they put some comments. In such a case, the other option is to have individual users generate a version after completing their changes (but ensuring that no other user makes a change at the same time by using editor lock and releasing the editor lock after generating a version).. Once the version is generated, u can use change management
    If the changes are made under different CR, then you can use version management as suggested by Rich.

  • Error 200617 on every usage of NI PXI-5610 that was previously working

    Hello,
    I have been trying to use the NI PXI-5610 RF upconverter (associated with a NI PXI-5441) with VIs that were previously working, including the NI GPS toolkit and RFSG example VIs. However, with the same setups as were working previously and the examples every time I now receive the following error
    niRFSG
    Commit.vi<ERR>The upconverter reported the following
    error:
    DAQmx Error
    -200617 occurred:
    Measurements: Output generation was aborted by the reverse
    power protection circuitry of the device. Either the output signal exceeded the
    output power limit, or power was being driven back into the output of the device
    by an external source. Correct the problem, then generate the signal
    again.
    I have tried restarting, allowing 30 minutes warmup time before trying to use the RFSG and ensuring that the chassis fans are set to high with no change in the result. 
    I have also tried doing a self test on the RFSG unit in MAX which fails with error 200020.

    Ok, well we can take the first steps via this forum post. If we need to RMA the PXI-board we can follow up on the created service request.
    First I want to start troubleshooting the Self Test error.
    I have attached a application which goes through all of the tests separately so we can determine where this error is coming from. Please run all of the test provide by this application.
    Thank you in advance.
    With kind regards,
    Casper
    Attachments:
    Self Test Internal.zip ‏32 KB

  • PXI-5610 Specification

    Hi,
    Does anyone have NI PXI-5610 Up-Converter Specification? The link on NI web http://sine.ni.com/nips/nisearchservlet?nistype=psrelcon&sortby=localesort&nid=13478&lang=US&filter=... seems to point to PXI-567X specification.
    Thanks.

    Luyy,
    Thanks for contacting National Instruments.  This is correct, the 5610 is part of the 567X RFSG unit.  Those specifications are the correct ones for that device.  There are 5610 specific data in that pdf.  Is there something in particular you are looking for?
    Regards,
    Kenn North
    Senior Product Manager - Search, Product Data
    http://ni.com/search

  • How am I to be able a user to grant access only to see a procedure / function without execute,compile,edit and drop?

    how am I to be able a user to grant access only to see a procedure / function without execute,compile,edit and drop?

    Sorry GregV but thank you, are you sure you can only be done by setting a PC? Can not by giving them certain privileges of a PC?
    PL \ SQL that we use a portable version.
    So actually like this, user A is only used by the X, user A wants to provide read-only access to user B is only used by Y (another PC) to the procedures / functions held user A. How do you?

  • Procedure won't execute after sql loader

    I have a procedure that uses DDE to call a bat file. The bat file uses sqlloader to load data from an excel sheet into Oracle. After the bat file is executed, I am doing a series of deletes and inserts.
    After this is done, I am calling another procedure. Whe I run the form, the second procedure is not executing. I put a message box and it does not pop up.
    When I comment the first procedure, only then the second procedure is called. Is this a sql loader issue or a DDE issue?
    thanks

    Thanks for the reply.
    There are no messages. Using Windows.
    As an example, C:\Program Files\Adobe\Adobe Lightroom 5.7.1 is the location for Lightroom 5.7.1. I have pinned it to the task bar. Click on the icon and 5.7.1 executes. If I double click on C:\Program Files\Adobe\Adobe Lightroom 5.7.1\lightroom.exe I can execute LR 5.7.1, just as I would expect.
    However, if I try the same with LR 6, which is located in C:\Program Files\Adobe\Adobe Lightroom, nothing happens. It is also pinned to the task bar. Click on the icon on the task bar, or on the desk top, and nothing happens. No message, nothing.
    Task Manager indicates the only program running is Firefox, my browser.

  • PXI 5610 Error -200437

    The card shows the error infomation:Error -200437,External calibration constants are invalid. Whether performing an external calibration would solve the problem?
    Thanks.
    Attachments:
    pxi 5610.jpg ‏124 KB

    Which version of NI-RFSG driver are you using?According to this KB, you can upgrade to the latest NI-RFSG driver and solve the problem.
    Thanks,
    NI AE Emily

Maybe you are looking for

  • Mass Upload via LSMW

    Hi All, I have got one client requirement in which client wants that whenever they run "Change in Employment" action for Contract extension/Change in Contract type, in IT0016 whatever contract end date comes should automatically populate in IT0041, i

  • Partially Paying AR Invoices using DTW

    I am trying to add a cash payment for $500 against a $1000 invoice using DTW by filling in the below fields. However, it is closing the invoice even though it hasn't been fully paid. Are there additional fields that I need to fill in to specify that

  • No ocijdbc8 in java.library.path (with Servlet under Tomcat)

    Hello, my problem is the following: I must use the OCI-Driver of the Oracle-Client 8.1.7 because I use special PL/SQL-functions. I get the error message: java.lang.UnsatisfiedLinkError: no ocijdbc8 in java.library.path when I try to start my servlet

  • How do I call get_ddic_reference?

    Hi, I am trying to call the get_ddic_reference method.  Here is my code:   TRY.       lr_column ?= lr_columns->get_column( columnname = 'CARRID' ).     CATCH cx_salv_not_found .   ENDTRY.   lr_column->get_ddic_reference( ). It runs, but does not do a

  • Will Adobe Premiere Work on this Mac?

    hello, I am just about to start editing school, I dont have much money and am considering buying the Mac below. Do you think it will work with the newest version of Adobe? Or will it possible lag a lot? I have a 2 tera Firewire Hardrive I plan on sav