Pxi-4071 Calibration
Hello,
Could you please let us know what the accuracy of the NI 4071 card will be if the Fluke 5700 calibrator is used instead of the Fluke 5720A? For example, will the accuracy of the 7 ½ digit DMM be reduced to 5 ½ or 4 ½ digits if the Fluke 5700 is used for calibration?
Thanks
Dan
Hi Dan,
Using the 5700 in place of the 5720A will not affect the digits of resolution of the DMM. However, because the 5700 is less accurate, the DMM may not be calibrated to its specified accuracy. This will depend on where the calibrator accuracy falls within its range, and as such, cannot be determined absolutely. Furthermore, the calibration may fail if the calibrator returns a ppm accuracy measurement outside of the range expected for the 5720.
Hope this helps.
David
Applications Engineer
National Instruments
Digital Multimeters
Similar Messages
-
PXI-4071 Calibration Date Update
Is there a process to have the internal Calibration Date update [using the Cal Executive software] without running the adjustment-reverify routines?
Hi sbdata,
In general, unless an actual adjustment is performed, then the EEPROM is not written to and the calibration dates will remain the same. That said, if you open an external calibration and then close it, the calibrated date will be set to the current system time.
Jeff Munn
Applications Engineer
National Instruments -
Dear sir,
Nice to meet you.
When I running the PXIe-4141 Calibration Function "niDCPower Cal Adjust Voltage Leve.vi" in niDCpower version 1.8,
it will show an error "calibration function is not supported by the device."(false cae)
But if I run the PXIe-4141 Calibration Function "niDCPower Cal Adjust Voltage Measurement.vi", (true case)
it will work.
See the attachments.
My Equiptment list:
WinXP SP3
Labview 2011
niDMM 3.0.5
niDCpower 1.8
PXI-1073 case with 2 cards.
slot2: PXIe-4141
slot6: PXI-4071
Could someone tell me the solution?
Solved!
Go to Solution.
Attachments:
4141-cal adjust voltage level.JPG 203 KB
4141 manual cal_20130722.vi 46 KBUpdate:
Error Code: xBFFA4A42
niDCPower Cal Adjust Voltage Level.vi<ERR>Requested calibration function is not supported by the device.
Status Code: -200445
See "voltagle level error.JPG"
And the same error for "niDCPower Cal Adjust Current Limit.vi".
See "current limit error.JPG"
Attachments:
voltagle level error.JPG 15 KB
current limit error.JPG 16 KB -
Pxi-4071 error on calibratio​n executive
hi!
i have this little problem with my NI calibration executive. The error says
" Error-1071001424 occurred at Error-1071001424 occurred at ***configcalibratorandwait.vi
Possible reason(s)
Current Compliant Voltage Exceeded".
Can you guys help me on this? I am calibrating this NI DMM PXI-4071. Please see attached captured error.
thanks.
atingpopoi,
Thanks for posting on the NI forums. I have a couple questions for you that should help us get to the bottom of the problem.
What version of Calibration Executive are you using? You should be able to find this by going to the Help Menu >> About Calibration Executive.
What Equipment are you using to perform the calibration? There is a list of equipment here.
Is this the only board that you are having difficulty calibrating with Calibration Executive?
Thanks!
Aaron W.
National Instruments
CLA, CTA and CPI -
Hi,
I am using NI LabVIEW, NI PXI 4071, and NI PXI 6070E to measure current through a variable resistance. Now, I am using one channel from SCB-68, but I want to add another channel in parallel so that I can have two resistors instead of one that I cam measure current through them.
I have attached a Pdf file showing the setup for hardware in use and LabVIEW code also.
Can anyone look at these files and give me guide lines or ideas that can help me resolving this issue, please.
Thanks in advance.
Best Regards,
Shaheen.
Solved!
Go to Solution.
Attachments:
IV copy for HS.vi 55 KB
Layout of NI Cards.pdf 248 KBYour 4071 can only do one measurement at a time. Your DAQ cannot measure resistance either nor has it got any analogue inputs.
However, you could use a multiplexer and multiplex your 4071 DMM. This wont give you simultaneous measurements but can acquire data one after the other, the speed is obviously dependent on the multiplexer you choose!
Hope this helps.
Beginner? Try LabVIEW Basics
Sharing bits of code? Try Snippets or LAVA Code Capture Tool
Have you tried Quick Drop?, Visit QD Community. -
We use 4110 voltage source for current measurements on pxi-4071. We supply high-ohm (1.5 GigaOhm) resistor with few volts and measure current on 4071. All cables are shielded on ground of 4110 voltage source, but cables are about 5 meters long. What can be reason for constant current 5 nA at zero supply voltage? We don't know whether it software or hardware problem, and i already asked this question in LabView forum, autozero enabled in max doesn't help. Could be a reason bad shielding of ground in pxi-4110 voltage source
I would focus on the hardware side initially. I would try the following steps.
1) Make a current measurement on the 4071 with the inputs open to see if there is an offset. If so we should look into the DMM configuration.
2) Connect your cabling and/or resistor without the 4110. If this adds the offset then it seems to be shielding/noise related.
3) Connect the cabling and/or resistor to the 4110 with the output disabled. If this adds the offset then it could be ground related or also related to an offset in the 4110 output.
Steve B -
Setting Input Resistance on the PXI-4071
The specifications for the PXI-4071 DMM indicate that the input resistance can be chosen as either 10 MOhm or 10 GOhm for the 100 mV, 1 V, and 10 V ranges. I see where this can be input on the instrument soft panel, but I have been unable to locate how to configure the input resistance when using the DMM in a VI. Suggestions?
Thanks!
Jeff
Jeffrey Zola
Solved!
Go to Solution.You need to use a property node.
There are only two ways to tell somebody thanks: Kudos and Marked Solutions
Unofficial Forum Rules and Guidelines
Attachments:
input resistance.png 13 KB -
Pxi-4071 aux I/O how to output a control signal?
HI, I am working on Pxi-4071 dmm card. I want to use aux I/O to out put a control signal for a relay on/off. Technically, I want to know how to do it with labview. Is any examples to start?
Thanks
LimingHello Liming,
The DMM is capable of sending TTL triggers out on both the PXI_Trig lines and on the AUX connector on the front connection. These signals are intended for controlling National Instruments switch modules and not powering/closing third party relays.
If you're looking to control a National Instruments Switch there are example programs that demonstrate synchronous scanning with DMMs and Switches. To locate these examples open LabVIEW and go to Help» Find Exampled... Then Navigate to Hardware Input and Output» Modular Instruments» NI-Switch (Switches).
If this isn't what you're looking to do, or if you have additional questions, please feel free to post back with more information.
Have a great weekend!
Brooks -
Error -1074118625 with PXI-4071 and PXI-2527
When using the LabVIEW IVI drivers for PXI-4071 and PXI-2527, I receive an error -1074118625 in TestStand. The sequence initializes the MUX, init DMM, connects the MUX, waits for debounce, and then upon reading from the DMM, I receive this error.
Error: niDMM Read Waveform (waveform data).viExplanation could not be found for the requested status code.
Verify that the requested status code is correct.
[Error Code: -1074118625, User-defined error code.]
This sequence of events is successfully used multiple times elsewhere in the TestStand. This error does not appear in any knowledgebase article nor in any help. Any explanation would be greatly appreciated.
Solved!
Go to Solution.I found this with google that seems like it might apply:
http://digital.ni.com/public.nsf/allkb/A593DEBFD86A69C68625727900748EEC -
Measure Power using Labview and two PXI-4071 DMMs
Hello,
I am trying to make a simple VI. I am measuring input AC voltage and Current using two PXI-4071 DMMs. I want to Calculate Power and Power Factor. I am able to read the voltage and current fine. I just dont know how to calculate the power and power factor. Im looking for real power not watt*amps. Please Help.
Scott
Attached is my VI
Attachments:
Measure AC Voltage and Current.vi 42 KBScott,
It appears that another applications engineer (Sean_N) has been assigned your initial forum post. I assure you that he is working on finding an answer to your question. We would like for our customers to limit questions to one forum so that we do not duplicate efforts on our end. I will contact Sean to remind him that you are awaiting a response. I will also point all future traffic from this post to your initial one.
Thanks,
Kareem W.
National Instruments
Web Product Manager -
Hi, we use pxi-4071 on pxi-1033 for measurments and we connected shielded cables few meters long. We wrapped foil cables from the kit. After that we connected all the shields to the ground from the power supply. But we still remain interferences about 1nA and DC component 5nA. If anyone faced a similar problem, is it possible that we are too often measured current. Can you advise the shilded cables for multimeter in this case.
alyast wrote:
Yes, we pretty sure in shielding, but we don't know why there is current constant component (5nA), while declared sensitivity - 0.1nA. Probably some bug with timing in LabView or something? I don't know what could be a reason for this behavior, and therefore ask noob questions.
Why would you suspect the software when all the evidence points to an issue with the HARDWARE? And if you're measuring in the nanoAmp region, you'd better be ABSOLUTLEY certain that your grounds are true and accurate. Probably grounding to your power supply is not good enough. Your power supply probably has to be grounded to a copper stake going - what is, it, six feet? - into the ground.
Bill
(Mid-Level minion.)
My support system ensures that I don't look totally incompetent.
Proud to say that I've progressed beyond knowing just enough to be dangerous. I now know enough to know that I have no clue about anything at all. -
Why is it so hard to find the ranges for the PXI-4071??????? I called National and they told me:
The ranges that are possible to use are:
+-.02
+-.20
+-2
+-25
+-250
Yet when I open the soft front panel there is a range up to 1000 volts. Where is this listed?? I need ranges for VDC, current dc and resistance ( opens...infinite) I'm using the DMM drivers that ships w LV 8.6.
ThanksAll of the range specifications will be found in the PXI-4071 user guide and specifications document.
If you search on ni.com for 4071 you will find the PXI-4071 product page. Going to the specifications tab, and then clicking "detailed specifications" opens this document. It contains all of the specifications and accuracy ratings of the device.
NI 4071 Specifications (Multilingual)
Page two specifies that for DC voltage the following are valid ranges:
100mV
1V
10V
100V
1000V
which also show up in the DMM soft front panel.
For DC current, see page 4, which states
1uA
10uA
100uA
1mA
10mA
100mA
1A
3A
Resistance, page 5,
100Ohm
1kOhm
10k
100k
1M
10M
30M
100M
5G
Thse pages also have the charts for the accuracy of the DMM at each one of these ratings.
Eric S.
AE Specialist | Global Support
National Instruments -
I have a PXI-4071 DMM and I am measuring a connection in the UUT. The value should be around 0.3 ohms. The program (exe) will run aprox. 20 times and during these 20 runs the measured value will progressively increase until is is over the limit of 1.5 ohms. If I shut down the exe and run the VI as a standalone (continuous loop), the resistance will gradually come down to 0.3 ohms. I then close LabView and restart my exe and the readings are normal (0.3) but then continue to increase until they are over 1.5 ohms. This process has been repeatable for the last week. I have attached the code I am running. What could be the problem with the DMM code?
Attachments:
DMM_MeasResistance2.vi 64 KBUninitialized shift registers are a good thing to check for such a problem. Do you see the same response with a soft panel and your DMM? if not then it is in code. Uninitialized shift registers would start at 0 (default is usually 0) and will retain their values from run to run. Just a possibility dont have LV up right now so I cant check the code sorry,
Paul
Paul Falkenstein
Coleman Technologies Inc.
CLA, CPI, AIA-Vision
Labview 4.0- 2013, RT, Vision, FPGA -
你好:
我有一张PXI-4071的DMM ,用MAX的NI-DMM SOFT FRONT PANEL程序时,显示OVLD V DC,正常的都是.显示数值0.00000000 V DC,请问这是什么引起的?是这张卡坏了吗?DMM线都没接,没有量测动作的时情况下,打开程序就直接OVLD,如下图所示
附件:
PXI DMM-4071.PNG 73 KB -
Setting up the sample rate on a PXI-4071
I am new to this and hoping that some one can help...
I have a DC power supply connected to a PXI-4071 that I am attempting to get some stable DC voltage measurements using LabView but
the measeurement values are all over the board so I am guessing that I am missing some sort of sample rate set up.
Any suggestions?
Thanks
Attachments:
My VoltageRead.vi 30 KBI figured it out..
I had my constant "2" wired to the function instead of the resolution...
Maybe you are looking for
-
Best Practises for File Sharing and Print Server on Windows Server 2012
Is it better to have File Sharing and Print Server installed on the same Windows 2012 Server ? Or, they should be installed on 2 different Windows Servers ? Hussien Sherief
-
Sync units with different Apple ID?
Our family's iPad is linked to my AppleID, and our MacBook Air is linked to my my wife's Apple ID. Is it at all possible that GarageBand on both units could sync through the same iCloud (mine)?
-
I have had no choice but tok reconfigure my computer and couldn't back up my itunes, how do i do this in new zealand, i have downloaded i cloud but this does not seem to work due to being outside the USA.
-
Ipod freezes in middle of loading itunes library
After reformatting the hard drive on my Powerbook G4 drive (per Apple instruction), I reconnected my ipod clickwheel and it hasn't been working since (using Tiger 10.4 and latest version of ipod updater), mostly showing the folder & exclamation point
-
Dear Friends, I have prob in Reporting total Stock o a given day. the Pro is We intialised the data using 2LIS_03_BX and the movements through 2LIS_03_BF and we are not using revaluations UM. Now for Ex the query is asking for inputs Material, Plant(