Cannot measure temperature

Im having troubles measuring temperature with thermocouples. When I measure the temperature of the air or places which are not with potential the card works well. But if I want to measure temperature for example in a transistor that is under a voltage of 13 volts the card shows wrong readings. If I´m not mistaken this card has an input voltage range of +/- 15 volts, so the voltage in the transistor is within the input voltage range.
I read in the manual that you can configure the input voltage for +-.625mv, +- 1.25, +- 2.5, +- 3.75, +- 7.5, +- 15 volts, but I cannot find the tool or the way to do this.
If the common mode voltage is < 2,5 volts I can measure the temperature, but if its bigger I get wrong readings from the card.
I already post this question in another forum, but nobody answered, I guess its more a software question! Im using the plug and play driver for the 435x devices.
Do you have any suggestion of what should I modify to make it work? I attached my code.
Thanks in advance,
Hernar
LabVIEW 8.2
PXI 4351
Attachments:
Temp_Messung.vi ‏140 KB

Hernar,
Thermocouples are made of electrically conductive wires.  If you connect them to an electrically energized point, you will measure a value which is determined by the voltage and impedance of the energized point as well as the thermocouple voltage and impedance.  Since the thermocouple voltages are usually quite small (mV), the voltage from the other circuit (your 13 V transistor) is likely to be much larger.
You need to use an electrically isolated thermocouple to measure the temperature of your energized transistor.  Or connect the thermocouple to the transistor through an electrically insulating, thermally conducting material such as the insulators used to mount power transistors to heatsinks.
Lynn 

Similar Messages

  • How do I run a J-type thermocouple to the usb-6363 DAQ and program labview to measure temperature?

    I have a usb-6363 DAQ and a J-type non-contact thermocouple that I am looking to connect and measure temperature through. However, the DAQ does not have any T/C inputs, which is needed to measure in the thermocouple temperature. I am connecting the thermouple to an analog input (+/-) and I am not looking to buy an amplifier, converter or any other hardware. I believe there is a way to program labview to read in the voltages of the thermocouple and convert it into accurate temperature readings. Any help/ideas?

    Hello George,
    This tutorial should step you through the basic process of configuring the device and connecting the thermocouple:
    Tutorial: Connect Thermocouples to a Data Acquisition (DAQ) Device
    http://www.ni.com/gettingstarted/setuphardware/dataacquisition/thermocouples.htm#Connecting a Thermocouple to Your Device
    From there, there are a number of things you can do- I'd recommend taking a look at the LabVIEW shipping examples (Help>>Find Examples...) as well as the DAQmx getting started tutorials:
    Getting Started with NI-DAQmx: Main Page
    http://www.ni.com/white-paper/5434/en
    At first glance, the 6363 you're using should have enough resolution to acquire usable data from a thermocouple- if you attempt reading raw voltages be sure that the acquisition range is configured for +/- 0.1V, though.
    Regards,
    Tom L.

  • Can I measure temperature using 6013 & 68LPR?

    Hello, from hardware I've got Multifunction DAQ (6013) and Therminal Block (68LPR). From software I've got VI Logger (1.0.1) and NI-DAQ (6.9.3). Using this, can I measure temperature with K-Type therminals? (Everything is installed and ready to go)
    If yes, how can I connect K-Type therminal to thermal block? Are there any instructions on how to use VI Logger to perform similar task to this one?
    Also what I want is, the temperature that has been read (for an hour, every 5mins) to be written in a file. Is it possible to write to a file using VI Logger?
    Thank you!
    Regards Slav.

    Hello:
    I answered this questions over here:
    http://exchange.ni.com/servlet/ProcessRequest?RHIVEID=101&RPAGEID=135&HOID=50650000000800000041CA0000&UCATEGORY_0=_30_%24_12_&UCATEGORY_S=0&USEARCHCONTEXT_QUESTION_0=%2B6013+%2Btemperature&USEARCHCONTEXT_QUESTION_S=0
    Sincerely,
    Brooks B
    Applications Engineer
    National Instruments

  • Measuring temperatur​e using thermocoup​le and SCB-68 box

    I used thermocouple and SCB-68 box to measure temperature. But the readings were fluctuating around the expected value. So I removed the thermocouple and short-circuited the differential input channels. As expected, the voltage reading was not zero and also fluctuating. It seems that there was something wrong with the SCB-68 box. What might lead to this and how to solve this problem? Thanks.

    The problem could be either the SCB-68 or the DAQ card you are using to make the measurement. You should first ensure that the DIP switches (S1 - S5) of your SCB-68 are in the correct position for your application. Please refer to the SCB-68 User Manual for more information on this subject.
    SCB-68 68-Pin Shielded Connector Block User Manual
    http://digital.ni.com/manuals.nsf/websearch/74C86A​DEF0E4813F86256C84007CB3AB?OpenDocument&node=1180_​US
    What do you mean by fluctuating? Full swings from 10V to 10V? Hitting one rail? Close to 0V but looks like noise in mV or uV? If your fluctuation is very slight, for example off by a few millivolts, than you could be picking up noise or your board could need to be calibrated. You can read more about accuracy and
    calibration from the following web page:
    Calibration Solutions from National Instruments
    http://www.ni.com/support/calibrat/
    If you have an E-Series Card you may be interested in running our Online E-Series Diagnostic Utility. This utility takes about 3 minutes to run, will perform a calibration, and will generate a report of anything that may be wrong with your board.
    Online E Series Diagnostic Utility
    http://www.ni.com/support/selftest/
    Regards,
    Justin Britten
    Applications Engineer
    National Instruments

  • Can NI myDAQ ai0 and ai1 be connected directly with thermocouples to measure temperature???

    Hi,
    I am trying to use myDAQ to realize both temperature measurement and digital output.
    By reading this page I have learned that how I can create a task for temperature measurement using thermocouple with a certain NI device.
    http://www.ni.com/white-paper/14338/en/
    What I found to be really interesting is that when I create a task for myDAQ temperature measurement using a thermocouple when my NI myDAQ is connected with my PC, two AI ports ai0 and ai1 both appears in the create task window!!
    I never found any information from NI myDAQ manual that "AI0 and AI1" can be connected directly with a thermocouple to measure temperature without using additional conditioning system. But this window suggests that myDAQ can be used in this way, is it?
    I am still trying to do it this way, and I'll post some results here again.
    I would be really appreciate if anyone could confirm or disprove my guess. Thank  you.
    PS:
    Attached is the 

    Attached is the screenshots for "create new task" in NI MAX
    Attachments:
    1.JPG ‏60 KB
    2.JPG ‏64 KB

  • How can i measure temperature using thermocouple using labview?

    please  list out the procedures to be followed to measure temperature using thermocouple using labview.

    Do a search in the LabVIEW Example Finder (Help->Find Examples).  There are a ton of simple Analog Input examples in there using DAQmx.  One you might be most interested is "Thermocouple - Continuous Input.vi".
    There are only two ways to tell somebody thanks: Kudos and Marked Solutions
    Unofficial Forum Rules and Guidelines

  • Problems on measuring temperature using thermocouples connected to SCB-68 terminal block

    I have problems on measuring temperature using thermocouples connected to SCB-68 terminal block.The thermocouples were connected to the differential channel 1 and a return path for the instrumentation amplifier bias currents and a 4 Hz low pass filter were both developed. But the readings were floating and jumped randomly between 0 and 200 Deg Cel abnormally.The CJC on the terminal block was disabled and a new one was developed by myself. However, the readings for the CJC were accurate and quite stable. So what is the reason for the floating of thermocouple readings? Did I miss anything that I need to do? Thanks.

    Are your TCs floating or grounded? Basically, if your TCs are mounted to a metal test article and the object is grounded to earth, then, the TCs are grounded -- and very likely contaminated with ground loops. You will need to reference the TC inputs accordingly. These ground loops will cause the TC data readings to float - recall that TC readings are in the millivolt level. It does not take much of a ground loop to ruin a TC measurement.
    Also, try this... first, set the filter to 10KHz and sample for a few seconds at 20 KHz (take 32768 data points). Now subtract the mean value from the trace and perform an FFT on the data. Look for the dominant frequency. Is it 60 Hz or some multiple? If yes, the likely cause is a ground reference problem.

  • Measuring temperature and acceleration simultaneously

    Hi,
    I am having problems getting a correct and stable temperature measurement, I think the acceleration measurements are correct. I am using a daq PCI 61110 and the temp sensor is a type-k thermocouple I tried switching the thermocouple with a probe but the result did not change and I have checked the examples to see if I have gotten something wrong in my code but I can't seem to find it. Please could somebody help me and the vi is attached below.
    Solved!
    Go to Solution.
    Attachments:
    MultipleChannels_Created.vi ‏70 KB

    I don't know your DAQ card, but I assume you are measuring in the voltage range for the accelerometer and them try to measure in the µV  from the TC.
    If you have channel crosstalk, your weak signal from the TC gets buried in noise. 
    Try to have some input channels connected to GND inbetween, maybe that helps.
    Wait, I took a look into the spec (BTW 6111 or 6110??) your minimum resolution is 98µV, that would equal around 2K resolution, accuracy is 353µV (around 8K with type K TC ... not quite the right card for a thermocouple reading!!)  Well and you don't have a CJ compensation....in other words: You don't have the right DAQ for thermocouple readings.
    NI has a USB thermocouple device that woul fit your needs and is not too costly... or use a B5 module or other thermocouple signal conditioner with a 10V output.
    Depending on your needs maybe a small NTC with a resistor as an voltage divider can be used to measure temperatur ...
    Greetings from Germany
    Henrik
    LV since v3.1
    “ground” is a convenient fantasy
    '˙˙˙˙uıɐƃɐ lɐıp puɐ °06 ǝuoɥd ɹnoʎ uɹnʇ ǝsɐǝld 'ʎɹɐuıƃɐɯı sı pǝlɐıp ǝʌɐɥ noʎ ɹǝqɯnu ǝɥʇ'

  • How do I measure temperature from 4 RTDs simultaneously?

    I am trying to set up a systerm that requires multiple temperature measurements. I am able to make one RTD measure. But if I mirror the same code to create two measurements, i am getting an Error - 50103 "the specified resource is reserved". I have just begun using LabVIEW, so any and all help is valuable. If there is already material available on this topic, would you direct me there please.
    Solved!
    Go to Solution.

    Yup read Multi-channel Single Sample with a single task.
    That being said this won't be "simultaneous".  That implies at the same instant in time all 4 will be read.  This just won't be true without a simultaneous sampling card.  Without that card your readings will likely be fraction of a milliseconds appart and for temperature that should be more than enough.  How much can temperature change in 0.001 seconds?  But for other signal types like trying to read a phase off set in a sine wave you will need a true simultaneous sampling card.
    Unofficial Forum Rules and Guidelines - Hooovahh - LabVIEW Overlord
    If 10 out of 10 experts in any field say something is bad, you should probably take their opinion seriously.

  • How to use labView to measure temperature using a thermistor

    I'm trying to create a labView that will give a temperature reading based on a circuit using a thermistor. Any ideas on the VI or how to construct the circuit would be helpful!

    Well, you need to generate a voltage. You also need a way to read that voltage. Do you have a meter, or a DAQ device?
    Temperature Measurements with Thermistors: How-To Guide
    As for other circuits, the internet is a good place to search. Have you tried? Here's a couple of places:
    http://www.discovercircuits.com/T/therm.htm
    http://www.facstaff.bucknell.edu/mastascu/elessonsHTML/Sensors/TempR.html
    http://www.ecircuitcenter.com/Circuits/therm_ckt1/therm_ckt1.htm

  • Measuring temperature with PCI-4351 & TC-2190

    Hi,
    I am a new LabView developer. I have a PCI-4351 data logger card and a TC-2190 thermocouple chassis. I cannot communicate with the 4351 cuz I don't know how. I've set up a channel with MAC and the hardware test works okay, but I can't find the right object to grab the measurement when I'm creating the vi. Do I use NI-DAQmx or VI Logger functions to take simple instantaneous temp. measurements?
    Is there some example code I can start with? I tried to run the ni_435x_thermocouple.vi example that I downloaded from the developer zone, but there are several sub-vi's not loaded on my machine (eg: AI Hardware Config.vi) that it tries to find upon launch. Looks like a vestiges of LabView 5 that aren't there anymore. Is there an easier way for this LabView 8 user?
    Sincerely,
    Jack Santori
    410-471-6262

    Hi Jack,
    The PCI-4351 only has support for the Traditional NI-DAQ functions and not NI-DAQmx.  It sounds like you don't currently have those drivers installed on the system, and this is why the Traditional NI-DAQ function AI Hardware Config.vi cannot be found.
    If you are running Windows 2000 or XP, the most current version of the Traditional DAQ drivers available for download is 7.4.1 and they can be downloaded here.  If you are not using one of these operating systems, Traditional DAQ driver versions for another OS can be found here.  Example programs are a great place to start, and the Developer Zone includes a wealth of information.  Installing the Traditional NI-DAQ drivers will also include example programs that you can open and run in LabVIEW 8.0 by going to Help>Find Examples.
    Regards,
    Andrew W
    National Instruments

  • Can not measure temperature and voltage simultaneously

    I am a beginner with labview. I want to measre temperature and voltage simultaneously. When I run
    the VI, I can get temperature or voltage, but not togther. I attach my VI, please give me suggestion on
    how to make it work. 3X
    Attachments:
    heatflux.vi ‏1069 KB

    Since I don't know your exact configuration I will make some basic assumptions based on how it appears that you have configured the DAQ Assistant Express VIs.
    Assumptions:
    1. You have only one DAQ board in your system.
    2. You want to scan continously
    3. You want to acquire 3 temperature channels at a rate of 1000S/s and take 100 readings at a time
    4. You want to acquire 2 voltage channels at a rate of 1000S/s and take 1000 readings at a time.
    Based on this configuration your first problem is that you have configured the DAQ board to acquire continously in the first call to the DAQ Assistant (your first frame of the sequence structure). This ties up all the analog acquisition resources without releasing them. When you make your second call to the DAQ Assistant (your second frame of the sequence structure) you are creating a conflict because the DAQ board is already busy running your first request. At this point you are probably receiving an error but you might not see this since you are not doing error checking in your code. This is also why you are only getting one set of data. Next iteration of the while loop the first call to the DAQ Assistance reconfigures the board and executes again and so the cycle repeats itself.
    I don't have a DAQ board installed so I can't confirm with certainty if I am correct but you can do this by simply changing the DAQ Assistant properties. In the 'Task Timing' tab change from 'Acquire Continuosly' to 'Acquire N Samples'.
    Assuming this works all you have done is confirmed that my assumptions are correct and technically your program should work. So now some programming advice.
    It's ok to scan all channels at once even though they might not be of the same type so go ahead and configure all your channels in one DAQ Assistant call and get rid of the sequence structure. Decide on one set of parameters for Scan Rate and Samples to Read, in your case I doubt this will be a problem. Since you are performing the same analysis on all channels you don't need to parse your data simply pass the 'data' from your DAQ Assistant into a single 'Amplitude and Level Measurements' Express VI. You will now have a single array with all your Mean values based on the order the channels are configured. If you want to plot the data in different graphs all you need to do is parse your channels using the 'Split Signals' or 'Select Signals' Express VI.
    Hope this makes some sense.
    -Christer

  • Measuring temperature Via Thermocouple and USB-6009?

    Hi,
    I am trying to perform temperature measurements using a k type thermocouple and the USB-6009 DAQ. I tried to use the Cont Acq Thermocouple Sample.vi from LabVIEW's example list, but unfortunately I don't know how to fit my DAQ and thermocouple in the vi code. I don't know how to create a thermocouple temperature measurement channel or how to get all the channels in the task, etc.
    Any help is appreciated.
    Thanks.

    bilbeigg,
    It should be noted that the USB-6009 is not well suited for thermocouple measurements.  It has 14 bits of precision with a minimum input range of +/-1V, which means it can only detect a voltage change as small as about 122 microvolts.  This is much too large a change to get a good temperature reading with most types of thermocouples.  Here is a discussion thread on the same topic (Note that the USB-6008 and 6009 are very similar):
    http://forums.ni.com/ni/board/message?board.id=170&message.id=294080#M294080
    Here is a knowledgebase that explains what you are seeing:
    http://digital.ni.com/public.nsf/allkb/8166630BD6965EEA86256FDD0052E89C?OpenDocument
    Chris M
    Message Edited by CMal on 03-12-2008 12:17 PM

  • Simple VI to measure Temperature using Agilent 34901A

    I am trying to measure/record temperatures using Agilent 34901A and 34970A. I have all the T type TCs connected. I am not familiar with Agilent drivers so if anyone can provide some guidance. I just need simple VI. I am familiar with typical DAQ Assist VI and acquire signal from NI Hardware but no experience with Agilent. I got the drivers for Agilent but it seems much more complex than I need it to be. 

    I have been messing with a variation of Agilent drivers for the last couple of days. Labview's built in Visa drivers don't handle Agilent stuff quite correctly, and so you need to install their driver version (with the Agilent IO Libraries Suite).  There is lots of documentation showing you how to configure LabVEW to work with their Visa drivers.  Once you get that taken care of, I'm sure there are sample programs you can model off of.  For example, try here:
    http://www.home.agilent.com/agilent/editorial.jspx?cc=US&lc=eng&ckey=1838058&nid=-11143.0.00&id=1838...
    And don't forget to apply the thermocouple coeffecients in to your program if you scan the voltages directly
    http://srdata.nist.gov/its90/type_j/jcoefficients_inverse.html
    Edit: Typo!

  • Measuring temperature with RTDs over a long distance

    I have created an application using the FP-RTD-122 to take RTD mearurements from multiple locations in a plant. The measurments are correct when the RTD leads are connected directly to the module (about 6ft). However when we wire the RTDs in the plant to the module, the measuremtns are incorrect. For instance the temerature is 79C at one location, but shows as between -193C and 19C in Measurment and Automation Explorer. The distance is about 150 ft. I have found a reference in the internet that suggests using a temperature transmitter that converts the RTD signal to a 4 to 20 mA signal. Does anyone have any experience with measuring RTD over long distances? Do I really need a temperature transitter? If so, do I need to use the FP-AI-111 module instead?

    Nate,
    RTDs allow temperature measurement by reading the resistance of a sensor. This is normally done by exciting the sensor with a known current and measuring the resulting voltage. There are three general types of RTDs; two-wire, three-wire, and four-wire. Two wire RTDs are less accurate because the wire resistance is indistinguishable from the sensor resistance. Three wire RTDs use two wires on one leg (one to carry the known current and one to measure the voltage) allowing one portion of the wire resistance to be nullified. 4 Wire RTDs use two wires on each leg (one for the excitation current) to nullify the wire resistance. As a result, 4 Wire RTDs are the best ones to use for high precision or long distance measurements. To measure 4 Wire RTDs with FieldPoint, you will need to use the [c]FP-RTD-124 module instead of the [c]FP-RTD-122. With that said, I do not know how well 4 wire RTDs will do at 150 feet. 4-20 mA transmitters will easily handle 150 feet so they will probably be your best option. To measure the 4-20 mA signal, you can use the [c]FP-AI-100, [c]FP-AI-110, [c]FP-AI-111, [c]FP-AIO-600, or [c]FP-AIO-610 modules.
    Regards,
    Aaron