Problem with advanced threshold in OCR - Vision Assistant 2013

I'm facing a problem with Vision Assistant 2013
The OCR charter set file advanced threshold data is always fixed at 255 upper value and optimize for speed is checked.
I edit them and reopen the file but no change.
anyone facing the same problem ?
Attachments:
Untitled.png ‏7 KB

Hi Paolo,
Thanks for your answer. Yeah I have seen the examples and I´m familiar with the use of the OCR VI. I have use it a couple of time already with good results.
My problem came last week. While I was trying to run OCR on my image in my LabVIEW code. The algorithm did not detected what I was expecting.
I decided to run the test on the vision assistant and it worked perfectly. I assumed my code had different parameters, so I decided to generate a VI from the vision asssistant and run it on LabVIEW for the same image to verify.
I did not change anything on the VI (all parameters are the same), and used the same image. Surprinsingly the results are different between the assistant and the VI. That strikes me a lot!
I´ll start a new thread as you recommended. Hope to find the solution soon. Thanks  
Regards,
Esteban

Similar Messages

  • Problem with Advanced Routing in WRT320N

    Hi, everybody.
    First of all, sorry for my poor English.
    I've a problem with Advanced Routing in router. My provider using routing and it looks like that:
    route add -p 172.16.2.0 mask 255.255.255.0 10.1.0.2
    route add -p 172.16.5.0 mask 255.255.255.0 10.1.0.2
    route add -p 10.0.0.0 mask 255.0.0.0 10.1.0.2
    Everything is working well in XP (or other OS), but then I try to setup a Routing in router:
    1.
    2.
    I'm realy tired to flashing and reflashing between 1.00.02 and 1.00.03, changing browsers, and hard reseting.
    Anybody'll gimme a shot trick to my tired brains?
    Solved!
    Go to Solution.

    1. Yes.
    2. netstat:
    C:\Users\dan>netstat -rn
    ===========================================================================
    Список интерфейсов
    31...........................inet
    12...00 40 f4 76 95 bd ......VIA Rhine III Fast Ethernet адаптер
    11...00 1d 92 2c 23 02 ......Realtek RTL8168B/8111B Family PCI-E Gigabit Ethern
    et NIC (NDIS 6.20)
    1...........................Software Loopback Interface 1
    15...00 00 00 00 00 00 00 e0 Адаптер Microsoft 6to4 #2
    21...00 00 00 00 00 00 00 e0 Адаптер Microsoft ISATAP #2
    13...00 00 00 00 00 00 00 e0 Teredo Tunneling Pseudo-Interface
    14...00 00 00 00 00 00 00 e0 Адаптер Microsoft 6to4
    19...00 00 00 00 00 00 00 e0 Адаптер Microsoft ISATAP #3
    18...00 00 00 00 00 00 00 e0 Адаптер Microsoft ISATAP
    16...00 00 00 00 00 00 00 e0 Адаптер Microsoft 6to4 #3
    20...00 00 00 00 00 00 00 e0 Адаптер Microsoft ISATAP #4
    ===========================================================================
    IPv4 таблица маршрута
    ===========================================================================
    Активные маршруты:
    Сетевой адрес Маска сети Адрес шлюза Интерфейс Метрика
    0.0.0.0 0.0.0.0 10.1.0.2 10.1.3.42 4501
    0.0.0.0 0.0.0.0 On-link 91.203.178.26 21
    10.0.0.0 255.0.0.0 10.1.0.2 10.1.3.42 4246
    10.1.0.0 255.255.0.0 On-link 10.1.3.42 4501
    10.1.3.42 255.255.255.255 On-link 10.1.3.42 4501
    10.1.255.255 255.255.255.255 On-link 10.1.3.42 4501
    91.203.178.26 255.255.255.255 On-link 91.203.178.26 276
    127.0.0.0 255.0.0.0 On-link 127.0.0.1 4531
    127.0.0.1 255.255.255.255 On-link 127.0.0.1 4531
    127.255.255.255 255.255.255.255 On-link 127.0.0.1 4531
    172.16.2.0 255.255.255.0 10.1.0.2 10.1.3.42 4246
    172.16.2.10 255.255.255.255 10.1.0.2 10.1.3.42 4246
    172.16.5.0 255.255.255.0 10.1.0.2 10.1.3.42 4246
    224.0.0.0 240.0.0.0 On-link 127.0.0.1 4531
    224.0.0.0 240.0.0.0 On-link 10.1.3.42 4502
    224.0.0.0 240.0.0.0 On-link 91.203.178.26 21
    255.255.255.255 255.255.255.255 On-link 127.0.0.1 4531
    255.255.255.255 255.255.255.255 On-link 10.1.3.42 4501
    255.255.255.255 255.255.255.255 On-link 91.203.178.26 276
    ===========================================================================
    Постоянные маршруты:
    Сетевой адрес Маска Адрес шлюза Метрика
    172.16.2.0 255.255.255.0 10.1.0.2 1
    172.16.5.0 255.255.255.0 10.1.0.2 1
    10.0.0.0 255.0.0.0 10.1.0.2 1
    ===========================================================================
    IPv6 таблица маршрута
    ===========================================================================
    Активные маршруты:
    Метрика Сетевой адрес Шлюз
    16 1125 ::/0 2002:c058:6301::c058:6301
    11 276 ::/0 fe80::f88b:9485:52cd:caaf
    1 306 ::1/128 On-link
    13 58 2001::/32 On-link
    13 306 2001:0:4137:9e50:2846:3ccb:a434:4de5/128
    On-link
    16 1025 2002::/16 On-link
    16 281 2002:5bcb:b21a::5bcb:b21a/128
    On-link
    11 276 fe80::/64 On-link
    13 306 fe80::/64 On-link
    13 306 fe80::2846:3ccb:a434:4de5/128
    On-link
    11 276 fe80::b9aa:b87b:89ea:fcdd/128
    On-link
    1 306 ff00::/8 On-link
    13 306 ff00::/8 On-link
    11 276 ff00::/8 On-link
    ===========================================================================
    Постоянные маршруты:
    Отсутствует
    C:\Users\dan>
     2. ipconfig:
    C:\Users\dan>ipconfig /all
    Настройка протокола IP для Windows
    Имя компьютера . . . . . . . . . : druh
    Основной DNS-суффикс . . . . . . :
    Тип узла. . . . . . . . . . . . . : Гибридный
    IP-маршрутизация включена . . . . : Нет
    WINS-прокси включен . . . . . . . : Нет
    Порядок просмотра суффиксов DNS . : mannet.lan
    Адаптер PPP inet:
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : inet
    Физический адрес. . . . . . . . . :
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    IPv4-адрес. . . . . . . . . . . . : 91.203.178.26(Основной)
    Маска подсети . . . . . . . . . . : 255.255.255.255
    Основной шлюз. . . . . . . . . : 0.0.0.0
    DNS-серверы. . . . . . . . . . . : 91.190.68.2
    91.203.177.4
    NetBios через TCP/IP. . . . . . . . : Включен
    Ethernet adapter lanHOME:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : VIA Rhine III Fast Ethernet адаптер
    Физический адрес. . . . . . . . . : 00-40-F4-76-95-BD
    DHCP включен. . . . . . . . . . . : Да
    Автонастройка включена. . . . . . : Да
    Ethernet adapter 1000LAN:
    DNS-суффикс подключения . . . . . : mannet.lan
    Описание. . . . . . . . . . . . . : Realtek RTL8168B/8111B Family PCI-E Gigab
    it Ethernet NIC (NDIS 6.20)
    Физический адрес. . . . . . . . . : 00-1D-92-2C-23-02
    DHCP включен. . . . . . . . . . . : Да
    Автонастройка включена. . . . . . : Да
    Локальный IPv6-адрес канала . . . : fe80::b9aa:b87b:89ea:fcdd%11(Основной)
    IPv4-адрес. . . . . . . . . . . . : 10.1.3.42(Основной)
    Маска подсети . . . . . . . . . . : 255.255.0.0
    Аренда получена. . . . . . . . . . : 17 октября 2009 г. 17:37:10
    Срок аренды истекает. . . . . . . . . . : 18 октября 2009 г. 17:37:09
    Основной шлюз. . . . . . . . . : fe80::f88b:9485:52cd:caaf%11
    10.1.0.2
    DHCP-сервер. . . . . . . . . . . : 172.16.2.2
    IAID DHCPv6 . . . . . . . . . . . : 234888594
    DUID клиента DHCPv6 . . . . . . . : 00-01-00-01-12-43-2B-71-00-1D-92-2C-23-02
    DNS-серверы. . . . . . . . . . . : 172.16.2.2
    Основной WINS-сервер. . . . . . . : 172.16.2.2
    NetBios через TCP/IP. . . . . . . . : Включен
    Туннельный адаптер Reusable Microsoft 6To4 Adapter:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft 6to4 #2
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    Туннельный адаптер Reusable ISATAP Interface {543B8BBE-84D2-471B-B8FB-F73600B4A8
    2B}:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft ISATAP #2
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    Туннельный адаптер Подключение по локальной сети* 5:
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Teredo Tunneling Pseudo-Interface
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    IPv6-адрес. . . . . . . . . . . . : 2001:0:4137:9e50:2846:3ccb:a434:4de5(Осно
    вной)
    Локальный IPv6-адрес канала . . . : fe80::2846:3ccb:a434:4de5%13(Основной)
    Основной шлюз. . . . . . . . . :
    NetBios через TCP/IP. . . . . . . . : Отключен
    Туннельный адаптер 6TO4 Adapter:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft 6to4
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    Туннельный адаптер isatap.mannet.lan:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . : mannet.lan
    Описание. . . . . . . . . . . . . : Адаптер Microsoft ISATAP #3
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    Туннельный адаптер isatap.{AE6B314F-94BA-478F-8751-74DEB5AB1888}:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft ISATAP
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    Туннельный адаптер Подключение по локальной сети*:
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft 6to4 #3
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    IPv6-адрес. . . . . . . . . . . . : 2002:5bcb:b21a::5bcb:b21a(Основной)
    Основной шлюз. . . . . . . . . : 2002:c058:6301::c058:6301
    DNS-серверы. . . . . . . . . . . : 91.190.68.2
    91.203.177.4
    NetBios через TCP/IP. . . . . . . . : Отключен
    Туннельный адаптер isatap.{8B513FFA-4355-406F-B999-1701EE42D5B9}:
    Состояние среды. . . . . . . . : Среда передачи недоступна.
    DNS-суффикс подключения . . . . . :
    Описание. . . . . . . . . . . . . : Адаптер Microsoft ISATAP #4
    Физический адрес. . . . . . . . . : 00-00-00-00-00-00-00-E0
    DHCP включен. . . . . . . . . . . : Нет
    Автонастройка включена. . . . . . : Да
    3. 
    4. without privider's cable:
    4. with provider's cable:

  • Vision Assistant 2013 returns faulty Kernel Sizes

    When playing around with custom kernels in Vision Assistant 2013 I noticed that in the function "Filters" unter "Convolution - Custom", the size shown in the "Kernel Size" Control does not match the size displayed in the matrix view on the right of the control, as displayed in the attached picture #1.
    Sizes larger that 5 grow to 17 instead of 7, 21 instead of 9, 25 instead of 11 and so on. When I hit OK and re-enter into the setup dialog, the Kernel size control shows "17" when I left it before with 7. I can then decrease the size from 17 to 7 and will finally get my 7x7 kernel.
    Also, when using "Convolution Custom", and entering the first "customized" value, the focus in the filter selection list suddenly jumps to "Convolution - Highlight Details". If I decide that this shall not disturb me much and keep on entering the values, all is fine. Until I hit ok and then realize, I might have made a mistake, which makes me doubleclick the Filter Setup again. this results in a total change of the kernel I entered into the standard "Convolution - Highlight Details" with all those -1 s and the high positive number in the middle, see attached pictures #2 and #3.
    Can anyone confirm this odd behavior?
    Attachments:
    20140214_VAS2013#1.png ‏14 KB
    20140214_VAS2013#2.png ‏61 KB
    20140214_VAS2013#3.png ‏75 KB

    Dear comrade,
    I reproduced this issue like you described and it definitely looks like a Bug in the Vision Assistant 2013 Software.
    So I will write a CAR which will be forwarded to R&D.
    Thank you very much for your support
    Kind Regards,
    Oleg Scherling, M.Eng | Applications Engineering | National Instruments | NIG |

  • Problems with the vba command sendkeys in Access 2013.

    Hi, I have problems with the vba command sendkeys in Access 2013.
    I use a text field (no control) in a form. By means of the VBA command SendKeys "+ {DELETE}", 1  I get the text in memory from the text field, then place (paste) the text in a Word document. The command line sendkeys
    is followed by moving the focus to a particular control. In Access 2003, which went smoothly. Now the sendkeys command is executed only in Access 2013 in the control that then receives focus. If the procedure ends with the sendkeys command then it works well.
    The procedure with the sendkeys command is activated by a button on a form.
    Who can help me?
    private user

    Hi H.Plmp,
    Based on the description, you want to call the sendkeys from a button.
    >> use a text field (no control) in a form. By means of the VBA command SendKeys "+ {DELETE}", 1  I get the text in memory from the text field<<
    What did you mean that you no control? And how did you get the text in memory from the text field? Also I am not able to understand the exact reason that the command doesn't works in Access 2013.
    And based on the test, sendkeys "+ {Delete}" works well for me when I set a focus for the textbox control. Here is th test code:
    Me.Controls("txtField1").SetFocus
    SendKeys "+ {DELETE}", 1
    If I misunderstood, please feel free to let me know.
    Regards & Fei
    We are trying to better understand customer views on social support experience, so your participation in this interview project would be greatly appreciated if you have time. Thanks for helping make community forums a great place.
    Click
    HERE to participate the survey.

  • Problem with image thresholding

    Hi guys
    I am trying to threshold 16 bit unsigned images, but at the time I apply threshoding nothing come out . Please let me know if I am doing right way or there is problem with my code.
    I will appreciate your help.
    Attachments:
    threshold_block diagram.jpg ‏59 KB
    THRES_front pannel.jpg ‏151 KB
    THRESHOLD.vi ‏48 KB

    Hi parag77,
    This links specify a little more about your problem and might be usefull in your case.
    http://forums.ni.com/t5/Machine-Vision/Error-1074396077-IMAQ-Vision-Incompatible-image-type-after/m-...
    http://digital.ni.com/public.nsf/allkb/C3C20E427DE91928862570DE00542B7D?OpenDocument
    http://zone.ni.com/reference/en-XX/help/370281U-01/imaqvision/imaqvisionlverrorcodes/
    Hope it helps you.

  • Problem with Advanced Healing Brush on LR 5

    I am experiencing a problem with the Advanced Healing Brush  on LR 5. Using, say, the Heal option, when I have marked the area I want to heal, and then move to the source I want to copy, I cannot get the colour of my selected source point to remove the heal spot.
    I have looked at a number of tutorials and working comments on using this function, and I am pretty sure that I am following the correct procedure for using the Advanced Healing Brush. I have the same problem when I try to use the Cloning option - the colour just does not fill the area I want to fix.
    Is there possibly a bug hanging around?  Can anyone give me any possible solutions to my problem?  I so want to try out this terrific new feature in LR 5.

    Thanks for your response
    I attach 2 screenshots
    Screenshot1  - See the orange tip of a feather behind the kid's head - I want to remove this with the heal option in the Spot Removal tool
    Screenshot2 - I have demarcated a piece of the feather with the spot tool and have chosen the source color of the wooden bench
    I was expecting the demarcated area to have adopted the chosen source color, but it hasn't, by a long shot. I've tried evry which way to get it to work, but no luck.
    I even tried using the Clone option, choosing the same source color and painting over the orange feather, but that also didn't work.
    I look forward to your comments
    Sorry - forgot to add the images
    Message was edited by: mandav1023

  • Problems with Advanced Queueing...

    Hi,
    Has anyone used AQ through OCI?
    We are having problems with OCISubscriptionRegister() the error reported back is 'two task conversions - overflow integer'. The code we are running is straight out of the published demos...
    Please HELP!!
    Thanks,
    Darryl Schallig

    There are discussion on this in the Advanced Queueing forum.

  • Problem with iPhoto 9.3.2: Any assistance or how to re-install?

    Hi all,
    I've got a problem with iPhoto 9.3.2 whereby upon startup, I get a spinning wheel which goes on forever (I've left the computer alone for 5 hrs and coming back, it's still spinning). Symptoms are it pulls up the Library with Events, Photos, Faces, even Albums. But I don't actually get any photos displayed on the right.
    I've tried :-
    1) reinstalling Mountain Lion
    2) using Disk Utility Repair Permissions
    3) Rebuilding thumbnails & library
    4) Creating new library
    5) starting up under a new user account (therefore not a problem with a config file in my homedir)
    6) Starting iPhoto up in Safe Mode
    Note during the spinning wheel, I can access the menu as such I can quit, import photos, etc.
    It actually works (worked) under the following conditions :-
    a) Until 31st July iPhoto 9.3.2 was working fine. I cannot remember what I did/installed/uninstalled/changed/deleted to cause it to have problems
    b) I can get access to the very same library with Aperture with no problems at all (i.e. no library issues)
    c) Photos will only display if I import a picture. Then all Events will appear. I can get into each event, I can even select the photo I imported and delete it. This is my current workaround to accessing photos in iPhoto. However I cannot switch to any of the albums I created. It will only display Events. Clicking on Photos, Faces, Places, and any of the Recent doesn't do anything.
    So finally now is to try to re-install iPhoto 9.3.2. Can any one please let me know how I can do this? I've tried Option-Click on Purchases tab in the App Store but I get a greyed out INSTALLED in iPhoto. Similarly I found a link that provided the iPhoto9.3.2Update.dmg but in attempting to installing that, it tells me it can only be done via the App Store.
    Any advice?
    Cheers

    @putnik, yes I believe the update came from that site.
    @mende1, perfect! That was easy! I thought I had to clean out plist files from my home directory, etc. Upon deletion and reinstallation, iPhoto started working again!
    Thank you so much!

  • Problem with advanced sidechaining in logic x

    probably best i explain what i'm trying to do first!
    I'm bussing 4 different audio tracks to  bus 1 (for example) which has eq / compressor etc and outputs to the master and also to bus 2
    bus 2 has a drastic high cut on it but has no output, it is intended to be used as a silent sidechain effector
    bus 3 has a compressor set to sidechain to bus 2, but instead of sidechaining to the high cut signal, it uses the dry one.
    i used to route the same thing the same way on logic 8 and 9 but on 10 i have encountered this problem, is there a setting i need to change or is it a problem with the softwar that i'll have to add extra busses to route around?
    thanks!

    When you send (or change the output) of a channel to a bus, an Aux is created automatically in the mixer for you.  You can create more Auxes manually and set their inputs to the same bus # if you need to.  For example you could create a submix of 4 gtrs by setting the output to bus 1.  In the mixer you will now have an Aux with bus 1 as its input.  You could create 2 more Auxes with bus 1 as their input.  So you have 3 channels all with the exact same submix of gtrs.  Now you could do some parallel processing:  1 Aux has distortion, 1 Aux is clean, etc.  Say you wanted to filter the crap out of the gtr submix and use it for a sidechain on something else, you take your third Aux and put your crazy filter on it, and change its output to Bus 2.  A new Aux is created, but you can either delete it or change its output to none.  Whatever you want to listen to the crazy filter in its sidechain listens to bus 2, where that filter has been sent (from the 3rd aux above).
    I'm bussing 4 different audio tracks to  bus 1 (for example) which has eq / compressor etc and outputs to the master and also to bus 2
    Does the AUX output to Master & send to Bus 2?  or is it bad grammar (not criticizing) and you mean the 4 audio tracks also go to Bus 2?
    bus 2 (You mean an AUX with bus 2 as its input) has a drastic high cut on it but has no output, it is intended to be used as a silent sidechain effector
    OK so far
    bus 3 has a compressor set to sidechain to bus 2, but instead of sidechaining to the high cut signal, it uses the dry one.
    I think what you want is to set the output of the AUX with the high cut to an open bus, and use that as the sidechain for whatever is on what "bus 3" is.
    Maybe think of it like this:  busses are pathways, they can even be potential pathways.  If you send something to a bus it's like you plugged in a cord to a jack.  The other end of that cord might not be plugged into anything.  Or it might be plugged into the sidechain of a compressor, or an AUX channel.  In Logic's virtual world, that cord is like a Y jack with, like, 250 ends on it.  If you plug one of those ends into a filter, and one into an AUX channel, and one into a sidechain, the sidechain isn't going to hear what came from the filter.   You need another cord (bus) to come out of the filter and then into the sidechain.
    If you want to be 'clean and conservative' with your busses (and who doesn't in this economy?), you could skip the intermediate bus/channel.  Unless the compression is critical to the silent sidechain signal, you could skip sending the compressed signal to bus 2.  Change the input of the drastic high cut channel to Bus 1, and set its output to Bus 2,  Make sure no AUXes are there with Bus 2 as their input.

  • Problems with black color creating PDF from Excel 2013

    Hi
    I want to create PDF from Excel 2013 using Adobe Acrobat distiller. When I do so I have problem with black which is 300 percent black instead of 100K. Problem is only with drawed lines from Excel's symbol menu (text is fine).
    I use Acrobat 9 Pro and few days ago switched from Office 2000 to 2013.....and problems started.
    Office 2000 and Acrobat 9 Pro worked ok and everything including drawed symbols like lines or squares were 100K black in PDF. Now in the case of Excel 2013 it is not so but when I use export function included in Excel and export file to PDF output is 100K. I am not sure if this kind of export is good for me. I use PDF's as inputs to InDesign. So I would like to have more control when create PDF from excel. But in Excel setup I can choose standard and minimum size and that is all.
    I imported Adobe PDF Settings from computer where these settings worked with Excel 2010 but they do not do same job for me in combination with Excel 2013 (I have 300 black).
    To explain my workflow:
    I am a publisher of crossword magazine and I use Excel to write crosswords. Then I create PDF's and import them into InDesign.   
    Should I use internal export included in Excel 2013? Are these PDF's ok for working in InDesign and finally for print purposes?
    Or If you have any suggestion how to create PDF via Adobe Acrobat distiller setup (which I prefer) please give me advice.
    Thank you in advance
    Juraj

    These are tough issues I suspect and the answer you need is probably trivial (it is finding it). I will make some suggestions, but I have no idea if any will work.
    -  In the printer under the Properties>Layout>Advanced>print quality, I use 300 dpi (default is 1200).
    - Under the printer preferences>settings>edit the settings file, I use again 300 dpi (default is 1200 dpi).
    These may not fix your issue, but migth be worth a try. You could also try using the preflight script to change to gray-scale. It may not make any changes, but again may be worth a try.
    All I am trying to suggest are the things I would be looking at. Maybe they will lead you someplace. Others may be by with better suggestions in time.

  • Problem with URL:s containing scandinavian characters (Office 2013 / Firefox / SP 2010)

    We are encountering problems with document opening when URL contains scandinavian characters (ä,ö). This issue occurs only with Office 2013 when using Firefox. With IE it works just fine. And Office 2010 works nicely as well.
    Error message: "We can't connect to 'https://intra.../Test powerpoint.pptx' Please make sure you're using the correct web address".
    This issue only occurs when document or document set name contains scandinavian characters. If I rename it, document opens without any problems. 
    Any idea how I could fix this?

    Hi Law, try changing the settings in the library to "open in client application." You should also read the following links for solutions:
    http://sharepoint.stackexchange.com/questions/16938/why-is-word-document-created-from-template-saved-locally-instead-of-to-the-docu
    http://social.technet.microsoft.com/Forums/sharepoint/en-US/ec048a1f-e6cc-481d-8f46-308823568b56/cannot-save-documents-to-sharepoint?forum=sharepointadminprevious
    cameron rautmann

  • Problem with 'Advance' entry

    Griends,
    I am getting this strange problem. Does it happens with you also? Any solution..
    The problem is as follows;
    1. Transaction PR05
    2. Enter only advance and save it with status Request Recorded/To Be Settled.
    3. OPen the trip again, Simulate, the advance is visiable.
    4. Click the 'Advance'  tab.
    5. As soon as you open the advance tab, the advance amount is grayed out (disabled)
    6. Simulate now and the advance is shown as 0.
    You see, only clickig the 'Advance' tab cause the adance to be ignored.
    Regars!

    Thanks Raynad,
    The poblem is solved now. Actually the problem was due the field 'Settlement amount' in Advance tab, which I made hidden. Upon clicking the 'Advance' tab, the system was moving 0 to this field. When I made this field visiable (still no input allowed), it started getting the corect valuse. I still feel it's a bug. Even if the field is not visible, it should contain coect value.
    Regads...

  • Problem with Advanced queueing

    Hi,
    I'm having the folowing problem:
    I'm trying to implement the Monitor feature from within VB. I used the sample code posted in: re:Rollback segments and odbc and VC++
    I created the class with the method NotifyMe, just as described.
    When I try to start the monitor, I get the following VB-error:
    Runtime error: -2147213295 (80042011)
    Method 'MonitorStart' of object '_IOraAQ' failed.
    I can't seem to find the problem, am I doing something wrong?
    B.t.w., normal sending and receiving works fine.
    We are using oracle database 8.1.7.0.0
    CLient: 9.0.1.0.1
    OO4O: 9.0.1.4.3
    Thanks in advance for your help!
    Kind regards,
    Leon

    Did you grant execute on DBMS_AQJMS and DBMS_AQIN?

  • Problem with advanced action and a counter.

    Hi!
    I got a problem on a project with a counter and advanced action.
    I got 7 words (3 wrongs and 4 rights), each associate with captions that shows on click
    When the students had clicked on the 4 rights (and their captions are shown), the 3 wrongs captions are shown.
    variable:
    v_count
    on wrong words:
    simple action: on click, show captions
    on good ones:
    advanced conditional actions:
    1) IF 1=1, THEN show > caption and v_count = v_count + 1
    2) IF v_count=4, THEN show> all captions of the wrongs
    The problem is: (Only when seen in browser) when I click on 3 rights (and not 4), all captions are shown! I have tried to put v_count=5, but then nothing happens.
    Really, I'm lost! Have an idea of what could cause that problem?
    Thanks
    Laurie-Anne

    You already posted this question, thought I asked you if I could have access to your file, because with this explanation only it is a bit difficult to answer. But I cannot find the thread anymore...
    Maybe I'm wrong, or you did not see this answer?
    Lilybiri

  • Mx_internal_uid problem with advanced datagrid

    I have a XML structure like this:
    <Projekti ...attributes>
    <Tehtava ....attributes />
    <Tehtava ...attributes />
    </Projekti>
    <Projekti...>
    ...etc
    And I'm populating advanced datagrid with this hierarchial
    data, so I can get a tree view in datagrid where there is Projekti
    -nodes and inside them is the Tehtava-nodes.
    Problem:
    When I bind the XML, and the hierarchialdata (that is created
    from the XML) flex automatically adds
    <mx_internal_uid>a324235351514......</mx_internal_uid>
    -nodes inside of the Projekti-nodes. And this causes empty folders
    added in a tree view when user clicks them.
    Is there a simple way to:
    a) prevent thos uid-nodes being created?
    b) filter thos <mx_internal_uid>-nodes out from the
    dataprovider of the advanced datagrid?

    I know this is way after-the-fact, but I just ran into this problem myself today. I found that it generated the mx_internal_uid if the node I was selecting was an XMLList instead of XML. A simple cast fixed the problem.