Pxi 6220 redefining terminal routes

Hello,
In MAX explorer, the pxi 6220 routes tab shows a direct route from PFI6 to the counter 1 gate (ctr 1 to be used for low freq counter) and an indirect route through counter 1 to PFI 4.  But from the terminal diagram for the card, it looks like PFI 4 IS the counter 1 gate.
However, in LabVIEW terminal rerouting VI, I can select PFI 6 as the source and PFI 4 as the destination (there is no option to select counter 1 input as destination), but the counter does not appear to be counting.  At the moment the input on PFI 6 is a "TTL" level square wave, but I don't get any counts.
Does anyone have any experience with this...any suggestions on what to try to determine where the problem is?
Best Regards,
Chris

Hi Chris,
If you are unable to see properties in your property node, I would first
check to see if you are filtering the properties according to the devices in
your system. There is a KnowledgeBase article outlining the property filtering
which can be found here.
Also, if you right-click the section of the property node where the
property is displayed, you should see a selection at the bottom titled “Select
Property…” If you click this, a browser should appear with all the properties
of DAQmx Channel.
If you are unable to select the property by right-clicking on the
property node, you can also navigate to View » Class Browser and select the
DAQmx object library and DAQmx Channel class. Hope this helps,
Daniel S.
National Instruments

Similar Messages

  • DC crosstalk between PXI 6220 channels due to high source impedance

    Dear all,
    I’m trying to digitize 10 single ended signals simultaneously with the PXI 6220. The source impedance for each signal is 1.5meg which should not be problem since the input impedance is >1G. Bias currents are taken in account and give only an offset error (assuming that they are constant). Crosstalk according datasheet is -75db at 100khz which is fairly okay for my app however the source impedance is not indicated for that specification. When I acquire data (Dc voltages) from the PXI6220 i see unexpected high crosstalks (again source impedance 1.5meg). The sample rate is 200Hz and I connected only one AIGND pin to the reference of the system. I verified whether the GND is okay by connecting one input pin directly to 5V (Rout=0) . This channel indicates 5V while the others (Z=1.5meg and grouded) give non -zero value of voltages (2V). DC leakage between channels in the cabling is not verified yet.  
    1)      Does anyone have a clue why I observe so much DC crosstalk at high source impedances.
    2)      Does the PXI6220 digitizer have 16 parallel channel or does it make use of an multiplexer.
    3)      Do I have to connect all AIGND’s to my system’s reference or is just 1 okay.
    Patrick

    Patrick,
    I think the culprit is probably the multiplexer and not the amplifier input impedance. Look at the settling time graph on page 2 of the speicifcations. At 100 ppm error (which several source impedance curves cross) the settling time increases exponentially with the source impedance. 4 us at 1kohm, 6 us at 2 kohm, 13 us at 5 kohm, and 25 us at 10 kohm.  It is probably not a reasonable to extrapolate that data more than two oreders of magnitude, but it is not unreaonable to expect some problems even at your slow sampling rate.
    The user manual has this to say (pg 4-7):
     M Series devices are designed to have fast settling times. However, several
    factors can increase the settling time which decreases the accu racy of you r
    measu rements. To ensu re fast settling times, you  shou ld do the following
    (in order of importance):
    1. Use Low Impedance Sources —To ensu re fast settling times, you r
    signal sou rces shou ld have an impedance of <1 kΩ . Large sou rce
    impedances increase the settling time of the NI-PGIA, and so decrease
    the accu racy at fast scanning rates.
    Settling times increase when scanning high-impedance signals du e to
    a phenomenon called charge injection. Mu ltiplexers contain switches,
    u su ally made of switched capacitors. When one of the channels, for
    example channel 0, is selected in a mu ltiplexer, those capacitors
    accu mu late charge. When the next channel, for example channel 1, is
    selected, the accu mu lated charge leaks backward throu gh channel 1. If
    the ou tpu t impedance of the sou rce connected to channel 1 is high
    enou gh, the resu lting reading of channel 1 can be partially affected by
    the voltage on channel 0. This effect is referred to as ghosting.
    If you r sou rce impedance is high, you  can decrease the scan rate to
    allow the NI-PGIA more time to settle. Another option is to u se a
    voltage follower circu it external to you r DAQ device to decrease the
    impedance seen by the DAQ device. Refer to the KnowledgeBase
    docu ment, Decreasing the Source Impedance of an Analog Input
    Signal , by going to ni.com/info  and entering the info code rdbbis .
    A buffer per sensor is the best fix.
    Lynn

  • Can we lock 80 MHz timebase to PXI_CLK10 on PXI-6220 M Series DAQ?

    I am using PXI-6220 to measure Frequency/Period of 32768 Hz clock signal, accuracy is very important. I have provided signal to measure on Gate input of Ctr 0 (PFI 9). I am using 80 Mhz Timebase. I want to Lock PLL to PXI back plane 10 MHz reference (PXI_CLK10) which in turn is locked onto the 10 MHz reference via PXI-5600 on Slot 2 (External 10 Mhz standard is connected). so far I don't see a way to lock PLL to PXI_CLK10. Is that possible at all? see the image attached to this message.
    Attachments:
    M Series DAQ Question.png ‏62 KB

    Hello Abhatti,
    Based on the diagram that you have attached, the M-Series card can PLL to a higher accuracy clock such as the PXI Clock_10.  The way to configure this change using the DAQmx driver is route the signals using the DAQmx Timing RefClk.Src Timing property node.  Once you place down this property node, and specify the RefClk.Src attribute, you select the PXI-Clk10 as your reference clock source.  This will discipline the 80 MHz Timebase of the DAQ card to the 10 MHz reference clock of the PXI chassis.  Which chassis are you using?  Also, how have you PLL'ed the 10 MHz backplane clock to the PXI-5600 Downconverter?
    Michael L.
    Applications Engineer
    National Instruments

  • Redirecting Voip call to another termination router

    Hello All
    The structure is very simple: There is one originating router (as5300 let say A) and
    one terminating router (as5300 let say B) with 1 pri line. I want to route some calls whose have specific destination pattern from B to another router (let say C) to be terminated.
    I have used "dial-peer voice xx voip" with "session target ipv4:x.x.x.x. on B.
    But it failed. When I debug it on B ,it gives " can not gateway with C".
    Do you think the idea is wrong for this structure . or how can I do this ( PS :I can administrate only termination routers )
    thanks in advance..

    to tapatel:
    First of all thank you for your assistance
    Router B is not making call origination, it is only termination router.
    Router A (I do not manage router A), is sending voip calls to B and I want to route some calls which has specific destination pattern from B to C. The configuration for B is like this:
    ! the dial-peer 101 and 99 below is working properly
    dial-peer voice 101 pots
    destination-pattern 3054T
    port 0:1
    prefix ,054
    dial-peer voice 99 voip
    destination pattern 3054T
    session target ipv4:10.2.2.2 (the ip address of router A)
    ! I want to route it to router C with the lines below
    dial-peer voice 102 voip
    destination-pattern 30549238.....
    session target ipv4:10.10.1.124 (this is the ip address of router C)
    May be I have to use gatekeeper for this project but i am not sure if it is necessary.

  • Repetative trigger from PXI-6733 to PXI-6220

    I am running a PXI-6733 in continuous mode and want to send a trigger pulse from the 6733 to a PXI-6220 at the start of each cycle. I will also want to do this in finite sample mode. I have used a DAQmx Connect Terminals VI to connect the trigger channels. But I cannot find how to set the 6733 StartTrigger as an output or how to synchronize it with the analog output. Any suggestions would be welcome.
    Bob

    Since your AWG needs to send the 6220 the trigger, the M Series (6220) correlated DIO feature doesn't help. Luckily, the 6733 also has the feature. I've attached a vi that shows how to set up the 6733 to output a companion digital waveform to the analog waveform. The first point in the digital waveform is a 1, the rest is 0. You will have to connect a physical wire from the digital line used to an input (PFI) of the 6220.
    If you can run the 6220 in a 'batch' mode (a series of finite tasks), you can use this digital line as a Start Trigger for your 6220. If your passes happen too quickly, you might need to run the 6220 continuously. Now, you might use a counter on the 6220 to generated the sample clock for the analog input task. The digital line from the 6733 would serve as the counter tasks's Start Trigger. The counter task would produce a series of pulses each time it got this trigger. This is getting a little complicated.
    Attachments:
    6733 Correlated AO and DO.vi ‏66 KB

  • About PXI-6220 PXI-6221

    We want to use PXI-6220 or PXI-6221 to design a system. And we will use the CVI or Labview, but we cannot find the PROGRAMMING MANUALS of PXI-6220 or PXI-6221.Where can I get them? Thanks!

    Hi wenyan,
    You can program both the PXI-6220 and the PXI-6221 using the NI-DAQmx driver. The following tutorial is a very helpful:
    Learn 10 Functions in NI-DAQmx and Solve 80% of Data Acquisition Applications
    Also, both LabVIEW and CVI come with various data acquisition shipping examples (Help >> Find Examples). I suggest using those as starting points.
    If you've installed the DAQmx driver, you should be able to find DAQmx Help from (assuming you're using WindowsXP) Start>>Programs>>National Instruments>>NI-DAQ
    More information on the DAQmx functions can be found:
    NI-DAQmx Help (User Manual)
    Hope this helps,
    Lesley Y.

  • PCI 6220 how to route I/O signals?

    Hi
    I am using a PCI 6220 connected with a SCC 2345, on labview 71 XP based.
    I have 24 V signal and I want to divide the frequency, and send out divided signal on RTSI 0.
    I tryed to use a SC-DI on the 2345, but I don't know how to route the DI signal to the CRT0 source signal?
    Any Ideas?
    Laurent

    There is an infodoc numbered 48430 which helps

  • PXI 6220 - problems with the counter

    Hello all,
    I try to make a buffered counting with counter0 of my PXI card 6220, to count the speed of a motor through a photocell and a sticker over the pulley of the motor.
    I have wired my photocell at PFI8 (source) and I generate a pulse train that I have wired in PFI9.
    The input to the counter is a pulse train 5V high level - 0V low level.
    The problem is that the counter counts sometimes more than 1 pulse for every turn of the motor. However, apparently the pulse train looks ok.
    I don´t know what´s the problem.
    The rising slope of the pulses looks very nice so, I think I could discard it.
    I can suppose that in between pulses there are some glitches at a very high frequency that I cannot see (I have sampled this signal only at 800kHz and at this frequency there are no glitches), but maybe the counter yes, because it works at 20MHz.
    Somebody has experienced this problem. Sometimes ago, using a PCI6221, I solved a similar problem by activating the digital filtering, but it seems it is not possible to activate for this cards. Could somebody confirm it?
    Any ideas will be appreciated.
    Thank you very much in advance!!

    Pete;
    I agree that is a little strange.We don't have such problem reported on our Data Bases.
    A workaround you can try is to make a sinle pulse goes to the source, before start the buffered acquisition, and see if that makes the counter keeps reporting a 1 value.
    You can also try the same thing on another counter of the board to see if that problem is related to the counter you are using.
    Regards
    Filipe A.
    Applications Engineer
    National Instruments

  • Regarding NI PXI-6602 - How to route PFI_33 to a grounded DIO Line.

    The PXI-6602 card with the TIO chip seems to be very robust on counter and DIO pin assignments and routing, but I have found that counter 1 has a pin (PFI_33 -AUX IN (UP/DOWN) ) that has been troublesome.  I need to either route this pin to another pin such as a DIO pin set as an output and pulled LOW or be able to manipulate the pin to a low state when I setup counter 1.  This is to ensure that noise is kept to a minimum due to the PXI-6602 being cabled to a board with a PLD on it.  Any solutions on how to tie PFI_33 low?
    -Tim

    B Spears,
    Looks like I'm heading down that path... We were looking to do it in software, but nothing has come to light.  Interesting enough, DAQmx has a good sub-vi (DAQmx Connect Terminals.vi) but I am dealing with code back from LABView 5 days utilizing Traditional DAQ.  Try writing a snipet utilizing both "Write to DIO.vi" and "DAQmx Connect Terminals.vi" and watch the error codes fly. 
    One option NI has offered was to re-write the DIO tasking in DAQmx instead of traditional... this maybe an option... but, wow!  That's a lot of code.
    Thanks for the reply B Spears,
    -Tim

  • PXI-6220 And Scan List

    I want to acquire mutiple singnels with PXI-6620, and I see that a Scan List memory of 4095 entries is indicated in the data sheet. I want to program the Scan List with CVI, and what should I do? I cannot find anything about it in the DAQmx mannul.
    Thanks!

    Hello,
    You can take a look at the examples we have that are included with NI-DAQ. You can find these examples in C:\Program Files\National Instruments\CVI70\samples\DAQmx. You can then program multiple scan lists depending on the application you want to do.
    Hope this helps!
    LA

  • Fiori extenstion toolkit - Redefine routes - How to?

    Hi,
    We are working with CRM Fiori apps SPS3.
    We want to redefine the routes defined in the Component.js
    How do I know what the syntax should be?
    In the extensive project
    The CRM My Opportunities app already have detailPageRoutes defined.
    I'm trying to add my custom route, but it is not getting called.
    Please see attached.

    I've figured it out thanks to a document I found on SCN "Use Cases for Extending the UI of SAP Fiori Apps.pdf"
    I've added this inside the metadata tag:
    routing : {
    "routes" : {
      "masterDetail" : {
       "subroutes" : {
        "master" : { // master is the name of the route
         "subroutes" : {
          "create2" : {
           "pattern" : "create2/{contextPath}/{processType}/{something}",
           "view" : "CustomView"

  • Cisco or Linksys DSL router that will work with Windows XP Pro?

    Any suggestions for a CISCO or LINKSYS DSL router to replace a Zyxel (Century Link) PK5001Z? System is Windows XP Pro, 20 wireless and wired devices connected (including 6 CISCO WVC210 IP cameras ). The Zyxel forwards as many ports as I want, but will only open a limited number of the ports. This limits my access thru the internet (local network sees all cameras). Zyxel support is unable to help.                  

    Hi Edward, if you're looking for a modem/router combo device, Cisco offers the SRP500 series. However this product is EOS/EOL. The small business product line does not have any other DSL termination router aside the SRP series. If your goal is to use a DSL modem and then have a router, you may want to consider RV325 router as it is very feature rich and quite robust.
    -Tom
    Please mark answered for helpful posts

  • Inter-VRF Route leakage

    Hi Guyz,
    I have 3 VRF's on VSS core.
    1) VRF A
    2) VRF B
    3) Global VRF.
    I have Firewall in L3 mode between these VRFs. Traffic between A & B have to cross firewall.
     i can use BGP or EVN to leak routes between VRFs,  but they leak only routes tht are present in  routing table.
    Now i need to leak specific route for eg 10.10.10.10/32 from VRF A to VRF B.
    10.10.10.0/24 is directly connected interface on VRF A. 
    i need to find a way where i can leake /32 route between VRFs.
    Thanks

    Changing the autonomous system number may be necessary when 2 separate BGP networks are combined under a single autonomous system. This typically occurs when one ISP purchases another ISP. The neighbor local-as command is used initially to configure BGP peers to support 2 local autonomous system numbers to maintain peering between 2 separate BGP networks. This configuration allows the ISP to immediately make the transition without any impact on existing customer configurations
    enable
    configure terminal
    router bgp as-number
    address-family {ipv4 | ipv6 | vpnv4| [multicast | unicast | vrf {vrf-name}]}

  • Query on NI PXI 6602 Timing I/O Module

    How to configure the PXI 6602 Timing I/O module to generate frequency on 8 lines and measure the frequency using the same card?

    To me, the pinout is pretty clear so yes, the PFI lines and P1,P2 are related.
    I can't advise you whether you should get the board repaired or make do. My concern would be whatever caused the failure of P0 would also cause the failure of whatever port you moved to.
    Attachments:
    pxi-6220.PNG ‏36 KB

  • PXI 7356 Read ADC Speed

    Hello,
    I am unable to achieve desired poll rates from the Read ADC vi on my PXI-7356 card.   I ran the Test_Motion_AI_Rate.vi form the article below and I am only getting 100 Hz update.  My system specs are better than what was indicated in the article.  I have a 3.16 GHz Intel Core 2 Duo with 4 GB RAM running Window XP.  I would like to achieve the reported 500 Hz update that was indicated in the article.
    http://digital.ni.com/public.nsf/allkb/41FDFB251D251C7F86256FE2007B412B
    Any suggestions?
    Adam

    The Read ADC rate depends on many factors like application complexity, processor speed, motherboard type and so on. The main purpose of the analog inputs on NI motion control boards is to provide analog feedback signals to the onboard control loop, but it was never meant to be a meaurement device. Due to the asynchronous nature of the data transfer you also will never be able to acquire signals with a deterministic timing.
    For fast and accurate measurements I strongly recommend a DAQ device like a PXI-6220. These devices also can be synchronized with your motor operations (e. g. correlate analog signals to motor positions).
    Kind regards,
    Jochen Klier
    National Instruments

Maybe you are looking for

  • IPod Cannot Update?  Unknown error (-50)

    Suddenly, my iPod won't finish updating and displays the following message: The iPod "ENZYME" cannot be updated. An unknown error occurred (-50). What is wrong with my iPod? Thanks for any assistance. 1.33 GHz 12" PowerBook G4 1.25GB ram "ANTIGEN"  

  • Ipod touch 2G: All App crash

    This just happened couple days ago when I noticed that all my apps crash. The only apps that doesn't crash are the App store, mail, music, safari...etc. But the other app such as games crash right away when I click on it. The only way I can solve the

  • Error after updating NuGet package installer

    I am getting an error on my OdatService app after downloading the new nugget package in visual studio 2012 Nu Get Packages are missing form this solution   Click Restore to restore form your online package sources. When I click Restore I get the erro

  • How to open a Word document formatted as .wps

    a friend sent an e-mail attachment composed in Word, using template features. It arrived with a .wps file extension. I can't open this, either in pages,keynote or numbers. I tried text-edit also to no avail. I tried changing the file extension to .pa

  • Force primaryKey based entity lookups

    Hi, When an entity has a OneToMany relation, and the "many" part changes - let's say one more related entity is added to the (many) list -, if the entity was previously retrieved via a find, the related list will not be updated when the entity is ret