LV 2010 and PCI-MIO-16​XE-50

Hi, I'm trying to use a PCI-MIO-16XE-50 with LabView 2010. Would it be possible?
I insert the DAQ card but I cannot see it from Measurements and Automation explorer. NI-DAQmx 9.3 is installed.
Any suggestion is welcome, thanks!
Solved!
Go to Solution.

That card will work fine with LabVIEW 2010. I sounds like it just isn't inslalled. This may have to be done when updgrading to a newer version of LV. 
When you do a hard reboot of the computer, look for the Found New Hardware wizard, and follow the instructions.Otherwise, try Add Hardware in Control Panel.
Richard

Similar Messages

  • How to reduce reading and recording data time/ Visual C++, PCI 6024E and PCI-MIO-16E?

    Right now I am working on a data communication programme. The time of reading digital signal and recording data is more than I expected, which is about 100 microsecond. I need to reduce the time. The software I am using is Visual C++ 6.0. I use two DAQ boards, which are PCI-6024E and PCI-MIO-16E. I have set the priority of Visual C++ to be realtime by task manager. The operating systme is Windows XP.
    I am wondering how to set the interrupt of reading or writing data to be the highest and stop any other kind of interrupt.
    Thanks in advance!

    The primary limitation you will encounter when trying to do high speed
    digital I/O using your E-Series hardware is that the digital lines are
    all software timed.  This means that for each and every digital
    read it is necessary for your software to issue a command to your
    hardware, read a single point, then receive the response from the
    board.  This prevents you from performing digital operations at
    very high speed.  In general, I would say that 100 microseconds is
    a very good response using software-timed digital I/O on a Windows
    computer.
    It is difficult to make a good recommendation without more detail about
    what you are trying to do.  You may need to consider switching to
    a Real-Time operating system, or upgrading your DAQ hardware to a board
    that will allow you to time your digital acquisitions on the board
    itself.  These devices, such as the m-Series DAQ or the PCI-653x
    series, would allow you to use hardware triggers for your digital
    acquisitions, change detection, and buffered data acquisitions to
    reduce the software overhead involved in a rapid acquisition.

  • Does pci-mio-16e-1 and pci-mio-16xe-10 function the same?

    Please excuse my lack of understanding for this hardware. I am not an engineer, just a computer tech tasked to build new pc's with this hardware.
    I need to know if the two cards can be interchanged. One is a 12 bit card, the other 16bit. The 12 bit card is the calibrated card to be used, replacing the 16 bit card.
    Thanks

    I'm not exactly sure what you mean by interchange, but both use the same driver, NI-DAQ. So if you have a version of NI-DAQ that is compatible with both devices (the last few versions are), the driver will recognize both devices. NI-DAQ will probably associate a different device number for the new board, when you put it in. So programs written for the other board might have to be changed slightly to account for the different device number. FYI - you can download (free) the most recent version of NI-DAQ here:
    http://digital.ni.com/softlib.nsf/webcategories/85256410006C055586256BBB002C128D?opendocument&node=132060_US
    -Alan A.

  • Can i measure resistance with a PCI-MIO-16X-10

    good day!
    In my work we measure inflator resistance with an external ohmeter and switch selector, for now we're working in update this method of measure resistance. The program used to activate the inflators was built in labview 7.0 and use a GPIB and PCI-MIO-16X-10 boards only.
    The question is if we can measure resistance with this board (PCI-MIO-16X-10), we can do it?
    thanks for all...

    Cordero,
    Not directly but with some auxiliary circuit elements you can.  To measure a resistance you need to know both the voltage and current associated with the resistor.
    One way you can do this with a DAQ device is to create a voltage divider using a known external resistor. By measuring the voltage drops across both the known and unknown resistances, you can calculate the value of the unknown resistance.  You can use the analog output or the board power supply as excitation voltage for the voltage divider. Use two analog input channels to measure the voltages across the resistors. As long as the values of the resistors are much lower than the input resistance of the analog input channels, you can get good results.  The accuracy is limited by how well you know the value of the "known" resistor, the resistance of the inputs, and the accuracy of the A/D converter.  The accuracy of the known resistor is probably the largest contributor to the overall error.
    Lynn

  • Data acquisition and generation with PCI-MIO-16E-4 card help!!!!

    Hello,
    I'm using a PCI-MIO-16E-4 card with labview7 (in using DAQmx)and I
    would like to draw a frequency response.
    My problem is: my program can run correctly in low frequency but, it
    can't work correctly in high frequency (after 2KHz the signal is
    false).
    For verifying my program: I only try to generate and acquire a sweep
    sine. I constate after 2KHz the sine I read dosn't resemble to a sine
    I wrote.
    Can you say me where is the problem?
    Thank you.
    Le Trong Thao.

    "Dr. Imad" wrote in message news:<[email protected]>...
    > my understanding is that you are generating a sine wave (on one of the
    > anlogoutput channels) and you are trying to read the signal.
    > Since you can verify the signal at low frequency, but not at high
    > frequency, it looks like you should increase the scan rate of the data
    > acquistion (if you can, set it to 10 times the signal frequency)
    > You can also use an oscilloscope to verify the signal is what you
    > really generated.
    Thank you for your response.
    I've tried to do it, and I've tried several other possibilities but it
    doesn't work correctly.
    In fact, I think the problem is in my program. Do you have a program
    that works correctly? I f
    ound examples in the help but they don't run
    correctly.
    Le Trong Thao.

  • Can I generate sine and square waves with NI PCI-MIO-16XE-50 card??

    Can I generate sine and square waves with NI PCI-MIO-16XE-50 card??
    How Do I generate those signals?

    Hello;
    You certainly can do that. The way to go about that is to use one of the Analog Output channels your board have available.
    If you are using Labview to program the board, you can find good examples at Search Examples->I/O Interfaces->Data Acquisition->Analog Output->Multiple Point (buffered) Analog Output.
    If you are using other Software language to program the board, you can fing examples at C:\Program Files\National Instruments\NI-DAQ\example\VisualC\AO.
    Hope this helps.
    Filipe

  • Problem integrating PCI-MIO-16XE-50 and AMUX-64 T multiplexer

    I can test the PCI-MIO-16XE-50 card in MAX but i can not test the virtual channels

    Hey Hiral,
    The AMUX-64T does not support virtual channels. You'll need to acquire the data and then perform the required scaling in software. The following document discusses how to access the appropriate channels.
    How do I Address an AMUX-64T with an MIO Board?
    This next document may also be useful. It discusses how to perform the necessary scaling for a thermocouple channel in LabVIEW.
    Taking Thermocouple Measurements Using the AMUX-64T
    I hope this helps!
    Sarah Miracle
    National Instruments

  • I am trying to generate a 50kHz sine wave for ten cycles, then aquire this waveform and store it in a file, I am using a PCI-Mio 16E-4 board, has anyone done anything like this?

    I am trying to generate a 50kHz sine wave output on the DAQ card and then simultaneously recieve. I am using a PCI-Mio 16E-4 series board to do this. Has anyone done anything similar. I can generate a 50k wave no problem, and recieve one, but I can not get these vis to run together? Any suggestiosn

    Jrod, yes you should be able to do this, actually there are quite a few example programs come with LabVIEW that handles simultaneous analog input and output, I check the specs of the board you use, the board should be able to handle the sampling/generation rate, but make sure for 50KHz sine wave, you would sample at least twice as fast, preferably ten times the generation rate, then this puts the sampling of the board to thew limit.
    The program you can try is attached here, good luck with the application,
    XD Gao
    Applications Engineering,
    National Instruments
    Attachments:
    Simul_AIAO_Buffer(E-series).vi ‏104 KB

  • Is it possible to start a PCI4472 and a PCI-MIO-16E-1 simultaneously using an analog trigger (with pre-trigger)?

    I would like to start several PCI 4472 and a PCI-MIO-16E-1 simultaneously. All boards are connected via a RTSI cable.
    My program works fine if I use software trigger, or an analog trigger from a PCI4472 channel. However, the analog trigger works only when I set pre-trigger (or pre-scan) to 0.
    Is it possible to start a PCI4472 and a PCI-MIO-16E-1 simultaneously using an analog trigger (with pre-trigger)?
    Thanks.
    Ian Ren

    Hi, Bill
    I think it is possible to set more than 38 pre-trigger scans on a single 4472 card. I've done this before. You can verify this by running the Labview example "Acquire N - Multi-Analog Hardware Trig.vi" which comes with LabView.
    What I try to do but without success/luck so far is to start data acquisitions of several 4472 cards and a PCI-MIO-16E-1 card using an anlog trigger (with pre-trigger).
    Thanks for your help.
    Ian

  • How to do a single channel DAQ using NI-DAQ driver software with a PCI-MIO-16XE-10 card

    Hi,
    I would like to find out how I could do a simple one channel Data Acquisition from a signal generator using the NI-DAQ driver software with a PCI-MIO-16XE-10 card.
    I have written some test problem but even when the signal generator is turned on/off I get back some weird values.
    Here is my code.
    CString sFunctionName("");
    double volt[OUTPUT_POINTS*2];
    double out[OUTPUT_POINTS*2];
    short timebase, ready, code, stopped;
    unsigned short sampleInterval;
    int i, status, count=0;
    unsigned long update, points;
    short* ai_buffer;
    short output_ch_vector[16];
    int local_ITERATIONS = 2;
    SAFEARRAYBOUND bound[1];
    double dataItem = 9.9;
    long j;
    long k;
    double* pTheValues;
    LPTSTR lpsz_ErrMsg;
    // Initialise device
    status = Init_DA_Brds (deviceNumber, deviceNumberCode)
    Initializes the hardware and software states of a National Instruments
    DAQ device to its default state and returns a numeric device code that
    corresponds to the type of device initialized
    Init_DA_Brds(DEVICE, &code);
    // Check return code from Init_DA_Brds
    Code return should be 204: PCI-MIO-16XE-10.
    if (code < 0)
    CString sError;
    sError.Format("Code error: %d", code);
    if (code == -1)
    sError = sError + ": No device found";
    LPTSTR lpsz = new TCHAR[sError.GetLength()+1];
    _tcscpy(lpsz, sError);
    AfxMessageBox(lpsz);
    delete lpsz;
    return S_FALSE;
    // Allocate memory for analog output and input arrays
    //ao_buffer = new short[OUTPUT_POINTS*2];
    ai_buffer = new short[OUTPUT_POINTS];
    // Set double-buffering
    status = DAQ_DB_Config (deviceNumber, DBmode)
    Enables or disables double-buffered DAQ operations.
    status = DAQ_DB_Config(DEVICE, 1);
    if (status < 0 )
    sFunctionName = "DAQ_DB_Config";
    goto TidyUp;
    // Get the rate parameters
    status = DAQ_Rate (rate, units, timebase, sampleInterval)
    Converts a DAQ rate into the timebase and sample-interval
    values needed to produce the rate you want.
    status = DAQ_Rate(RATE, 0, &timebase, &sampleInterval);
    if (status < 0 )
    sFunctionName = "DAQ_Rate";
    goto TidyUp;
    // Setup scan
    status = SCAN_Setup (deviceNumber, numChans, chanVector, gainVector)
    Initializes circuitry for a scanned data acquisition operation.
    Initialization includes storing a table of the channel sequence
    and gain setting for each channel to be digitized
    status = SCAN_Setup(DEVICE, 1, ai_channels, gain);
    if (status < 0 )
    sFunctionName = "SCAN_Setup";
    goto TidyUp;
    status = SCAN_Start (deviceNumber, buffer, count, sampTimebase,
    sampInterval, scanTimebase, scanInterval)
    Initiates a multiple-channel scanned data acquisition operation,
    with or without interval scanning, and stores its input in an array
    status = SCAN_Start(DEVICE, ai_buffer, OUTPUT_POINTS, timebase, sampleInterval, timebase, 1000);
    if (status < 0 )
    sFunctionName = "SCAN_Start";
    goto TidyUp;
    while(count < local_ITERATIONS)
    // Check whether we are ready to input another half-buffer
    status = DAQ_DB_HalfReady(DEVICE, &ready, &stopped);
    if (status < 0 )
    sFunctionName = "DAQ_DB_HalfReady";
    goto TidyUp;
    if (ready == 1)
    status = DAQ_DB_Transfer(DEVICE, ai_buffer, &points, &stopped);
    if (status < 0 )
    sFunctionName = "DAQ_DB_Transfer";
    goto TidyUp;
    count++;
    // Clear the analog input
    status = DAQ_Clear (deviceNumber)
    Cancels the current DAQ operation
    (both single-channel and multiple-channel scanned) and reinitializes the DAQ circuitry.
    status = DAQ_Clear(DEVICE);
    if (status < 0 )
    sFunctionName = "DAQ_Clear";
    goto TidyUp;
    status = SCAN_Demux (buffer, count, numChans, numMuxBrds)
    Rearranges, or demultiplexes, data acquired by a SCAN operation
    into row-major order, that is, each row of the array holding the
    data corresponds to a scanned channel
    status = SCAN_Demux(ai_buffer, OUTPUT_POINTS * 2, 2, 0);
    if (status < 0 )
    sFunctionName = "SCAN_Demux";
    goto TidyUp;
    //Convert binary values to voltages (Doesn't actually take a reading from board)
    status = DAQ_VScale (deviceNumber, chan, gain, gainAdjust, offset, count, binArray, voltArray)
    Converts the values of an array of acquired binary data and the gain setting for that data
    to actual input voltages measured.
    status = DAQ_VScale (1, 0, 1, 1.0, 0.0, OUTPUT_POINTS , ai_buffer, volt);
    if (status < 0 )
    sFunctionName = "DAQ_VScale";
    goto TidyUp;

    Hello,
    Please take a look at lots of examples available at :
    1. www.ni.com >> NI Developer Zone >> Development Library >> Measurement Hardware
    2. C:\program files\national instruments\ni-daq\examples\visualc
    Sincerely,
    Sastry V.
    Applications Engineer
    National Instruments

  • How do I add a multiplexer accessory device to my PCI-MIO-16XE-10

    I've been told I have to add a multiplexer to my PCI-MIO-16XE-10 card (so I can read more than the 10 chanels it provides for at the same time).
    How do I do this in the Measurement&Automation Explorer?
    Thanks for any help!

    Brain&Body,
    The card can function as either 8 channels in differential mode or 16 channels (Referenced single ended or None referenced single ended - later uses system ground). The single ended mode(s) will allow you get access to all the channels, though it would mean that you would expect to see more noise on the signal than in differential mode.
    You can select these modes within MAX under the properties for the specific DAQ card. Look under analogue input (AI) there you will see a pull down for the modes please select the mode you want. I would recommend that you read the user manual regarding the benefits and limitations of the modes. If using a developement tool (LV,CVI, VB or VC++) please make sure that you are reading single ended signals.
    If you wish to expand the system further you might want to consider purchasing a multiplexer system such as an SCXI system (reference page 184 of the 2005 catalogue). I would recommend that you contact your local sales office and speak with one of our Internal Sales Engineers - UK general number of +44 1635 572400.
    Kind regards
    JoeW
    NI-UK

  • Question about setup_MITE in the PCI-MIO-16E-1

    I have a quick question about setup_MITE in the PCI-MIO-16E-1. In one of the manuals, it states that if you're using Dell Optiplex GX series for our hardware configuration, is it true we need to re-write the MITE functions to correlate with the hardware? Please if you can clarify what this means, I would really appreciate it. Thank you in advance.

    National Instruments is starting a new newsgroup to provide a forum for customers to discuss register level programming issues and ask questions. This newsgroup is not a direct support channel for register level programming, but may be monitored by National Instruments engineers. Please post you question with additional detail in:
    natinst.public.daq.driver-development.general
    Christian
    Christian Loew, CLA
    Principal Systems Engineer, National Instruments
    Please tip your answer providers with kudos.
    Any attached Code is provided As Is. It has not been tested or validated as a product, for use in a deployed application or system,
    or for use in hazardous environments. You assume all risks for use of the Code and use of the Code is subject
    to the Sample Code License Terms which can be found at: http://ni.com/samplecodelicense

  • Pulse Train Generation using PCI-MIO-16E-4

    I need to generate pulse trains of changing pulse widths using PCI-MIO-16E-4 daq card using Igor Pro.
    I have tried to use the CTR functions (but always keep getting errors and it doesn't work). I also used the general purpose counters (GPCTR functions) but cannot visualise them on a CRO. If someone can help in regards to the programming aspect, it will be greatly appreciated.
    Thanks
    Vakul

    Hi Vakul,
    Some questions that might help me to get you a solution.
    Are you using Traditional DAQ, or DAQmx and which versions are you using?
    What operating system are you running?
    When you say you can't visualize them on CRO, I assume that you are hooking the counter out to a cathode ray oscilloscope and can't see the pulse train?
    Are you going to need more than two different pulse widths for your pulse train?
    There are several online knowledge bases describing how to change the frequency of a pulse train on the fly.
    This link explains how to use software to create a new pulse width.
    http://digital.ni.com/public.nsf/websearch/7B1B0427C39FE33086256CEE00752133?OpenDocument
    This link discusses using frequency shift keying, which will only allow you to have 2 different pulse widths.
    http://digital.ni.com/public.nsf/websearch/EC1F089252F33B248625696E006FB6FC?OpenDocument
    I assume that you are writing this program to compile in Igor Pro using C. I can help you with the C code development, but don't know anything about the Igor Pro API.
    Please get back to me about those questions and let me know if you need further help.
    Michael Denton

  • Glitch Energy for pci-mio-16xe-10

    The E-Series user manual does not provide the "glitch energy" for the pci-mio-16xe-10 board's analog output. It DOES give the glitch energy for the other boards (pci-mio-16E-1, pci-mio-16xe-50, etc.). Why the ommision for my board? I know that my board does not allow reglitching, but I still need the normal glitch energy magnitude and duration. I need this information ASAP please!

    Hi,
    The card mentioned, unlike the other E series cards does not have a reglitching circuitry. Hence its specification is not done because it is too small to be noticeable, particularly for 16 bit boards.
    Please do let me know if you have any questions.
    Sincerely,
    Sastry Vadlamani
    Applications Engineer
    National Instruments.

  • Installing two PCI-MIO-16XE-10 DAQs

    I have two PCI-MIO-16XE-10 boards, which work fine independently. I now have put them into the SAME computer, and the Windows XP has successfully found them and installed drivers for them. But when troubleshooting using the NI Measurement&Automation program, neither card is working anymore. The self-test fails, and the error message does not explain why. How can I use both DAQs together in the same computer?
    thanks,
    SAK

    Hi SAK-
    There should be no problem with using both cards in your system.  A few things to check are:
    Make sure that NI-DAQmx and Traditional NI-DAQ are not trying to access the board at the same time.  If you use the board with Traditional NI-DAQ you will need to reset the Traditional NI-DAQ driver before using the board with NI-DAQmx again.
    Make sure that you have administrative rights on the PC you're using or the driver may not function properly.
    Make sure that no other applications are currently accessing the boards- the driver will lock out access when the board is currently in use by another application.
    You can further troubleshoot the issue by removing all other PCI cards from the system to try and identify the problem as a resource conflict between your DAQ boards and any other peripherals installed on the system.
    Hopefully this helps-
    Tom W
    National Instruments

Maybe you are looking for

  • Open external links in new window

    I can't seem to make firefox open external links, e.g. links opened from other applications, in a new window. This is particularly troublesome on multiple desktops since a new tab may be opened in a firefox window on a desktop other than the active o

  • Reinstall ScanjetG4010

    Tried to reinstall G4010 to new computer, windows 7.  Pretty sure I got all current drivers and ran the HP install sequence from their current website.  The scan light does not come on or make its normal pass over the scanning surface.  I uninstalled

  • Clearing email addresses using Comcast

    When I begin typing an email address into (Comcast email form) the "To" bar, an "intuitive" box pops up with possible email addresses to select from. Some of these addresses are old, and I want to clear them out. When I go to Safari's Preferences and

  • BRF+: No context assigned to DB Lookup

    Hello expert, I am trying to create a BRF+ routing rule but I am facing since hour a problem. I am refering to http://scn.sap.com/thread/3324445 When trying to create rules I got the message: No context assigned to DB Lookup 'Z_GET_ROLEID' yet. Query

  • Whenever I try to buy songs on my phone it goes through my credit card authentication process. When I select done it gives a message saying session timed out. What do I do?

    Whenever I try to buy songs on my phone it goes through my credit card authentication process. When I select done it gives a message saying session timed out. This has happened mulitiple times. What do I do?