Usb 6009 acquisition fall after 15min

Hello,
Like it's said on the name of my post i have some problem with my new usb 6009. When i start it all work good but after a while my two AI fall to 0V. I put my code in joint pieces. I use AI0 and AI1 with a 220ohm resistance.
Thank you
Samuel Rioux
Attachments:
test.vi ‏30 KB

Hi skling,
DON'T use an additional Wait function when you have set your DAQmxRead to use a certain sample rate!
It will wait on it's own and after some time you will miss to read samples in time...
Best regards,
GerdW
CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
Kudos are welcome

Similar Messages

  • Usb-6009 high-speed continuous acquisition for slow control loop application

    Hi,
    After watching a few Youtube videos, I successfully built a control loop in LabVIEW using my USB-6009. Currently, this loop is limited by the speed my actuator can take and respond to commands, topping out at 200 Hz. This is completely sufficient for my application. My question is more regarding the wasted potential of the USB-6009 to acquire samples at a much higher rate than what would be obtained if I used a simple single sample-on-demand during every iteration of the loop.
    It seems to me that having the USB-6009 continuously acquire samples and having my control loop periodically check in for the latest data when it is ready to execute would have the advantage of allowing me to average groups of samples together to lower the overall noise associated with the measurement process (I am reading voltages from an amplified photodiode).
    1) Does anyone know how I would instruct the USB-6009 to continuously acquire at say 10 kHz, so that on each loop I will have 50 or so samples that can be averaged together for a more accurate measurement of my control variable? Unfortunately, I'm new to this and not too familiar with DAQmx Timing.vi.
    2) I've heard that reading into a buffer can increase latency for moving data into the computer. Given the low speed of my control loop, I don't foresee this as being a problem. Is this a correct assumption? I can probably live with an extra millisecond of latency.
    3) I goofed around briefly with DAQmx Timing and managed to see a buffer overflow error. Certain rare events will by necessity cause my control loop to take much longer to execute, so I cannot guarantee that occasionally during normal operation I will not overflow the buffer. Can I suppress this error so that the dialog box does not pop up and the program continues to run uninterrupted? If the buffer overflows and old data gets overwritten it isn't a big deal for me since the control loop is only concerned with the latest data anyway.
    Thanks for any info!

    Howdy Patrick!
    Here is a picture of the front panel:
    Here is the block diagram:
    And I've attached a copy of the VI saved for LabVIEW 7.1.
    Regards,
    Barron
    Applications Engineering
    National Instruments
    Attachments:
    most recent samples.vi ‏73 KB

  • USB-6009 Temperature Acquisition

    Hi All
    I seem to have some brain fade today. I need to use the USB-6009 to with a J type thermocouple. When I hook it up the temperature is not correct and widely varies. I am getting reading from 0 to 60 C at room temperature. I need to incorporate this into another circuit for testing. I assume that I am overlooking the simple and obvious. I have tried using a DAQ Assistant and a DAQmx. Any help will be appreciated.
    Thanks
    Gerald

    Hello Gerald,
    You don't have a brain fade! I recently got in a few usb-6009's and am seeing a similar response when trying to measure temperature with thermocouples.
    I don't believe this is noise as much as resolution on the usb-6009. At this point I have added resistors to bring down noise but am seeing the same "waver" that you describe (although not quite as extreme). I am looking into building a signal amplifier or alternatively purchasing a  TAC80 thermocouple-to-analog converter for use with the usb-6009.
    Have you had any progress with your issue?
    Adzel

  • Usb 6009 continuous analog output

    Hi
    I'm pretty new to Labview, so you may have to bear with me. I have a USB-6009 and I want to generate a continuos analog output voltage. I've got the example you have posted 'Gen Mult Volt Updates-SW Timed_LV7.1 (0 to 4).vi' working and it outputs a continuous sine wave.
    In a similar manner I need to generate:
    1. Ramp up voltage (and ramp down voltage): starting with a continuos constant voltage, which after a period of time linearly increases to another constant voltage, and which allows me to specify the 3 time intervals and the rate of increase.
    2. Pulsations: whereby I can specify the amplitude and time interval of each pulsation
    This is so that I can control voltage to a valve which regulates my pipe flow. In the example I have (one stated above), I do not understand specifically:
    i. Why I need both a 'Software Loop Time (ms)' as well as 'Timeout'
    ii. The 'Sinusoidal Pattern' input which seems to generate data for a sine wave- I can't get a description of that and there are no similar VIs for other waveforms; and what is the function of the two numbers it has?
    iv. How do I change amplitude and time period for the wave?
    iii. What does 'Index Array' do?
    thank you

    Hi there,
    I'm guessing the VI you are using is the one from this KB:
    http://digital.ni.com/public.nsf/allkb/6F2C2B49A89D685C8625711D007BDD64
    i. The software timed loop control is to control the rate at which you change the voltage output.  The timeout on the DAQmx Write VI is the maximum time in seconds the VI will wait to output a sample (eg. if the write buffer is full, the Write VI will wait for it to become available for 'timeout' seconds before outputting an error).
    ii. The sine pattern is just an array constant of doubles that make up the sine wave voltage values, and the for loop adds an offset (of two volts) to every single one of those values. You might want to replace that entire array constant and for loop with a Simulate Signal Express VI (just search for Simulate Sig or look in the Functions >> Input palette) and convert the dynamic data output to an array of doubles.  You can configure the type of waveform, amplitude, and time period from that express VI as well.
    iii.  Refer to previous answer.
    iv.  Index array returns an element of an array based on the index input.  You can turn on the context help and move your mouse over functions to get more help on them.
    I hope that helps!
    Way S.
    NI UK Applications Engineer

  • USB-6009 digital input change detection

    Hi,
      I am trying to acquire digital signals from an ni usb-6009 device using daqmx on labview 8.0. I need to write the signals from a port of the device to an output file only when there is a change in the signals and I am not able to do this successfully. I created a digital task in MAX and was able to select acquisition mode as '1 sample (On Demand) ' only. Any other selection for acquisition mode would result in error 200077 with the following message -
    Measurements: Requested value is not a supported value for this property.
    Property: SampTimingType
    You Have Requested: Sample Clock
    You Can Select: On Demand
    I selected '1 Sample (On Demand)' acquisition mode and was able to see the channel values on running it in MAX as I rotated a device connected to the usb-6009 device. As expected the hex values on MAX were showing up as FF,7F,3F,BF and this pattern was repeated everytime I provided the input by rotating the device.
    Next, I created a simple vi to use the digital task I setup on MAX. I want to be able to record the signals only when there is a change in the digital pattern from the channels.That is when there is atransition from FF to 7F to 3F and so on, the vi should write to an output file and if no change happens, it should not write to file. To do this, I explored the Timing vi and tried using the 'Change detection' mode, but that produced the same error as above. 
    Can anyone please suggest a way to accomplish what I am trying to do ? Any guidance will be a hugely appreciated.
    Thanks
    ~Sej

    All of that effort with sample rates, timing, change detection, etc ciuld have been avoided if you had read the specs. As you finally found out, the device only supports sftware timed i/o and no change detection. You wll have to do your change detection in software this is done with a shift register/feedback loop.

  • USB-6009 Analog Waveform Output

    I'm trying to create an analog waveform from a USB-6009 device through one of the analog output channels. However, through the DAQ Assistant, I can only select the 1 Sample (On Demand) setting. When I test the device, I can only push a constant voltage out. If anyone has any advice on how to create an analog voltage waveform using this device, I would like to know.

    >I am only a novice at LabVIEW and data acquisition in general.
    I don't use Labview (I write in C++) so I can't help you with that. I try to answer questions about C/C++ and general hardware questions.
    >What did you mean by software-timed waveform generation?
    That's where the software driving the device sends a sample to the DAQ device, waits for the prescribed time interval, then sends the next sample. So the timing of the waveform is driven by the software, using the computer's clock, instead of by the DAQ hardware. Software-timed is less reliable and has (often) lots more jitter because of competition for CPU resources.
    >Does this mean that I cannot generate a waveform that is more than 150 Hz?
    It's worse than that. It means you can't generate a waveform faster than 150/N Hz, where N is the number of samples (data points) that define the waveform.
    John Weeks
    WaveMetrics, Inc.
    Phone (503) 620-3001
    Fax (503) 620-6754
    www.wavemetrics.com

  • Usb 6009 not recognized in simulink/matlab 2012b

    i have matlab 2012b and when connect DAQ ni usb 6009 matlab recognize it but when i open simulink library and Data Acquisition Toolbox and when add analog input it belongs "winsound Michrophone...'
    so any body can help me to simulink/matlab recognize usb 6009
    thanks

    Hello,
    The Data Acquisition Toolbox that you are using is supported by MathWorks®. It is supported under the version of the software that you are using. Please contact The Mathworks, Inc. for further troubleshooting steps.
    MATLAB® and Simulink® are registered trademarks of The MathWorks, Inc.  All other trademarks are the property of their respective owners.
    Wear
    National Instruments
    Product Support Engineer

  • USB 6009 - Calibration of Analog Input (Mac)

    Hi,
    I have recently acquired a NI USB 6009. I'm using it under labview 7.1 on Mac OS X 10.4 (Tiger).
    I can run the software examples and the datalogger without problems. However, the device seems
    not to be properly calibrated, and I can't figure out how to do it from the manual. When I connect
    pin 1&2 (analog in 0) to pin 32 and 31 (GND and 5V ref voltage) the software claims that the
    measured voltage it ~3.7V (a multimeter confirms that it is in fact 5.0V). Can anyone help
    trouble shooting this ?
    Thank you in advance,
    Niels

    A little update : After thinking some more I realized that it's because
    the USB 6009 is in differential mode, and I hadn't connected one pin. Now
    however, I have trouble figuring out where to change/set this. None of
    the programs (e.q. Acquire One Voltage.vi) seems to define this, so where
    is this defined ?
    Thanks,
    Niels

  • USB 6009 Data Acquisitio​n noise

    When I acquire the voltage from USB 6009, It gives some fluctuation/Noise (Please check the image, green color).
    What is the reason for that?. How to overcome this noise? One way is averaging, is there any other ways to do or overcome this problem
    Could you please suggest me?
    Thank you,
    Best Regards,
    Rathan
    Attachments:
    Noise.GIF ‏16 KB

    Dear Rathan
    The correct Terminal configuration depends on if your power supply is floating or grounded. Please have a look at tabel below.
    As to the filter, which type of filter to use depends on the frequency of your signal. If your frequency (or frequencies) is low a lowpass with a cutoff frequency a bit over your frequency would be best. Conversly if your frequency is high a highpass with a lower cutoff frequency would do the job. If you are interested in only a single or small spann of frequencies a bandpass could be used.
    You can filter your signal either via a hardware filter before you acquire it via your DAQ, but that would require that you have or can build a hardware filter that fits your needs. Else you can filter the signal in software after you've aquired it. The filters in LabVIEW can be found in the functions palette under Signal Processing -> Filters. I think a butterworth filter would do nicely. You can find an example through the LV Help.
    Best Regards
    David
    Message Edited by DavidEk on 07-29-2009 05:03 AM
    Attachments:
    Connection Configs.PNG ‏66 KB

  • USB 6009 - mysterious problem

    Have mysterious problems with my new USB 6009.
    I'm using LabView 7 on MacOS-X 10.3.8.
    All vi's of DAQmx show "ERROR 15 - Ressource not found".
    Then I tested:
    - "lsdaq" shows the device
    - "nidatalogger" works
    - "mxbaseconfig" doesn't recocgnize it!
    I realy have no idea know what to do now?
    Thank you for helping.
    Greetings,
    Stephan

    I happend to stumble over the same issue and found a solution.
    The problem really seems to be the firmware in the USB-device. There is
    only one way (I think) to make USB600x devices work with linux:
    You have to start MAX (on a Windows-box) , go to tools - NI-DAQmx configuration - NI-DAQmx USB Driver Selector.
    Press "Switch" and select the BASE-Driver for your USB-Device. This
    seems to write something into the device's flash memory. After this
    procedure you can use the device with linux and DAQmxBase
    Funny thing is, mxbaseconfig still can't see it, but nidatalogger finds it.
    Hope that helps, even though this thread seems to be rather old...
    -DB

  • Usb 6009 freeze

    Hi all,
    When using the USB 6009 interface for reading out two sensors (temperature and pressure, using the Voltage they generate) and sending signals to valves, my program keeps working.
    But after a while the USB module crashes.
    Most of the time both input and output don't seem to work anymore. I don't get any error, it just doesn't work anymore. When I pull out the USB device and insert it again in the USB port, it works again perfectly
    FYI (It happened once or twice that the outputs still worked, but I got no input from the devices.)
    Does anyone have similar problems or knows how to fix this?
    Thank you in advance.
    Regards,
    Bart
    Solved!
    Go to Solution.

    Here is the code.
    Attachments:
    Leakage_test_v1.1.vi ‏355 KB

  • USB 6009 DAQ to use in Simulink at External mode with DAQ toolbox

    hi
    Currently I am using usb 6009 daq board for data acquisition in using simulink. I have created a model with taking analog input from data acquisition toolbox. Then when i want to run in external mode and try to connet to target, the following error is coming up:
    Could not execute target data map file 'untitled_rtwin_win32\untitled_targ_data_map' or it does not exist. Stop the target, delete the untitled executable, rebuild the code,and try again. Note that execution of external mode requires the build directory to be present...
    How can I solve this problem? For your info, I am using 32-bit matlab, version: 2012a
    For your info, a snapshot of the error is attached,
    Any help would be appreciated Thanks
    Attachments:
    Untitled.jpg ‏130 KB

    Hi Mahmudur,
    what does MathWorks say about this error message?
    Best regards,
    GerdW
    CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
    Kudos are welcome

  • Usb - 6009 Device Recongition Problems

    Hi,
    I just got a USB-6009 DAQ. I have labview 7.0, drivers software and the supplied software installed with no problems. In the measurment nad automation explorer (MAX) the device is recongnised and declared problem free. I tried to test the device in labview using help tutorials which suggests the DAQ assitant block. After selecting Analogue input and then voltage I am told that no device is detected i.e. Where I should be selecting my terminal(s) I am not giving any options.
    If anyone has had the same problems or know how to fix them I would be thankful of your help.
    Kyle

    Hi Kyle,
    The USB-6009 device works on DAQmx base. This is a restricted version of NI-DAQmx which is used on some of our products. Unfortunetly it is incompatible with DAQmx and the DAQ Assistant uses DAQmx. Therefore you can't use the UDB-6009 with the DAQ Assistant.
    To use the device, load up the USB-600x Interactive Control which should be located in the following directory.
    C:\Program Files\National Instruments\LabVIEW 7.0\examples\daqmxbase
    Take a look at this and figure out how the code works. If you want to configure it to do your own specific functions then I would suggest that you use this program as a base and customise it for your purposes.
    Best regards,
    Peter H.
    Applications Engineer.
    NIUK.

  • USB-6009 Daqmx Base 3.3 Cannot creat 2 or more channels on Windows mobile 6

    Hello,
    I have USB-6009, PDA iPAQ 214, WM6, Daqmx base 3.3, Labview 2009. In order to create new mobile project I use wizard and standart dynamic example: Cont Acq&Graph Voltage-Int Clk.vi. All works ok, but if I add more chanels in VI "Create Channels" or place new VI block and enter "Dev1/ai1", exe on my PDA cannot work correctly. Exception after RUN. I have no idea how build correct application with 3 channels without this exeption...  On PC all works. Whay right way to create channels? I attach VI which don't work, for examle. I read daqmxbase cannot support WM6, but 1 channel work ideal!
    Attachments:
    diagramm.jpg ‏113 KB

    Vital000,
    I forgot to mention that, yes, Windows Mobile 6 is not supported along side of DAQmx Base. That means you might be able to get some  of the same functionality(limited) that you would with Windows Mobile 5, but most likely not 100% functionality. This could explain why 1 channel works and multiple channels do not. I will try and help you out, but as this is not supported my ability to help is limited.
    Ben Sisney
    FlexRIO V&V Engineer
    National Instruments

  • USB-6009 analog input at certain voltage

    i am using USB 6009
    i want to use a certain voltage to activate something.
    i want to activate a pop up vi inside a case structure at a voltage of 2.9V.
    i want to de activate the case structure at a voltage of 0.13mV - 0.16mV . then the sub vi will close.
    How to setup the DAQMX?
    can provide any help ??
    example will be good.
    pls show where to find the icons. thank you

    Hi Ace,
    There are several examples from the LabVIEW Example Finder (Help»Find Examples...) that describe using limits. I searched "limits" in the Example Finder and found several examples that might help, such as Limit Testing Measurement.vi, Waveform Scalar Limit Comparison example.vi, and Limit Example.vi. In each, you would have to modify it to include data acquisition, but it shows how to take sampled data as you would get from data acquisition and perform a limit test. The Express VI that Jennifer O. suggested will probably be the easiest route, but there are other VIs that could suit your needs, as seen in the other examples. 
    As far as state machines, there is a template built into LabVIEW that can give you a starting point for creating your own state machine. Under the File»New... menu, you can choose the state machine template under VI»From Template»Frameworks»Design Patterns»Standard State Machine template.
    Please let us know if you have any specific questions we can answer about configuring your data acquisition or about a specific VI. I hope this helps!