"Read Single Point" Keithley 2400 Problems

Hello!
I am recentely having some problems with the "Read Single Point Measurements" function in Labview 2012. I have already sucessfully used my Keithley 2400 in other programs to sweep the voltage, but now I am trying to use it also to read the current across my sample. 
I wrote a very simple program, which resemble the one you can find among Labview examples (see 1st attachment). The only differences are a sweep-subVI and a "for" cycle. My problem occurs at the "Read" Keithley function. On the Keithley's screen I have the following errors:
- 113 : undefined header;
- 230: data corrupt or stale;
- 420: query unterminated.
While in the Block diagram, when the system gets to the "read" function, the error -1074000000 shows up.
I have found a lot of posts on this topic, but unfortunately I could not find the any solution for me. I also tried the 2nd attached program, to check the comunication to the instrument, but errors still occur. In "Measurements and Automatic", it says that the instrument is working properly. 
Do you have any ideas?
Thanks  
Attachments:
Keithley 24XX Sweep and Acquire Measurements.vi ‏26 KB
Basic Serial Write and Read (1).vi ‏26 KB

The following video demonstrates how to check the firmware revision: http://www.keithley.nl/centralized_display?mn=2400&assetid=55934.
In order to update the firmware you will be need to download a flash program and the firmware file. gAfter obtaining the Flash program and the Firmware file from Keithley Application support staff, unzip the three files in a folder of your choice and run the "setup.exe" program.  Follow the instructions to install the program.  When the installation is complete, launch the program from your Windows Start menu under Programs/Keithley Instruments. The Application support staff will also provide the file for the latest firmware revision. Put that file in a folder where you can find it and then run the Flash Wizard32 program. The program will autodetect the instrument and request to specify the firmware file.
This link contains the flash program: http://www.keithley.com/base_download?dassetid=52609
 See attactment for the firmware revision.
Attachments:
2400c30.zip ‏358 KB

Similar Messages

  • Keithley 2400 Problems

    Hi,
    I'm trying to run a Keithley 2400 SourceMeter with LabView using the latest driver (Keithley 24XX) and the example VI "Keithley 24XX Read Single.vi".
    After selecting the GPIB Adress I run the program and Labview throws the following error message:
    Error -1074000000 occurred at Keithley 24XX.lvlib:Error Query.vi
    Possible reason(s):
    Instrument reports:
    -113,"Undefined header"
    -230,"Data corrupt or stale"
    -420,"Query UNTERMINATED"
    0,"No error"
    Complete call chain:
    Keithley 24XX.lvlib:Error Query.vi
    Keithley 24XX.lvlib:Read (Multiple Points).vi
    Keithley 24XX.lvlib:Read (Single Point).vi
    Keithley 24XX Read Single.vi
    I also have a Keithley 2000 Multimeter up and runing with it's latest driver and the example program and it works just fine.
    What could be the problem?
    Thanks in advance!
    Solved!
    Go to Solution.

    There should be examples with those drivers.  If not, then go through the driver palette and think about what you would do manually.  Find those functions to do what you would do manually.
    There are only two ways to tell somebody thanks: Kudos and Marked Solutions

  • KEITHLEY 2400 code problem

    Hello
    I am trying to acquire data from keithley 2400 using labview 9, NI VISA 4.5.1,  and serial RS 232 interface.
    I used the attached "Keithley 24XX Read Single.vi"  ( obtained from NI certified drivers )   and got this message-
    Error -1073807346
    occurred at VISA Write in Keithley 24XX.lvlib:Error Query.vi->Keithley 24XX.lvlib:Configure Measurement.vi->Keithley 24XX Read Single.vi
    The keithley meter indicated "REM", as soon as VI was run, which means "Remote" operation and froze up.
    Under MAX, i can see COM1  and device status is "Device is working properly"
    Kindly  help in debugging
    1. Should i use GPIB?  Are the drivers good for serial interface? I used the  first NI certified driver indicating serial/GPIB compatibility.
    2.Is keithley meter being recognized by MAX?
    3. Any errors in the codes?
    Regards
    poli
    Attachments:
    Keithley 24XX Read Single.vi ‏24 KB

    you  guys are amazing!
    I switched off all other vi and tested the "Basic serial read and write.vi". It worked fine.
    So the first step regarding MAX communication  is ok.
    Now, i tested the attached "Read (Single Point).vi and it showed the following error, after some time.
    code = -1073807339
    VISA Read in Keithley 24XX.lvlib:Error Query.vi->Keithley 24XX.lvlib:Read (Single Point).vi
    If i get this to work, that will solve the problem.
    Attachments:
    Read (Single Point).vi ‏17 KB
    Keithley 24XX.lvlib ‏12 KB

  • Error using the Keithley 2400

    Hi, I am pretty new to LabVIEW and right now I am just trying to set up a simple test (source current, measure voltage) using the Keithley 2400 and Labview 8.2
    I got the driver for this instrument off of NI's website at
    http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=25B255F3AA83660EE0440003BA7CCD...
    In the Measurement and Automator it sees the Keithley (GPIB0::14::INSTR) and I also gave it an alias so it is easier to find amungst the other hardware current set up.
    I wrote a simple program using some of the VIs in the driver to source a current and read a voltage.  When the program runs, it errors out, and then after it errors the keithley still runs the test.  I can see the measurement on the front panel of the instrument however it does not send anything back to the computer.
    I then tried to just run an example found in the driver and I get the exact same error.
    Error -1074000000 occurred at Keithley 24XX.lvlib:Error Query.vi
    Possible reason(s):
    Instrument reports:
    -113,"Undefined header"
    0,"No error"
    Complete call chain:
         Keithley 24XX.lvlib:Error Query.vi
         Keithley 24XX.lvlib:Read (Multiple Points).vi
         Keithley 24XX.lvlib:Read (Single Point).vi
         Keithley 24XX Read Single.vi
    Anyone have any idea's as to what might be causing this error?  Any help would be greatly appreciated
    Nolan

    Hello nblue:
    Can you communicate with the Keithley 2400 in Measurement and Automation? Do you know if there is a query command for the instrument, if so can you test to see if you can query the instrument to make sure it is working.
    Regards,
    Roland A.
    Applications Engineer
    National Instruments
    Check out VI Roadshow

  • Keithley 2400 errors

    Hi ,
    I want to do sweep measurement using keithley 2400 with labview 2010. Here are the errors I came across.
    error running read single.vi
    Error -1074000000 occurred at Keithley 24XX.lvlib:Error Query.vi
    Possible reason(s):
    Instrument reports:
    -113,"Undefined header"
    -230,"Data corrupt or stale"
    -420,"Query UNTERMINATED"
    0,"No error"
    Complete call chain:
         Keithley 24XX.lvlib:Error Query.vi
         Keithley 24XX.lvlib:Read (Multiple Points).vi
         Keithley 24XX.lvlib:Read (Single Point).vi
         Keithley 24XX Read Single.vi
    running error query.vi from keithley24xx lib----
    code
    d-1073807346
    VISA Write in Keithley 24XX.lvlib:Error Query.vi
    I downloaded keithley 2400 driver in NI website.
    http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=25B255F3AA83660EE0440003BA7CCD...
    I am very new to labview and keithley 2400. Kindly help me in this regard. Please give a detailed reply.

    Thanks for your reply. I have posted the VI and a screen print of MAX communication.
    Attachments:
    Keithley 24XX Sweep and Acquire Measurements.vi ‏26 KB
    NI- MAX.JPG ‏111 KB

  • Keithley 2400 Meter Read dans une boucle - probleme de status byte read

    Bonjour,
    j'utilise un Keithley 2400 pour mesurer un potentiel. J'ai débuté en utilisant le modèle Read.vi du driver. Dans mon cas, la portion "read single" se trouve dans une boucle for et après un certain nombre de mesures (cela semble aléatoire), la fonction gèle et en remontant les étapes, on découvre qu'il s'agit de la lecture de "status byte" qui ne reçoit aucune réponse en amont. Il faut alors redémarrer complètement le programme. Celà survient à chaque fois et après un nombre d'itérations variable.
    J'ai même tenté de mettre une boucle while directement dans l'exemple "read single" (le .vi) et le problème est exactement le même, même si on ajoute un wait à la boucle. Après un certain temps, ça gèle et c'est le status read qui ne reçoit aucune réponse.
    Merci.

    Bonsoir PG123,
    Ce que je vois directement dans votre VI c'est le fait que vous exécutiez à la vitesse maximum votre boucle While sans le moindre cadencement ne laissant ainsi jamais de temps disponible au processeur pour effectuer d'autres tâches. Il est propable de plus que l'instrument ne supporte pas un nombre de requêtes aussi élevée saturant peut être son buffer ou ne laissant pas le temps à l'instrument d'écrire correctement les valeurs.
    De plus, je pense que vous n'avez pas besoin de récupérer à la vitesse maximum toutes les valeurs possibles provenant de votre instrument. C'est pourquoi je pense qu'une lecture toutes les 200ms suffirait déjà amplement.
    Voici donc à quoi devrait ressembler l'intérieur de votre boucle While :
    Essayez avec un cadencement d'1 seconde (1000ms en entrée de la fonction) dans un premier temps et voyez si le problème de gel se produit. Si ce n'est pas le cas, vous pourrez alors progressivement diminuer la valeur de cadencement (jusqu'à 200ms minimum, celà devrait suffire comme déjà dit au dessus.
    Cordialement,
    Jérémy C.
    National Instruments France
    #adMrkt{text-align: center;font-size:11px; font-weight: bold;} #adMrkt a {text-decoration: none;} #adMrkt a:hover{font-size: 9px;} #adMrkt a span{display: none;} #adMrkt a:hover span{display: block;}
    Travaux Pratiques d'initiation à LabVIEW et à la mesure
    Du 2 au 23 octobre, partout en France

  • How to read current using labview 7.1 from Keithley 2400

    Hi, All,
    I am trying to read current out from labview 7.1. The program I wrote is attached (program.gif). But when I run this program, the error message poped out when it reach the function of GPIB write (error.gif). Can anybody help to point out what's wrong with this program?
    Thanks a bunch!
    Joyce
    Attachments:
    PROGRAM.gif ‏40 KB
    ERROR.gif ‏45 KB

    Hi, Dennis,
    Thanks for replying. You reminded me of putting 24, which is the GPIB address of Keithley 2400, in the front panel. Now this error doesn't occur, but it failed when it's trying to scan the result.
    I attached the program in this post. I attached the figure because i thought people couldn't open the code if they don't have the right version of LV.
    Thanks,
    Joyce
    Attachments:
    Keithleyreadcurrent.vi ‏63 KB

  • Keithley 2400 record data problem

    Hi,
     I use Keithley 2400 to Measure I-V curve, at a sweep mode, like scan the V from 0 to 0.2 V, at low NPLC(eg. 0.1), the data can be recorded very successful and save to data.txt. But when I increase the NPLC (eg. 5 or 10,), this vi usually cause some problem. When the scan votage reach the 0.2 V, the keithley display panel stops at 0.2V and shows this current, the VI still shows at working, and no data recorded in the data.txt. In the correct process, Keithley display panel shoud show the "OFF" and the VI also stop working, and the data recorder correctly. Could some one help me out of this? Thank you very much.
    Attachments:
    24xx_Swp-V_Meas-I_gpib.llb ‏373 KB

    Why can't you stick to a single thread for same topic? Please see here and here

  • Problem with Keithley 2400 source meter

    Hello all,
    I tried to controll a keithley 2400 source meter with Labtracer using a NI GPIB-USB-HI card but when I run the test is send me this error message: "There was ab error reported on Visa Read in Query. Test aborted. Error code: -1073807339"
    is this card (NI GPIB-USB-HI) able to controll keithley 2400 source meter ?
    the keithley 2400 source meter screen show "REM TALK" during the running of the program,
    I installed NI488227 in my computer (windows XP) and also "Agilent Connection Expert", this later also show no possibilty to connect with the Keithley meter, I attached a file for  "Agilent Connection Expert".
    Any help is appreciated
    Thank you by advance
    Attachments:
    ACE.docx ‏133 KB

    Follow the preliminary steps to first check if the device is able to communicate via the GPIB-USB convertor.
    navigate to: Start>>National Instruments>>NI 448.2>>Interactive control.  Use this utility to query the GPIB device for some standard commands for e.g.: *IDN?
    There is also a GPIB analyzer in the above path which you can use to verify the device connectivity to the system.
    Also for communicating the with 3rd party GPIB devices always enable " NIVisa Tulip.dll" which will help recognising the device in max.  Look at the picture to enable this option in max.
    post back for queries
    With regards,
    JK
    (Certified LabVIEW Developer)
    Give Kudos for Good Answers, and Mark it a solution if your problem is solved.
    Attachments:
    tulip.jpg ‏175 KB

  • Pl/sql block reading reading table data from single point in time

    I am trying to figure out whether several cursors within a PL/SQL block are executed from within a Single Point In Time, and thus do not see any updates to tables made by other processes or procedures running at the same time.
    The reason I am asking is since I have a block of code making some data extraction, with some initial Sanity Checks before the code executes. However, if some other procedure would be modifying the data in between, then the Sanity Check is invalid. So I am basically trying to figure out if there is some read consistency within a PL/SQL, preventing updates from other processes to be seen.
    Anyone having an idea?.
    BR,
    Cenk

    "Transaction-Level Read Consistency
    Oracle also offers the option of enforcing transaction-level read consistency. When a transaction runs in serializable mode, all data accesses reflect the state of the database as of the time the transaction began. *This means that the data seen by all queries within the same transaction is consistent with respect to a single point in time, except that queries made by a serializable transaction do see changes made by the transaction itself*. Transaction-level read consistency produces repeatable reads and does not expose a query to phantoms."
    http://www.oracle.com/pls/db102/search?remark=quick_search&word=read+consistency&tab_id=&format=ranked

  • I do not use or have not signed up for a MobileMe Account but I am getting duplicates in my iCal for some reason.  All discussions I read are pointing to the MobileMe being the problem but it is not.  Please help me.  There must be another reason.

    I do not use or have not signed up for a MobileMe Account but I am getting duplicates in my iCal for some reason.  All discussions I read are pointing to the MobileMe being the problem but it is not.  Please help me.  There must be another reason all my entries appear twice. 

    If you wish to submit comments to Apple, the best way is to use their feedback pages.
    http://www.apple.com/feedback/itunesapp.html
    It's not likely that anyone from Apple will see your comments here among the thousands of other posts.
    Regards.

  • How can I measure above 100 point by Keithley 2400?

    Hello!
    I am using Keithley 2400 and Labview 6.0. I have some questions.
    1) In my program, I cannot measure above 100 points. The limit that I can measure is only 100 points. In order to measure 1000 point, what should I do?
    2) I want to make "Sweep Voltage" by Step Voltage [start:0V, stop: 5V, step: 0.1V. I have already implemented this, but my program is not good. Could you modify my progrma?
    Thank you!

    I don't have 6.0 installed anymore - I hope this image helps.
    Attachments:
    answer.gif ‏132 KB

  • Programmed vs measured voltage on Keithley 2400 and other things

    Hello everyone,
    I have two issues with our Keithley 2400, which we control through Labview, and I would be very grateful for some help/advice. The issues are listed below:
    1) We use the 2400 for 4-point probe measurements, so we plot graphs of current vs voltage to get the resistance from the slope of the curve. However, when we look up the data generated during a measurement, the voltage sweep values that we program often, but not always, do not match the measured voltage values. For example, what was meant to be 1 V, might actually end up being 1.2 V or maybe 0.8 V, this can be both positive and negative with respect to the programmed values and is not constant. This also leads to the fact that our curves, although straight (so Ohmic conductor), do not go through the (0, 0) point on the graph.
    2) Second problem is that we would like to determine what is the time between successive measurements in a voltage sweep. We have tried some example vi files that by default come with the Labview, e.g. Keithley 24XX Read Multiple - SW Trigger or Keithley 24XX Read Multiple. If someone knew how to find the time separation, then we would be grateful. We only have control over the number of invididual readings taken and the timeout, which we guess is the maximum time allowed for the measurement.
    Many thanks!

    Good Afternoon PiotrW,
    1. Could you please provide the code you are using to program the voltage sweeps? 
    2. You could use the timestamp output of the Keithley 24XX.lvlib:Read (Multiple Points).vi to allow you to find the time difference between successive samples. It should be a simple case of subtracting succesive measurements from each other to find the time difference between data points.
    Regards,
    Al 
    Regards,
    Al C
    National Instruments,
    Application Engineer

  • Keithley 2400 VISA driver limited resolution

    Hi,
    I have a Keithley 2400 sourcemeter, what I use in the +-1mA range, as a current-source. This is the driving part of a PID control.
    I use the official VISA drivers, and everything is okey, except that, it seems I cannot use the specified resolution from LabView.
    In this low +-1 milliAmp range, the usable resolution is 50 nanoAmperes. In manual mode, everything is ok, I can set for example an output current to be:
    -0.245450 mAmperes.
    But when I use LabView with the Configure Output .vi from the official llvlib package (i attached it) it rounds for microAmps:
    For example the above double value will become: -0.245 mAmperes.
    This is really bad for my project, since I need the 50 nanoAmp resolution for the really fine control.
    Can someone help me?
    Thanks in advance!
    Solved!
    Go to Solution.
    Attachments:
    Configure Output.vi ‏18 KB

    Thanks very much!
    Now I learnt that, I should be always suspicious even with "official" libraries and drivers.
    Actually there are two LabView drivers for Keithley 2400, one from Keithley, it is an exe installer type, and the other from National Instruments. I became curious about this bug, and I checked the Keithley VISA drivers, they are okey, they use everywhere double precision variables. 
    But if you have a look at the NI Keithley 24XX.lvlib:VI Tree.vi, where is the full list for available VIs, there are many problems I think. At many places (voltage sensing, sourcing, etc) the programmer just used single precision variables to pass information.
    I think these problems should be reported toward NI, but I do not know what is the way to do this...

  • Keithley 2400 Trigger 2000

    Hi,
    I have tried for a couple of days to let a Keithley 2400 source current and measure voltage and at the same time trigger a DMM2000 on which a current will be measured. This is for use in a LIV characterization. I�m using Labview but prefer to write directly to the bus instead of using the Labview drivers, which I find a bit complicated to use.
    First I configure the 2400 with the following commands
    *RST;
    TRAC:CLE;
    SOUR:FUNC:MODE CURR;
    SOURWEPAC LIN;
    SOUR:CURRTAR 0;
    SOUR:CURRTOP 0.1;
    SOUR:CURRTEP 0.0001;
    TRIG:COUN 100;
    SOUR:CURR:MODE SWE;
    SENS:FUNC "VOLT";
    FORM:ELEM CURR,VOLT;
    TRIGOUR TLINK;
    TRIG:INP SOUR;
    TRIGUTP SENS;
    Then I write the following to the DMM 2000 to initialize
    *RST;
    TRAC:CLE;
    SE
    NS:FUNC "CURR";
    TRIGOUR EXT;
    TRIG:COUN 100;
    Now the two instruments should trigger each other.
    I then write to the 2400 to turn on and initiate the sweep.
    OUTP ON;
    INIT;
    And finally an INIT is sent to the DMM2000 to initiate this instrument
    This starts the measurement ONLY if the instruments have been turned on just before the program is executed.
    If I run the full program again it doesn�t start. What is wrong?
    If I run the program from the initialization of the 2000 it works well. But if the intialization of the 2400 is included it just wont start the sweep.
    Any suggestions?
    Regards
    Francis

    You may wish to try this. I struggled with a 2400 for some time. After a single reading, the buffer in the 2400 contains current, voltage, ohms, a timestamp and another variable. You have to read the entire buffer of the 2400 or it ignors future commands. I struggled with the 2400 drivers also. The help files of the drivers did not describe one of the variable inputs. Finally, I figured out that it was leaving data in the buffer. I increased the variable to read all five values and it finally worked!
    You will have a different solution because you are not using the drivers. But I suspect you may have the same problem. You will have to write commands to be certain you extract all measurment information for the reading. Perhaps the DMM may also have a simi
    lar feature. If you do not read all the data, the systems will hang.

Maybe you are looking for

  • How to delete the attachments of several ancient emails simultaneously?

    In order to diminish the space used by Thunderbird, I'd like to delete the attachments, without deleting the email. How to do this for 50 emails all together? I only found a way to do it one by one, email after email, which takes much time- Thank you

  • Constant Notifications from Adobe CC on Mac OS 10.9.3

    I just updated the applications manager for CC to the latest version 1.6.0.393 released on 05/27/2014.  Since then, every time I restart my computer I get constant notifications flashing in the top right hand corner of my monitor as shown below: The

  • Which iPod do you recommend?

    Hi! I'm thinking about getting an iPod. I don't have tons of music (about 2 gigs), and I'm trying to decide between a Shuffle or a Nano. I don't need all my music on the iPod at one time, but I wouldn't mind it either. I've heard about some screen pr

  • My iPhone shows 1.0GB of memory used by Mail.  I've emptied all mailboxes and folders but still show 1.0 GB

    After upgrading to iOS7, I noted my iPhone 5 shows 1.0GB for Mail in the Usage list.  I've cleaned out all mailboxes and folders but still show the Mail app as using 1.0 GB.  Am I missing something?

  • Flash FLV Video File

    I have a Flash Video on one of my pages. Also on the page is a menu selection to hide/show some layers. One fo the layers has the Flash FLV video and plays ok when the page is loaded. But if you select another menu item I want to stop the video from