PXI-4462 problem...

Hi All!
I have PXI-1036 DC with 8101RT+6653+6682+4462.
It seems thereis a problem with PXI-4462 driver:
I format disk and install fresh system software (step by step). When I install DAQmx 9.7 with DSA, after reboot I see message after "Welcome to LabVIEW Real-Time 12.0f4" :
ing the driver might fix the issue. Ohterwise, contact National Instruments technical support."
PXI-4462 appears in MAX with some duplicated device (Dup.jpg). When I call Test Panel, I see error (see TestPanel_Error). After "Yes" and press "Start" I see error as on TestPanel_Opened.jpg.
What's happened?
Jury
Attachments:
Dup.JPG ‏176 KB
TestPanel_Error.JPG ‏188 KB
TestPanel_Opened.JPG ‏198 KB

Hi Jury,
Have you tried to Reset the device in MAX?  I assume you've already tried rebooting the system, right?  What about moving the device to a different PXI slot?  Or even a different PXI system, if available?
If none of those steps work, try deleting the device from the Windows Device Manager, and allow the operating system to redetect it and install a fresh version of device drivers.
See this article, for reference.  If none of these steps yield any change in behavior, it may be necessary to send the hardware in for repair.
Kyle B  |  Product Support Engineer  |  ni.com/support

Similar Messages

  • PXI-4496 vs PXI-4462 resolution

    Is it possible that for a same voltange range, the PXI-4462 card has a lower resolution that the PXI-4496, even if they are both 24-bit?

    Both cards are 24-bits.
    Using the PXI-4496, on a +/-1 Volt range, the 24 bits seems to be dictributed over à 2.2 V range (-1.1 to +1.1).
    Using the PXI-4462, on the equivalent +/-1 Volt range, the 24 bits seems to be distributes over a 3.8 V range (-1.9 to +1.9).
    By resolution, I mean the smalles voltage increment we can measure, i.e. Q= (Vrangemax-Vrangemin)/(2^24).
     

  • PXI-4462 - Radio Sound Environment

    Hi, everyone.
    I have a PXI-4462 audio card.
    Is it possible to use the card to analyze and measure the Sound Environment Settings (eg, Jazz, Rock, Classical, and etc) of a radio audio outputs?
    20Hz and 20kHz is the range of the interest frequency for plotting a graph of Gain (dBr) vs freq.
    As I know, I may need to sweep the freq range of 20Hz to 20kHz for obtaining the curve.
    Hope someone can help to advise.
    Attached is the example of the EQ measurement in graph.
    Thanks in advance. 
    Attachments:
    800px-Peaking-eq.svg.png ‏20 KB

    Hi Karthik,
    Can you share the image of the noise signal if you have?
    One possible cause i can think of is the High frequency noise source nearby. Have you analyzed the Noise characteristic? You can perform simple FFT to identify the noise source by switching ON and OFF of the nearby lights or machines etc...
    One more way is to move the measurement setup out of the current location.
    I guess there shall be a better way of identifying the Microphone connectivity rather than comparing the floating signal RMS with the sound pressure threshold level for identification. 
    Thanks,
    Vijay.

  • PXI 5610 problem

    When running the PXI-5610 adjustment procedure in Calibration Executive , the main power suddenly shut off . then what problem will happen to the PXI 5610?

    Thanks a lot for your reply.
    After the PXI controller 's suddenly power off when performing adjustment ,  we restart the OS and found the PXI 5610's error infomation in the MAX : Error -200437,External calibration constants are invalid,and he card cannot be used in our Auto-Test-System anymore. I guess the calibration data in the EEPROM was missed or uncompleted,but i have no relevant softwre tools to read or write PXI-5610's EEPROM.
    Could you give me some suggestions to solve this BIG problem? Thanks.

  • PXI 6220 - problems with the counter

    Hello all,
    I try to make a buffered counting with counter0 of my PXI card 6220, to count the speed of a motor through a photocell and a sticker over the pulley of the motor.
    I have wired my photocell at PFI8 (source) and I generate a pulse train that I have wired in PFI9.
    The input to the counter is a pulse train 5V high level - 0V low level.
    The problem is that the counter counts sometimes more than 1 pulse for every turn of the motor. However, apparently the pulse train looks ok.
    I don´t know what´s the problem.
    The rising slope of the pulses looks very nice so, I think I could discard it.
    I can suppose that in between pulses there are some glitches at a very high frequency that I cannot see (I have sampled this signal only at 800kHz and at this frequency there are no glitches), but maybe the counter yes, because it works at 20MHz.
    Somebody has experienced this problem. Sometimes ago, using a PCI6221, I solved a similar problem by activating the digital filtering, but it seems it is not possible to activate for this cards. Could somebody confirm it?
    Any ideas will be appreciated.
    Thank you very much in advance!!

    Pete;
    I agree that is a little strange.We don't have such problem reported on our Data Bases.
    A workaround you can try is to make a sinle pulse goes to the source, before start the buffered acquisition, and see if that makes the counter keeps reporting a 1 value.
    You can also try the same thing on another counter of the board to see if that problem is related to the counter you are using.
    Regards
    Filipe A.
    Applications Engineer
    National Instruments

  • PXI-4472 Problem: Several channels always oscillating between +/- 10 V

    Hello,
    Hoping that someone has seen a similar problem. This just recently came up. Observable when running LabView VI's as well as in MAX test panels.
    Any ideas on the cause? Any ideas on the fix?
    Would appreciate your help,
    Hunter

    Hello, Shea
    To answer your questions:
    Yes, we see this on top as well as when no signals are connected.
    We are looking at voltages from eddy current prox probes in the 0 to -5 V range, using DC coupling.
    Yes, as I stated earlier, this oscillation are seen both in LabView and in MAX.
    This board is mounted in a PXI-1002 chassis. The chassis incorporates a MXI-3 board and three 4472 boards.
    This problem board is located in the master slot beside the MXI-3. The other two 4472 boards seem to be operating fine.
    Look forward to hearing your thoughts ASAP.
    Thanks,
    Hunter

  • Pxi-6704 Problem

    If I short the DAC outputs ( Analog voltage )on the Pxi-6704 ,, Is there any problem ? Because i see that the device does not respond for any further voltage changes. But when i do a resource test in MAX it passes..Please suggest.

    Hello;
    It's generally not a good idea to short the output lines anyway.
    If you need the output channels to output a steady zero voltage, the best way to accomplish that is to insert a pull down resistor to ground instead.
    Hope this helps.
    Filipe A.
    Applications Engineer
    National Instruments

  • PXI-4462 settings...

    Hi All!
    Is it possible:
    1. Set gain for particular channel using DAQmx Channel -> AI.Gain properties instead max/min voltage?
    2. Set DC offset for particular channel using DAQmx Channel -> AI.DCOffset properties? What units of this offset?
    3. What properties control LP analog filter? Or this filter always ON? May be this filter controlled by writing F/T to the AI.Lowpass.Enable properties?
    Jury

    Hi Jury,
    Yes.  This article should be helpful.
    DSA modules like the 4462 will remove a DC offset automatically when AC coupling is selected.  You'll select AI.Coupling as the property to write in your DAQmx Channel property node.  You will not be able to enter a value for the offset.  Related info, here.
    The anti-aliasing filters on DSA devices are not configurable through a property node.  They automatically adjust with sampling rate to follow the Nyquist frequency.  The AI.Lowpass.Enable is not a property supported by these modules.
    Kyle B  |  Product Support Engineer  |  ni.com/support

  • PXI-4462 actual sample rate and delay...

    Hi All!
    Is the DAQmx properties, which give me actual sample rate and filter delay? Or I must manualy calculate this values, as described in DSA manual?
    Jury

    Hi Jury,
    I am not sure if I fully understand your question, but there are DAQmx properties that do return the actual sample rate, but there is not one for filter delay. If you place a DAQmx Timing property node down, the actual sample rate will be returned by selecting Sample Clock » Rate.
    Regards,
    Jason D
    Applications Engineer
    National Instruments

  • Pxi mx2 problem with vt1432

    Hello,
    I have a PCI MX-2 card installed on Windows XP SP3 on a Dell precision T3400.  When I run resman it detects the cards.  If i run vt1432 afterwards, I get an error VI_ERROR_SYSTEM_ERROR.  
    I installed the card on another machine with the same OS and it worked.  Any ideas?  I implemented the latest drivers. version 3.6.0.0.
    thanks in advance.

    Hello,
    The other computer is it a DELL?
    Regards,
    Nacer M. | Certified LabVIEW Architecte

  • Conductanc​e measuremen​t using NI PXIe 4492

    Hello,
    I am applying AC voltage using NI PXI 4462 card and measuring the voltage output from my device using NI PXIe 4492. I use Lockin program to measure the amplitude voltage of the signal acquired. I need to convert this output voltage/amplitude into conductance but don't know how.
    Any assistance to solve this problem is appreciated.
    Thanks,
    Mangesh

    Hello Mangesh,
    The 4496 is a 16 channel AI module and the 4462 is a 4 channel AI module, so you will need a module to source the signal. Where is the AC signal coming from? You could set one of the channels to measure current with a shunt resistor and another channel to measure voltage. If you have those two measurments synchronized, you can calculate conductivity from there (assuming the modules have enough accuracy for the application). The delta-sigma ADCs of the 4492 and 4462 may not be ideal for measuring conductance, but will work for the lock-in amplifier.
    A DMM would be the best in terms of DC accuracy. If you're just looking for DC conductance, you would use a DMM to measure resistance and take the inverse for conductance. 
    Jake H | Applications Engineer | National Instruments

  • How do I get started with PXI timing?

    I'm working in a lab that is trying to start to use a National Instruments PXI/VXI hybrid system.  I currently know very little about timing and triggering.  Basically all I know is that the timers we are using (PXI-6653) generate a square wave at 10 MHz and I can route it to the CLK10_in.  I'm able to confirm that the clock is being routed there by using some of the example VIs and using a scope to see that the square wave is being outputted through the BNC connector on the back.  I can see it's not the regular PXI 10 MHz reference clock because there is a difference in 10 MHz signals between before I hit the "GO" button on the routing VI and after I hit the "GO" button.  So all in all, I can confirm I'm getting the 10 MHz signal routed to the CLK10 backplane signal.
    However, my main question is how I can use that CLK10 signal on my DAQ cards?  We are using PXI-4462 and PXI-6123 cards.  I'm trying to use some of the example VIs given with NI-DAQmx.  Is the "onboardclock" a timer that's on the DAQ card itsself?  What about this sampleclock?
    I've read through the examples and tutorials on the NI website, and they aren't very "newbie-friendly". 
    Thanks,
    Michael 

    Hello Michael,
    Thanks for posting on the forums!  What you are trying to do is called reference clock synchronization.  This is also referred to as Phase Lock Looping (PLL) to a reference clock.  You will access this reference clock with a DAQmx property node as shown below.
    Your PXI-4462 is a DSA device.  This needs to be the Master device.  Take a look at the Synchronization examples that ship with LabVIEW.  Launch the NI Example Finder by going to Help»Find Examples.  Browse by task to Hardware Input and Output»DAQmx»Synchronization»Multi-Device.  The image above comes straight from the Multi-Device Sync-AI and AO-Shared Timebase & Trig-DSA.vi  example.  You can modify this example to suit your application.  But remember, the DSA device needs to be the master.
    You may be wondering what the Sync Pulse is for.  In addition the the NI Example Finder, the NI-DAQmx Help is a great resource.  Do a search for Sync Pulse, and you will find quite a bit of information about it as well as other useful things to know and consider with DSA synchronization.
    Message Edited by RT4CY on 10-25-2008 03:06 PM
    Rod T.
    Attachments:
    refclock.jpg ‏110 KB

  • Function control of the PXI system

    Hi All!
    We order  NI PXI-1036DC + PXI-4462 + PXI-6682 GPS + PXI-6653 + PXI-8101.
    Is it possible to control:
    1. DC voltage and current?
    2. GPS coordinates of the system?
    3. Accuracy (or quality) of the 10 MHz clock disciplining by the 6682 GPS? It may be accuracy of the GPS positioning or number of satellites.
    4. Temperature? It may be processor temperature.
    Jury

    Hello Jury,
    Thank you for bringing this issue to the forums.
    Regarding your questions:
    1. What do you want to control the DC voltage and current of? Which card exactly?
    2. I am not exactly sure what you mean by controlling the GPS coordinates of the system. Can you please explain further? 
    3. You can find the specifications of 10 MHz clock and accuracy of the GPS on page A-5 and A-6 of the PXI-6682 GPS Manual: http://www.ni.com/pdf/manuals/372292b.pdf
    The 10 MHz clock accuracy is +/- 1 ppm, you cannot change this value. The accuracy of the GPS is within 15 ns, you cannot change this value as well. Once again all this information can be found on page A-5 and A-6 of the PXI-6682 GPS Manual. 
    About the number of satellites, the more satellites available to the receiver, the more precisely it can determine time and location. Therefore, the location of the antenna should be such that it will receive signals from the greatest number of satellites possible.
    4. The +/- 1 ppm accuracy of PXI-6682 is specified in the range of 0 to 55 C.
    Thank you,
    Vimal Fernandez
    Applications Engineer
    National Instruments

  • PXI 4472 DAQmx Start VI taking a long time.

    I have an issue with DAQmx Start taking up to a secoond to to execute and it's killing my cycle time.
    I have a PXI 8360 with 3 PXI 4462's and 3 PXI 4472's.  I configure DAQ Task and Commit at the begining of the sequence however every time i do a DAQmx Start it takes about a second to execute.  I need to do about 5 in each Test sequence so it's eating up about 5 seconds of cycle time.  I am using DAQmx 9.7.5 drivers and Labview 2013.  We have the exact same hardware and software setup in another location but we are not seeing this issue there.
    Any ideas what might be causing this would be appreciated.
    Attachments:
    DAMmx Start.png ‏4 KB

    Hi Nathan,
    I verified this by disabling the Start Task VI.
    I've attached the commit code.
    Thanks
    Attachments:
    PXI 4472 Acquire.png ‏8 KB
    PXI 4472 and 4462 Commit.png ‏13 KB

  • PXI-8106 shuts down shortly after boot, the power light is flashing.

    How can I tell if this is a board power supply problem in the NI PXI 8106 or a PXI crate problem?   The time between shut downs decreases to the point where it will not boot at all.

    Hi dlknies-
         Take a look at the PXI-8106 User Manual on page 3-14.  It says that a blinking Power LED indicates that one of the PXI or onboard supplies is operating outside of the normal limits or is not functioning.  You can attempt the repair yourself or you can have it sent back to National Instruments for an RMA.
         I hope this helps.  Best of luck to you!
    Gary P.
    Applications Engineer
    National Instruments
    Visit ni.com/gettingstarted for step-by-step help in setting up your system.

Maybe you are looking for

  • How to  work with java in linux system

    Hi, java 1.5.0_04 program is not running in linux o.s.But the same program is working fine windows . can any body let me know what environment should i set to run it in linux system. test.java:19: error:Cannot find class "JLabel" [JLS 8] thanks.

  • How to export report to excel and to text file in previewer

    How to export report to excel? And, it takes long time for generate report to rtf or pdf file in previewer. How to speed up the process? or, it's better generate report to text file directly by Run_product? Please help!

  • Can my power book g4 run on leopard????

    hey im trying to update my powerbook, i dont know to much about apple in trying to see if i cant put a new software.. My itunes wont accept my ipod my iphone. I tried updating my itunes and it said my software font allow it.. Just wanna know what kin

  • Photos distorted when imported in Leopard, yet ok in Tiger - why??

    I use a Powermac G5 dual processor (non-intel), and find that any photos I import are distorted when viewed or printed. Yet if I boot from an external HD with OS X 10.4 those same photos are fine, and remain so even after restarting from the Powermac

  • List of company for Consolidate detail in Inter-company

    Hi all, Do you have any idea how i could not see the list of company for Consolidate in inter-company? As the user guide from SAP, there is more field but my own is not.