Hello.  I am having difficulty accurately acquiring an AC voltage signal using a NI 9206 and cRIO.  I'm attempting to acquire voltage signals from two types of Magnelab Current Transformers: split core and rope.  In Labview, I first populate an array of size 2,500 with the sensor signal (DIFF mode), then compute the RMS of the array.  For the split core CTs, I am able to accurately acquire the correct voltage reading (verified by measuring the corresponding amp on the line using a Fluke 434 Power Analyzer.  For the rope CT, however, using the same array/RMS method, I am unable to accurately acquire the correct voltage reading.  Measuring the amp on the line using the Fluke 434 PA, the correct voltage from the rope CT should be 0.05v.  Using the 9206 (DIFF mode), the RMS from the array gives a voltage reading of 0.071.  Now the interesting part is when I measure the line voltage using two different True RMS DMMs, I get two different readings.  One multi-meter, a Klein CL2000, reads the correct voltage at 0.05v.  Another multi-meter, a Fluke 189, reads incorrectly at 0.071, the same as I get using LabView and the NI 9206.  I suppose the question is how is the Klein interpreting the signal differently than the Fluke 189 and the NI 9206 via LabView.  One difference between the split-core and rope CTs is that the rope CTs require an external power supply 12-30v AC or DC.  I am providing them with 12v DC.  I have tried multiple voltages in both AC and DC and still get the same incorrect reading.  I'm quite certain it is not a power supply issue, although perhaps the integrator in the rope is creating a unique signal.  Any ideas?  I appreciate any input.
Thanks,
J.Grant
Solved!
Go to Solution.

AK2DM:
Update - solution found
While the Butterworth low-pass filter sort of worked, it was NOT the solution.  I spent some time analyzing the rope CT signals by sampling at a very high rate.  The AC signal coming off the rope CTs is triangular NOT sinusoidal.  The RMS function in LabView assumes a sinusoidal waveform and essentially calculates the RMS as the average of voltage peaks (for each cycle) / sq.rt (2).  Since the AC voltage single emanating from a regular split-core CT is basically sinusoidal, the LabView RMS of the signal gives an accurate reading of the voltage signal.  Unable to find a RMS function in LabView for a triangular AC voltage signal, I ended up developing my own custom triangular RMS function in LabView.  My function analyzes the signal by finding the peak of every cycle within the signal, takes the average of these peaks, then divides this average peak by sq.rt (3).  see this link for formulas:
http://www.optics.arizona.edu/nofziger/OPTI%20380B/Lab%201/AVG%20and%20RMS%20Values%20of%20Periodic%...
Using the triangular method, I am able to accurately acquire the signal off the rope CTs. 
Thanks for your help!

Similar Messages

  • How can i find the gain/amplification factor used when acquiring a voltage signal using PCI 6036E

    in short, i am interested in find the sensor gain for used in feedback loop simulation

    Greetings!
    The gain setting defaults to unity gain, but it is selectable to x10 or x100, also. (This is your pre-digitizing gain, of course...after it's in digital form, you can tweak the gain to anything you want in your VI.
    Hope this helps.
    Eric
    Eric P. Nichols
    P.O. Box 56235
    North Pole, AK 99705

  • How to send TTL output AND acquire AI voltage data using USB-6211

    Hello,
    I am relatively new to Labview, so please bear with me.  I have a research application involving 2 pressure transducers and a high-speed camera.  I wish to acquire analog voltage data from the 2 pressure transducers.  However, at the start of the acquisition, I will need to send a single TTL output to trigger the camera.  This TTL pulse must be sent out at exactly the same time that the AI acquisition begins, in order to ensure that my 2 pressure measurements and camera images are 'synchronized' in time.
    Is this possible on the USB-6211 running with LabView 8.20?  I currently have a fairly simple LabVIEW vi that uses a software trigger to start an AI acquisition - I have attached it with hopes that it may help anyone willing to assist me.  I would prefer to be able to simply add something to it so that it will output a TTL pulse at the start of the acquisition.  
    Thank you in advance.
    Regards, Larry
    Message Edited by Larry_6211 on 12-19-2008 11:24 AM
    Attachments:
    USB6211_v1.vi ‏212 KB

    Hi All,
    I'd like to clear a few things up. First, you'll find that if you try to set the delay from ai start trigger and delay from ai sample clock to 0, you'll get an error. Due to hardware synchronization and delays, the min you can set is two. Note that when I say two, I am referring to two tick of the AI Sample clock timebase, which for most acquisitions is the 20MHz timebase. I modified a shipping example so you can play around with those delays if you want to - I find that exporting the signals and looking at them with a scope helps me visualize what is going on. The Manual has some good timing diagrams as well but it looks like you've already hit that. The defaults would give you a delay of  250ns from the start trigger - is this too high for your situation? What is an acceptable delay? I tend to think that "exactly the same time" is a measure of how precise rather than an absolute (think delays in cable length making a difference.)
    With all that in mind, I see a few options:
    Start your camera off of the AI start trigger (an internal signal) and just know it is 250 ns before your first convert. 
    Export the convert clock to use as a trigger. This assumes your camera can ignore the next set of convert clocks.
    More complicated option: Internally you have an ai start trigger, sample clock and convert clock. From your start trigger to the first convert is 250ns but if you export your convert clock you're going to get future convert clocks as well. One option would be to generate a single triggered pulse using a counter (start with the  Gen Dig Pulse-Dig Start.vi example) with the AI start trigger as the trigger for the counter, an initial delay of 250 ns, and a high time of whatever you want it to be. This should give you a singe pulse at very close to same time (on the order of path delays) as your first convert clock. 
    Hope this helps, 
    Andrew S
    MIO DAQ Product Support Engineer
    Getting Started with NI-DAQmx
    Measurement Fundamentals
    Attachments:
    Acq&Graph Voltage-Int Clk.vi ‏37 KB

  • Acquiring DC Voltage Signal from HP3457

    I have a question that I believe to be easy to solve. I am using LabVIEW to acquire DC Voltage from a HP3457A DMM. I can acquire the signal OK, but when I write the data to an lvm file only about every 8th measurement is correct (with the others being way off!!) I am using a modified VI from LabVIEW for the 3457. I have modified it to sample every second, plot the data, and write it to an lvm file. Anyone have any experience with this or any idea what may be going wrong here. Thanks in advance for all of your help.
    Attachments:
    3457.vi ‏132 KB

    Hello,
    I checked the VI that you sent and the only thing that I could see is not proper is that the LVM file is open and close every time you write to it but that doesn�t seem to be the problem.
    I removed the GPIB part and sent a constant number and it worked ok in the LVM, so I believe the problem is acquiring the data.
    I would suggest you to remove right now the LVM part and first make sure that the data acquisition is working ok. Just extend the output of the converter from string to number outside the loop, enable indexing and create an indicator. Stop the vi and make sure that the values that you see are the correct ones.
    Hope this troubleshoot helps.
    Ricardo S.
    National Instruments.

  • I am trying to log data from 4 voltage input signal using labview, but when i use DAQ assistant i am able to log data from one signal at a time only.

    I am trying to log data from 4 voltage input signal using labview, but when i use DAQ assiatant i am able to log data from one signal at a time only.I am trying to get all 4 input data to logged in a single file againt time. I am new to Labview, I need to sample this data within a couple of days can someone help please.

    Naveen
    Check out the info in the Analog Input section of the document linked below.  (Ignore the part about Global Channels.)  In Figure 5, notice that you can select multiple channels while holding <Ctrl> or <Shift>.
    Developer Zone Tutorial: NI-DAQmx Express VI Tutorial
    Kyle B  |  Product Support Engineer  |  ni.com/support

  • Using the DAQ-9178 and 9205 Voltage Module, how do I build a 16 Channel Voltmeter with a range of 0--50Vdc?

    I have already built an 8 channel 10:1 voltage divider using 1% resistors.  My problem is there is a lot of AC noise on the DC signal and the voltages don't match my Fluke DVM which is sitting beside.
    Solved!
    Go to Solution.
    Attachments:
    8 Channel Voltage Screen.JPG ‏172 KB

    A little more info... Only Channels one and two have signal going to them.  Ch1, (a0), is connected to a 9volt battery, and ch2, (a1), is connected to a variable bench supply set for 17Vdc.
    The DVM is rock steady, however the LabVIEW indicators show what I assume is a product of the AC noise.
    Any help would be appreciated
    Thanks so much
    -Matt
    Attachments:
    8 Channel Voltage Block Diag.JPG ‏269 KB

  • How to use voltage signals as input signals into USB-6211

    Hello everyone,
    I have a USB-6211 and a PDQ80A -quadrant photo detectot from thorlabs, i am trying to take three voltage signals from the PDQ80A into my PC using USB-6211 from NI. THese voltage signals are X,Y and sum (X+Y).  
    Looking forward to your reply.
    Thanks in advance,
    _Perseus 

    Technically, the question was in the title.  But of course, the answer to that is... Connect them using WIRES.   It's designed for voltage signals, just wire your cell outputs to whatever DAQ input channels you're going to read.
    Using LabVIEW: 7.1.1, 8.5.1 & 2013

  • I recenty ran into a problem when measuring 6 equal voltage signals. When using a g

    ood quality digital meter all voltage signals where equal and also using the labview test panel all voltage signals where equal. When measured by my vi i found that the voltage signals where not the same 2 of the channels displayed almost twice the amount.After changing the channel order using the I/O tabs on the right of the control panel I found i got the correct voltage readings. What caused my vi to diplay incorrect voltage readings due to the order of my I/O channels?I am using Labview 6.0 with a SCB-68 shielded I/O Connector Block interfaced with a NI-DAQ Multifunction I/O board PCI-MIO-16E-4.
    Attachments:
    6_Channel_Data_Logger-Rev0.vi ‏1 KB

    ood quality digital meter all voltage signals where equal and also using the labview test panel all voltage signals where equal. When measured by my vi i found that the voltage signals where not the same 2 of the channels displayed almost twice the amount.Hello;
    I couldn't open the attachment you posted. However, you probably set the wrong input limits for your gains or had a too small interchannel delay in between channels that wasn't letting the instrumentaion amplifier to settle down properly.
    A good way to make sure your VI is the cause for the problem is to run one of the Labview shipping examples to do the acquisition. The best one to try is the one named Continuous Acquisition and Chart(buffered).vi.
    Regards
    Filipe A.
    Applications Engineer
    National Instruments

  • How can I use Labview to create a graph of digital voltage signal vs. Time?

    I am measuring a digital voltage signal and I need to generate a graph vs. Time and be able to add user controls to start/stop the graphing and be able to export the graph to other programs such as MS Excel. Any help you could give would be much appreciated. Thank you.

    You probably want to use a chart. Place it inside a while loop that runs continuously. The chart should be in a case structure so it is bypassed depending on the boolean pause/continue switch.
    To save the data, get the "history data" of the chart via a propery node and save it to a file. It is probably easiest to save it to an ASCII file. Most other programs will be able to read it.
    See how far you get. If you get stuck, post your VI and we'll help you.
    LabVIEW Champion . Do more with less code and in less time .

  • Acquiring signal using DAQmx 6221

    I'm new to using DAQ. I need some pointers on how to go about the following
    issue. I want to acquire a sine wave through DAQmx (PCI 6221). I need to
    acquire the signal at a particular rate and store it as a binary and ASCII
    file. Also I want to know how to output a signal through DAQmx. Could you
    please give some pointers and example VIs if any?
    Thanks,
    Sukanya

    Go to Help>Find Examples. Then expand Hardware Input and Output>DAQmx. Also look at Getting Started With DAQmx.

  • Whether could I generate one AC signal and acquire one DC signal and one AC signal together using NI myDAQ?

    Dear Sir:
    I would like to know whether I can generate one AC signal and acquire one DC signal and one AC signal together using NI myDAQ? If I can, could you please tell me the easiest way to achieve the goal? If I cannot, please tell me whether I can use both white box to do this because I have got another white box USB-6008. Cheers
    Kind Regards
    Jeff 

    duplicate post
    Don't create a new thread. In the thread above, you asked for someone to review your code but you did not post a modified VI. Your original one is obviously wrong as I noted.

  • Acquiring multiple voltage and current sine waveforms, and log them along with the respective time stamp in a file."

    "I am doing a data acquisition project using a 16 channel N.I. series 6020E card where I want to capture respective current and voltage sinusoidal signals. I am planning on capturing four pairs of matching current and voltage signals. I planned on taking one reading across the channels every fifteen minutes. This reading will need to be 60 data points per period or 3600 readings per second per channel in order for me to recreate the sine signal to determine the phase angle difference between the voltage and current. I was able to get one channel to capture a signal but could not get more than one working at a time. I am also trying to sycronize the data collection so all samples are with respect to each other. Thanks for any help."
    -posted by cappy on 2/11/2002
    John N on 2/12/2002 answered:
    "Hi Cappy,
    Could you post this question to the Hardware > Counter/Timer or Hardware > Multifunction I/O forums? I think you will get more responses there. Here is a link.
    The hardware is usually a better differentiation since software can support such a huge range of functionality.
    The data rates you are talking about sound reasonable. Are you using the USB or the AT interface?
    One other thing to be aware of is that the samples taken will not be simultaneous due to the multiplexed nature of the MIO board measurements. However if you know the channel clock rate(output from one of the VI's), then you can calculate the time difference between each measurement and correct your phase measurements by that amount.
    If you have more questions, please post this to the other forum and copy all of our discussion to let everyone know what we have talked about already.
    Best Regards,
    John Nieri
    Applications Engineer
    National Instruments"
    I am using the USB connection. If anyone can help with or has a similar VI for this, it owuld be appreciated so I can work on the larger aspect of the total project.

    The easiest VI to use for this acquisition is AI Acquire Waveforms in the data acquisition palette in LabVIEW. This VI allows you to perform a finite acquisition from multiple channels on your DAQ board. As John mentioned the acquisition from individual channels will not be truly simultaneous. There will be a slight time shift from one channel to the next on the order of the maximum acquisition speed of the card, 100kHz or 10 microseconds in your case. By accounting for this 10 microsecond delay between consecutive channels you can properly calculate the phase shift between voltage and current.
    Christian L
    NI Consulting Services
    Christian Loew, CLA
    Principal Systems Engineer, National Instruments
    Please tip your answer providers with kudos.
    Any attached Code is provided As Is. It has not been tested or validated as a product, for use in a deployed application or system,
    or for use in hazardous environments. You assume all risks for use of the Code and use of the Code is subject
    to the Sample Code License Terms which can be found at: http://ni.com/samplecodelicense

  • How to input/output a digital signal and acquire an analog signal at the same time?

    Dasylab, version: 8.0.04
    Acquirement Card: PCI1002L
    When I use DasyLab to acquire the analog signals is no problem without digital inputs and outputs,
    and when I use DasyLab to input or output a digital signal is no problem also, but when I do that at the
     same time, DasyLab tell me the rate is too high and stop.
    so, I searched the manual book1 (user guide) for that, it showed me :
    To internally equalize measurement time and system time in the analog input, digital input and counter
    hardware modules, use the following settings:
       Synchronization: PC Clock
       Sampling rate: <= 5Hz
       Block size: =1
    the problem is, if I set the Sampling rate to 5Hz, the speed of the acquirement datas is not enough for my
    application.
    so, how to improve it? who can give me a example programm for me. thanks!
    by the way, I come from China, my English isn't good, I'm sorry.
    Allen, China.

    Hi,
    Have things changed over the years?
    I need to syncronise a digital output (Modul NI9474) and an analoge input (AI-Modul NI9203) module. I need to measure time intervals from a flank in signal A to a flank in signal B. I would like accuracies of the order of 1 ms. Currently, the signals are not synchronised, with errors of the order of 2 times the block length (block size x sample rate), sometimes much higher. The best I got so far was a block size of around 20 with a sample rate of 1 kHz.
    If I use the master and slave settings on the RTSL settings, my program doesn't run properly.
    If I use digital signals for input and output, I can syncronise them with RTSL settings and everything is good, but I can't always do that.
    Also, if I do anything in the GUI (such as scrollowing something or going to another window), my output gets screwed up properly.
    1. What can be done to synchronise AI with DO?
    2. Is there something that can be done to avoid messing up the output when something happens in the user interface? (I know that I am messing up the outputs as they make some valves switch and that is loud).
    Thanks in advance!

  • Can you monitor signals using frame api on a PDA?

    Hi all
    I'm pretty new to LV all together, but I'm working on writing a program to monitor information off our CAN network, like oil pump speed, water temp, batt voltage, etc.  Is there a way to monitor these signals using the frame api, opposed to channel api, and display them continuously on a gauge/meter?  Since I'm using PDA module, I have to use the frame api, and not channel.  Frame api seems a little more vague to me than channel api.  Any advice or tips are welcome.
    Also on a side note, I copied/pasted a tab that i changed the background color to yellow from a windows targeted program to a PocketPC one.  The tabs stayed yellow when I built and transfered the .exe to the pda, but i don't have any control now on its color because PDA module doesn't have Property Nodes.  Every now and then, the tabs will go to default grey, and i have to undo the changes to get it back to yellow.  Is there some way to control the color?  I don't think the Change Control Color works in this case.
    Thanks ahead for any comments and advice.

    ....By the way, i'm using a PCMCIA-CAN/2 (HS) card on a PDA running WinCE5.0 and soon to be one running WinCE4.2...

  • Acquire 2 digital signal with custom scaling (Engg Units)

    I am a newbie to DIO world.
    I have write a VI to acquire 2 digital signals. one from a load cell and other for motor rpm (pulses). I need to acquire these two signals and then convert them into engg units using custom scaling and write them to a txt file with time stamp.
    Please suggest best ways to accomplish this task.
    Thanks
    DAQG
    Solved!
    Go to Solution.

    Look at the examples on DAQmx in the example finder.
    You wouldn't actually be acquiring 2 digital signals.  You would be acquiring an analog signal from the load cell.  A digital or counter acquisition is what you would acquire from the motor.  Search for analog measurements and counter measurements in the example finder.  Some of the example should show you how to apply the custom scaling.