NI-DAQmx false duplicate channel error
I have an SCXI chassis w/ a SCXI-1600 and a SCXI-1102 w/ SCXI-1303
accessory. I created a series of NI-DAQmx global channels for
thermocouples, calling them Air Temp0 through 17. Using NI-DAQmx
subvi's I read them. See a very simplified VI attachment.
The channels individually all check out, but when I try to read the
entire group, I get an error -200072 and it says that Air Temp15 and 16
reference the same physical channel, but when I check the channels in
MAX, they are indeed different. Likewise, if you read one at a
time and stimulate the various thermocouples, the displayed response is
always consistent with the physical channel. I.E., if you heat the
right thermocouple, the reading goes up, but if you heat the wrong
thermocouple, the reading doesn't go up.
Using the simple error handler, the error message is:
"Error -200072 occurred at DAQmx Read (Analog 1D Wfm NChan 1Samp).vi
Possible reason(s):
Duplicate channels in the list of physical channels are not supported by this device.
Channel Name: Air Temp15
Duplicate Channel: Air Temp16
Device: SC2Mod2_Mini
Task Name: Air Temp15:16"
Any ideas?
Attachments:
diagnosis.vi 59 KB
Thank you for your input. It does seem to have been a corruption of the configuration somehow.
This is one of the things I hate about MAX. The unexplained and
unexplainable corruptions. And then the usual solution:
"Oh, reinstall everything and try again." Grr. This has
been going on for years now.
The station had not crashed or otherwise displayed any
misbehavior. The channels were configured by doing the first one
and then duplicating it using the MAX "Duplicate" feature. I then
made a set of channels by individually programming each one, with a
different name. I then deleted the troublesome channels and then
renamed the new ones to the desired names. Whaddya know, they
work now. Is this a MAX bug, known or otherwise? I have
'way too many other things on my plate to be spending 6 hours chasing
corruptions like this. Virtual Instrumentation is supposed to
save time.
Similar Messages
-
DAQmx physical channel error message
Hi,
I am new to Labview but I managed to create a new analogue voltage input task in Measurement and Automation Explorer. I am using DAQmx create virtual channel VI but I am getting an error message saying that "required input physical channel is not wired". Any idea why I am getting this error message. Also what is the difference between using DAQmx Assistant and using VI's like DAQmx create virtual channel, DAQmx timing etc.Any help is appreciated.
Thanks
Duran
Attachments:
error2.JPG 171 KBThe first VI 'Create Virtual Cannel' has an input called physical channel. This needs an input. Right click on the terminal and select Create>Constant. Then you can select the channel (something like Dev1/ai0).
To your 2nd Q: The assistant is a fast way to configure your measurement. You can later also convert it to the VIs you use now. Using the VIs gives you full programmatically control of the task. (The meaningful phrases are 'configure' and 'full programmatically').
Felix
www.aescusoft.de
My latest community nugget on producer/consumer design
My current blog: A journey through uml -
ORA-03113: end-of-file on communication channel ERROR
After applying Database patch 3095277 to my 9.2.0.1.0 installation
in order to update it to 9.2.0.4.0,I'm getting ORA-03113: end-of-file on communication channel
ERRORS when attempting various tasks. This is a local installation running on a Windows XP Workstation.
I've even tried creating a new test database but the result is the same.
The reason for moving up to 9.2.0.4.0 is so I can start using Advanced Queuing, which I understand had problems in earlier releases.
However before the update I could at least create the JMS queue tables without ERROR whereas now I get the following errors when attempting to create a queue table in SQL Plus from within Enterprise Manager Console.
SQL*Plus: Release 9.2.0.4.0 - Production on Thu Sep 18 08:20:08 2003
Copyright (c) 1982, 2002, Oracle Corporation. All rights reserved.
Connected.
BEGIN SYS.DBMS_AQADM.CREATE_QUEUE_TABLE(Queue_table=> 'JMSQTBL',Queue_payload_type=>'SYS.AQ$_JMS_STREAM_MESSAGE',multiple_consumers=> false); END;
ERROR at line 1:
ORA-03113: end-of-file on communication channel
BEGIN SYS.DBMS_AQADM.CREATE_QUEUE(Queue_name=> 'JMSQUE',queue_table=> 'JMSQTBL'); END;
ERROR at line 1:
ORA-03114: not connected to ORACLE
I would be grateful of any pointers as to how to resolve this problem.
Regards
Gary.Chris, Thanks for the suggestion,
I took a look in the trace log and found the following entries that point to a ORA-00600 as you mentioned earlier.
This is only a test database so there's no great loss if
I have to re create it, but in saying that I did create a new test02 from scratch and had the same problem.
Is it posible something went wrong when applying the 9.2.0.4.0 patch ?.
Regards
Gary.
Dump file c:\oracle\admin\test01\udump\test01_ora_2556.trc
Wed Sep 17 16:25:14 2003
ORACLE V9.2.0.4.0 - Production vsnsta=0
vsnsql=12 vsnxtr=3
Windows 2000 Version 5.1 Service Pack 1, CPU type 586
Oracle9i Enterprise Edition Release 9.2.0.4.0 - Production
With the Partitioning, OLAP and Oracle Data Mining options
JServer Release 9.2.0.4.0 - Production
Windows 2000 Version 5.1 Service Pack 1, CPU type 586
Instance name: test01
Redo thread mounted by this instance: 1
Oracle process number: 10
Windows thread id: 2556, image: ORACLE.EXE
*** 2003-09-17 16:25:14.046
*** SESSION ID:(7.17) 2003-09-17 16:25:14.031
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** Object defn.s out of sync w/ data
*** 2003-09-17 16:25:14.281
ksedmp: internal or fatal error
ORA-00600: internal error code, arguments: [qmxiUnpPacked2], [121], [], [], [], [], [], []
Current SQL statement for this session:
select 1
from resource_view
where any_path = '/xdbconfig.xml'
----- PL/SQL Call Stack -----
object line object
handle number name
6A4FA348 6 package body XDB.DBMS_REGXDB
6A504D88 1 anonymous block
6A576AB8 639 package body SYS.DBMS_REGISTRY
6A1DBB6C 1 anonymous block -
Daqmx Create Virtual channel timeout
Why almost a minute and a half to get an error message from daqMX Create Virtual channel VI.
I need to find a way to have my executable bleep immediately if the underlying drivers are not installed.
For example, I have LV at home and supposedly have the drivers installed too but no hardware connected.
Sometimes I run a that uses daq, however when it encounter Creat Virtual Channel VI it appears to hang for a minute and a half before finally giving an error.
Anyone else been there, done that?
ThanksThanks for replying D-cubed.
Actually Im calling create virtual channel in a state machine to set up the data acquisition.
When using on a machine that has a daq device installed its a piece of cake. no sweat.
However, when I am at home I have LabVIEW but possibly either no daq drivers or no daq hardware.
I believe it may be the second case since I can place a task on the BD so, I will assume the drivers are installed correctly.
There is just no daq hardware in the system.
WHen the create virtual channels Vi is called the PC appear to hang because that VI is executing internally (looking for a device I presume).
The timeout takes 1.5 mins and if not for me waiting the PC would appear to be hung up.
Finally some 'Mig error' message appear contact Ni about and so on.
WHat I am trying to do is have my program error out immediately instead of 1.5 mins later that there is no Daq device in the system.
Thanks again -
Hi,
I have seen several discussion regarding this topic, however it seems that none of the solutions are working for me. Here is my situation, i am using LabView 7.1 which I use to generate my .EXE, when I generate my .EXE I use the Tab Intaller Settings to generate the Setup.msi. I then take the folder which contains the .EXE and the installer and bring it over to another machine which is running VM Ware with a freshly installed windows XP. (to make sure nothing is on the machine and then i get surprise at the custormer's site)
Once this was done, i would start by lauching the installer and then the .EXE the result was the following, I would get the following message:
Missing subVI DAQmx Create Task.vi in VI instance 1 3tst
.vi.
Missing subvI DAQmx Create channel (TEDS-AI-Voltage-Basic).vi
in VI Instance 1 3tst.vi.
Missing subVI DAQmx Create Channel( AI-Voltage-Basic).vi in VI
instance 1 3tst.vi
Missing subVI DAQmx Timing(Sample Clock).vi in VI instance
1 3 tst.vi
Missing subVI DAQmx Read(Analog 1D Wfm NChan nSamp).vi in
Vi instance 1 3tst
then as suggested in other discussions of this forum I download (537MB) and installed
on the target machine (VM Ware) the nidaq710 which i downloaded from
http://digital.ni.com/softlib.nsf/websearch/C1F802E457EA042586256DD00001B0C9?opendocument&node=132070_US
This did solve some problems BUT I am still getting the following message when I launch the EXE
Missing subVI DAQmx Create AI Channel TEDS(sub).vi in VI DAQmx Create
Channel(TEDS-AI-Voltage-Basic).vi.
Thank you
MaldiniHi Maldini,
Thank you for posting to the NI forums. Is the VM Ware PC using the same operating system as the development PC? Also, are you using DAQmx 7.1 on the development computer, or are you using a more recent version? When distributing applications, the driver version on the deployment PCs must match the driver version of the development PC.
If this is already the case, then you shouldn’t be receiving any of these error messages. There are few things that may fix this problem.
Reinstall the DAQmx driver.
Perform a repair install the DAQmx driver on the deployment computer.
Upgrade the DAQmx driver on both the development and the deployment PCs.
With Windows XP and LabVIEW 7.1, you are able to use the latest version of DAQmx, which is currently DAQmx 8.5.
After going through these steps, please post back if you continue to have problems.
Ed W.
Applications Engineer
National Instruments -
NFS Commnication channel errors for deleting and placing
Hi All,
I was trying File to file scenario.
I am picking the file from App server(AL11) using NFS Adapter and the processing Mode is Delete
In comm channel monitor it showing an error as
Sender channel error:
Error: File '/staging/INTERFACE/SAP/ca/kar370.txt' - deleting after processing failed
Same way while placing the file
Receiver channel error:
Error: Message processing failed: FileNotFoundException: /staging/INTERFACE/SAP/ca/test120110906-122051-980.txt (Permission denied)
Whether i dont have the permission to delete and place the file from and to the App server(AL11)?
What is the procedure in order to make it possible.
Thanks
SaiHi All,
Thanks to every one.
The problem is solved ,
Basis Guy has given the permissions, and i am able to place and delete the files .
Thanks
Sai -
Receiving a false paper jam error message for a 6310 all in one printer
I have an hp 6310 all in one printer. It is showing a false paper jam error message with no paper in the printer. I have tried all of the fixes that have been suggested on the website as well as on you tube to no avail. Nothing works, I have cleaned all of the parts that have been suggested to clean and reset the printer. I am still receiving the same paper jam message with no paper in the printer. Does anyone have any suggestions to try to fix this printer.
Do you here any clicks or other sounds when the 6310 first starts up?
I'm trying to fix false paper jam error on Sharp Fax/copier,F0-4450.
When researching the problem, found suggestion that a bad solenoid could cause that error.
After completely disassembling the copier found that wasn't the problem.
I here 2 clicks during start up that sounds like gear slipping a tooth, as if something is preventing a shaft from turning.
Back to the drawing board.
UPDATE!!! Took fuser out, turned gears with screwdriver til 81/2x11 paper went thru, lubricated ends of fuser roller with powdered graphite lube.
3 page fax came thru. -
XI Communication Channel Error Explanation
Hi All,
Does anyone happen to know :-
1) Is it possible to use CCMS to email the errors coming from a communication channel error Explanation field
2) Are these messages stored in a table I can check
The reason I need to know this is because we sometimes encounter an error such as "Conversion of file content to XML failed at position 838: java.lang.Exception: ERROR converting document line no. 839 according to structure 'XXX_XXX_XXX':java.lang.Exception: ERROR in configuration / structure 'XXX_XXX_XXX.': More elements in file csv structure than field names specified!
This tells us exactly where the error occurred but using Communication Channel Monitoring the error message only remains for a little while before the polling messages remove it from the display. Since the process runs early hours of the morning we never get to see these messages.
Any help appreciated.
Thanks
CraigHi Raquel,
Let me try to clear out some things in this thread . First of all you do your alert configuration which you say you have already done. Thats good . Now you ask that you want only to monitor the communication channels and you want specific guide . Here is the specific guide.
If you have configured alerts you will know that when you create an alert rule you have the option to choose radio button where did the error occur it could be no restriction , integration engine or adapter engine. You need to choose adapter engine.
Now whenever the colour of the lights change to red. You will get your alerts.
Regards
joel
Edited by: joel trinidade on Apr 9, 2009 9:41 AM -
BI 7.0 - Duplicate Record Error while loading master data
I am working on BI 7.0 and I am trying to load master data to an info object.
I created an Infopackage and loaded into PSA.
I created transformation and DTP and I get an error after I execute the DTP about duplicate records.
I have read all previous threads about duplicate record error while loading master data and most of the them suggested to check 'Ignore duplicate records' option in the infopackage. But in 7.0, I can only load to PSA with infopackage and it doesn't have any option for me to ignore duplicate records.
My data is getting loaded to PSA fine and I get this error while loading to info object using DTP.
I would appreciate your help to resolve this issue.
Regards,
Ram.Hi,
Refer:
http://help.sap.com/saphelp_nw2004s/helpdata/en/45/2a8430131e03c3e10000000a1553f6/frameset.htm
http://help.sap.com/saphelp_nw2004s/helpdata/en/d0/538f3b294a7f2de10000000a11402f/frameset.htm
With rgds,
Anil Kumar Sharma .P -
Getting end-of-file on communication channel error on this query
Hi guys,
I'm getting the end-of-file on communcation channel error when running the following query in SQL PLUS, If run it in a stored procedure called through code of JDBC, it gave "no more data to read from socket" error. Any idea what went wrong? kind of frustrated now. We are using the Oralce 9.0.1.4.0, which suppose to fix some sdo_uion and sdo_buffer bugs. Thanks a lot!!
select SDO_GEOM.SDO_BUFFER((SELECT SDO_AGGR_UNION(MDSYS.SDOAGGRTYPE(GEOLOC, 0.011119487)) FROM GEOTEL_SOURCE where MSA='5000' and COMPANY_NAME ='AMERICAN FIBER SYSTEMS'),1.0,5,'unit=MILE arc_tolerance=0.05') from GEOTEL_SOURCE a where a.MI_PRINX = 1;Hi Daniel,
Just tried the 9.2.0.2 patch, seems to work better now. At least I don't get the end-of-file on communication channel error. But how about the speed thing? I isolated the problem, and find out that sdo_aggr_union is taking too long to finish( 1 Hour!!! for 2300 lines with the same city. Any clue to optimize it, I tried some hint like /*+ordered*/ No_Merge and something like that, does not improve though. Thanks! -
Hi there,
Could anyone explain me why we get hundreds of sub-vi's when I use the
vi "DAQmx Create Virtual Channel"? This can easily be seen by exploring
the hierarchy of Labview examples such as "Acq&Graph Voltage - Ext
Clk.vi"???
This multitude of vi's considerably slows down my computer, especially at loading. Is there any way to get rid off that???
Thanks in advance for you help!
JonasHi,
the easier you have to program, more subVI´s needed for your kind of application. Because the VI´s from DAQmx are polymorph, so many SubVI´s are included in the VI´s. In traditional DAQ you have several palettes for each kind of measuement method. Analog In, Analog Out, Digital, Counter, ...! In DAQmx you have one VI and that can all.
But, the performance of you application at runtime is not affected. Of course it takes some time to load, but just at development environment not as an application.
Figure 3. Polymorphic VIs simplify the NI-DAQmx API.
Link to Document:
http://zone.ni.com/devzone/conceptd.nsf/webmain/EE47B125BB9E053686256FBC0014C384
Hope this helps.
Alex
NI Switzerland -
Adobe Acrobat 8 Form Duplicate text error
Hello World!
I work for a school with 200+ computers installed with **. The problem I'm having is when a user downloads a PDF form from our web page and goes to start inserting data in the fields the text just duplicates in all the boxes. If you try and delete the text from any of the boxes it delets everything. I've tried repairing the install by running the original installer file. I've even reinstalled a test system and still duplicate this error. Please let me know if there is any other information I can provide that will help you help me.
Windows XP SP3
Office 2007
Adobe Acrobat 8Hi,
This forum is not for Acrobat product.
Plaese post your question again in http://forums.adobe.com/community/acrobat if you haven't.
Thanks,
-- Hiroshi -
Receiver JDBC Channel Error : Incorrect syntax near 'OR'
hi,
we are facing the below error at the receiver communication channel:
Error while parsing or executing XML-SQL document: Error processing request in sax parser: Error when executing statement for table/stored proc. 'table' (structure 'statement'): com.microsoft.sqlserver.jdbc.SQLServerException: Incorrect syntax near 'OR'.
When we again reprocess the message,it shows the exact error at the channel.
Any ideas why the first error shows up?Hi, Anu
The most probable it's that your target XML it's not correct because it haven't a correct format.
Please, check your target XML and see
http://help.sap.com/saphelp_nwpi711/helpdata/en/48/465581929734d9e10000000a42189d/frameset.htm
Regards,
Carme -
Communication Channel error (Device busy)
HI all,
i have a problem with sender communication channel (CC) (type file) on PI 7.1
The CC has follow error:
Delivering the message to the application using connection File_http://sap.com/xi/XI/System failed, due to: com.sap.engine.interfaces.messaging.api.exception.MessagingException: java.io.FileNotFoundException: /IFR_PI/prd/out/TYPEB/SEDE_PV/PV_ART/VARGPV_I003 (Device busy).
Anyone can help me to understand this error?
Thanks in advance
PMHi,
The file size is a drawback of file adapter.
Check the note Note 821267 - FAQ: XI 3.0 / PI 7.0/ PI 7.1 File Adapter
See point 14. Memory Requirements
Indeed the error is IO Error and it is because the Adapter engine was not able to send the message to Integration server. But it happens due to memory/heap size issues.
Look at these thread, they are having the same problem. Please try the remedy measures suggested by them
Mail to Proxy scenario with attachment. AF channel error.
Error with huge file
problem with big file in file-to-proxy scenario
regards,
ganesh. -
JDBC to JDBC Scenario - Communications Channel Error
I am following the published JDBC to JDBC Scenario:
https://www.sdn.sap.com/irj/sdn/wiki?path=/display/xi/step-by-step+guides
I am a 3rd Party Vendor, and I'm using NX7: SAP Exchange Infrastructure. My goal is to have a JDBC Sender that sends MS Access data from my PC to NX7, and have a JDBC Receiver that receives MS Access data, and map the data from one database to the other via XI per the scenario, with both databases on my PC.
I set up the NX7 SAP XI per the scenario, but I'm getting an error on the NX7 Communications Channel:
Error during database connection to the database URL 'jdbc:odbc:Driver={Microsoft Access Driver(.mdb)};DBQ=//204.245.44.232/C:/Documents and Settings/All Users/Documents/XI_FILES/SENDER/STUDENT_Database.mdb' using the JDBC driver 'sun.jdbc.odbc.JdbcOdbcDriver': 'com.sap.aii.adapter.jdbc.sql.DriverManagerException: Cannot establish connection to URL 'jdbc:odbc:Driver={Microsoft Access Driver(.mdb)};DBQ=//204.245.44.232/C:/Documents and Settings/All Users/Documents/XI_FILES/SENDER/STUDENT_Database.mdb': SQLException: [Microsoft][ODBC Driver Manager] Data source name not found and no default driver specified'
The data source is in
C:/Documents and Settings/All Users/Documents/XI_FILES/SENDER/STUDENT_Database.mdb on my PC and it is a shared folder. My PC IP 204.245.44.232 is a public IP address. It should be using the JDBC-ODBC bridge driver in the JDK.
Why am I getting the error?
Thanks,
Ed CulnaneHi
Connection Parameters are of given below check
Transport Protocol: JDBC 2.0
Message Protocol: XML SQL Format
Adapter Engine: Integration Server
JDBC Driver: jdbc.odbc.JdbcOdbcDriver
Connection: jdbc:odbc:Driver={Microsoft Access Driver (*.mdb)};DBQ=//xiserver/xitemp/db/TestDB.mdb
5) In my case there is no userid and password required, but if you have userid and password, then the parameter for Connection becomes,
jdbc:odbc:Driver={Microsoft Access Driver (*.mdb)};DBQ=//xiserver/xitemp/db/TestDB.mdb;Uid=yourid;Pwd=yourpwd;
Well ms access file should be in the XI Server path or it should be able to accessed by the XI
Try this above option of keeping u r file in the XI Server path instead of the local
rgd
srini
Maybe you are looking for
-
I need to figure out how to unlock my locked account. I just got a new iPod touch 5th generation and since it's my first purchase I have to answer security questions. When I answer I am wrong I guess and it doesn't show me a straight forward way on h
-
Recommended server specifications for IOP 11.1.2.0 server?
In the iop_11120000_readme.pdf it references the following document: "Information about system requirements for EPM System products is now available in a spreadsheet format in the Oracle Hyperion Enterprise Performance Management System Certification
-
Newbie needs help about Data Encryption
bought a Wireless G router Saturday, followed the instruction on the cd and hooked up the router to the Desktop (ME) and the cable modem. I am happy that my new laptop (XP) can surf the Internet anywhere in the house. Can anyone see what I can doing
-
Oracle Drive, dav_portal and accents
Hi everyone. I am using the DAV protocol to have access to some of my documentation in Portal. The thing is when I use some accent in the archives names like "documentaciónuno.doc" and try to access to the DAV with Oracle Drive, the archive is listed
-
Dahdi Package from AUR - makepkg error
Unfortunately I have no easy way to copy/paste (it is on a different machine...). This is the exact error: $ makepkg -s -i ==> Starting build()... patching file Makefile Hunk #1 succeeded at 132 (offset 4 lines) You do not appear to have the sources