NI USB-6509 Self-Test

Hello all,
Does anyone know what the built-in self-test of the NI USB-6509 hardware or any similar digitial I/O hardware does? Does it just confirm the hardware can communicate, or it actually tries to write and read certain digitial I/O lines?
I just want to have some idea about what it means when the self-test passes or fails.
Thanks,
Wenhao

The self-test is just a diagnostic that checks that the onboard chips are powered properly.  The Test Panels actually check I/O.
AC - LV Solver

Similar Messages

  • MyDAQ fails Self-Test and will not work with any software

    I cannot get the myDAQ to communicate properly. I did everything I could to make sure that the device was detected by the computer. It shows up in Device Manager under Data Acquistition Devices as NI myDAQ and in NI MAX as "myDAQ1", but I receive errors when trying to use it with any NI program.
    Trying to self-test or reset the device in MAX gives these errors:
    -The self test failed. -88705 The specified device is not present or is not active in the system. The device may not be installed on this system, may have been unplugged, or may not be installed correctly.
    -There was a problem resetting the device. Error code: 80040383.
    It also gives this error when trying to use the Elvis Oscilloscope:
    -Error -200324 occurred at ELVISmx_DAQmx Create Channel (AI-Voltage-Basic).vi:1 
    Possible reason(s):
    NI-DAQmx is unable to communicate with the device. Make sure the device is present in and accessible to the system, is not currently being reset, and is not reserved by another driver such as Traditional NI-DAQ (Legacy).
    I have uninstalled and reinstalled and repaired all installed NI products twice. I updated to NI-DAQmx 9.8.0 and am using NI Elvis 4.5. I have checked  NI Device Loader and mDNS Responder Service in Services and everything is running properly. I have also tried Reset Configuration Data in NI MAX several times. I turned Fast Startup off as suggested in one article. I have tried plugging it into every single USB port on my computer; all 8 of them. The blue led is on when plugged in, but I cannot figure out why it will not work with any of the software. I am running Windows 8 Pro 64-bit. Also, it completely works on the school computers running Windows 7 without any problems. I would like to have it functioning with all the NI software purchased, but all I want to do is use the oscilloscope software for now. It has honestly been way too much work and a huge loss of time for something that should work out of the box. Please help me resolve this, thanks.
    Attachments:
    ni_support.zip ‏341 KB

    Hi chaka123,
    I had already read that article so it wasn't anything I hadn't tried. I did, however, uninstall both NI DAQmx and NI ELVISmx, ran the registry cleaner and then plugged my device in. Now it shows up as USB Flash Firmware Updater. This is actually what it showed the first time I plugged this in out of the box. After much troubleshooting I was finally able to get it to at least recognize it as a myDAQ. This is problem I know others have had.
    To answer your questions:
    Does the Test Panels work at all?
    I am not sure, I didn't try it before uninstalling. Now the Test Panels isn't an option because it is recognized as the Flash Firmware Updater.
    Have you tried to delete the device in MAX?
    Once again, I cannot because it shows up as Flash Firmware Updater.
    Which is the color of the icon on the left side of “myDAQ1”?
    It was green before the uninstall. Now it shows the USB symbol.
    I tried deleting the driver. When I plugged the device back it, Windows started automatically trying to load a driver for it. Now it shows as NI myDAQ with a yellow exclamation symbol. In MAX it shows as NI myDAQ with a USB symbol. Under Driver Details it says "Windows does not have a driver associated with your device." 
    What folder can I manually point Device Manager to in order to load the correct driver?
    Attachments:
    devicemanager.PNG ‏40 KB

  • To check ink levels I am told to print self test. page from cpu or printer.

    There is no way to print a test page on my" psc 2410 all in one" connected to a HP cpu with Win 7, 64 bit. Is there a workaround/

    The page here shows how to print a self-test report from the front panel, but it does not appear to have ink level information.  I have not used a PSC 2410 for years, but the following may give you what you need:  press the # and * buttons together on the printer.  The screen will say "Enter special key combination".  Press 1 2 3.  This should bring up a service menu.  Use the right arrow button repeatedly to bring up a "Special Reports" screen, OK, then right arrow to select "Cartridges" or some similar, then press OK. A page will print with ink level information shown about the middle of the page.
    Bob Headrick,  HP Expert
    I am not an employee of HP, I am a volunteer posting here on my own time.
    If your problem is solved please click the "Accept as Solution" button ------------V
    If my answer was helpful please click the "Thumbs Up" to say "Thank You"--V

  • Software Updates for Remote PXI System/DAQ-mx Self Test

    1) In MAX, I'm attempting to run the Labview Realtime Software Installation Wizard. When I select the software for install, I get an error "Cannot Process Install because other Dependencies are required: Labview RT 7.1.1". What should I do to correct the problem? I'm trying to update both DAQMx and Traditional DAQ from 7.2 to 7.3. The wizard actually indicates that the remote system only has 7.2, and that it sees that 7.3 is available.
    2) Also in MAX in the Remote System Tree, I get an error when trying to perform a self test on my PXI-6025 card, or when I attempt to reset it. The error code is 88705. I noticed on another post that there might be the need to perform a repair of the installation. This was done for both DAQMx and Tranditional DAQ, neither of which fixed the problem.
    3) In MAX I have used the Tools>>Traditional NI-DAQ Configuration>>Remote DAQ Configuration to configure my PXI-6025 DAQ channels, accessory, device #, etc. I then proceed to run a Analog Input VI on the remote system by correctly targeting it and selecting i.e. channel 0. No errors occur. However when I try and select "Browse" under the channel control, the "Select Names" window pops up empty. I would assume that this window would pop up with the DAQ channel names assigned during the channel configuration of the PX-6025??? Whats missing here?
    Thanks
    Pete.

    Hi Pete,
    I'm assuming you are using LabVIEW RT 7.1, since that is a requirement for using DAQmx. Here are my responses for your three issues:
    1) You are correct in assuming you can upgrade the DAQ drivers in the manner you outlined, normally that works fine. However, as the following KnowledgeBase points out, the list of dependencies vary for different installs.
    http://digital.ni.com/public.nsf/websearch/F5D80FBE806B4B9D86256F0F005630CF?OpenDocument
    Do you have LabVIEW RT 7.1.1 installed, which appears to be the required dependency? If not, you may download this from ni.com>>Drivers and Updates
    http://digital.ni.com/softlib.nsf/MainPage?ReadForm&node=132010_US
    If so, but the problem persists, please post all versions of your software so I can determine where the incompatibility lies.
    2) After we take care of upgrading the DAQ drivers, let's see if the 88705 error goes away, it might take care of it.
    3) Again, all the steps you described are correct. It might be that all these problems are related and due to a bad install of NI-DAQ.
    When a number of software issues like this arise all at once, it might be best to just re-format the controller and perform a fresh install of everything. Often times that will correct the erroneous behavior.
    Regards,
    Jeff M.
    Applications Engineering
    National Instruments

  • USB-6509 Stops being recognized by Windows 7 device manager.

    I am currently using a USB-6509 under Windows 7.
    On a freshly formatted machine I have no trouble setting up the device using the National Insturments Measurement & Automation Software version 9.4.
    Some windows updates were pushed to the computer that my USB-6509 was connected to, causing the computer to reboot in the middle of use. When the machine came back up the USB device couldn't be found by the Measurement & Automation Software, nor by Windows Device Manager. I did a complete uninstall of the software and reinstalled, this did not fix the issue. I uninstalled 9.4 and installed 9.5 and this too has not resolved the issue. 
    Inside device manager windows will show "Static Digital Firmware Loader" under "Data Acquistion Devices". But there is nothing installed pointing to the USB-6509. 
    The green light on the back has yet to come back on. I would prefer not to reformat if at all possible. Any suggestions?

    I was having this exact issue.  "NI Device Loader" was running on the machine, but stopping and restarting this service by itself was what finally caused the device to start working again after lots of rebooting and other troubleshooting steps.  Thanks for the tip.

  • Ipod nano starts with a self-test after sleep

    My new Ipod nano runs a kind of self-test if it was at sleep for a couple of hours.
    Does this have to be? If not how can I determine specifically what is wrong? Thanks

    Gert,
    When iPods are not used for a period of 24 hours of more then they go into something called "Sleep Mode" or "Deep Sleep".
    You'll see the Apple logo when you try to turn your Nano back on which is normal, you are "waking up" the iPod.
    I've never heard or seen of such self-test, is it still on self-test? Or you want suggestions on how to get out of that?

  • Hard Drive failed both postive read and smart short self test

    Hello, my Lenovo Think Pad's Hard Drive failed both targeted read  test (error code WHD20-HJF) and the smart self test (error code WHD16-KPY). I admit that I'm an older user and I have no idea what this means and what is the possible fix for it.
    I am open to all suggestions;
    THANKS,
    Octavious ONeal

    Hi, Octavious
    When the hard drive fails those tests, it's an indication that the drive is dying. It's recommended to back up any valuble data on the hard drive and then replace it soon. The drive may still work for a little while, but it will likely stop working sometime soon.
    Hope it helps,
    Adam
    Did someone help you today? Press the star on the left to thank them with a Kudo!
    If you find a post helpful and it answers your question, please mark it as an "Accepted Solution!" This will help the rest of the community with similar issues identify the verified solution and benefit from it.

  • HDD replaced with SSDHDD & Drive Self Test failed hardware scan by LSC

    Hello! Happy new year!
    I bought a new SDDHDD instead of the HDD that came with the laptop when I bought it because it was faulty.
    I want to make sure that the new drive isn't faulty when running the hardware scan using Lenovo Solution Center (LSC), I got this result code: WHD01V000-HK9V1D - Warning for the SMART Drive Self Test. However I tried testing using HD Sentinel and SeaTools but they showed no errors.
    Note: I'm taking LSC seriously because it showed warnings then critical alerts before my first HDD failed so I don't want to ignore it.
    Please let me know what it means and if I should worry or how to fix the problem? Or does it simply not support SDD?
    Thanks for reading this long message! (Sorry)

    Hi, quocphong
    It sounds like the hard drive is running okay overall, but it has some bad sectors in it. You will not be able to read or write to these sectors. That being said, I would first try running a ChkDsk on the hard drive and see if any of the sectors can be repaired. To run a ChkDsk, go to Computer and right-click on the C: drive and select Properties. Navigate to the Tools tab and select the box that says Check now... This should scan the drive for errors and attempt to fix them if possible. The test can take a while, so I would do it when you will not need access to your computer.
    If the test is not able to fix the errors, then I would contact Lenovo at 120-11072 (Toll Free) or +(84) 8 3521 8000 (Standard Charges Apply) and discuss a hard drive replacement with them. As for Windows, you will need to ask for recovery discs as well. Lastly, I would backup everything on the current hard drive so that it is not lost.
    Hope it helps,
    Adam
    Did someone help you today? Press the star on the left to thank them with a Kudo!
    If you find a post helpful and it answers your question, please mark it as an "Accepted Solution!" This will help the rest of the community with similar issues identify the verified solution and benefit from it.

  • XI 2.0 WorkBench File Adapter Self-Test Error

    I have installed the file adapter on our XI serve and when I perform the Self-Test I get the following error HTTP request failed. Error code: "401". Error message: "Authorization Required".  I have the user  XIRWBUSER defined to the Adapter Engine.  What else am I missing?

    Hi Dan,
    You are thanking a poster for his answer.
    Please next time also give him points.
    See: /people/mark.finnern/blog/2004/08/10/spread-the-love
    This time I did it for you.
    Remember: When you ask a question you also assume the responsibility to assign points.
    Please check your other questions and if the answers were good.
    Spread the love and give some points.
    Thanks in advance, Mark.

  • PCI-MI0-16XE-50 Failing Self-test

    I am using a PCI-MI0-16XE-50 which is failing the self-test in MAX 3.1.1.3004.  I have used the E Series Diagnostic Utility which gives the error "Error 10809 occurred at cwdaq:  Starting DAQ operation Attempted to start a pulse width measurement with the pulse in the phase to be measured (e.g., high phase for high-level gating)."  Windows XP detects the device fine, as does MAX.  The obvious answer is that the device is broken, but is it feasible or economical to have the device repaired?  What tends to cause this type of problem and/or how can it be corrected?
    Thanks,
    Brandon

    Hello C.
    You will need to contact National Instruments for a quote for repairing
    you board. If you bought less than a year ago, it may still be under
    warranty in which case it would not cost you anything to repair.
    However, I would suggest in case you need to pay for repair to consider
    getting one of the new improved and more performent yet low cost M Series devices.
    Hope this helps.

  • Help! Ipod doing some crazy self test thing and now wants some remote test.

    well i turned my ipod mini on this morning to plug into my headunit but it came up with some screen about memory and restarts and selftests etc (i couldnt get out of this screen) anyways i mustve pressed something and it went through this self test then started prompting me to do certain things (like check scrolls etc) this was okay then it came to a remote check. i dont have a remote for my ipod. It says i need to check volume, pause, play, next song etc and nothing i press on my ipod does anything. Can anyone help??? what should i do??
    Thanks all.

    thanks for reply but i tried that earlier and ipod wont respond to any commands from ipod buttons.

  • RWB - Integration Engine self test - web service security and proxy

    Hi,
    I am working with a new installation of PI 7.0. In the runtime workbench, under self test for integration engine, there is this error/warning:
    ""Details for 'Is Web service security available?'
    Communication error: Proxy calls are not permitted on sender or receiver side on the IS (client)""
    What exactly is the problem? Is there any additional configuration needed within PI to use proxies? We do not have the cryptographic toolkit installed. Is that nesseccary to work with proxies? We have done several other scenarios with RFC, MAIL, HTTP, etc and they work fine. If anyone else had this problem and managed to fix it, please let me know..
    Thanks,
    Lasya

    You can ignore this error. It is  simply a warning that says message level security has not been configured. Without message security too, you can do proxy communication.
    But, if you want to configure messag level security, go through XI Config guide section 12.4.
    Message was edited by: Jay

  • Video recording with USB 6509

    I am working on one project. In which I have CCD image sensor with EIA B/W video camera and its output is connected to Analog Front End(AFE). The output of AFE is 12 Bit Digital data at the rate of 9 MHz. I am thinking to use USB 6509 for data acquatision from AFE. Actually I want to produce one video file(e.g. 30 second video) from the AFE that is the reason I am using USB 6509 to get the digital data in my computer. 
    I know USB 6509 is compatible with LABView SignalExpress software. 
    Now, My question is, as I have data to my computer using USB 6509. How can I generate video file (preferrable lossless compression) from the digital data using the SignalExpress or Do I need any other software?
    FYI: I am new to LABView.

    Hi erhk,
    SignalExpress will not be able to create a video/image from the data; though, it will allow you to acquire it. If you can get the data into a 2D array of pixel values, you can use LabVIEW with the Vision Development Module to create a video with it. There also may be other programs out there that can convert the 2D array to a video/picture file.
    Bill E. | Applications Engineer | National Instruments

  • PCI-6221 fails self-test in MAX but device manager says it's working

    Hi everyone,
    I bought a few PCI-6221 (37-pin) data acquisition cards and have been able to get one to work but 3 others don't work.
    Device Manager sees them and says they work.  MAX sees them and when I run Self Test, they fail with error number
    -50002.  Motherboards in non-working systems are PC Chips M957G and P25G.  OS is Windows XP Pro with SP2.
    I already tried uninstalling SW and the board, reinstalling SW and the
    board...autodetect etc works fine.  I even reformatted
    the HD on one of the computers and reinstalled the OS.  Anyone have any ideas?  Thanks in advance.  Halden
    [email protected]

    Thanks Tom for your suggestion to try NI-DAQ 8.0.  I just saw it today.  Yesterday,
    I was able to isolate 3 PCI-6221-37 boards as defective because they all wouldn't work in another computer while
    one other identical model card did work in that same computer.  I called to arrange to return them to NI and learned
    that for these 3, I needed NI-DAQ 8.0.  The fourth one was okay
    with NI-DAQ 7.5.  Apparently, this was a known problem. 
    I wish NI would have simply notified me that my boards would
    not work with the software sent with them and sent the NIDAQ 8.0 CD as soon as they knew rather than have me
    waste a day of my engineering time trying to figure it all out.  Booo NI.
    One of the boards has worked for almost a day under NIDAQ 8.0.  I was able to make some progress on the
    system I am building.  But then it stopped working.  LabVIEW
    (version 7.0 executable) programs would not open completely
    or would open slowly.  One that accessed the COM port only (no
    data acquisition) did work.  MAX opened very slowly or not at all
    and when I click on
    Devices and Interfaces, the hourglass is up for a long time and finally the + icon goes away indicating that MAX is not
    aware of any devices being present.  The Windows Device Manager says the board is present and working.  Windows
    Task Manager lists LabVIEW Runtime and MAX when only MAX is open and indicates that MAX is Not Responding.
    I shut down the computer, removed the PCI-6221 board, booted it up again, and MAX is the same way.  I uninstalled
    MAX and NI-DAQ LabVIEW 7.0 support and then reinstalled them.
    Now the MAX loading window is up for more than a minute and NIMAX.exe is listed in Task Manager's processes.  After
    the first minute, LabVIEW Runtime appears in the Task Manager's
    Applications list.  After a while longer, the MAX window finally
    opens
    and MAX is listed in the Applications.  I click on the + icon by Devices and Interfaces in MAX and the hourglass appears
    and Windows Task Manager says MAX is Not Responding.  Eventually, the + icon and the hourglass go away and
    the Task Manager lists LabVIEW Runtime and MAX as Running.
    I shut the computer off for several minutes, started it up again, and MAX is still takes a long, long time to open. 
    I haven't put the PCI-6221 card back in because MAX still isn't working.
    What do you suggest I do next?
    Thanks
    Halden

  • Satellite A300 does not pass BIOS self-test

    Hi!
    My laptop does not pass self-test in the BIOS. Inscription "Toshiba Leading Inovation" does not disappear. So what to do?
    The buttons are not active.

    Hi
    That means your screen is completely black and you can see nothing? Thats strange but you can try this workaround:
    -Try removing the battery and the AC adaptor
    -Wait at least one hour or one night (no battery and AC adaptor should be connected)
    -Connect battery and AC adaptor again
    -Try to start the notebook as usual
    If the same problem occurs you should contact an authorized service provider. It seems to be a hardware problem :(

Maybe you are looking for

  • Error while comitting a transaction in oSB. The following is the error stack

    Error while comitting a transaction in OSB. The following is the error stack <Apr 20, 2015 12:00:15 AM MDT> <Error> <EJB> <BEA-010026> <Exception occurred during commit of transaction Xid=BEA1-1AE41F1CAE45F2B146FD(296700848),Status=Rolled back. [Reas

  • Sharing a library between multiple user accounts on the same PC

    I have an XP pc setup for the kids and each has their own login so they can have their own bookmarks, shortcuts, desktop, etc. They each have an iPod and use iTunes. I want them to share the same music library so that when one downloads or rips a son

  • Saving Message Payload

    Hi, i want to do the following: When i receive a message in XI, i want to save the message payload temporally in an abap table. So i will define a rfc to call from XI and give the message payload as an import parameter. Have you any idea how to do th

  • The query is performing very very slow, can any one please help me tue this

    SELECT h.ROWID row_id, e.preparer_user_id FROM arsarc.ars_sv_reconciliation e, arsarc.ars_sv_reconciliation_accts a, arsarc.ars_sv_acct_status_history h WHERE e.rec_id = :b1 AND e.rec_id = a.rec_id AND e.ledger_id = h.ledger_id AND a.ccid = h.ccid AN

  • Add Customer_PO number to SQR Invoice

    I need to add customer po number to our invoice.I am very new to PeopleSoft and SQR when I look at biivcpn.sqr and bisubp00.sqc I can't tell where(What Table) it pulls from How do I tell that? How do I need to add the Customer PO line to bisubp00.sqc