Pxie-1073 trigger

Can the PXIe-1073 drive/monitor the PXI trigger lines?

Hello Robert,
I have not seen any software that can monitor and display the information of the PXI Trigger Bus that is in the backplane of the PXI. A workaround is using an external card that can be used to export this lines to the front panel of the PXI. Some of the cards that can help with this are the 6652 or the 6653 timing and synchronization cards.
On the following link you should be able to see an example on how to export a Trigger to a PFI line:
https://decibel.ni.com/content/docs/DOC-2350
Also, I would like to show you some articles that speaks how to manipulate the trigger lines:
Using PXI Timing and Triggering Functionality
http://digital.ni.com/public.nsf/allkb/BD37A128B1DF27F3862570FC0072359D
Routing PXI Trigger Lines Across the Buses of Multisegment PXI Chassis
http://digital.ni.com/public.nsf/allkb/CA9A49FA50393C4186256DC200774E2B
Signal Routing for PXI Devices - NI Digital Waveform Generator/Analyzer Help
http://zone.ni.com/reference/en-XX/help/370520J-01/hsdio/hsignal_routing_pxi/
I really hope this answer properly your question. If not, please let me know so I can see how I can assist you further.
Regards,
Luis C.
 

Similar Messages

  • Can i use PXIe-1073 with an embedded PXIe controller PXIe-8108

    I have a PXI express chassis PXIe-1073 with an integrated controller MXI-Express controller PCIe-8361. I now want to use the chassis as a standalone system with an embedded controller.
    (a)Can i use PXIe-1073 chassis with PXIe-8108 embedded controller
    (b)Can i use a windows based laptop to control the PXIe-1073 via the PCIe-8361 integrated controller.

    Hello kanad,
    The PXIe-1073 is a MXI-integrate chassis.  This means that the MXI card that would normally go into the controller slot (slot 1) is built into the chassis.  The PCIe-8361 or Expresscard8360 are placed into the desktop or laptop, respectively, and connect to the chassis.  The cards in the PXIe-1073 then appear as if they are part of the desktop of laptop.
    Since the controller is built into the chassis, the PXIe-1073 does not have a slot 1.  Controllers, such as the 8108 can only be plugged into slot 1.  Therefore, the PXIe-1073 can only be controlled by the desktop or laptop MXI link.  It cannot run standalone with an embedded controller.
    Consider reading page 1-3 and 1-6 in the manual below:
    NI PXIe-1073 User Manual and Specifications
    Regards, 
    George T.
    Applications Engineering Specialist
    National Instruments UK and Ireland

  • How to determine what card is in what slot of a PXIe-1073 unit programmab​le

    How can i programmable determine what card is in what slot of a PXIe-1073 unit I have a PXI-4063 and a PXI-2503 in a 1073 chassis. I use the DAQmx Device Product Type to to find the 2503 matrix card but it woun't return the 4065 product type.

    Hi cjp1,
    The KnowledgeBase article Programmtic PXI Slot Detection walks you through the steps to do this. There is also a community example called Read PXI an PXIe Card Serial Numbers Programatically that demonstrates the type of LabVIEW architecture you can use to do it. 
    You should first ensure that all of your modules can be identified and self-test in MAX. 
    In case the above hyperlinks don't work, you can copy-paste the addresses below:
    http://digital.ni.com/public.nsf/allkb/2342900AED0​2975A86256BD800764C8B
    https://decibel.ni.com/content/docs/DOC-20207
    Jeff Munn
    Applications Engineer
    National Instruments

  • How to use PXI Star Trigger for PXIe-5663 in PXIe-1075 chassis

    HI all,
    I have this sytem configuration:
    PXIe-8135 controller. Windows 7 64-bit, RFSA 2.7.5. NI-SYNC 3.4.1
    PXIe-1075 chassis
    PXIe-5663 (2x)
    PXIe-6672 Timing & Sync Card (slot #10)
    I want to trigger the recording of my Digitizer with an external trigger.
    The External Trigger is connected to PFI0 of the PXIe-6672 Timing card.
    Then, the PXIE-6672 card routes the trigger to the backplane of the PXIe-1072 (Destination "NISYNC_VAL_PXITRIG0")
    The PXIe-5663 are triggered with “NIRFSA_VAL_PXI_TRIG0_STR” as the source.
    The trigger fires my PXIe-5663 correctly, but the timing is not tight (> 5ns).
    I would like to use the PXI Star trigger instead, I think that I should be able to acheive much better synchronization with this.
    But NI-RFSA won't let me do this:
    When I try to call
    "niRFSA_ConfigureDigitalEdgeStartTrigger(rfsa_sess​ion, NIRFSA_VAL_PXI_STAR_STR , NIRFSA_VAL_RISING_EDGE)", I get the error:
    "Specified Route Cannot Be Satisfied, Because the Hardware Does Not Support It"
    I don't understand why the PXIe-5663E would not be able to use that Route.
    Any idea?
    Regards,
    Serge
    Serge Malo, ing.
    Concepteur logiciel
    Software Developer
    T (514) 842-7577 x648 | [email protected]

    That explanation isn't quite right. Usually, even PXIe modules have a connection to PXI_Star. The PXIe standard added the PXIe_DStar trigger buses, and it also preserved the PXI_Star bus from the PXI standard.
    However, there is an additional twist in this situation. I'm assuming that your PXIe-5663 includes a PXIe-5622 as the digitizer. It turns out that a synchronization technique called NI-TClk has eliminated the need for our more recent digitizers to rely on triggering from PXI_Star. I was able to find some documentation that includes this information, here and here. Given that, I think you have two options that should result in better synchronization.
    The first option is to use TClk; I found an example program that demonstrates using TClk to acheive phase coherent signal acquisition across two 5663s. The second option is to use cables of matched length to connect two PFI front panel terminals of the timing board (6672) to the PFI1 front panel terminals of the digitizers (5622). The timing board would accept the external trigger on PFI0 and then issue triggers on PFI1 and PFI2 with around 500 ps of skew (manual, page A-4) . The digitizers would use NIRFSA_VAL_PFI1_STR as the trigger. I hope one of these solutions will meet the demands of your particular application.
    I will also follow up with the owners of the RFSA product documentation to see if we can include a note about why PXI_Star is not supported in some cases.
    James Blair
    NI R&D

  • PXIe-1073 sytem not detected

    I connected a PXIe-1073 system (PXIe-6363 and PXI-2567 connected) to the PC (running Windows 7) and the PC does not detect the chassis. It started happening when I recently upgraded to LabVIEW 2013 and NI-DAQmx 9.7.5.
    I also installed the MXI BIOS compatibility software Ver 1.4, changed the DIP switch1 to ON position on the MXI express card 8361 and rebooted the computer. That did not solve the problem.
    Then, I upgraded to PXI Platform services 3.2.3 and restarted the computer. I still don't see PXI system recognized in NI MAX. When the computer was restarted, the 8361's MXI card's switch 1 was still in ON position.
    Any ideas on how to fix this issue ?
    Thanks
    Jeet

    Hi Matt
    Sorry for delay in replying.
    After a lot of uninstallation/installation/repair of different driver versions, i have currently got the following software (see attached pic): NI-DAQmx 9.5.5, Ni PXI platform services 3.2.1.
    In the picture, you can  see that all my cards are detected in NI-DAQmx devices as Dev1, Dev2 etc.
    My PXIe-1073 is not detected or listed in MAX. I expect to see chassis controller detected and all the cards listed under the chassis controller as "PXI2slot2" device or "PXI2slot4" device etc.
    The attached pic shows NI-MAX and device manager snap shot.
    Please advise.
    Thanks
    Jeet
    Attachments:
    nimax dev man.jpg ‏316 KB

  • Where can I find 3D model step files for PXIe-1073 chassis?

    I need 3D model step files for the PXIe-1073 chassis. Does anyone know where I can get them?

    Eduardo,
    I again apologize for the extremely long delay.  Our R&D department has completed your request, and it can be found online, here:  http://digital.ni.com/hardref.nsf/websearch/99B2D43123FBABE1862578B60077848F?OpenDocument&node=20053...
    Paul Davidson
    Sound and Vibration Software Staff Product Support Engineer
    National Instruments

  • PXIe-1073 not detected

    I'm trying to install a PXIe-1073 chassis on a new Dell system running Windows 7 x64.  I have the latest version of Labview, NI-DAQmx, etc.  The PCI express controller shows as properly installed in the Device Manager ("PCI standard PCI-to-PCI bridge").  The chassis and installed DAQ card are not detected in MAX.
    I am actually transferring this system from an older computer.  I recall having exactly the same problem before, and I'm fairly certain I solved it by installing the "MXI-Express BIOS Compatability Software".  There was never any way to actually identify the "PXI System" as anything other than "External PC" even when it was working properly.
    After installing the current version of the MXI-Express software, I rebooted and was able to load MAX briefly before Windows BSODed with an error involving "niraptrk.dll".  This proceeded to occur every time I tried to load Windows.  I was able to start in safe mode, but attempting to uninstall any NI software gave me an "error 1601".  I was finally able to load the last known good configuration and uninstall the software.
    When I checked MAX again, the chassis was still unrecognized, but there were multiple broken entries for the DAQ card installed in it, so I assume it was detected at some point.  It looks like the MXI-Express software worked, but it's obviously unusable if it causes the computer to BSOD every time Windows starts.  Why is the computer unable to recognize the chassis without it?  What else can I do to solve the problem?

    Hello Bob_mcbob,
    After looking over this issue, I think we have a problem with the Windows installer on your machine. I have found a KnowledgeBase about the Error 1601 stating that this is a Microsoft Windows installer error and that the installer will need to be reinstalled. The Knowledgebase explains how to do this.
    As for the BSOD, do you have any more information that it provided when this occurred like the address line, etc? Did you do something in MAX before it crashed?
    Since accessing MAX now gives you a BSOD, I think it would be best to uninstall/reinstall the drivers to fix the issue. What cards do you have in this system?
    Jim St
    National Instruments
    RF Product Support Engineer

  • PXIe 1073 Input Grounding

    Hello,
    I was wondering how the grounds from the hybrid slots of a PXIe 1073 chassis are connected within the chassis. I'm using 3 PXI 6255 cards, though I don't believe that makes a difference. What I'm looking for is whether they are tied together or not, or similar information. Thank you.
    Solved!
    Go to Solution.

    BVeezy wrote:
    UPDATE: I was incorrect. Please refer to the following KB:
    http://digital.ni.com/public.nsf/allkb/02A451C0A88​A3EAE8625702F0059C98E
    Regards,
    Brandon V.
    Applications Engineer
    National Instruments 
    Yeah, I discovered that not long ago.  Every ground on DAQ cards are tied to the chassis ground.  It is just that each type of ground (DGND, AGND) have their own planes on the card itself.  And chassis ground will be going to Earth Ground, in case that matters to you.
    There are only two ways to tell somebody thanks: Kudos and Marked Solutions
    Unofficial Forum Rules and Guidelines

  • PXI-6541 Trigger Interrupt

    Hello,
    I'm using the PXI6541 to perform a simple digital output based on an interrupt input.
    As seen in the attached diagram I drew up, I want one of the PXI's trigger lines to listen for a falling edge from my hardware.
    Once the CONVST falling edge occurs, I want to wait x ns before I set one of the DIO pins (EN) to go high until CONVST goes high.
    I've been modifying the HSDIO "Dynamic Generation of Multiple Waveforms" example VI in LabVIEW, trying to add triggering to it. (See attached VI). 
    My problem is that it doesn't appear to wait for a trigger, but instead times out after a number of seconds. I've attached my modified VI to this post.
    Also, what happens if I do not enable "Voltage Configuration"? What's the logic level it will work at by default?
    Many thanks in advance.
    Attachments:
    Dynamic Generation + Trigger.vi ‏33 KB

    Hi guys,
    To answer each of your questions in turn.
    Regarding the trigger lines, what is the difference between say PFI 0 and PXI Trigger Line 0/RTSI 0? I had been using the latter pin from the breakout cable. The hardware I am testing is just a bench top eval board for the part. I hadn't realised there was access to PFI 0 via the SMB connectors.
    Regarding timing, I require a minimum of 15ns, so a 20ns interval would be fine for this. The constraint is to allow the part time to wakeup and setup, so having it slightly longer is ok.
    Hope this clarifies the questions.

  • PXIe 1073 Power On Start

    Hello.,
    I have a PXIe 1073   system
    I want to start it automatically when there the Mains  Power is connected.
    any clues !!

    Dear aparab,
    as far as I understand your inquiry, you'd like to boot a pxi controller without pressing the on-switch. The PXI chassis has three different methods on how to power on: Wake-On-LAN, an inbuilt switch and a BIOS setting. Details you can find in the following article.
    Remote Power of a PXI Chassis
    http://digital.ni.com/public.nsf/allkb/FF5AB8BB6A1157DB8625756D00502D55
    with regards
    AndGar

  • AC Recovery PXIe-1073

    Is there any way to set how the PXIe-1073 behave when AC power is restored after AC power loss?
    I want the PXIe-1073 to power on after AX power loss.
    I'm using the PXIe-1073 in a long time test system (1000h) and it must start-up automatically after a power loss.

    Hi,
    The NI PXIe-1073 is lacking support of remote power control. For high priority applications or long time test systems I recommend upgrading to one of the supported chassis. 
    When on a supported chassi, this Knowledge Base provides some more details and other related features of PXI and PXIe chassis: Remote Power of a PXI Chassis
    Best regards,
    Robert P-F
    Applications Engineer
    National Instruments Sweden

  • PXIe-1065 Trigger lines

    Hi all
    I'm trying to acquire data on 7 PXIe-6358 boards. All these boards are put in a PXIe-1065 chassis which is connecte to a RMC 8354 throught a MXI 8360. In order to synchronize all boards togethet i use a PXI 6682H board which will generate a clock and a start trigger on two PXI trigger lines.
    I made some test on a PXIe 1075 chassis and everything seems to work fine. I moved then on a PXIe-1065 chassis and now i'm having some problem when i try to acquire data on the 7 boards. When i have more than 5 boards, it seems at least one board doesn' receive either the clock or the start trigger.
    Then trigger are configured in MAX to go through all 3 bus. When i configure only 5 boards (wherever in the slot) i works fine. But when i declare more than 5 boards i have the problem again.... It seems you cannot connect more than 5 boards to the trigger lines.
    Does someone have an idea...

    Ok, i found the problem, it has nothing to do with the PXIe 1065 chassis.
    In my application i use the PXI 6682H to generate the start trigger and a clock. The acquisition tasks are waiting for the trigger to start the acquisition. In my case when i switched the chassis, the PXI 6682H becomes a master (in the old configuration it was parameter as a slave). When it is configured as a slave i check that the board is synchronized with the master (this can takes a little time). When i am the master i doesn't wait (cause i am already synchronized). Therefore, in that case the start Trigger is generated before all instances of the PXIe-6358 are started, and they wait indefinitively the trigger ....
    I modify the code in order to wait a certain amount of time (even if the PXI6682 is configured as the matser) , and now its works properly ...

  • Setting up a PXI-5412 Trigger

    Hi,
    I have a PXI-5412 FGen and a PXI-5105 Digitizer. I was wondering if anyone could explain how to internally and/or externally set up a trigger. The idea is that I'm sending a signal with a pulse repetition frequency (PRF of 500Hz for example) from my Fgen to one of the channels on my, but I would like the digitizer to acquire at that same PRF so that my signal is not all over the place.
    Both these PXIs are installed on a PXI-1044 chassis which I'm assuming was correctly set up!
    I know this should be relatively simple, but I don't know where to start. I've been seeing a lot of the NI-TClk VIs in the examples and discussions forums. Should I be looking to use those? And will I only be using one set of VIs for the Fgen or do I have to play around with the scope as well?
    If need be, I can provide more details on the way I set up the Fgen and the scope.
    Thanks,
    Marc

    Hi Marc,
    You're on the right track. TClk is one of the ways to synchronize your cards, and makes the most sense for your application.  You will need both a NI-SCOPE and NI-FGEN session with NI-TCLK tying the two together. Refer to this example:
    NI-TCLK Synchronize AWG and High-Speed Digitizer
    Let me know if you have further questions.
    Cheers,
    Aaron
    National Instruments

  • PXI 5422 Trigger output control

    How can the trigger output of the Digital Data and Control be controlled?
    Can it produce either a trigger pulse or digital value change on one of the PFI outputs through software instead of just at the start of the generation?

    Hi,
    Could you explain a little more about your application and what you are trying to achive,  as well as what events you want your trigger to be output on?  What is this tigger going to be driving?
    Thanks
    Matt Surridge
    National Instruments

  • I am using NI PXIe-1073 with Labview-2014.After deploying the VI how can i read the data through host computer?

    sankar,new delhi

    PIB. I want to read the data from the GPIB after the operation complete. I am using *opc? command, which should set the status register bit after the completion of the operation, but this is not working. How to know that the Operation is complete"-Thanks.The NI-488 global status variables ibsta, ibcnt, and iberr is what you are looking for. Look into the 4882 help file for details. Also NI-Spy, http://www.ni.com/support/gpib/nispy.htm, is a good debug util. There is a website that lists common GPIB error codes and solutions. You may check there for some things to try. You'll find the link at ni.com > Support > Troubleshooting > Installation/Getting Started > GPIB, titled "GPIB Error Codes and Common Solutions".
    You could find a driver for this instrument at http://www.ni.com/devzone/idnet/default.htm . If it's not listed there, it leaves you with one of a couple options. First, I would like you to submit a request for this driver at: http://zone.ni.com/idnet97.nsf/instrumentdriverrequest/
    We develop dri
    vers based on demand and popularity so the more requests we have for it, the greater the possibility that we will develop one.
    If you would like to try developing your own instrument driver (or modify the existing one), we have documentation, model instrument drivers, and driver templates to help at :
    http://www.ni.com/devzone/idnet/development.htm
    We also have a syndicate of third party vendors that specialize in National Instruments' products and services. Some of the vendors specialize in driver development. I would suggest contacting one of the Alliance members at:
    http://www.ni.com/alliance
    Hope this helps.

Maybe you are looking for

  • ADF-af:inputListOfValues  not working properly when used in inside table

    Hi All, i am using jdev version 11.1.1.5.0 . In my usecase i have create one eo based vo where in empname attribute i have add lov and drag this vo as adf table on jsff page. and i have create row on table and select empname from lov. lov working pro

  • Want to Receive Response Back From SOAP

    Hi , My Scenario is IDOC-XI--SOAP Iam sending the IDOC from R/3 and want to hold the Response message from SOAP like "Success or Fail". I want to the response message in XI itself, no need to reach the response even to R/3. How can we achive this Reg

  • Concatenating the stings

    Hi friends , This is a scenerio from idoc to file . I need to concatnate 3 fields from the same segment . I concatenate the fields and map to the target . What I found is in source payload the field combination mapping does not exist . If I look at t

  • ITunes Library cannot be saved. Unknown Error has Occurred (-54)

    I have been having the following message appear in iTunes: iTunes Library cannot be saved. Unknown Error has Occurred (-54) I created a new library which seemed to solve the problem.... until I plugged in my iPod Classic or tried to Sync my iPhone. W

  • Unable to invoke EJB from a BPEL process.

    Hello all I am running BPEL process manager over weblogic 8.1 App server. I am trying to invoke a simple stateless session bean from a synchronous BPEL process. I have created a WSDL for this EJB with appropriate binding and service elements. I am ab