Counting TTL pulses from 10 Hz to 3000000Hz with LAB-PC-1200

I want to count TTL pulses from an electron analyser. How can I use the DAQ wizard to connect the analyser to the PC-1200 and look if it works?

Augerspectro,
The counter chip used on the Lab-PC-1200 is the 8253. Thus, you will be able to use any of the 8253-counter examples available in LabVIEW. To count pulses, I would suggest that you examine the Count Events (8253).vi example. With regards to the required hardware connection, this is discussed in the device's user manual (link available below).
Lab-PC-1200/AI User Manual
Good luck with your application.
Spencer S.

Similar Messages

  • Counting TTL pulses from A between TTL pulses from B

    I have two sources of TTL pulses. I need to count, from a c/c++ script, how many TTL pulses are received from source A between each of the  TTL pulses received from source B.
    I have a PCI-6229 connected to a BNC 2090A, and both of my TTL pulses sources are connected via BNC plugs.
    Is there any function that can do that?
    I tried to get an example script using the DAQ assistant from LabWindows, selecting acquire signals -> counter input -> Edge count. The function that I got as a result was DAQmxCreateCICountEdgesChan, and I am afraid that it only takes the counter as an argument and not the second digital input that marks the intervals.

    From your description it seems to me that the best option is to use two-edge separation measurement. I am not on a PC with CVI installed so I cannot point you to a specific example but this tutorial gives you a brief explanation and some guidance on examples to look at.
    Proud to use LW/CVI from 3.1 on.
    My contributions to the Developer Zone Community
    If I have helped you, why not giving me a kudos?

  • Counting ttl pulses

    Hello Everyone,
    I am attempting to continuously record (i.e. time stamp) only positive (i.e. 'Logic High' event) TTL pulses from a PMT using a PCI-6602 board via a BNC-2121.
    From reading the LabVIEW tutorials/examples it seems that I should be able to get very high precision (15 digits) from the "Get Date/Time In Seconds" VI. I need to be able to record times to the microsecond (at least). Yet when I look at the data it appears that for every digit less than a millisecond either a 9 or a 0 is recorded, which leads me to believe that I can't get better time resolution than a millisecond. Is there anything I can do to change this?
    Assuming that I can get this to work I would also like to simultaneously monitor and record a second PMT channel. Any hints or tips would be great!
    I've attached my very basic VI below.
    Cheers,
    CJ
    Message Edited by CJKS on 05-22-2006 11:53 AM
    Attachments:
    PMTcounter.vi ‏126 KB

    Yes the simpliest way to do what you want is to count the time between rising edges using a counter (the 6602 has 8 80 Mhz 32-bit counters).  Yo will not use digital io but time counting.  The gate of your counter will be the TTL/Cmos pulse from your PMT and the source of the counter will be the 6602 master clock 80Mhz, so you can count clock ticks between photons, now you have a 12.5ns time resolution.  You can use the system clock to get a time zero if you would like to use absolute time.  The olny thing to keep in mind is that you will be limited to a maximum time between pulses of 2^32/80M or 53 seconds, you will get an overflow of your counter but usually this is not a problem with photon counting.  You will be also limited to a maximum photon rate of about 20Mphoton events.sec this is also not a problem since the interesting photon events seem to happen between these time scales.  if you want a longer time scale you cvan always casceds 2 counters to get 64bit resolution but this is probablu not necessary.  Look at the buffered period measurment examples.  You can easily scale this to 8 counters with 8 PMTs if desired with your curent hardware but this gets expensive.  Goodluck.
    Paul
    Paul Falkenstein
    Coleman Technologies Inc.
    CLA, CPI, AIA-Vision
    Labview 4.0- 2013, RT, Vision, FPGA

  • How to count TTL pulse by DAQ through BNC 2090?

    I'm trying to count the TTL pulse from a APD by DAQ. There are some very nice discussions about this issue in this board. However, I am a novice and not sure about the hardware connection. The signal from APD is delievered out through a BNC cable and I'm trying to deliever this signal to DAQ through BNC 2090. I tried just simply connect the BNC cable to ACH0 and tried running the vi's, such as "Count_Edges.vi" and "Count_Events.vi". But I got no counts. I noticed that both these vis are actually reading signal from PFI port. That could explain why I got not counts.
    So now the problem is how should I connect my signal in a BNC cable to the DAQ and have the counting working? Thank you for your help.

    Hi Opal,
    Here is what you should do for the hardware connections with these devices:
    - connect your external signal to the BNC-2090
    - connect the 68 pin connector on the BNC-2090 to the 68 pin connector on your DAQ board. You should have a 68 to 68 pin cable for this.
    That is all that is involved with the hardware connections. After this, you should be able to read your signal. Here is the link to a helpful DAQ manual with in depth getting started information: DAQ Getting Started Guide. I hope that you find this information helfpul.
    Regards,
    Hal L.

  • How to count TTL pulses during 100ms ?

    I need to count TTL pulses coming out of an instrument (particule
    counter). The pulses appear randomly. I must count how many pulses
    appeared in the last 100ms, and so on, every 100ms. I don't want
    cumulative counts, but the number of events every 100ms.
    I need a hardware clock, so that the counting pace cannot be disturbed
    by Windows making some weird, unrequested, CPU-hungry task.
    I fact, I need to do that for 2 identical instruments of the same kind,
    and I have 2 cards with 2 counters each, total 4 counters. My idea
    would be to use 1 counter as the clock for 2 other counters, but how to
    trigger the counter value read ?
    Thanks.

    Hi,
    In the example I gave you, you need to supply the TTL you want to count the pulse (connected on pin 37) and a TTL "clock"  which is connected to PFI0. The TTL on PFI0 permits a hardware timing (100ms for example) to count the edges that occur on the TTL source.
    If you do not need hardware timing, have a look at the example attached. You only have to connect the TTL source and the VI counts every 100ms (software timed) the edges that occur durind those 100 ms.
    Hope it will help you,
    Regards,
    David D.
    Application Engineer - NI France
    01-02-2006 01:26 PM
    Attachments:
    Compatge des fronts toutes les 100 ms.vi ‏61 KB

  • Counting TTL pulses at high speed

    Hi all,
    I am using PCI-6221 board with DAQmx to count the number of TTL pulses (which varies in its frequency between 0Hz to 10MHz) at a high speed (200,000 samples/sec.) and I am having a problem when the TTL pulse frequency drops below a certain level.
    I am using CTR0 to generate continuous pulse train at 200kHz frequency to feed to CTR1 Gate input. I verified that the pulse train is being generated fine.
    I am using CRT1 with buffered counting to collect the count for 200,000 samples at a time (duration of 1 sec.). I got the example code (Cnt-Buf-Cont-ExtClk) and pretty much used it as is.
    CTR1 Gate is coming from CTR0 Out, which is 200kHz pulse train with 50% duty cycle, and CTR1 Source is the TTL signal that I am trying to count. At first, I thought that everything was working fine with the Source signal being at around 5MHz. Then, when I had the Source signal down below about 300kHz, I noticed that the program is taking longer than 1 sec. to collect the same 200k samples. Then, when I got the Source signal down to 0Hz, the program timed out.
    I am guessing that somehow the counter is not reading for the next sample when there has been no change to the count, but I cannot figure out why and how.
    Any information on this and a way to get around would be greatly appreciated.
    Kwang

    One thing you can try is to set the counter input CI.DupCounterPrevention property, this setting will filter the input, it is possible that when the ctr 0 is slow then many of the values you are counting become zero as well and are filtered out, since they are nolonger points, the counter will not collect enough points before the time-out occurs and the counter input read times out.  I am not sure if this is your issue but I found out the hard way that this occurs when I switched to daqMX where this feature was added.  Let me know if it worked,
    Paul
    Paul Falkenstein
    Coleman Technologies Inc.
    CLA, CPI, AIA-Vision
    Labview 4.0- 2013, RT, Vision, FPGA

  • Counting TTL pulses within specified integration time

    Dear all,
    I have a PCI-6014 board, and I use Labview 7.1 and would like to do the following task with DAQmx7.4:
    Analog source:  analog square wave (or = 1Hz TTL pulses) produced by a function generator
    Signal: photon signals from a photomultiplier (random TTL pulses)
    I would like to count the photon signals at 100ms integration time, repetitively. But I don't want cumulative counts, just the number of photons (TTL pulses)  in every 100ms.  I would also like to make sure that I synchronize the counter with the analog source, i.e. to make sure I get exactly 10 data points in each analog cycle. ( I can't use the computer to start the function generator and it just keeps running., so I want to make sure I start counting on the starting edge of the analog wave.)
    I used to do this with the traditional DAQ and connected the analog source to the "GATE" pin of the counter , so i could count "while gate is high" ( or while gate low, or count on rising edge/ falling edge. ) However I found some problems and I think my counter wasn't counting correctly, and I would like to rewrite the task  with DAQmx anyway.  Ideally I would like to separate the signals that occur in the first 1/2 and the second 1/2 of each analog cycle. But if that is not quite achievable, I can do the separation later when I process the data, provided that I know each 10 data points from the counter correspond to each analog cycle.
    I have read several threads on the discussion board, but I still couldn't configure my applications using those examples.
    Also I wonder if the "Arm Start Trigger" function is not available on my PCI-6014 card ? Because I can not find it in the Start Trigger property node.
    Thank you very much for your reply. Any hint or advice would be greatly appreciated.
    Joyce

    ...(Continued from prev post due to 500-char limit)
    Now, here's a proposal for a method that should be able to work, though it'll require more data and more processing.  One key area to watch out for under DAQmx is the property node for "Duplicate Count Prevention."  The behavior changed between 7.1 and 7.3 (on E and/or M series, but not on TIO-based counters), then the defaults changed in 7.4, and something changed yet again in 7.5
    Bottom line: you can set it to True or False.  One setting will allow you to buffer up 0 values for intervals where there are no Source signals (photon events) within a sampling period (10 Hz hw clock).  The other setting will not record anything in the buffer for those intervals.  Clearly, you need the former setting, so you can always experiment to see how it behaves.  I kinda gave up on trying to interpret it, understand it, or keep up with the differences by DAQmx version and DAQ board.
    Ok, here goes:  I'm going to propose that you actually oversample by about 10x and that you capture the Analog square wave in sync with your photon counts.  Then your post-processing can determine for sure which samples correspond to Analog Square High and which to Low.  You'll also be able to adjust for times when you get 1 too many or one too few samples in a square wave cycle, due to having independent un-sync'ed timebases.
    So on one counter you generate a 100 Hz clock.  The other one is programmed for buffered period measurement using the 100 Hz clock as a SampleClock and the photon pulses as the Source signal (this must be set up using a DAQmx Channel property node I believe).  Remeber to be careful also about the DuplicateCountPrevention property.
    You would further setup an AI task using the same 100 Hz clock as a Sample clock.  Be sure that the 100 Hz clock task is started after starting both the other tasks.  Also, be sure to always read the same # of samples from both tasks to keep the counter period and AI data in sync.
    Voila!  You now generate a data record of # of photon pulses for each 10 msec interval along with the voltage of the analog square wave at the end of that interval.  A bit of post-processing and you're covered.  If you're not required to manipulate data while the acq tasks are running, you can surely afford to bump up the oversampling rate considerably.  The main advantage is to gain resolution on the time of transition of the analog square wave.  For any interval ending with a TTL state change, you won't know where within that interval the transition occurred.
    -Kevin P.

  • How can I control 3 stepper motors w/ amplifier by sending TTL pulses from DIO96

    We have NI PCI-DIO-96 board and we have 3 stepper motors with amplifier and encoders. We want to control the motors by sending TTL pulses directly from the DIO-96 board to the amplifiers. Is this possible? What would you recommend if this is not a good approach? Thanks in advance.
    Roman Zeylikovich

    Roman,
    Thank you for contacting National Instruments. While using a motion controller would be the recommended approach for any type of motion application, you may be able to use your DIO board to generate a TTL pulse train to control the step and direction inputs of your drive and motor. You will need to make sure that the current sinking and sourcing specification for the DIO-96 fits your stepper motor. That board is not designed to source very much current at all so this is one issue you will need to verify. Also, this board does not have any counter/timer logic that can handle quadrature encoder inputs.
    Again, this digital device is probably not the best solution to control a stepper motor, but, depending on your hardware, could be configured to work proper
    ly. The PCI-7334 is a low-cost stepper motion controller that can control up to 4 axes and is designed to easily accommodate these types of applications. You can browse through more information on our Motion controller boards at the following website:
    http://sine.ni.com/apps/we/nioc.vp?cid=3809〈=US
    Let us know if you have any more questions or comments.
    Regards,
    Michael
    Applications Engineer
    National Instruments

  • Count ttl pulse only over 2 V

    Hi I have a problem while I was trying to count a ttl pulse by using the DAQ Assistant. The problem may be simple but I just cannot solve that since I am new to LabView.
    There is a TTL pulse generated from our APD which is about 30 ns width, 5V. I want to use LabView to do the photon counting. However due to the noise, I would like to set a count limit of 2V, so the software only count the TTL pulse which is greater than 2V. But I cannot do that. There is no option to set the limit.
    And our hardware is BNC-2110 and PCIe-6323.
    Can someone help me with this? Thanks.
    Solved!
    Go to Solution.

    Okay so perform a finite or continous read of your analog input.  Then you can perform a threshold, where all values below 2 become 0.  This can be done many ways but I'd suggest an auto indexing for loop and using a select function in the for loop.  Then you can perform any normal edge detection you normally would to count pulses.
    EDIT: Post some code if you can.
    Unofficial Forum Rules and Guidelines - Hooovahh - LabVIEW Overlord
    If 10 out of 10 experts in any field say something is bad, you should probably take their opinion seriously.

  • How do I count arbitrary pulses from an oscilloscope with labview

    I am looking at arbitrary pulses on a TEK 2024B oscilloscope.  I want to use LabView to count those pulses and record a time stamp for each pulse (perhaps to an excel spreadsheat).  The duration of the measurement is over ten minutes and the pulses, on the average, occur once per millisecond.  How do I count/record these pulses with LabView?

    Jay,
    I would start by downloading the Instrument Driver for the 2024B from ni.com/idnet It should have the functions that the scope is capable of implemented there. If you have problems using the driver please post back.
    -Marshall R
    National Instruments
    Applications Engineer
    One stop for all your NI-VISA Support
    GPIB Support has a new homepage

  • How do i generate the TTL pulse from PCI 6024E DAQ card

    Hi, I'm trying to write a program which generates a single TTL pulse when it is run. My goal is to generate the pulse, send it to the output channel of my PCI 6024E DAQ card, and then use a BNC cable, to display the signal on the oscilloscope. I have written the code and I think it appears fine but when I try to view the signal on the scope I am unable to see the pulse. I have tried slowing down the sweep on the scope and have adjusted the voltage and time scales but have not seen any response. After examing my DAQ card, I noticed that none of the PFI/TRIG terminals on the card are wired; the output channel connects to terminals for DAC0OUT and Ground. I'm wondering if this pin needs to be wired in orde
    r to generate the pulse since PFI0/TRIG1 is the start trigger terminal. Any help would be appreciated. Thanks.
    Attachments:
    ttl.vi ‏26 KB

    Hi,
    You are looking for a retriggerable pulse generation. Please check the Generate Retriggerable Pulse (DAQ-STC).vi example that ships with LabVIEW.
    Regards,
    RamziH.

  • Counting TTL pulses using USB 6008

    Hi Folks,
    I am trying to setup a performance monitoring VI for a manufacturing line.
    I am going to use a sensor at the end of the line to count the number/rate of boxes that come of the line and perform a calculation to give us the productivity.
    The problem I have is that I would like to be able to scroll back over time to view historical productivity. Also I would like to be able to insert reason codes for downtime and print off a report at the end of the week showing a productivity graph and downtime associated with the line.
    I am new to Labview and I am not sure if this can be done, any help would be appreciated.
    Thanks
    Tony.
    Attachments:
    K2 Productivity-s.vi ‏60 KB

    Search 'count'.  Count Digital Events.vi is the most basic one of interest.
    Of course you won't find an example of counting pulses, analyzing the data and creating a report.  You'll have to put all the pieces together yourself.  This is just a start.  The good news is that LabVIEW is designed to do just the type of thing you're needing and your app will be quite simple to contrive once you get the hang of LabVIEW.  I suggest going through the many tutorials to get the basics down before you try to architect your program.
    Using LabVIEW: 7.1.1, 8.5.1 & 2013

  • How to count the pulses of an incrementa​l encoder with PCI-6010?

    Hi everyone,
    I am trying to use an incremental encoder and PCI-6010 card to measure the speed of a rotary shaft in Labview 2011. I need to use the counter to count the numbers of the pulses that the encoder generates but I have some problems there to do so.
    I connected the encoder signal to the counter source (PFI0) and built a daq assistant block in the Labview (Acquire signals -> Counter Input -> Edge Count -> ctr0) and it is set as shown in the figure. The problem is that no matter how I turn the shaft the measured value will always be 0 although the pulses can be seen very clearly on the O-scope. 
    I suppose I must have made a very simple mistake. Could anyone help me on that please?
    Thank you.

    Hi YShZh,
    The current Acquisition Mode setting is set to "1 Sample (On Demand)" and will therefore only ever give you the first count value.  The fist count value is equal to the initial count value, which in your case is 0.
    Try changing the acquisition mode to either N Samples or Continuous Samples.
    Kind regards,
    Marshall B
    Applications Engineer
    National Instruments UK & Ireland

  • How to measure pulses from 24v flow meter sensor with NI 9411

    Hi,
    I have a 24v pulses flow meter sensor, how is it possible to connect it to My NI 9411 ?
    Regards

    duplicate
    Steve M.
    National Instruments France
    #adMrkt{text-align: center;font-size:11px; font-weight: bold;} #adMrkt a {text-decoration: none;} #adMrkt a:hover{font-size: 9px;} #adMrkt a span{display: none;} #adMrkt a:hover span{display: block;}
    >> Vidéo-t'chats de l'été : présentations techniques et ingénieurs pour répondre à vos questions

  • Counting pulses from an encoder 0-7,5 V

    This is my VI and my signal measurement:
    [IMG]http://i.imgur.com/mfttanz.png[/IMG]
    How can I count the pulses from the encoder with this VI? I want to count it like in this VI:
    [IMG]http://i.imgur.com/xRzz0Lr.jpg[/IMG]
    So that I can measure the angle of a motor-driven frame using a 24 VDc motor.

    You will need some external hardware to clean the signal. The best might be to connect the signal to a comparator with some hysteresis and an output voltage compatible with your counter, typically 0 - 5 V, although some counters only accept 3.3 V.
    The signal you showed in the image on the first post in this thread should NOT be connected to a logic level counter. both the low and high voltages are outside the acceptable range for TTL compatible logic devices and coude damage the counter.
    I just looked at your VI and see that you are using an analog input. That should be fine for digitzing the signal.
    1. I recommend that you switch to a Producer/Consumer architecture for your program. Do the data acquisition in one loop (the Producer) and the calculations and analysis in the other loop (the Consumer). This will let you run the acquisition loop fast enough that you never  miss any pulses.
    2. Your signal is probably clean enough to use a simple comparison to a 2 or 3 V threshold. Then you can count those logical transitions. If you sometimes have more noise than shown in the image you could add some hysteresis to the comparisons or you might need a more sophisticated algoroithm to eliminate spurious counts.
    3. Do you have some means of determining when to start and stop the counting?
    Lynn

Maybe you are looking for

  • My Creative Zen Sleek is bro

    My Creti've Zen Sleek got dropped, so only half of the LCD display currently works. I have sent emails to Creative through the website asking for the necessary details so that it can be fixed, but have received absolutely no response. I understand th

  • Apple tv lost connection

    My apple TV has been working ok and suddently it can see many options like movies, etc, but when I try to actually donload and look at one it is not there. Somehow it can get a look aat some of the movies, but when I tryo to purchse will not download

  • Pdf parameters not working

    Firefox does not seem to support PDF parameters in links. e.g. mypdf.pdf#page=6&view=fit It works in IE and Chrome, but only goes to the first page of the pdf in Firefox.

  • .mov files are grey and not able to import

    Help.  I download files from my Contour helmet cam into my Imac but can't import them into Imovie because they're grey.  Some earlier files are black and "loadable" ... what's up?  Is there some importing ritual I screwed up?

  • Cropping LR4 occasional strange behavior experienced by others?

    The LR4 cropping process has become unstable during use on several occasions.  The LR4 cropping tool seems to get confused. During the cropping process, unwanted changes to position and leveling of the cropped area happen upon exiting.  The cropped i