Sudden spike in strain measurement

Hi!
I have a PXI 1062Q Chassis, 8108 controller (Windows OS), one PXI 4472 module(Slot 2) & two PXI 4331 modules(Slot 3 & Slot 5). The test scenario involves measurement from strain gauges mounted on a plate which is subject to some impact. But when I try to acquire strain data in the quarter bridge configuration from all the 16 channels at 100 K sampling rate, I find that there is a sudden spike in the data apart from the one observed due to the impact (Please refer attached pics). This spike is seen at exactly the same instant across all the channels. The following are the settings used for strain channel acquisition.
Acquisition Mode :Continuous Samples
Sampling Rate :100K
Samples to Read :10 K
Maximum Limit :1.5m
Minimum Limit : -1.5m
Strain Configuration :Quarter Bridge I
Voltage Excitation Source :Internal
Excitation Voltage :2.5V
Gage Resistance :120
Gage Factor :2
Note: Offset nulling & Shunt calibration are enabled (100K shunt resistor in R3 location)
The application was built using LabVIEW 2011 and DAQmx driver version 9.3.5
Can somebody please tell me what could be the possible reason behind this phenomenon?
Thanks in advance!
Shweta JP
Attachments:
Data with spike_1.png ‏34 KB
Data with spike_2.png ‏36 KB

Shweta_JP wrote:
Hey Lynn!
It's just a 1 sample spike in the positive & negative direction. Could an electrical transient result in such a small duration spike??
Shweta JP
EMC spikes could be of any length....
A good software spike filter is the median filter, here a filter rank of +-3 should do the job...
Looking at your picktures it seems that they appear in a close relation to your trigger ...  
Crystal_ball_mode: You wrote impact test.... is there some sort of solenoid/relay/inductive release involved ?   Without snubber they can be a source for such spikes.... 
Spoiler (Highlight to read)
Greetings from Germany
Henrik
LV since v3.1
“ground” is a convenient fantasy
'˙˙˙˙uıɐƃɐ lɐıp puɐ °06 ǝuoɥd ɹnoʎ uɹnʇ ǝsɐǝld 'ʎɹɐuıƃɐɯı sı pǝlɐıp ǝʌɐɥ noʎ ɹǝqɯnu ǝɥʇ'

Similar Messages

  • Why does my track volume on selected track suddenly spike?

    Why does my track volume, on the SELECTED TRACK, suddenly spike to max as if I had volume automation set. It never did this before a few weeks ago. Very frustrating. It happens during playback and loop recording. It started around the same time I started using a new audio interface (steinberg UR22) but I don't think that is causing the issue because I wen't back and used my MBOX to be sure, and it did the same thing. I have never had any use for automation or a master volume track either. It doesn't matter if I have a guitar, bass or mic going direct or through a preamp. It seems to be internal to GB settings. The only change I've made to my macbook recently was I increased my RAM from 2 to 4gb.
    It's like Garageband has a mind of it's own and it feels as though there isn't enough signal reach the track to it boosts it itself. What am I missing?
    Please help I can't get anything done because it spike in the middle of a take and I have to start over.
    I'm using GB 11 version 6.0.5

    Hello 3mb2d,
    It sounds like you are experiencing some unexpected behaviour when you are trying to use Logic Pro X. I recommend starting at step 3 of the following article and working your way down as needed:
    Logic Pro X: Troubleshooting basics
    http://support.apple.com/kb/HT5859
    Ensure Logic Pro X is up to date
    Installing the latest update ensures that your software has the latest improvements and enhancements.
    Choose Apple () menu > Software Update and make sure that you have the most recent update for Logic Pro X.
    You can also download and install updates from the Mac App Store. See Mac App Store: Update OS X and appsfor more information.
    Thank you for using Apple Support Communities.
    Take care,
    Sterling

  • Strain measurement using 9237 and cDAQ 9172

    I am trying to measure strain using a 9237 and cDAQ 9172.
    The strain gage is connected to 9945 (Red wire to EX+, White to IN+ and Black to QTR terminal also tried connected the white and black to IN+ terminal).
    The 9945 is connected to 9237 quarter bridge strain measurement module on a cDAQ 9172. I have two 9237 on the cDAQ 9172.
    The device passes the test in MAX.
    The problem is I cannot read the strain value.
    Can anyone help me please?
    Attachments:
    Strain measure.vi ‏19 KB

    Hi jujare,
    In the VI you attached, you don't have the bridge configuration set up. The default is full bridge I, so if you have a quarter or half bridge wired up, that is one reason you could be getting bad data. I'd suggest you take a look at "Cont Acq Strain Samples(With Calibration)-NI 9237.VI" in the Example Finder. It shows the configuration for bridge type and has some handy wiring diagrams. You could also add controls to your vi to set the bridge configuration if you'd prefer.
    If your strain gauge only has three wires, it's probably a half or quarter bridge. A full bridge needs at least four wires.
    I hope this helps!
    Regards,

  • Configuring NI9219 for strain measurement

    Hi!
    I am trying to set-up a strain measurement VI using Labview. I am using NI 9219 module on NI cDAQ 9172 chassis.  My strain gage is configured in a quarter-bridge arrangement and I have made strain gage connection accordingly (to terminals 3 & 5 in NI 9219). However, I do not understand what are TEDS terminals for and what to connect to those terminals. Can I use virtual TEDS for this? I couldn't find a virtual TEDS for NI 9219. Any information in resolving this issue would be highly appreciated.
    Best,
    Nyquist99
    Solved!
    Go to Solution.

    Hi, the TEDS terminals are only for transducers, like a strain gauge, that support TEDS (Transducer Electronic Data Sheet).  The TEDS standard is just the ability of a transducer to save information about itself in internal memory.  This information includes things like calibration constants and dates, scaling factors, electrical characteristics, etc.  It is stored in a standard template which can be read by computers and DAQ cards via those TEDS lines.  For you application, you should just leave them unwired, and setup up your strain gauge normally.
    Chris Bakker
    National Instruments
    Applications Engineer
    Check out LabVIEW 2009 and the New X-series DAQ!

  • Load cells and non contact strain measurement

    I am designing a bi-axial machine which meant for a fatigue test.
    Could someone recommend me where to get a dynamic load cells and as well non-contact strain measurement?
    Thanks.

    Hi Alvin788,
    Do you presently use any hardware to measure these or are you enquiring as to what hardware/software you will need? Are you planning to use LabVIEW? The following three links talk about measuring Load Cells as well as Strain Gauges 
    Continuous Acquisition from a Smart TEDS Load Cell Sensor
    How do I Configure a Load Cell in MAX?
    Strain Measurement with a Strain Gauge
    This post will also be highly helpful.

  • Strain measurements using labview and DL750 scopecorder

    Hi
    I want to use labview to do strain measurements using DL750 Yokogawa Scopecorder. I have installed the DL750 drivers and I  can access them from labview, but I am a bit lost as to getting to communicate with the scopecorder, setting up measurements and so on. I am all new to strain measurements and to labview, is there a link or a tutorial that anyone would recommend I refer to?
    Regards, tickle

    Hi Tickle
    It looks like the driver you are using is supported by a third party. I took a look at the driver you are using and it has a couple of examples included with it. I would recommend taking a look at these examples as they are normally a good starting point for communicating with your device. Has anyone else used this device in LabVIEW?
    Best Regards,
    Adam G 
    National Instruments
    Applications Engineer

  • Strain measurment and write to file with NI9237

    Hopefully this lab view VI can help some of the novices like myself to acquire strain using the NI9237 and write to a file.  The example VI available from NI "Cont Acq Strain Samples (with calibration)" is fine for sampling data and viewing on a chart, but you need to add your own elements if you want to logg data or write to a file.
    Since aquistion is continuous, reading and writing data must be done at discrete intervals rather than continuously, or, you will end up with a very large file that can not be easily handles in excel. Also, data gets acquired in "packets" of several samples for each channel, so even if you put a write function within a case structre, as I have done ,and you don't convert the data you get that whole "packet" every time the comand writes to file. Again, you will end up with a lot more data points than you would like.
    Though the top level VI is pretty simple, it took me some time and help from many folks for me to figure out what was going on.  I thought I would share.
    Thanks to Sandra T, Jarrod B, Kyle B, Marc A, Herbert E and Jesse O. for your help
    Next goal is to acquire voltage from a different Daq card, convert to load and graph and write load vs. displacement.
    hopefully the attached files are the correct method so the VI fuctions with its sub-VI's.  If not let me know.
    Regards,
    Alfonso
    A. Lopez
    Attachments:
    Cont Acq Strain-Qbridge-I-write to file-NI9237.vi ‏199 KB
    Measure Strain.llb ‏238 KB

    Alfonso,
    I'm glad you got your application running.  It is also nice that you are wanting to share your code.  NI is currently testing a new system that allows community members to share code.  If you like, you can also submit your code here.
    Regards,
    Jesse O.
    Applications Engineering
    National Instruments
    Jesse O. | National Instruments R&D

  • Errors in strain measurement system

    I am trying to measure strain using SCC SG04, SC2345, and M series PCI 6221. The system is saying that it is not installed properly even though MAX is recognizing the devices. If anyone has set up a similar system could they please help out. Project deadline is getting very close and i am in a bad way

    Dr. Imad wrote:
    what is not installed properly ? the M card or the SC2345 ?
    Once the card is installed, go to max and verify that you see it in Max. Once this is done
    under
    devices and intefaces
    right click---create new
    select DaqMx SCC-connector block-----SC2345
    Then you configure the SC2345, Type, where SG04 is installed (J1....) etc
    then you should be able to access it from LabView. You may want to use Daq assisitant first.
    Dr. Imad wrote:
    what is not installed properly ? the M card or the SC2345 ?
    Once the card is installed, go to max and verify that you see it in Max. Once this is done
    under
    devices and intefaces
    right click---create new
    select DaqMx SCC-connector block-----SC2345
    Then you configure the SC2345, Type, where SG04 is installed (J1....) etc
    then you should be able to access it from LabView. You may want to use Daq assisitant first.
    I have done as you recommended. The Daq board does show up in max. I have added the connector block and the strain module. That seems ok until i do a self test and it fails. It says in the error message that the device may not be working properly or is not properly installed. I have removed the whole system and reinstalled it only to get the same error message. In the device manager in windows 2000 it says that the DAQ card is working properly. When i go to the drivers tab it says that the driver provider is National Instruments but the date and version are unknown. It is not digitally signed. I have tried to install a driver for this namely NIDAQ7.3CD1.zip and NIDAQ7.3CD2.zip but this doesn't make any difference. The DAQ assistant didn't get it working either.

  • 1520 Strain Measurement

    Hello All,
    I am using SCXI 1520, Terminal Block 1314 and DAQ 6035 E Card, for
    Strain Gauge Measurement.
    We have got a Strain Gauge of 120 Ohm with the NI Hardware which was
    connected to CH02(One of the 8 channels provided), Now my problem is
    that if I connect the Strain Gauge to some other channel(except Channel
    2) I am not able to get any readings , and if I reconnect it back to
    Channel 2 it gives prorper results.
    I have connected the strain Gauge in
    a Quarter bridge Configuration(obviously for a Single Strain Gauge) as
    per the circuit diagram provided correctly.
    I have Pentium -4 , 256MB RAM, Win XP SP2 machine.
    Could anybody please try to explain where I am going wrong.
    With Regards
    Sandeep

    I agree, please check this measurement in MAX by setting up a DAQmx Strain Gauge Task, or by using the test panels associated with the chassis if you are using Traditional DAQ.  What happens when you change the channel you are using and test it using MAX?  Is the signal responsive, but inaccurate?  Is it just displaying 0, 10, -10? 
    If you are able to measure correctly, then just make sure you are changing the channel in your program.
    If the signal is there and responsive with what is happening to the strain gauge, but inaccurate, you may have an offset error that can be corrected with the calibration resistor.  The completion resistors of the 1520 should be checked as well.  Make sure they are the correct values. 
    http://digital.ni.com/manuals.nsf/websearch/56E77CCD7B41AA8386256968005BBB73
    The DAQ Driver version would also be helpful if you are still having problems.

  • Angle and Strain Measurement

    Hey Guys,
    I'm a LabView newbie and want to program a LabView-VI which should write the measurement of a Heidenhaim angle measurement device and multiple strain gauge strips in one data file like this:
    angle | strain1 | strain2 | strain3 ...
    The strian gauge strips are connected to the Laptop with NI9237 modules and a NI chassis via USB. The Heidenhaim system is also connected via USB.
    Can LabView identify the angle measurement system or do I need additional drivers?
    I found this post (http://forums.ni.com/t5/Instrument-Control-GPIB-Serial/Heidenhain-ND-287-connect-with-VISA/td-p/2372... and tried to figure out the given example "Advanced Serial Write and Read.vi". But I would be happy if anybody has more information, examples, codes, etc.
    Thanks for your help

    Hi,
    here is an example to measure the angle with a Heidenhaim device :
    http://forums.ni.com/t5/LabVIEW/Angle-Measurement-Using-Absolute-Encoder-and-LabVIEW-FPGA/td-p/14114...
    Aurelie

  • Dynamic strain measurement

    Sir,
    My experiment involves the measurement of strain on the specimen on the application of torque.
    I'm using SC-2043SG for the quarter bridge ,with an internal excitation of 4.13V.
    The problem is that when i apply 1.5 Nm and 2Nm torque, both are showing the same strain, 243 microstrain.
    Whereas in the digital multimeter they show .7mV and .9mV respectively.
    The strain is supposed to vary , which is not happening for every 0.5Nm. The multimeter is showing a
    variation of .1mV for every 0.5Nm increase in torque. But this is not being shown in the strain output. Only
    when it reaches 3Nm it shows 486 Nm. But this remains the same even for 4Nm.
    external signal conditioning is also not helping.

    thanks for replying ,
    The DAQ device i m using is PCI -MIO-16E4
    The DAQmx Device driver version is 7.5 0f 1
    im attaching the program i made for my application.kindly see through it
    thanks a lot
    Attachments:
    indu new.vi ‏131 KB

  • Strain measurement with nı 9235

    hello everyone
    I am user of NI 9235 and  my question is about measurement of strain gages. I attached the gages to material propertly and connect to nı9235  with embeded gage factor and ets. then I calibrate just with using button of calibration as attached pictures. I am not sure am I measure right data. I checked NI website, there is something mention about shunt calibration but it doesnt offered at NI9235.
    Could anyone give some information about my problem?
    Attachments:
    IMG_20150323_121018.jpg ‏5751 KB
    IMG_20150323_121005.jpg ‏5594 KB

    Hi,
    Here is a National Instruments white paper http://www.ni.com/tutorial/7130/en/ that walks you through taking a strain gauge measurement. This should assist you in setting up your measurememnt. What about the data makes you think it is not measuring correctly? 
    I have found some other resources that may prove helpful http://www.ni.com/white-paper/11488/en/
    http://forums.ni.com/t5/Signal-Conditioning/shunt-calibration-NI-9235/m-p/2460276
    Please let me know if these are helpful or if you are still having problems after this. Also it appears you are trying to use a DAQ Assisitant and from what I have read you cannot due a shunt calibration using DAQ assistant.
    Matt P.
    Applications Engineer
    National Instruments

  • Ni 3214 strain measurements

    I am currently trying to create a LabVIEW program for the night 3214 that measures and detects strain placed on an object for my senior design project. The order has been placed for the node and while my group waits for it to arrive we are trying to set up the code using the simulated components provided in the LabVIEW wsn module. using the example program provided in the manual for the node I have questions on how to set up a data base for logging data, as well as how to set strain limits for the node so that the warning system will go off.
    Thanks in advance

    Westbury,
    I would suggest posting this to the Wireless Sensor Networks forum over at http://forums.ni.com/t5/Wireless-Sensor-Networks/bd-p/430
    The people who view that forum are much more likely to be able to help.
    Matt J
    Professional Googler and Kudo Addict
    National Instruments

  • Preview eats up CPU usage/sudden spikes

    Hello I'm an owner of MBP 13" 2011 10.7.2
    When running Preview to open large pdf files, one of the mdworker processes spikes up to huge CPU usage, as well as Preview process.
    Thus, I'm having hard time reading pdf files. Is there something I can do to prevent this?
    Thanks.

    It looks like you have more than enough RAM, and your system should be able to handle most documents. But 384MB is quite large; in many cases, authors may choose to chapterize a large document in order to cut down on really large PDFs. For scale, a 900 page PDF book is often no bigger than 35MB.
    In short, the graphics haven't been optimized, and you'll have difficulties opening a document like this on a desktop computer.
    My suggestion: chapterize upcoming documents (if you can), or save as another type of document in Preview/PDF/Save to Postscript.

  • SUDDEN SPIKE IN DATA USE?  ACCURACY OF DATA OVERAGE ALERTS?

    Our data usage over the past 6 months averaged 2.25 GB  Top usage during that period was 2.99 GB in May
    Since I had been paying for 6 GB and never using it, I switched to 4 GB in May and signed up for Data Usage Overage Alerts.
    Then...all "heck" seemed to break loose???
    May 23 Received data alert via text: Have Used 75% of 3 GB--Added 1 GB
    May 27 Received data alert via text: Have Used 90% of 4 GB
    May 29 Received data alert via text: Have Used 100% of 4 GB-Added 2 GB
    June 1  Received data alert via text: Have Used 75% of 6 GB
    June 7  Received data alert via text: Have Used 90% of 6 GB
    OUR BILLING CYCLE ENDED JUNE 9
    June 11 Received data alert via text: Have Used 100% of 6 GB (???)
    Stunned that between the May 10 - June 9 billing cycle my reported data usage suddenly went over 6 GB for the first time ever, and not doing anything differently at all, I called Verizon Customer Service...
    The rep asked if I had been doing something differently with my phone (apps, photos, etc)
    I replied that my usage had not changed in any way...
    To which she responded that she was unable to track where/why my data use was suddenly so high...???  She had "No way of knowing"
    NEED HELP FROM VERIZON AND/OR COMMUNITY: 
    1.     ARE VERIZON DATA ALERTS UNRELIABLE?
    2.     HOW/WHY WOULD DATA USAGE INCREASE SO DRAMATICALLY WITH NO WAY FOR CUSTOMER TO CHECK/VERIFY?
    MANY THANKS...

    Like radar guns used by local police departments, I firmly believe that without some sort of independent oversight like cable companies use to check the meters Verizon wireless can tell you anything and you the consumer gets hit with high data use.
    My best solution for you is to reset the data counter on the phone starting with the first day of your billing cycle. Then check your online data used and see if they match.
    I also recommend you contact the FCC or even keep taking them to small claims or a superior court to have them produce records of data meter accuracy. At present no known outside concern can monitor what Verizon is claiming you are using.
    Get a good lawyer and take them to court.
    Good Luck

Maybe you are looking for

  • Windows 7 pro problem with redirected folder for 1 computer in the domain only.

    Hi. We have a computer in the domain we have that started to behave strangely. We have redirected folder on the GPO for the Desktop, favorites, and my documents folder. The server is 2012 essentials Sometimes the customer want to save it's document a

  • Changing audio

    I have the audio from the original video but I want to silence3 this and put my own mp3 on - just downloaded the trial version - adn having trouble finding out how to do this. I was able to add something called smart audio (or something like that) -

  • Displaying Custom Infotype in the Overview Section of LSO_PV10 and LSO_PV1A

    Dear experts, We would like to display the custom IT in the Overview Section of Create Course Run (LSO_PV10) and Change Course Run (LSO_PV1A) screen. User currently can create and edit some standard ITs from this overview section. Currently, the cust

  • Can have two iCloud different account in the same device?

    Can I have a two different iCloud account in the same device?

  • Transfer Binary JPG from Table to AL11

    Hi folks, I have the problem, that I can not transfer a JPG-File from a Table (stored in bin-format) to AL11 directory properly. When I transfer the files to my local harddrive, everything is fine and I can open the picture. When I tranfer it to Al11