Can source and gate used to measure 2 signals using one counter?

I have one PCI 6602 card to measure measure the frequency. PCI 6602 has 10 channels, in which channel 5 and 10 doesn't have 'source' terminal.
We have to measure frequency from 16 signals using one PCI 6602.
From channel 1 to 4 and 6 to 9, we have 8 'Gate' and 8 'Source' terminals.
We have connected 16 signals to these 16 terminals.
At any given point of the time we want to measure either 'Gate' signals(8) or  'Source' signals(8).
Can any provide any guidance to make a program in labview to measure frequency (period) as per above given points?
Regards
Vijay

Hello Vijay,
It takes one counter for each digital frequency you want to measure and the PCI-6602 has 8 counters total.  This means that the maximum number of digital frequencies you can measure is 8.  If you want to measure 16 frequencies you'll need 16 counters.
Single counter frequency measurements are made by comparing the board's time base (80 MHz) to an external frequency.  The board counts the number of time base periods that occur during one period of your incoming frequency.  When you configure the board to take a single counter frequency measurement the card routes the time base to the source of the counter and then you connect the frequency you want to measure to the gate of the counter.
Here are a couple of articles that explain how these measurements are taken and calculated.  (The second article is on period measurements so for frequency measurements you measure period and take the inverse):
Making Accurate Frequency Measurements
Period Measurements with a Counter
There are example programs available in the LabVIEW example finder.  You can locate these by going to Help>> Find Examples... The examples you want is:  Hardware Input and Output>> DAQmx>> Counter Measurements>> Digital Frequency>> Measure Digital Frequency- Low Freq 1 Ctr.vi.
I hope this explains how counters work and why you need one counter per measurement.
Cheers,
Brooks

Similar Messages

  • Is there a way we can edit and improve existing youtube videos by using premiere pro? Just simple changes like having professional looking opening and closing credits?

    Is there a way we can edit and improve existing youtube videos by using premiere pro? Just simple changes like having professional looking opening and closing credits?

    Assuming you do not have access to the "original" video/project and the YouTube video is the only copy, there is a method. Try "YouTube Downloader" which is a free software that allows you to download clips from YouTube as .mp4 files. You can then edit with Premiere, and re-encode again for upload to YouTube. Just keep in mind that you are re-compressing the video so there will be some quality loss. The better the quality of the original, the better the copy will hold up.
    Thanks
    Jeff Pulera
    Safe Harbor Computeres

  • HT201320 Unable to step my outlook account keep getting warning unable to verify account I can send and receive mail on my iMac using the outlook account HELP

    Unable to setup my ipad for my outlook account I keep getting the message unable to verify, although I can send and receive mail from my iMac using the same outlook account.

    Confirm you did this: https://support.google.com/mail/answer/1173270?hl=en
    You might also try removing all gmail passwords from your keychain in Keychain Access. Then connect again and enter the password code given by Google.

  • I want to transmit and receive the transmitted audio signal using labview with usrp ...can i get some help to do soo...thanks in advance

    plz help me

    this example receives frames on the same port. By CAN design you cannot receive the frame you are transmitting. If you want to receive what you transmit in addition there are 2 options:
    - use the second port of your CAN device for Receiving the frames you transmit
    - use the Self Receiption mode (refer to the manual for further details ho w to use this mode)

  • How can I output frequency readings of my signals using FFT?

    Hi there, I need help in regard to the topic stated.
    Currently I have accquired a sine wave with frequency that vary, using one of the accquistion card by NI. I had converted it to a digital signal so that labview can read it. Now, I will like labview to output the measurement of the frequency.
    For example : sine wave of 5KHZ is accquire ,  labview will use FFT to convert time domain to frequency domain, and then output the frequency of the sine wave, in the case, 5KHZ.
    What function block should I be using? I had explored the function blocks of all FFT. Most of them output either real FFT, complex FFT, magniture or phase. But none of them output the reading of the frequency.
    Maybe I had missed out some function block, so how do I go about output a frequency reading after using FFT?
    Thanks a lot in advance.

    After Converting your signal from time domain to frequency domain, u cant expect there is only one frequency component,they should have different frequency components according to your input signal.Use extract single tone information to get the output reading of the frequency.
    Balaji PK (CLA)
    Ever tried. Ever failed. No matter. Try again. Fail again. Fail better
    Don't forget Kudos for Good Answers, and Mark a solution if your problem is solved.
    Attachments:
    Detect Freq.vi ‏15 KB

  • Can source and target tables be the same

    Hi there,
    I need to restate profit center column in the delivery fact table. Historical Profit Centers need to be populated since today is the first time we brought in Profit Center field from LIPS table. The source for the delivery fact table is SAP LIPS table. so what I did is:
    joined the LIPS and delivery fact table on (LIPS.delivery number = deliveryfacttable.delivery number & LIPS.delivery line item number = deliveryfacttable.delivery line item number) to get the profit center.
    So the source and the target delivery fact table are the same. Is this a good practice becuase Data services warns me that the source and target tables are the same.
    Please let me know a better alternative approach to this OR a better approach to restate fields for historical data. Thanks in advance.
    Regards,
    samqiue

    Arun,
    actually the LIPS resides on another server and in order to fetch data I have to write an R3 ABAP data flow, so I cant use it directly in a lookup.
    Except what Im thinking based on your reply to use a lookup table is: that I will extract 3 colums (delivery number, delivery line item number and profit center) to a table and then use the lookup.
    Thanks.
    Regards,
    samique

  • Can I and how do I set up using to ipods on one itunes account.

    Hello,
    I have an itunes account and today my wife was given a shuffle, can I use the same account or do I have to set her up a new one. If so how to a download some of the songs I have on my account onto her ipod.
    Thank you.

    can I use the same account
    Yes, if you wish.
    or do I have to set her up a new one.
    Only if you want to.
    If so how to a download some of the songs I have on my account onto her ipod.
    Plug it in and sync it.

  • Can start and stop in the inspector be used 2 delay when music start/stops?

    I figured out how to add music to my site (home page) and it loops which is great. However, I want
    a 3 second delay at the start of the loop, while keeping the music on AUTO, so it starts when visitors come to my page Can this be done? Thanks!

    Yeah...I've worked out an alternative, but it would be very nice if there was a feature like what I'm talking about. Basically, a start and end memory location that is multi configurable to whatever you want. It can loop, or just go back to the start location and stop. And every press of rewind takes you to the start.
    I've come up with the key command, go to left locator or right locator, and don't enable cycle mode.

  • Can swipe and pinch to zoom functionality be used in creating an interactive pdf through indesign?

    A client is telling me that there is a way to add these functionalities to an existing indesign file but I cannot figure out how. They would like to have the exact same features that an iBook has, but on a standalone pdf. Is this possible through indesign, or do I need a 3rd party software? I did manage to speak to a company that does this, and all they told me was that they take the indesign file, which is a horizontal layout of a book, and add the swipe and zoom functionality that exactly mimics those of ibooks.

    Not in PDF. That sounds like DPS.

  • Measure rpm and PWM signals using digital input of NI9401 module

    Hi, is there a simple way how measure rpm (to 5000 rpm) and PWM (about f=3kHz) signals using digital input of NI9401 (ultrahigh-speed digital I/O)  module plugged in NI 9172 ?  I will wellcome any suggestions and VI examples (bether still). I'm begginer using LView 8.2. I apologize for my poor English. Many thanks.

    Hi CUA,
    Since you have already the right tools, you just need to get started with the software. The NI-9401 will give you access to the counters that are inside the cDAQ chassis 9172.
    You need to place the module in slots 5 or 6. You can read in how they work and how to configure them in the counters section of the user guide found here.
    Now to measure rpm, you can either measure the period of the signal (seconds) and multiply by 60 to have minutes and the number of ticks per revolution. You can find a shipped example to read period called “Meas Dig Freq-Buffered-Cont-Large Range 2 Ctr.vi” found under Hardware Input and Output » DAQmx » Counter Mesurements
    To measure PWM, it is similar procedure and you can read the semiperiod (high time or low time). There is also a shipped example called “Meas Semi-Period.vi” found under the same location.
    There is great information for these kind of task at this page or just post back if you need to.
    Hope this helps.
    Gerardo O.
    RF Systems Engineering
    National Instruments

  • 6602 on Traditional DAQ : can't assign gate(0) to source input of counter (0)?

    We are replacing a system that used PC-TIO-10 boards with 6602 boards.
    In an attempt to minimize system changes, the new system has been wired
    identically to the old system, which used the gate lines to count by
    opening the gate on the rising edge of the gate line and counted on the
    falling edge.
    Apparently, we are prevented from using the gate line as source to a
    counter. It is confusing enough that the manual says each of the fourty
    counter I/O lines can be assigned to any of the counters and any of the
    counter functional inputs: I still don't know if this is true or if it
    is an improper interpretation of the manual.
    Additionally, using DAQ 7.4, we are specifically limted to using source(x) lines as source inputs to the counters.
    a) Is it really not possible to connect a gate() line to the source input of the same counter?
    b) If so, is it possible to connect the gate(x) input to the source
    input of counter(x+1) or (x-1) without affecting counter (x), which is
    how it appears in the VI's?
    c) If worse comes to worse, we are driving the gate inputs with the
    output of an MC14490 hex contact bounce eliminator. Can we tie the gate
    and source inputs to a single counter together, and in software set the
    source as input and set gate on, and count events that way?
    and I guess,
    d) we're hopelessly  crazy and misinformed, and have to remake the
    interface boards and/or completely rewrite the software to use DAQmx?
    ray

    We are replacing a system that used PC-TIO-10 boards with 6602 boards. In an attempt to minimize system changes, the new system has been wired identically to the old system, which used the gate lines to count by opening the gate on the rising edge of the gate line and counted on the falling edge.
    I don't understand the function you were accomplishing.  What do mean by "opening the gate" on the rising edge?
    It is confusing enough that the manual says each of the fourty counter I/O lines can be assigned to any of the counters and any of the counter functional inputs: I still don't know if this is true or if it is an improper interpretation of the manual.
    No, the assignments are not that flexible.  However, you certainly can assign any counter's default source pin to be used as a source by any other counter's task (or by several tasks at once).  There are a few other special-case assignments, but for the most part they are flexible within their own category.  Any source pin can be the source for any counter, any gate pin can be the gate for any counter, etc.
    a) Is it really not possible to connect a gate() line to the source input of the same counter?
    b) If so, is it possible to connect the gate(x) input to the source input of counter(x+1) or (x-1) without affecting counter (x), which is how it appears in the VI's?
    c) If worse comes to worse, we are driving the gate inputs with the output of an MC14490 hex contact bounce eliminator. Can we tie the gate and source inputs to a single counter together, and in software set the source as input and set gate on, and count events that way?
    a) I'm not at hw now and can't recall for sure about using a counter's own gate signal as a source.  See b).
    b) I'm pretty sure a counter can use as its source the gate signal from its paired counter (x+1 or x-1).
    c) Um, it isn't clear to me why you need both source and gate connected this way if you are just counting events.
    d) we're hopelessly  crazy and misinformed, and have to remake the interface boards and/or completely rewrite the software to use DAQmx?
    I don't think so.  If the PC-TIO-10 was just performing event (edge) counting, then you shouldn't need to worry about the gate inputs on the 6602.   The new counters can simply count source edges without regard to a gate input.
    Can you describe how the source and gate inputs were used to accomplish the measurement you needed on the PC-TIO-10?  It's been a long time since I used one and I only ever used it for period measurement...
    -Kevin P.

  • HPCM custom driver use source and destination

    Hi All,
    I'm using HPCM 12.1.2, I need your help to create one custom driver that uses information from source and destination. I need to use 1 dimension of source and 2 of destination.
    I have the situation bellow:
    STAGE1 (SOURCE)
    Dim1 Machine
    Dim2 Account
    STAGE2 (DESTINATION)
    Dim1 Process
    Dim2 Operation
    Dim 3 Account
    Measure: Time used in the Machine
    The infromation is extracted:
    Machine1 - Process1 - Operation1 = 10 Hours
    Machine1 - Process1 - Operation2 = 9 Hours
    Machine2 - Process1 - Operation1 = 13 Hours
    How can I create one Driver that I could use these informations? It is possible?
    Thanks in Advance
    Diogo
    Edited by: user10432898 on 15/05/2012 14:21

    Hi Alex,
    You have to increase the Max Row size limit in Xcelsius.
    Goto >> File menu >> Preferences >> Increase the "Max Row Sixe"  by default it will be 500 you can increase it as per the requirement.
    Regards,
    AnjaniKumar C.A.

  • Can't Drag and Drop with a Double Tap using Single Finger in Lion?

    I installed the Mac OS X Lion two days. But I can't drag and drop with a Double tap using Single Finger, which I used to do in Snow Leopard.
    If I choose Three Finger Move, then Swap Between Pages goes back to "Scroll left or right with Two Fingers" that only works with Safari, but doesn't work with Chrome or Firefox.
    I tried literally every setting in Trackpad and also checked if there was a override settings and there is no overrride setting.
    Would be thankful, if you guys could suggest me what to do?
    ======================================
    My Machine Configuration:
    Macbook Pro 13" (mid 2010)
    Intel Core 2 Duo 2.4GHz
    4GB Ram
    Nvidia Geforce 320m
    Mac OS X Lion (Installe on a updated Snow Leopard 10.6.8)

    Hey Serge many thanks for telling me how to do. now its working like previously
    I just cheked the Dragging under Trackpad option and unchecked Three Finger Moves. now I can drag and drop with a double tap using single finger.
    Thank you guys. You are the best

  • SQL Server Source and Flat File Target in OWB

    Hello All,
    I have a question, is it possible to have SQL Server as Source and Flat File as Target without using any intermediate Oracle table(s).
    like
    SQL Server --> ETL Operators --> Flat file?
    as i m getting errors in data type conversions here. but if i replace flat file with Oracle table it works fine. (though i have used some conversion functions for data types as well in Flat file loading but it is giving me errors)
    Thanks in advance.
    Tayyeb

    Yes this should be fine. Are you using the gateway to access SQLServer or code template mappings? You need to track which columns are throwing the data type conversion errors.
    Cheers
    David

  • How to free rotate an area that has been measured using the measurement tool.

    I've measured and area using the measurement tool on one pdf drawing and now want to copy and paste it in to another pdf drawing and rotate it in to position. Is this possible?

    Object>Expand Appearance. Then proceed to Ungroup, release clipping paths, etc., to further dissect it. Exact steps will depend upon the specific artwork.
    JET

Maybe you are looking for

  • How to recover the for loop or seq() from a sequence result

    assume generate a sequence with code for i in 1..2 do     append 1 to a list to represent outer loop     for j in 1..3 do            append 2 to list to represent inner loop result in [1,2,2,2,1,2,2,2] is it possible to recover for loop code and seq(

  • Help with httprequests and responses

    it is my first time writing a servlet, but i need one to work with a project i'm working on. I have a class that will do all the methods on the server side of the project, but my problem now which i need to resolve, is how to get my client side app (

  • Does File Delivery support files to be stored in more than one format?

    Hi All, Can we have internal delivery channel configured to store files in more than one format. For eg. We want the files to be stored simultaneously in both xml and txt format as per requirement. Is this possible in Oracle 10g B2B.

  • 802.1X on switch 2950

    Hi, I tried to configure 802.1X on a 2950 switch, i can't connect from a pc. That's what i did on: * ACS aaa client ip IP_Switch authenticate : radius (ietf) key : xxxxx a certificate has been created on a ca server and installed on the acs (on same

  • Profile parameter name

    Hi Gurus, What is the profile parameter name which is used to increase the character length of user name? Thanks in advance.