6025e MAX test panel

I'm using MAX test panel to test my digital I/O pins, however i couldn't find any manual explaining the function of the test panel. there are 2 tab on DIO. what are the operation/function in the 2 DIO tab?
First Digital IO tab
a. which port does it test?
b. why can't i select the input state?
For the second Digital IO tab
a. i can only select port 2,3,4, what does the port 2,3,4 refer to? coz for 6025E, i have only port 0,1,2,3
b. why can't i set the input i want for the input port?
c. how i check if the digital IO port is reponding?
thks & best regards

For the PCI-6025E, the number of the digital ports are :
0: DIO
2: PA
3: PB
4: PC
There is a big difference between DIO and the 3 other digital ports. While each line of DIO can be individually set to be a DI or a DO, all 8 lines of PA, PB or PC must have the same direction.
See the user manual for more information.
The first tab is used for the DIO port (DIO0 to DIO7). You can set the direction of each single line.
The second tab allows you to define one input port and one output port (eg PA (2) = DI and PB (3) = DO). You can then see the state of each line of the input port and set the state of each line of the output port.
You can easily test your hardware by wiring a DO line to a DI line (eg PA0 to PB0). Change state of DO and check if DI is co
rrect.

Similar Messages

  • Why can i NOT use "differential" on the MAX test panel for channels 8-16??

    I have a USB-6218
    This is a 32 channel device with 16 Differential inputs.
    Why can't i use the MAX test panel to view any channel beyond AI-7 using the Differential setting? It shades out differential even though my device has this capability.
    WHY???? 

    I am using MAX 4.1.0.3001 in conjunction with LV 8.0 and the USB-6218.
    I go to MAX, Expand devices, expand MX devices, Right click on the USB-6218 (device2), and go to the test panels.
    From there i am unable to view any differential pair beyond 8.
    How am i supposed to Thru-Calibrate my equipment? Generally to achieve a valid calibration i use MAX itself to record the raw voltages the board "sees" as i apply known loads to the sensors.
    The reason i need to check all this out is because DIFF pairs > 8 also are not working in my VI! Im am not sure if this is related or not.

  • For a 3.33v input, DAQ & MAX Test panels read 018v. What is missing in set up?

    Hi All:
    I am resurrecting a Labview setup that has not been used for a few years.
    The goal is to measure strain.
    Software is:
            Labview 6.0
     DAQ V6.8.0f9
     MAX 2.0.3.6.
    Hardware is:
     PCI-MIO-16E-1
     SCXI-1000 Chassis
     SCXI-1121 Module
     SCXI-1320 Terminal Block
    As a fist step I decided to test the hardware using MAX.
    I reset all of the SCXI-1121 jumpers to factory settings except the Filter jumpers.
    Filter jumpers were set at 10kHz  and  I left them at that setting.
    In MAX I followed the software procedures for setting up the PCI-MIO-16E, SCXI-1000 Chassis,
    and SCXI-1121 Module (Module was found using the auto detect). MAX verifies the PCI-MIO-16E
    card, and the SCXI-1000 chassis.
    I verified that EX 0 thru 3  were 3.33v using a digital meter.
    I then wired  EX 0 to CH 0 and using the MAX Test Panel and the set up string of
    ob0!sc1!md1!0
    in the channel box on the Analog Input tab.
    I expected to see a 3.33v reading on the test panel, but got a reading of .0184v
    (I assume this means zero or is noise). The result is the same on channels 1, 2, and 3.
    As far as my bifocals can tell, all cables are connected properly, and there are no bent pins.
    I have re-verified the jumper settings per the SCXI-1121 User Manual. I have printed the
    “Knowledge Base”, “SCXI Trouble Shooting Resources” and walked through them step by step
    to verify the set up of hardware and software.
    I have set up a virtual channel for each channel and used the DAQ virtual test channel
    test panel, and I get the same result .0184v.
    Anybody have any suggestions or clues as to what I have missed?
    Thank You
    Peter Waltz

    Hello Peter,
    If you are reading 0.0184V on your analog input, then it probably means that you are reading nothing at all.
    This seems like a driver issue more than anything else. Make sure that you are the latest possible driver installed for this setup. Traditional DAQ 7.0 or 7.1 are compatible with LabVIEW 6.0 and should work fine.
    Once you have the right driver, try connecting your analog output to your analog input. Output a sinewave or a specific voltage and see what the AI reads.
    Most likely the driver upgrade would solve your problem.
    You can download the drivers from here.
    Chetan K
    Application Engineer
    NI
    Message Edited by CKap on 05-08-2006 02:57 PM

  • My daq card PCI6025e' digital ports PA,PB,PC are in high state and cannot be changed even in MAX 2.1 testing panel and continuously output 5 volts even set for output.

    my digital ports PA,PB,PC are in high state and cannot be set low even if it is set for outputs.
    Thanks

    The MAX utility is the closest to the driver level and will eliminate software configuration as a possible issue. Please disconnect anything that you have connected to the DIO lines.
    Use the MAX test panel for your DIO lines, configure them as outputs, and set them to a low state. Use a multimeter (DMM) to observe the line state. If it is still high then you may have a problem with your hardware. If this is the case, I advise calling National Instruments support and investigate a possible RMA (repair).
    Best Regards,
    Justin Britten
    Applications Engineer
    National Instruments

  • Has to type each character twice while supplying command in Open VISA test panel

    has a set of commands to work witha system connected to the MS windows PC. when I trype the commands in Hyperterminal (with/without ) they are yielding the expected results.
    in LabVIEW, to get the expected results I MUST has to type 'each character' in the command Twice(except blank space) in the Open VISA test panel (viWrite/viRead). If I give the command  is given the usual way(with each character typed only once), the result always is 'illegal command'.....the same works fine in the Hyperterminal.

    What confused Ashwin is that you said "In LabVIEW". OK, so you're using the MAX test panel. There's one thing that's not clear. You said "when I trype the commands in Hyperterminal (with/without )". With/without what? How is your Hyperterminal session configured? Have you tried something like PortMon to see what's actually being sent from the serial port driver?

  • DIO won't work until I use the Test Panel

    Here is some strangeness.  A program that works fine all of a sudden has an issue on a clean install.  I am using a DIO-24 and after the PC has been wiped and and the software including the OS and drivers are reinstalled I get an error trying to access the DIO.  Once I open the MAX test panel for the device (DIO-24), wiggle some lines, and close it all is fine.  The program will continue to function fine.  What "Magic" is happening when I open the DIO-24 in MAX?
    Matt
    Matthew Fitzsimons
    Certified LabVIEW Architect
    LabVIEW 6.1 ... 2013, LVOOP, GOOP, TestStand, DAQ, and Vison

    I am saying:
    Are you saying that after installing the driver and your device you had to open MAX and use test panels before your device would work with your application, and now you can always use your application (even after restarting your computer) without opening MAX?  - Yes
    Does this behavior only occur with your software, or does it occur with a shipping example too? - I Have not tried to run example after a fresh install so don't know
    Are you using LabVIEW to communicate with your device?  - Yes LabVIEW 8.0
    If so, what version and what driver have you installed. If not, what programming language are you using? - NI DAQmx 8.0 for Feb 2006 Disk
    Is this driver version the same that you had installed when you developed your application? - No, previous version
    What operating system are you working on? - XP 
    Do you need MAX open to run your application, or do you just need to run it before your application? - Just run once and close before running application
    Any additional information you can provide about what you are experiencing would be helpful.
    I spoke the the engineer who gave me the information and he said it happens only some of the time.  Most systems come up fine but a few require that MAX before the app will run correclty.
    Matt
    Matthew Fitzsimons
    Certified LabVIEW Architect
    LabVIEW 6.1 ... 2013, LVOOP, GOOP, TestStand, DAQ, and Vison

  • Test Panel in VI format

    I am looking for the Measurement & Automation Explorer (MAX) Test Panel in a VI format. It has the ability to write or read from all lines in the DAQ. Each line is modeled as an LED.
    I know that the test panel is available in the nitpan32.exe. I need to make some slight modifications to the file and was looking for it in VI format.

    Greetings,
    I am not aware of the existence of the Test Panel as a VI, but the example "Dig Multi-line Direction (E).vi" does accomplish what you have described. You can assign each line's direction, and they are represented by an LED (input) and a switch (output). This example is installed with LabVIEW.
    Good luck with your application.
    Spencer S.

  • NI-DAQmx task works in MAX or DAQ Assistant test panel but not in LabVIEW

    I am attempting to read a single AI channel from a PCI-6024E card via an SCB-68. I have created a NI-DAQmx Analog Input Voltage Task in MAX for this channel, sampling in contiuous aquisition mode at 100 kHz, 10000 samples at a time, with RSE terminal config. If I use the Test feature from MAX, the channel acquires data as expected.
    In LabVIEW, I call this task using an DAQmx Task Name Constant. If I right-click this constant and select "Edit Task", the Daq Assistant opens and I can use the Test feature from the DAQ Assistant to see that the data is still being acquired as expected.
    However, when I try to programmatically read this channel in LabVIEW using the VI "DAQmx Read (Analog Wfm 1Chan NSamp).vi", the VI returns a constant DC value of 500 mV, which I know is incorrect (I can monitor the signal across the two terminals in the SCB-68 with a DMM to know that the signal coming in varies as expected, and as I read using the test panels). This erroneous reading occurs even if I make a new VI, drop the task name constant on the diagram, right-click the task name constant and select "Generate Code - Example" and let LabVIEW create its own acquisition loop (which is very similar to what I already set up, using the "DAQmx Read" VI).
    Any ideas why the Test Panels work correctly but the LabVIEW code does not?

    Hello bentnail,
    I'm not sure why the test panels are readin the value correcly but the LabVIEW code does not, but there are a couple of things we can try.
    1) What happens if you just use the DAQ Assistant and place it on the block diagram? Does it read out the correct values?
    2) Try running a shipping example that comes with LabVIEW. "Acq&Graph Voltage-Int Clk.vi" should work well.
    3) What kind of signal are you expecting to read (peak to peak voltage, freqeuncy, etc.)?
    Thanks,
    E.Lee
    Eric
    DE For Life!

  • MAX 4.2.1.3001 Task Test Panel Button Missing

    I just upgraded to LabView 8.2.1, along with the drivers, NI-DAQmx, MAX, etc., and now when I use MAX to view my existing tasks, they no longer have a "Test Panels" button.  They appear to now have an integrated test graph.  That is nice, but before I was able to press the "Test Panels" button, select a type of "Values" instead of a graph type, and this would allow me to see the raw values that each channel was returning.  What happend to this functionality in 8.2.1.3001?  I know that I can go to each individual channel and see its raw values, but that is not acceptable.  I need to see all of the values of each of my task channels together in one place, the same way I was before "upgrading" to the new MAX.  Am I missing something?
    Thanks for listening!
    Rick Howard
    Exquadrum, Inc.

    This post and also this post, which are the same, have been replied to on this DAQ forums post, on which you can continue posting.  Please post only once on the most relevant forum.
    Regards,
    Raajit L
    National Instruments

  • In Max, click Test Panel - Get Executable version (7.1.1) doesn't match resource file (7.1)

    Hi,
    I haven't used LabVIEW for a while.
    I plugged in a USB-6251 and ran Measurment and Automation explorer and saw that the USB-6251 was in the NI-DAQmx Device and was green.
    I clicked on the Test Panels button and got a window with the message:
    Get Executable version (7.1.1) doesn't match resource file (7.1)
    When I clicked OK  MAX was locked up.
    I found this:
    http://digital.ni.com/public.nsf/websearch/680E61A4D02158A186256F7A0073C228?OpenDocument
    which told me to repair the LabVIEW Runtime thingy.  I did that and it's still broken.
    My MAX version is 4.2.0.3001
    I assume that's a pretty old version but I would have thought it should have been updated when I upgraded to LabVIEW 8.2.1.  But I don't see a MAX folder in the 8.2.1 install.

    I thought I had replied to this but somehow it didn't get there.  In answer to the question about running the device driver CD after installing LV 8.2 I don't know the answer. 
    Anyway I installed LV8.6.1 and now it goes into test panels OK so I should be able to progress.

  • Simulated device in MAX, self tests without error and has working Test Panels, but doesn't show up in DAQ assistant.

    I'm trying to create a development machine where we can test new code without using our physical hardware. I've followed this guide in setting up a simulated device. I can get to step 3.2b, but the device does not show up in the DAQ assistant. In MAX, the device self tests and self calibrates successfully, and when I open the test panels, I see some sort of signal. I assume this is a default simulated input since I haven't told the device to look for anything? Note that the two devices I'm trying to create show up in the Devices and Interfaces section, but that even after running Self-Calibrate, the Self-Calibration date is still unspecified.
    When I try to test the device and create a voltage input according to the guide, I am unable to see either device in the DAQ task creator.
    Steps 1 and 2 of this guide are obviously satisfied. Step 3 is not, but this is unsurprising since a simulated device wouldn't be found in the Device Manager anyways. Also, I am not running RT, so step 4 is satisfied.
    Does anyone have any ideas?
    Solved!
    Go to Solution.

    That would be because the PXI 5124 is a digitizer not a analog input device.  You need to use the NI SCOPE driver not NI DAQmx
    Jeff

  • MAX 3 locks up when using test panels

    I am using MXI-3 fiber link to connect to PXI-1000B chassis from my win XP PC. I have a PXI-6115 and PXI-6533 in slot 3 and 5 respectively.
    Both devices shows up in max and window device manager.
    But when I try to check the properties test resources on PXI-6115 then nothing happens.If I run test panel on max or in virtual channel, its locks up the MAX.
    Runing the test resouces in MAX on PXI-6533, I get the message that "This device is not responding to selected base address.
    I have also tried this setup with windows 2000. When I runs MAX the TEst Panels and Test Resources buttons are grayed out.
    I have changing diffrent setting for IRQ's and DMA without ant results.

    When you install a MXI-3 card in the PC and the PXI chassis, there is a MXI Initialization disk that you get which installs the MXI-3 Optimization program on your PC, which must run on bootup of the PC.
    Also, your PXI chassis must be powered on FIRST before your PC is powered up.
    Have you done both of these tasks first?
    Mark

  • Can not send command via VISA test panel in MAX

    Hello all,
    I am having a small issue using the VISA test panel in MAX.  I am trying to send a simple command to a device and then retrieve the response data.  I am able to do this successfully via Labview, but for whatever reason I can not get a good response in MAX.  The command that I am sending is simply a letter (address of the device) followed by an end of line character (\r\n in ascii).  In MAX, I simply append the \r\n onto the string sent in the buffer (Send End on Writes == FALSE).  All serial settings are the same as what I have in LV (19.2, 8-n-1).  Does anyone have any thoughts?  This is kind of frustrating as I just want to do something simple and don't want to have to build a whole routine in LV to do this.
    Cheers, Matt
    Matt Richardson
    Certified LabVIEW Developer
    MSR Consulting, LLC
    Solved!
    Go to Solution.

    Thanks, Broken.  I am getting a time out error.  Clearing the buffers doesn't seem to be making a difference. This is baffling to me.  I must be missing something fundamental.
    Something that I didn't add before is that I seem to be able to send the command via labview without a read, and then read it via MAX.  This must have something to do with way it is sending the command.  Arrrrgh!
    Matt
    Matt Richardson
    Certified LabVIEW Developer
    MSR Consulting, LLC

  • CAN 8473 Not working with Test Panel in Max

    Hello all.  I am working with a USB-8473s CAN unit.  I have it connected to the bus and can see the can traffic through the bus monitor. So everything is working properly there.  I am trying to send a certain message so I set up a message and channel in MAX but when I go to the test panel I get an error saying that the "Test Panel cannot run since no CAN ports are available. Make sure you have at least one NI CAN card inserted into your computer".  What is going on? I thought that the USB can unit was the CAN card?  I'm on Win XP.

    You're not going to like the answer.  The 8473 only supports the frame API and not the channel API.  No messages in MAX.
    http://sine.ni.com/nips/cds/view/p/lang/en/nid/203​384  >> specification tab >> API Support
    I ran into this same problem.
    You'll either have to use the frame API or purchase a different unit that supports the Channel API.
    Robert C. Mortensen
    Certified LabVIEW Architect
    Certified LabVIEW Embedded Systems Developer
    Endigit

  • MAX visa test panel viread error 0xBFFF0015

    I am running labview 8.5 and VISA 4.2. When I go to MAX and open the VISA test panel, the viread malfuctions. When I test it, it returns a timeout error (Hex= 0xBFFF0015). The viwrite tab also gives me the same timeout error sometimes. What can I do to correct this error?
    Also, In the labview program, we are using an RS232 serial port to collect data. Using the visa configure serial port function, we have set the timeout to 10 seconds, and we have set the VISA read byte count to 1. In our code, we have three subVIs in a visa session. The first one works fine, the second and third malfunction, because the bytes at the port show a value of 0.  Is there any relation between this problem and the previous one? Thanks for any help!

    I do not know if you are familiar with the instrument, it is a CSAT3 3D sonic anemometer. I have attached the three subVIs, since I am not sure exactly what the command is. I am very new to labview, so I am not too familiar with the commands.
    The initialize VI is the first one executed, the second is the SA sync, and the read anemometer is the third one.
    Message Edited by Ning123 on 06-20-2008 10:22 AM
    Attachments:
    Read_Anemometer_steve.vi ‏153 KB
    Initialize_SA_Comm.vi ‏136 KB
    SA_sync_steve.vi ‏39 KB

Maybe you are looking for

  • Mail gone, wont quit or display anything after update...

    Hi Folks, In the process of helping my dad upgrade his G4 mini to 10.5.6 from 10.4.something. He's on dialup still so screen sharing to get him going is not really an option and I"m trying to help him with this over the phone and it's slow going. The

  • Why can't i cut and paste

    Downloaded the new Mountain Lion and now I can't cut and paste. Tried the keyboard shortcuts and the drop down menu with no luck. Also can't accomplish this from right clicking. Any ideas?

  • Storedge 3510 failover configuration with 1 host

    I people. I'm new to the storedge configuration. I have a Sun storedge 3510 - 2 controllers with 2 x host port fc and 1x drive port each. I want to do a simple configuration - connect 1 host to the storedge with failover. It's correct to connect the

  • Image dimensions and file size, why the change?

    Here's a Friday puzzler for ya I just ran a batch on a couple of thousand orthophoto images (RGB + 1 alpha channel geo-tif) to add some typographic information to the bottom of each image. All the files were saved in PS as flattened, standard uncompr

  • ASA5510 Enable password not working

    Hi all, I have a problem with an ASA5510 (8.0.4) firewall in South Africa (I'm in the UK). It's a replacement firewall that I am trying to configure remotely through a serial device with an internet facing connection, but the enable password is not w