Configurat​ion check request

we need test 10 pieces of thermocouples ,3 LVDT signals and 3 strain guage at one time.we both want the configuration of the signal conditioning and DAQ device meet our requirements and the low cost,so we select the following configuration:
SCXI-1000
SCXI-1520
SCXI-1330(instead of SCXI-1314)
SCXI-1540
SCXI-1330(instead of SCXI-1315)
SCXI-1102
SCXI-1330(instead of SCXI-1300)
PCI-6023E(12 bit is enough)
SSH6868
SCXI-1360,SCXI-1361
we hope these selections can make a whole DAQ system,could you check the configuration and please tell us if there is existing any mistake,and welcom better configuration. thanks

Instead of the SSH6868, you will need a shielded cable assembly to connect your 60xxE (E-series card) to the SCXI chassis. You will require the SCXI-1349 cable assembly. It has two parts 1)SH6868 cable and 2) an adapter to connect to the SCXI backplane.
As for the choice of the DAQ card,you may find the low cost approach severly limits the system bandwidth and overall accuracy. The low cost E-series boards only have 512 samples of onboard memory, as opposed to 8192 for the regular E-series. This makes controlling buffered analog input more difficult. In addition, you need to understand that the rated sampling rate for the board is based upon a single data channel. As you have 16 channels, your "best" sampling rate would be approximately 12.5 KHz per channel. The actual rate will be less. You have made no mention of the type of data you want to acquire (steady-state, dynamic, time domain, frequency domain), so, I can not give any better insight into the "best" sampling rate you could achieve.
Now, considering that you are mixing thermocouple (TC) readings with other inputs on the same DAQ call, you may find the 12-bit selection aggravates your data quality. When you acquire data using a multiplexer, the quoted accuracy for the device assumes that all of the voltage levels on the various channels are very near the same value. This allows a relatively uniform settling time for the input signal. It is quite possible that the TC inputs will be on the order of 2-10 mV while the other signals could range as high as 0.5 volts. Such a difference in input levels can lead to settling errors that are not obvious to the user, but can cause gross errors in the measured data levels. If you set the input range (only 4 for the low cost e-series) to +0.5/-0.5 volts, your resolution for the TCs will be about +/- 0.24 mV or about 6-10 degrees of error. (ouch!!). So you will be required to set different input ranges for the different input channels (0.05/-0.05 for the TC and +5/-5 or +.5/-.5 for the remaining channels. These multiple range calls can further limit your bandwidth by slowing down the "round-robin" interval between the data range changes. I know that your goal is low cost; however, you may find that the data quality is unacceptable. You will want to consider, at the very least, upgrading to the 6034E board. It is a 16-bit device (still only four input ranges); however, if you configure the board for only 1 input range, +0.5/-0.5 volts, your bandwith will improve and the TC resolution will improve by a factor of 16 times to approximately 0.015 mV (under 1 degree).

Similar Messages

  • Auto Spell Check Requested! Or just move the button, please.

    Auto Spell Check Requested! Or just move the button, please.

    Thanks for the suggestion.  We'll make sure to pass it along.
    If a forum member gives an answer you like, give them the Kudos they deserve. If a member gives you the answer to your question, mark the answer that solved your issue as the accepted solution.

  • In RSRT - Is it possible to check request wise data in RSRT only.

    Hi,
    In RSRT - Is it possible to check request wise data in RSRT only.
    Kindly advise me on the same.
    Thanks
    Bujji

    Saveen,
    Here is my problem.
    I have a infocube in which Material No and Base Unit are there.
    I have a DSO in which Material, (Alternate unit, Numerator and Denominator) which are associated to that material are there .
    I need to match the same material number from both infocube and DSO and load the associated Alternate unit, Numerator and denominator in the infocube.
    Since the infocube is non cumulative, I am not able to build Infoset.
    So I added the Infoobjects of (Alternate unit, Numerator and Denominator) to the cube.
    Now the cube has Material no, base unit-------For Both data is filled.
             and extra Alt unit, Numerator and Denominator- For these data empty.
    I need to load the alt unit Numerator and denominator from the DSO for which the Material no matches with the Infocube.
    I am not very good in explanation.Hope u understand. Pls adjust with the long text...
    Pls help me.
    Thanks.
    Guru

  • Create NSAPI SAF plug-in equivalent to check-request-limits

    Hi Masters,
    GoodDay. Okay, here's the problem:
    I'm a newbie in writing a custom SAF's, & what I'd like to do is to write a SAF which is closer to this one:
    +PathCheck fn="check-request-limits" monitor="$ip" max-rps="10" specifically the check-request-limits fn.+
    I believe source code of the server is not available so the question is:
    How can I get total request of a specific client in a given time (secs for example) [ip->TOTAL_REQUEST_IN_1_SEC]?
    char *ip = pblock_findval("ip", sn->client);
    //NEED HELP HERE
    if (ip->TOTAL_REQUEST_IN_1_SEC < 10 ) {
         //my logic goes here
    }Any help would do! Thanks in advance!

    Source code is out. Check this thread :
    Announcing Open Web Server [http://forums.sun.com/thread.jspa?forumID=759&threadID=5360537|http://forums.sun.com/thread.jspa?forumID=759&threadID=5360537]

  • To check request scoped component in dyn/admin

    How do i check request scoped component in dyn/admin console?
    I have enabled loggingDebug for my formHandler but none of the debug messages appear in the console.
    I need to check the dyn/admin to see if loggingDebug=true.

    1004856 wrote:
    How do i check request scoped component in dyn/admin console?
    Request/Session scoped components will not appears in component browser. You can find them like this:
    [1]. In dyn/admin click on Component Browser [http://localhost:8080/dyn/admin/nucleus]
    [2]. Suppose you want to open CartModifierFormHandler that is request scoped. then directly append full component path in url after config like below:
    http://localhost:8080/dyn/admin/nucleus/atg/commerce/order/purchase/CartModifierFormHandler/
    And Gurvinder already mentioned about enabling loggingDebug for such components
    -RMishra

  • Finance for Wire , Check request ,Fixed Assets for the Capital Expenditure

    Hi,
    We are trying integrate the Finance for Wire , Check request ,Fixed Assets for the Capital Expenditure with adobe forms.
    Where would i find more information regarding the same?
    what tables are used and what are related t-codes which can give more information?
    any documentation is appreciated.
    rgds
    vara

    Hi,
    We are trying integrate the Finance for Wire , Check request ,Fixed Assets for the Capital Expenditure with adobe forms.
    Where would i find more information regarding the same?
    what tables are used and what are related t-codes which can give more information?
    any documentation is appreciated.
    rgds
    vara

  • Checking requesting cost center to a company code

    Hi All,
    How to check a requesting cost centre which is belong to a company code ?
    Thank you.
    Nies

    You need to use validation along with a user exit.
    User Exit is some where you write the logic that the company code (BUKRS) of the requesting cost center (AKSTL) should be same as the company code on the WBS (PBUKR)
    Take help of ABAPer in writing the logic of the user exit.
    Regards
    Sreenivas
    Pls close the post if satisfied.

  • When is the average rps computed? (check-request-limits throttling)

    Hi all,
    The documentation indicates that the default interval for computing the average request per second is 30 seconds. My question is: When does this 30 second start and end? Is it based on the server clock? That is, it computes the average rps @ 01:01:00 then @ 01:01:30 then @ 01:02:00 etc? Or is it based on the server's up-time? Or perhaps based on another algorithm all together?
    I simply need to know the frame of reference for computing the average rps. The documentation and the blog posts by Jyri Virkki are not clear about this.
    Thank you in advance,

    IIRC, the interval is computed on demand when needed for a given rule. The interval is a minimum interval, if no matching requests arrive then the rps won't be computed again until needed.
    FYI Sun open-sourced most of the Web Server code, so you can check the implementation directly if you wish.
    Looks like the relevant part is starting on line 308 of http://heliod.svn.sourceforge.net/viewvc/heliod/trunk/src/server/safs/reqlimit.cpp?revision=2&view=markup

  • The "write key" configurat​ion file vi use of "trim string" prior to writing the data can modify any string data written.

    I tried to use the config VIs to record some front-panel settings for later restoration, one of which could be a single space character (part of a string parsing system).
    I soon discovered that whenever I tried to save that single-space value to an INI file, only a null string was saved.
    After doing some digging I discovered that buried in the Write Key vi is a worker vi called Config Data Modify that uses Trim String on the string data before it is written to the file and that's what was eating my string character. I don't know whether this is a bug or a feature but there are at least three ways to fix it.
    1) Assuming you want to leave the library VIs alone, you can pre-process any stings sent to "write key" to replace all spaces with "\20" and then post-process all strings read using "read key" to replace all instances of \20 with spaces.
      and if you don't mind modifying the library VIs, either to save/use under a different name or to stick back into the library in a modified state (caution - can cause problems when you move code to another machine with an un-modified library) then...
    2) You can yank the trim-string out of the Config Data Modify vi and hope that it does not have any undesirable side effects with regards to the other routines that use Config Data Modify (so far I have not found any in my limited testing)
    or
    3)  You can modify the string pre-processing vi, Remove Unprintable Chars, to add the space character to the list of characters that get swapped out automatically.
    Note that both option #1 (as suggested above) and option #3 will produce an INI file data entry that looks like    key="\20Hello\20World\20"   while option #2 produces an entry that looks like   key=" Hello World "
    The attached PDF contains screenshots of all this.
    Attachments:
    Binder1.pdf ‏2507 KB

    Hi Warren,
    there's a 4th option:
    Simply set the "write raw string" input of the write key function to TRUE
    This option only appears when a string is wired to that function!
    Just re-checked:
    I think it's a limitation of the config file format. It's text based and (leading) spaces in the value are "overseen" as whitespaces. So your next option would be to use quotes around your string with spaces...
    Message Edited by GerdW on 05-02-2009 08:32 PM
    Best regards,
    GerdW
    CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
    Kudos are welcome

  • How to get the configurat​ion of a bank (not with FP get configurat​ion info)?

    Hi everybody!
        I need my fieldPoint application to be able to check
    for the filedpoint structure, that is I want my application
    to return the number of FP nodes, and for each node how many FP IO modules are connected, and of which
    type. 
       I'd like not to use "FP get configuration info", 
    which is based on locating the .iak file.  I have my .iak file
    correctly generated, but I do not want to use it. 
       At the end, I'd like to have a variable configuration, and when I change module order and number, I don't want
    to re-generate a .iak file, (because I don't want to use MAX... ).
       Should I query my modules until I reach an error?......
    graziano

    First of all, thanks, Khalid!
       I want to use the COM port. I gave a look to the "Find
    FP-100x example", based on optomux commands, and I found it very
    interesting, to my purpose. But I have some questions about it.
       First of all, in the example I have to provide explicitly
    an address (the fp1001 address, 0x0 for me), so I should build a loop,
    in which I scan all possible addresses (that is 0, 10, 20,... ) which
    generate many errors, in normal conditions.
        After this, in the example I get a Module ID, which is:
       FP1001 -----> ID: 2
       FP-DI-301 --> ID 105
       FP-DO-401 --> ID 106
       FP-DI-301 --> ID 105
       FP-DO-401 --> ID 106.
       The list actually matches the order of my FP chain, but
    I'd like to know if  these IDs are given to the modules when
    they're configured for the first time in MAX, or are set in HW?
       Thanks!
    graziano

  • Not respond to IRQ level and configurat​ion problem

    Hi
    Problem1: I have a Windows XP desktop, and a PCI 6036E. I tried to use measurement and automation to test the PCI 6036E device. The error shows that not respond to IRQ level. 
    Problem2: I have a Labview program. It can read in the sensor signal from PCI card. The signal can be tuned by external amplifier. But even I tuned the signal to 0. The program still reads around 2 (voltage). I suspect the problem might be due to the configuration of my card which I didn't do. So how can I config or calibrate my card?
    Thanks

    Hi RobinX, 
    Below is the Knowledge Base I have been referring to on this problem, which is why I asked if you could use later version of the DAQ drivers where this has been addressed.  Please take a moment to read through the document that goes into more detail on the problem.  As to why you never saw a problem with the other machine I am still looking into.  Is it possible it is using a different operating system and/or has a dual core processor that handles interupts differently?  The KB below shows you how to check what form of interupts the computer is using.  This may be the main cause of the IRQ issue instead of different MAX versions.  
    KB: 373FLOA6: PC Card (PCMCIA) Device Can Exhibit Interupt Failures on Windows 2000 and XP
    http://digital.ni.com/public.nsf/allkb/14DFB1A1788​D1F3286256E4D007CA783?OpenDocument 
    Regards, 
    Kyle S
    Applications Engineer
    National Instruments

  • 4.6 sync configurat​ion error code 0x3b21

    Hi.  I've finally got desktop manager 4.6 running and my blackberry 8330 backed up.  When I try to configure syncronization for Outlook 2000, the error "Warning encountered--warning code 0x3b21 check documentation" comes up.  I'm having trouble tracking down the code/problem.  Any suggestions?
    Thanks!

    Hi,
    From what I can see in the documentation for 4.6 it does not support Outlook 2000.
    I suggest 3.6. Outlook 2000 was discontinued in 4.0
    Download 3.6 from here:
    https://www.blackberry.com/Downloads/entry.do?code​=A8BAA56554F96369AB93E4F3BB068C22
    Please let me know how it goes.
    Thanks
    Click Accept as Solution for posts that have solved your issue(s)!
    Be sure to click Like! for those who have helped you.
    Install BlackBerry Protect it's a free application designed to help find your lost BlackBerry smartphone, and keep the information on it secure.

  • EPrint Configurat​ion: Option to remove document URL

    Hi,
    thank you very much for providing this useful app.
    I am having trouble configuring the printer-settings when printing out a document that is stored online (e.g. Google Docs or iWorks, when documents are opened through the ePrint browser): I cannot print the document without it's URL being displayed on top.
    So far the ePrint app for the iPad only allows me to display or remove the URL on the bottom of the document, however it is always displayed on top of the document (even in bold letters).
    It is imperative for this app to be able to print out documents whithout bold URLs on top of them in order to make ePrint useful for printing regular documents such as letters etc.
    Could you please offer the option to remove the URL on top of a page in the next Update of ePrint for the iPad?
    Thank you very much!

    Hi Warren,
    there's a 4th option:
    Simply set the "write raw string" input of the write key function to TRUE
    This option only appears when a string is wired to that function!
    Just re-checked:
    I think it's a limitation of the config file format. It's text based and (leading) spaces in the value are "overseen" as whitespaces. So your next option would be to use quotes around your string with spaces...
    Message Edited by GerdW on 05-02-2009 08:32 PM
    Best regards,
    GerdW
    CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
    Kudos are welcome

  • How to disable automatic launch of WIE in startup configurat​ion with no options to do so?

    Thinkpad T400 2764CTO running Windows 7 Home Premium 64
    PC only 2 weeks old, very little on it, but getting slower and slower. Finally found the configuration for the startup menu and disabled several things I knew I didn't want at startup. One problem that I can't seem to figure out, is how to disable Windows Internet Explorer from being launched at startup? I have tried disabling every Microsoft option, and it still launches. Is there a specific Lenovo option [that does not obviously have anything to to with WIE] in the startup menu that, if disabled, will stop the WIE browser from being launched? How do I know which options are not safe to disable when attempting to slim down startup configuration for faster/more reliable performance?
    Thanks,
    Ann

    Either I *completely* misread your instructions; that fix is major overkill for what I was trying to do; that fix does not apply to my problem; that fix is only for XP or Vista and not W7; or my PC is already so messed up (despite clean hardware diagnostics, no viruses, no malware, and it is only 2 weeks old), that it performed unexpected and unwanted actions when trying to use the Windows features on/off program control panel options.
    Opened Control Panel, went to Programs, and finally found the option on the left for turning Windows features on and off. Unchecked the box for IE8, which required a restart. The restart time was 4-5 times longer than usual, even when launching superfluous programs from the startup menu, since it had to "reconfigure Windows." After restarting, WIE did not automatically launch, and its toolbar icon was gone too. Tried to turn it back on like you said, which also required a restart, which didn't make sense to me, since I expected the browser to launch again following another extra long restart once the box was re-selected. Although, surprisingly, it didn't, so for a moment thought it had somehow worked. Then I realized that my WIE icon was still missing from the toolbar, so there was no way to launch it even if I wanted to. Went back into the Control panel and checked every Windows feature box possible; another extra-long restart; still no toolbar icon for WIE. Had to go into the computer files to find WIE so that I could create a desktop shortcut icon so I could manually launch the browser from that. However, when I clicked on the shortcut, it went into a browser setup mode, as if it was being set up for the first time. Finished setup and tried to go to a webpage, but kept getting a connection error.
    Diagnosed it, and it said that the Local Area Connection had no valid IP address. Waited on hold with Mediacom for a hour before being disconnected, trying to find out what the IP address was and where to re-enter it, since unchecking the IE8 box in the Windows features part of the Control Panel, doesn't just temporarily disable the function, but apparently actually deletes the toolbar icon, as well as the IP address. Took several more attempts over several hours to get ahold of Mediacom reset the signal to get the IP address back and get back online. So that really does not seem like a viable solution for trying to save time and resources by removing something from the startup menu, if it just quadruples startup time, deletes WIE toolbar icon and IP address permanently, requires creating a desktop shortcut, and then requires several more hours on phone with ISP for signal reset to get the IP address back. Would have been much faster and done less apparent damage than just going ahead and letting it launch at startup and then just closing it. *sigh*
    I know that just having an icon in the toolbar does not mean that that particular program will launch at startup, as I have several others that have toolbar icons that I have been able to uncheck in the startup menu, so I can use the toolbar icons to launch each program manually. There just doesn't seem to be any program active in the startup menu that would be launching the browser, and still concerned about unchecking something necessary in startup that could screw it up even more.
    Is there any way to get my toolbar WIE icon back without a full system restore, so that I have less shortcut clutter on my desktop?

  • Rescue and Recovery with T40 booting the system (XP) without restoring factory configurat​ion

    (plus bas en français)
    Hi,
    I've a T40 which I want to restore the factory configuration.
    I've installed the up-to-date version of Rescue and Recovery but :
    - when launching Rescue & Recovery from XP, it does not offer to restore the factory configuration
    - at least, this is feasible when I press the "Access IBM" blue button when starting the computer...
    - ... even if factory configuration restore is available there, it does not work. Indeed, it asks me to confirm and then starts the computer again without restoring anything.
    Can someone help me please ?
    Bonjour,
    J'ai un T40 pour lequel je souhaite restaurer la configuration usine.
    J'ai installé la toute dernière version de Rescue and Recovery mais :
    - quand je lance Rescue & Recovery depuis XP, je n'ai pas la possibilité de restaurer la configuration usine.
    - à la rigueur, c'est pas grave car quand j'appuie sur le bouton bleu "Access IBM" lors du démarrage du pc, Rescue & Recovery me propose de restaurer la configuration d'usine... mais ne fonctionne pas. En effet, après avoir confirmé l'opération, le système redémarre et lance XP sans avoir restaurer quoi que ce soit.
    Quelqu'un peut m'aider svp ?

    hey sylvain,
    for the Rescue and Recovery to work, did you make a backup of the system via recovery media ?
    WW Social Media
    Important Note: If you need help, post your question in the forum, and include your system type, model number and OS. Do not post your serial number.
    Did someone help you today? Press the star on the left to thank them with a Kudo!
    If you find a post helpful and it answers your question, please mark it as an "Accepted Solution"!
    Follow @LenovoForums on Twitter!
    Have you checked out the Community Knowledgebase yet?!
    How to send a private message? --> Check out this article.

Maybe you are looking for