No BNC-2090 Accessory Option in MAX/No Signal

I'm using an AT-MIO-64E-3 card for A/D signal collection, with the card connected to a BNC-2090 signal box. The Measurement and Automation Explorer doesn't have an accessory selection for the BNC-2090, when I try to run the test panel I get no signal, and VI's give the error -10003 from AI Hardware Config. I'm running Windows 2000, and this problem arose when I upgraded from NT to 2000. Is there an easy fix for this, or will I need a newer A/D card? Signal is getting to the BNC-2090, it's just getting lost somewhere between the 2090 and Labview.

Hello Dstockw,
Are you using AC Coupling with this application? If so, E-series cards do not support it and will give error -10003 from AI Hardware Config.
What version of the NI-DAQ device driver are you using? NI-DAQ 7.0 is a fairly recent version that supports that board (using Traditional DAQ). If you are using an older version of NI-DAQ, try updating it and run the test panel again. NOTE: If you update the driver, be sure to uninstall your previous version of NI-DAQ first.
Check to be sure that you have properly installed the BNC 2090. Information about installing and configuring this device can be found at:
http://www.ni.com/pdf/manuals/321183a.pdf
Also, be sure that you are using a proper cable to connect to your AT-MIO-64E-3 card. Th
e recommended cable is the SH 1006868 shielded cable. Try double-checking the installation/configuration of this DAQ card by looking at:
http://www.ni.com/pdf/manuals/320517g.pdf
You may also want to try running the E-Series Diagnostic Tool. This tool will test the functionality of your card and run a self-calibration on it. You can find this test at:
http://www.ni.com/support/selftest/
Try these steps and let me know if you have any further questions.
Scott Romine
National Instruments

Similar Messages

  • Compatibility between BNC-2090 and PXI-6541

    Can we use BNC-2090 terminal with HS-DIO  PXI-6541 ?
    John Doe

    Hello,
    As it said in the BNC-2090 documentation :
    BNC-2090 – A shielded, rack-mountable adapter with 22 signallabeled
    BNC connectors and 28 spring terminal blocks to simplify
    connections to your I/O signals. The BNC-2090 includes component
    locations for passive signal conditioning and connects to any 68-pin
    M Series, S Series, E Series, or B Series DAQ device.
    So your card isn't a DAQ card but a HS-DIO card, so it doesn't appear in the supported cards list.
    So it shouldn't be compatible.
    Please contact your commercial contact at NI France for more informations.
    Regards,
    Pierre R...
    Certified LabVIEW Developer

  • Cross talk in BNC 2090

    Thanks for the information provided. I followed your advice by connecting
    DGND1 with a wire to USER1 and DGND2 to USER2. But I still get cross talk.
    It also puzzles with a fact that when I connect a signal source (from a
    functional generator) to ACH1 but in fact the signals show up on the
    Biobench screen as ACH0 as defined. So my hunch is that something wrong
    happens to BNC 2090 or the PCI-MIO-16XE-50 board installed in the computer.
    Would it be possible for me to send in the BNC2090 for a check? If so, what
    are the procedures to follow.

    We handle all repair requests over the phone. The procedure is as follows:
    1. Go to www.ni.com/ask to create your own Service Request number (SR#). Choose the Phone NI option. Fill in the form with the appropriate information.
    2. Call the support number that is given to you and enter in your SR#. You will be directed to an Applications Engineer (AE) that will assist you through the process. The AE will be able to answer any questions you may have about the repair process.
    Have a great day.

  • Bnc 2090 offset and other errors

    I have a bnc 2090 plugged into a pci 6070e. The BNC is currently plugged into a DC variable power supply. Right now, when I change the voltage on the power supply, the other channels reflect this change even though it is only plugged into one channel. I have read about ghosting and even when I lower the sampling rate, the problem is not fixed. There is also an offset that is showing up on the MAX test panels and in my VI. This offset steadily decreases and seems to arbitrarily change. Whats going on here?

    About the offset another point, I assume you use a measurment on one coax connector (+ and - in one connector), in that case your switch should be set to RSE.
    In MAX this should be set as well (and in your VI) instead of the default differential measurement.
    Ton
    Free Code Capture Tool! Version 2.1.3 with comments, web-upload, back-save and snippets!
    Nederlandse LabVIEW user groep www.lvug.nl
    My LabVIEW Ideas
    LabVIEW, programming like it should be!

  • Using PFI1-PFI9 on BNC-2090 to trigger recording on a PCI-MIO-16-XE-50

    Hi there,
    I've got another question about triggering using the BNC-2090. Now I can get an output of a particular digital line from the DIO-96 I'm using using a ribbon connector connected to CB-50 Connector block. This connector block has individual line outputs, so is it possible to take one of these individual lines, plug it into PFI1-PFI9 of the digital connection area of the BNC-2090 and have the PCI-MIO-16-XE-50 use the appropriate digital PFI line as a trigger? I ask this because I assume the default trigger line whey I click "Use external trigger" in NIDAQ Tools for Igor Pro is PFI0, so I was wondering if there's a way to change that. Thanks a lot!
    Arul Thangavel

    Arul,
    Yes, you can connect the dio line of the dio-96 to any of the PFI lines that are located in the pinout of the bnc-2090. The default line to be used for triggering an acquisition is PFI0,but you can modify that.
    In LabVIEW, you can select the proper PFI line by configuring it in the "AI Start.VI", or by using the "AI Trigger Config.VI" before using the latter VI.
    In LW/CVI, you can also select the proper PFI line by configuring it with the "DAQ_Config" function call and with the "Select Signal" function call.
    Hope this information is helpful.
    L Aguila
    Applications Engineer
    National Instruments

  • How to count TTL pulse by DAQ through BNC 2090?

    I'm trying to count the TTL pulse from a APD by DAQ. There are some very nice discussions about this issue in this board. However, I am a novice and not sure about the hardware connection. The signal from APD is delievered out through a BNC cable and I'm trying to deliever this signal to DAQ through BNC 2090. I tried just simply connect the BNC cable to ACH0 and tried running the vi's, such as "Count_Edges.vi" and "Count_Events.vi". But I got no counts. I noticed that both these vis are actually reading signal from PFI port. That could explain why I got not counts.
    So now the problem is how should I connect my signal in a BNC cable to the DAQ and have the counting working? Thank you for your help.

    Hi Opal,
    Here is what you should do for the hardware connections with these devices:
    - connect your external signal to the BNC-2090
    - connect the 68 pin connector on the BNC-2090 to the 68 pin connector on your DAQ board. You should have a 68 to 68 pin cable for this.
    That is all that is involved with the hardware connections. After this, you should be able to read your signal. Here is the link to a helpful DAQ manual with in depth getting started information: DAQ Getting Started Guide. I hope that you find this information helfpul.
    Regards,
    Hal L.

  • Strange voltage observed in NI BNC 2090

    Hi all, I have a BNC 2090 and AT-MIO-16O-2 DAQ board. I am recently
    (starting this morning) experiencing strange voltage readings from the
    ACH1, 2,3,4and 5. The voltage input from ACH0 is about -.2V; and range
    from -1 to -3 in other channels. These happens even when nothing is
    plugged into the BNC other than the its connection to the DAQ in the
    computer.
    any suggestion on what might be happening?
    Roshan.

    This is actually very common. When nothing is connected to the pins the voltage will charge up on the leads. Sense the lines are not connected to anything and the voltage has no where to go the lines will normally completely saturate the capacitors on the board.
    This is not a bad thing and it should not affect your measurements, if you do not scan the channels that are not connected. If you are worried about the unused floating signals, you can wire the unused inputs to ground.
    If you are performing measurements and notice that the inputs are floating you should check your grounding methods. In general use RSE for floating sources, NRSE and DIFF for grounded signal sources. If you want to use NRSE or DIFF on a floating signal you will need to add a 10-
    100kOhm bias resistor from the negative lead to AIGND to reference the signal. Refer to this KB for more help.
    http://digital.ni.com/public.nsf/websearch/DF70FCE​03D397BFE86256A3700780FFA?OpenDocument
    I hope this helps,
    Joshua

  • Option GT Max Card + MBP = Speaker Noise...Help!

    HI All,
    I bought the new Option GT Max card, and it works on both the Mac side and the Windows side (I'm running Parallels). I use the downloaded driver on the Mac side and the native cingular windows drivers suplied with the card on the XP side. My issue is that when active, the modem transmissions interfere with my speakers....the speakers make a buzzing and grating sound, even when turned off. This makes me think it's an electromagnetic interference issue. Has anyone else come across this issue, and if so, been able to resolve it?
    MBP   Mac OS X (10.4.9)  

    Hello atom6. Welcome to the Apple Discussions!
    My question is how do I get it to work with the Base Station and my Apple TV? Is there a card reader or some sort of adapter that I can use to connect the Option GT Max 3.6 Express Card to the base station like a DSL or Cable modem?
    The 802.11n AirPort Extreme Base Station (AEBSn) requires an Ethernet connection for Internet access. Whether that be an existing Ethernet network or a Satellite/DSL/Cable modem. I'm afraid that I'm not aware of any type of Ethernet adapter that will work with your Option GT Max card to do what you are looking for.

  • BNC-2090 & PCI-6036E: MIO power or LPM power?

    Which power mode do I choose?
    There's a choice of "MIO-16" and "LPM-16" on the back of the BNC-2090.
    Neither the string "MIO" nor the string "LPM" appears in the owner's manual for the PCI-6036E [322339c.pdf].

    The PCI-6036 is an MIO board so you'll want to use the "MIO-16" selection. The MIO boards are our newer series of boards whereas LPM series is our older legacy boards.
    Erick D.
    NI Applications Engineer

  • Floating voltage in BNC 2090, PCI 6024E

    Hi experts, I have a single ended connection on a BNC 2090 connected to PCI 6024E board. However, when I leave it connected(without any input) I see that the voltage gradually increases to +10V. Even when I input a signal, I see that the 10V forms a baseline, which hinders any measurement. Is this some sort of grounding problem? Kindly post any helpful thoughts on this matter. thanks. Roshan.

    Roshan,
    I believe that the 2090 has a switch for each channel which configures the inputs for either Single Ended configuration (AI channel is read with respect to your 6024's Analog Input Ground) or Differnetial configuration (AI channel reads signal on BNC connector differentially). Ensure that the terminal configuration you have programmed for you channel matches the terminal configuration selected by the switch on the 2090. If you do not make any explicit setting, the 6024 will default to differential mode.
    I hope this helps,
    Dan

  • 68-pin connector on BNC-2090

    Can either of the 68-pin connectors on the BNC - 2090 be used for input?

    The 68-pin connectors essentially feedthrough the BNC-2090 and the BNC channels are tied to these pins in parallel. This document explains this in a little more detail:
    http://digital.ni.com/public.nsf/websearch/FDE04C8A1E4481B486257199005D71A7?OpenDocument
    So you do have access to those pins on the front and the back of the 2090. You just have to be careful that you don't double-drive the same line.
    Garrett H
    National Instruments

  • Can I use a SH68-68EPM cable to connect a PXI-6704 with a BNC-2090?

    Hi,
    I was wondering if I could use a SH68-68-EPM cable to connect a PXI-6704 (sitting in a PXI-1002 Chassis and talking to my (Windows 7) PC via a PCI-8331) with a BNC-2090-box.The manual suggests to use a D1-cable but maybe it is possible to still get the analog outputs from the BNC-box although the cable is different?
    Thanks for your help!

    Hi cars10,
    The BNC-2090 is not compatible with the NI 6704 due to pin mapping.  You can use BNC connections via the CA-1000 or use a card from the analog output series family with the BNC-2090A or BNC-2110 as described in the user manual. 
    Regards,
    h_baker
    National Instruments
    Applications Engineer
    Digital Multimeter Resources

  • Error -201342 "Unsupport​ed Accessory" when using MAX to create tasks PXIe-4330 & TB-4330.

    Hi all!
    I have an -18 slot PXI chassis with multiple PXIe-4330 and TB-4330 installed.   When attaching any TB-4330 to the -4330 module within slot 11 and generating a task within MAX, the system returns the following error: "-201342, unsupported accessory". Any task can read all other slots without generating an error.
    I don't expect this to be SN based as the TB is interchangable with -4330s in other slots though MAX will not allow me to add a TB-4330 to the configuration.
    How can i help MAX see the TB-4330 within slot 11?  Is there a driver update that corrects this?
    Thank you!

    Hi Mlvca,
    Our support team is currently working on this issue. If there are any other suggestions, please feel free to post.
    Sunaina K.
    Product Manager
    Embedded Systems

  • Can't find PXIe 8135 as a option in MAX

    Hi
    I am trying to setup a project with the PXIe 8135 controller and the PXIe 1085 chassis but can't find it in MAX. I am currently downloading the latest DAQmx drivers 9.6.1, so hopefully that helps. I am trying to run the setup the project with simulated hardware until we get the PXI that's on order.

    NO Luck... Look DAQmx 9.6 is not installed yet and it look like I dont have support for the PXIe8135 yet I am running LABVIEW 2011
    Attachments:
    PXI8135.png ‏140 KB

  • No option to set tv signal in video

    Trying to connect ipad2 to tv but there's no tv options in video setup in settings. The only thing I've got there is captioning and start playing. Please help!!

    In the screen capture you provided earlier you say your Apple ID is listed below PJ O'CONNELL?
    Because it's never done that for me and I don't know why you would want it to. 

Maybe you are looking for

  • Creation of a collection based on specific version of software installed.

    Hi...This is pushkar and from the last few days I was trying to create a collection of 64bit machines which has 32Bit Office installed in sccm2007, But am not able to create such collection. Could somebody please help me with this issue.  

  • Group by not working for me

    say I had a list of part numbers: 0104589 0119493 0119787 0134490 0245849 0204056 0287694 0294886 0308956 0388690 0406990 0588006 0590880 but I just wanted a list of the first two numbers using the query: <cfquery name="qrySeriesList" datasource="erp

  • S/MIME Support for BIS

    Hello we enabled S/MIME support on our BES which works fine with S/MIME enabled devices. However, on one device we're using BIS instead. S/MIME support does not seem to be available for BIS (hosted by vodafone.de). Does anyone know if S/MIME can be u

  • Maintain custom page size with bleed

    I use CS4 and when working in Illustrator ( and later taking into a book in InDesign) how do I maintain custom page size with bleed in Illustrator when saving as PDF . I have been told my files are not print ready and have no bleed even though I set

  • Not able to connect remotely via TNS entry

    Hello, I need to do clone, so wanted to have remote connectivity to my test database. I brought the test database in nomount. Put the right entries in tnsnames.ora file of the same, yet when i login directly it shows it is in NOMOUNT but when remotel