How to connect 2-wire strain gage to NI 9237 by quarter bridge

I have a Strain Gage: KYOWA KFG-20-120-C1-11 (2 wire) http://www.kyowa-ei.co.jp/english/products/gages/pdf/kfg-01.pdf, NI 9237 on a Compact DAQ (cDAQ 9174) and RJ-50 (10-pin modular plug to pigtail wires).
Strain Gage features:
Gage Factor: 2.08 +/- 1.0%
Gage Resistance: 120.2 +/- 0.2 Ohms
I'd like to connect them to measure the strain with quarter bridge configuration but I have some doubts about which line goes to which terminal.
Thanks.
Solved!
Go to Solution.

You can find the manual for the NI 9237 module here
It's a good idea to buy an NI 9944 or NI 9945 Quarter Bridge Completion Accessory if you're planning on using this module with a quarter bridge configuration, but  you can look at the following diagrams to see what kind of internal connections are implemented: Connecting Strain Gauges and Shunt Resistors to the NI 9237
PS Posting in the DAQ or signal conditioning forums instead of Labview may get you better/faster answers. 
Cheers
Message Edited by mishkin on 03-11-2010 02:40 PM
Misha

Similar Messages

  • Please tell me how to connect 3-wire-strang gage to NI 1314 (SCXI1520) by 1/4 bridges

    I'd like to connect 3 wire strain gage to NI-1314 (SCXI-1520) to measure the strain with quarter bridge configuration
    Please show me which line goes to which terminal..

    Hello blackburn,
    All of the wiring information is included in the SCXI-1520 user manual. Navigate to Chapter 2 >> Connecting Bridge Sensor Signals >> Quarter-Bridge Type I. The link below will take you to a PDF version of the manual.
    http://www.ni.com/pdf/manuals/322583b.pdf
    I hope this helps!
    Matthew C
    Applications Engineer
    National Instruments

  • Why do I need to connect QTR and S+ directly on the strain gage?

    I connected two wires with a 2" 120-ohm strain gage. Later, connect these wires to a strip connector. Now connect the red wire from NI system (SCXI-1314 which is already converted for 120-ohm) to red in the strip. This red is P+ in SCXI-1314. Now I connect the black (QTR in SCXI-1314) and white (S+ in SCXI-1314) to the white wire on the strip connector. Unfortunately this did not work. I got a constant strain output. But when  I connect the black (QTR in SCXI-1314) and white (S+ in SCXI-1314) to the strain gage directly it works fine.
    I tried with 4 strain gages and got the same result.
    It will be great if any one can tell me what went wrong?

    Hi Shaurav,
                Sorry, in my last post, I meant to say "SCXI-1520", which I am assuming you have  because you have the 1314 terminal block.  The SCXI-1520 has programmable gain, as mentioned HERE.  Also, I recommend looking at THIS discussion forum post becauseit talks about the different ways to set that gain.  Have you tried applying a gain in Measurement and Automation Explorer?  How are you programming this?  Thanks!
    aNItaB
    Applications Engineer
    National Instruments
    Digital Multimeters

  • How do I connect 16 strain gauges in quarter bridge arrangement to a SCXI 1322 connecter block?

    Specifically, do I connect 4 bridge completion resistors to each VEX- channel or do I need to wire them together somehow first?

    I have attached a bitmap image that explains how to wire a quarter-bridge strain gauge to the SCXI-1122 (with 1322 terminal block.) The 1122 only provides half-bridge completion, so you have to wire an extra resistor into the circuit to complete the bridge.
    You can find this image in the SCXI-1122 User Manual, which is available for download at:
    http://search.ni.com/search/query.html?col=alldocs&qp=%2Bcontenttype%3Aproductmanual&lk=2&layout=ProductManual&qt=1122
    Attachments:
    1122_strain.bmp ‏920 KB

  • How to I import MAX 4.6 config files into MAX 5.5? Missing Data Neighborhood. Strain gage set up is different.

    I have been trying to import my config file from MAX 4.6 into a new system running MAX 5.5 and daq 9.8.
    The data neighborhood does not appear, however my application sees the information. How can this be remedied? Start over?  I have a lot of inputs and would rather not have to recreate all of them.
    What is the difference between the strain gage settings in MAX 4.6 and MAX 5.5? The selection box for the bridge in 4.6 has a selection for "let niDAQ choose". Seems that is necessary to use the 4.6 config data. Try though I might, the readings in 5.5 do not match those in 4.6.     Is there a work around fo this? I don't want to have to perform an entire system calibration.
    Thanks
    Chris
    Solved!
    Go to Solution.

    Attached JPG's.
    Here is more information regarding the MAX Strain Set Up.
    Attached are screen shots from MAX 4.6 and MAX 5.5. 
    Hardware:  PXI chassis with bridge to a PC.  SCXI-1520 input card.
    Software: Old system:  Windows XP SP2 running MAX 4.6 with niDAQmx 9.0.2,   programming in LabView 2011
                       New system: Windows 7 running MAX 5.5 with niDAQmx 9.8.  Programming in LabView 2013
    5.5:  What is the gage factor?  Is that the mv/v sensitivity of the load cell?
    5.5:  Asks for Gage resistance, easy to look up and I entered the data, 350.
    5.5:  How do I determine the Initial Voltage?
    5.5:  No choice of using V Excitation for scaling
    5.5:  Is Full Bridge I the same as 4.6 Full Bridge?
    5.5:  Requires use of Strain in scale factor
    4.6:  Requires use of Volts in scale factor (flags an error if I try Strain and requests Volts)
    4.6:  Choosing to use Vex for scaling makes readings look similar to 5.5 readings, but still not correct. ie around -79 in 5.5 and around -50 in 4.6 vs. 5.2 without.
    Is there a way to set MAX 5.5 so that it provides the same starin set  up function as 4.6? If so, what is it?
    Thanks,
    Attachments:
    Max 4.6 Strain Set Up.jpg ‏32 KB
    Max 5.5 Strain Setup.jpg ‏23 KB
    MAX 4.6 Strain Scale 2.JPG ‏165 KB

  • How can I connect 1-wire sensor using DAQ?

    Hi,
    How can I connect 1-wire sensor using DIO in DAQ? I didn't find anything about it on this site.
    Thanks.

    Hello Depablo,
    When making any voltage measurement you need to provide two wires and signal and a reference voltage.  If you have a digital sensor or device with a number of outputs there should also be a reference or ground voltage output so that you're measurement device can read the correct value.  If you don't connect the ground/reference voltage to the DAQ device then you'll be measuring a floating voltage and it won't be accurate.  Voltage is a relative measurement--your sensor may be outputting 5 volts relative to its internal ground, but that ground may be 10 volts above your DAQ cards ground.  If you measure the 5 volt line with your DAQ card you'll read 15 volts and not 5 volts unless you also connect the sensors ground to the DAQ card.  If you're unsure which wire to use as a reference you can either contact your sensor manufacturer, or if you like you can post back with a link to your sensor's data sheet and I can see if I can identify the correct wire.
    I hope this helps, and have a great day!
    Brooks

  • Can I measure Vibrating Wire Strain Gauges using Field Point Inputs?

    Is wish to connect a Vibrating Wire Strain Gauge using Field Point.
    The sensor requires a swept frequency excitation (typically 400 to 1500 Hz) and then generates a frequency output proportional to the strain applied on the vibrating element. The frequency (roughly upto 1kHz) needs to be measured.

    What sort of resolution is needed on the frequency sweeping? There are currently four FieldPoint modules that can be used for generating pulsed outputs; the FP-CTR-500, FP-CTR-502, FP-PWM-520 and FP-PG-522. For the remained of this posting, I will use FP-CTR-50X to represent both the FP-CTR-500 and FP-CTR-502 as they are identical in functionality, different in their input/output stages.
    The FP-PWM-520 is a Pulse Width Modulation output with 12 bit resolution on duty cycle and 16 bit resolution on the pulse period. The period is adjustable from 1 to 655535 milliseconds, which makes the module slightly less than ideal for frequency sweeping as it's highest frequency is 1 kHz and it's second highest frequency is 500 Hz.
    The FP-PG-522 is a Pulse Generator module that can operate in an continuous pulse mode. The On Time, Off Time and resolution (time base) are adjustable. The smallest time base is 100 ms, for a maximum output frequency of 5 kHz. Like the FP-PWM-520, the FP-PG-522 adjusts it's output rates by changing the time rather than the frequncy, making it difficult to smoothly sweep across frequencies.
    Finally, the FP-CTR-50X modules may be used. The FP-CTR-50X modules have 8 16 bit counters, 4 Gates and 4 outputs. The module has two internal clocks, 1 kHz ad 32 kHz that can be used to feed the counters. The outputs can be configured to toggle each time an adjustable terminal count is reached, allowing output pulse generation of up to 16 kHz (32 kHz internal clock divided by 2). The FP-CTR-50X modules have the advantage over the above modules in that the counters can be cascaded and set for differing terminal counts allowing multiple pre-scalers which may allow better resolution of generated frequncies.
    For measuring the output frequency generated by the gauge, the only current modules that will allow you to directly measure the output are the FP-CTR-50X modules. However, as they were primarily designed for event counting, it takes a little bit of extra programming to allow them to be used to measure frequencies. The effectiveness of frequency measurement on the FP-CTR-50X modules varies on the sampling period, the programming and the network module. There are several KnowledgeBase and Developer Exchange articles talking about how to use a FP-CTR-50X module to measure frequencies, so I will not duplicate the information in this post.
    Regards,
    Aaron

  • How to connect my dv camera to my mac and make movies?

    i just got my new macbook and dont no how to connect my dv camera to it...i got a software disc for my dv cam but it doesnt run automatically and i dont no how to install it....it is usb connection...do i need fire wire?...help me plz help...getting frustrated ...is the any sites that help with macs ..cuz im use to pc...dont no anything bout macs...thanks so much!!!

    Hello,
    It would be helpful to know what camera you are trying to connect and what exactly you are trying to do with it.
    In regards to firewire vs usb i only really have experience with firewire so i can not help you there unfortunatetly.
    If you are just trying to get the video footage off the camera you could try and open up iMovie from within you applications folder. Attached the camera, my iMac is currently in for repairs so i can quite tell you the whole thing but look for options of attached cameras.
    The reason why the cd doesnt autorun is that it is probably window software, does the disk specifically state Mac OS X?. If you double click on the cd it should open up in a window. Have a look around for any Mac folders or any obvious applications that you can run.

  • How to connect fp mic on msi G31M3-L V2 ?

    Hi ,Does anybody knows how to connect front panel mic jack or audio jack, the manual for MSI G31M3-L V2 doesnt says anything about JFP2.  I have already connected the three jacks with 5.1 home theater speaker system, and so i do not have free mic jack for my web cam, or one channel for 5.1 speaker system
    Free wires are mic_bias, mic_in, Gnd, Spk .. return L, return R, Thanks!

    Hi,
    the JFP1 and JFP2 connectors cannot be used for connecting the case's front panel audio ports.  You have to use the JAUD1 Connector.
    The Pinouts of the JAUD1 Connector are described in the user manual.  If you want a more detailed pin definition, check the Intel Front Panel I/O Connectivity Design Guide, page 22 (Intel® High Definition Audio):
    http://www.formfactors.org/developer%5Cspecs%5CA2928604.pdf
    Quote
    Free wires are mic_bias, mic_in, Gnd, Spk .. return L, return R
    Looks like your case's wiring is not fully compatible with the Intel HD Audio/Azalia Connectivity Standard, so don't be surprised that the JAUD1 Connector has more pins than you have wires. Connecting the microphone should be pretty easy anyway.  There are basically just two possible positions. The GND-Wire goes to the GND-Pin (2).  The two MIC-Wires go to the two MIC-Pins (1 and 3).  Try both positions, one of them should work.  When you test the microphone and it does not work, make sure you try the following things:
    1) Disable the Front Panel Audio Jack Detection in the Realtek HD Audio Manager
    2) Run the microphone calibration utility (called miccal.exe (it is in the Realtek Installation folder))
    3) Enable Microphone Boost

  • How to connect your Verizon 4510L to your home Wifi Network

    Hello there! 
    I recently set my mom up with one of these guys and ran into the problem of having more than 5 devices that need to connect to the Internet and/or each other.  I couldn't find any resources online that described how this can be done, or even if it is allowed.  So I figured I would finally contribute to the Online community and share my success with all of you!  As a disclaimer:  I am not Verizon/Linksys/etc technical support and provide the following information freely and without warranty. The instructions below are specific to my configuration, and while yours WILL work using the same general principles, the exact steps required to make it work may be different.  In other words, YMMV   Good Luck!
    First off, my environment doesn't equal your environment, and what I will describe below is one of the more complex scenarios.  You can disregard any extra routers if you only use 1.  Here is the environment I had to work with:
    A large 2 story home with a basement area as well.  The wifi signal does not travel well from floor to floor, so we had to have 1 wireless router per floor.  We DID have the luxury of using home wiring to to phsycally connect both routers together, if you don't have that luxury, I might suggest using a "Powerline" adapter to help wire your two routers together.  Some devices required wired connectivity on both floors as well, so wireless repeaters would not be sufficient.  We purchased 2 routers to accomplish the following task.  Brand of router should not matter.  If your home is a 1 story, or have adequate coverage from a single router, you can disregard the 2nd (SLAVE) router instructions (4-5) below. 
    First off, what you will need to make this work:
    One or more Wireless Routers (we used Linksys Wireless-G or Wireless-N 4-port routers)
         *EDIT: You need only 1 router, and if an additional access point is needed, purchase an Access Point or repeater with wired/wireless capabilities*
    Verizon JetPack 4510L MiFI device
    One Wireless Bridge (we used Linksys WET610 1-port Entertainment Bridge)
    Next, lets setup the wireless/wired home network to connect all devices with one another:
    1) Choose one router to be the MASTER, and the other to simply be your SLAVE/repeater for the other floor.
    2) Before physically connecting the MASTER and SLAVE routers together perform the following:
    3) On the MASTER router:
         - Change the router IP address to 192.168.10.1
              - The 10 is random, and you can make it whatever you want (as long as it isn't "1"), just remember it and keep it consistent.
         - Configure DHCP (usually on by default) but change the subnet from 192.168.1.x to 192.168.10.x.
              - The reason for changing the DHCP subnet is because your 4510L device uses DHCP as well, and it is configured to use 192.168.1.x
                   - They can't be the same or it won't work
         - Configure your wireless SSID and security (preferably WPA2)
    4) On the SLAVE router *EDIT: or repeater or access point*:
         - Change the router IP address something outside the DHCP range 1-100, but on the same subnet as the master router (ie. 192.168.10.101)
         - Disable DHCP on the router, as all address assignment is going to come from the MASTER
              - You may want to tape over the WAN port as well to remind yourself that it will not be used
         - Configure your wireless SSID and security to match that of the MASTER router
              - This will make it easier for your clients to use both access points and move throughout your home without having an issue
    5) Connect both routers together with Ethernet (either direct cable, home ethernet wiring, powerline adapter, etc)
         - You want to connect both routers using one of the built in network ports (1-4)
    There, now you have both floors of your home setup to accept either wired or wireless clients who will be assigned 192.168.10.x IP addresses.  However, they will not have access to the Internet...yet!
    The MiFi device by default has a long SSID, and the password to connect is on a label on the bottom of it.  Power it on in the room next to a computer (probably a windows computer), and view your wireless networks to connect to from the desktop.  You will probably see the wireless network name you just created on the routers and the "Verizon ..." MiFi name.  Simply make a note of the EXACT SSID being broadcasted by the MiFi device.  If you changed the default WiFi password before attempting this, make a note of that new password as well.
    Next, lets use the key to making this thing work...the Wireless Bridge.  The wireless bridge first needs to be physically connected to either router on ports 1-4.  The bridge will get an IP address assigned to it and you will need to find that address out so that you can access it from a web browser.  The easiest way to find that out is to connect to the MASTER routers web-based administration page (http://192.168.10.1), click on Status menu->Local network, and then click the "DHCP client table" button.  You will find the bridge name and IP address there.  Keep that administration window open. 
    Next, connect your web browser with the bridges IP address so that you can manually configure it to access the MiFi device.  The bridge will ask if you want auto or manual configuration, and you want manual.  When prompted for the SSID and security type, type in the SSID of the MiFi device EXACTLY as you wrote it down.  Next, under security, change the type to WPA2-Personal, and type in the password to access the Verizon MiFi device (either on the sticker under it, or whatever you previously changed it to).  Once you complete that setup, the status page should tell you that the bridge is connected and have an IP address (192.168.1.x) assigned to it.  Once that is done, you can close your web browser to the wireless bridge.
    Disconnect the bridge from the routers network ports and now connect it to the MASTER routers WAN port.
    Next, return to the MASTER routers web-based administration page (http://192.168.10.1), click on the Status menu and view the WAN IP information.  If everything was connected correctly:
    MiFi wirelessly -> Wireless Bridge wired -> Master router WAN port (then optionally wired from port 1) -> Slave router port 1 
    Then, you should see the WAN IP address of 192.168.1.x assigned to your MASTER router.
    If you do see that, then congratulations, open a web browser from any of your wired/wireless clients connected to your home network, and you should be able to access the Internet!  You will be able to support as many clients as your Routers will allow you to directly connect (104 if I am not mistaken)...Yay!
    I hope my experience helps you all in achieving the same results!
    Troubleshooting:
    If you do not see a WAN IP address, you might want to power off the MASTER router, the Wireless Bridge and MiFi device.  Then in the following order power up each one, one at a time, giving the unit time to fully configure and connect to their respective networks:  First MiFi, then the Bridge, then the MASTER router.
    After the Master router is fully powered on, test a webpage again from any client connected to your new home network and it should work. 
    If it _still_ doesn't work, I would recheck the steps listed above to see if anything was missed.  Particularly, making sure you changed the router IP and DHCP subnet of the MASTER router to something other than 192.168.1.x.
    Good Luck!

    Q: "In your example of using the wireless repeater, assuming it is on a different floor than the main router, how are the clients connecting to the network on that floor?"
    A: However you like, both devices support thier own cabled and wireless connections. 
    I assume you are only interested in wireless client connections since cable connections are straight forward.  Depending on how you setup the SSID's clients will connect to either the router or the repeater and they should be overlapping eachother a little. 
    Same SSID:
    If the router and repeater share the same SSID then the client will decide on thier own which one to connect to.  The client will roam between the two on its own criteria as the user moves around the enviornment.  Roaming and transitioning will happen seamlessly for you and you should not notice a break in your connection unless you have something really specific connected to the internet.  In most cases you cannot influence this behavior on your devices as they all have thier own criteria for choosing one SSID over another with a similar name. 
    Different SSID:
    If the SSID's are different then the user must manually connect to one network or the other.  Roaming will not be seamless and as the user walks out of range the client will be disconnected from the device.  This provides the best reliability and control but requires more user intervention.
    Q:"So correct me if I am wrong, but in your setup, there would be 2 wifi network, and the soft limitation of 4 devices on the repeater floor.
    A. Nope, the repeater will support as many devices as it can, depending on the model you purchase.  Even though the repeater needs a wireless connection with the MiFi, the Repeater rebroadcasts and acts as its own independant wireless router.  The MiFi could care lesss how many people are connected to the Repeater, all it see's is one connection.  If you removed the MiFi from the repeater then the repeater will continue to broadcast its own SSID.  Clients connected to the repeater will not be disconnected from the repeater SSID but they would not recieve internet until the MiFi returns.
    With the repeater enviornment you could have 10 people on the repeater, 10 people on the wireless router and 4 people directly conneted to the MiFi all at the same time.  I dont suggest it running like that for performance reasons but hopefully you get the idea.

  • Speaker wire cut, want to fix myself - which is the best way to connect the wires back together?

    Some how, one of the wires going to my left-center speaker got cut in half. Not sure if it was the vacuum or my dog........anyway, I want to fix it, so - would you recommend butt-connectors, soldering the wire back together, or some other method?
    Also, does it matter which wires connect back together? The wires are identicle (smooth covering) and both wires appear to be copper........ I was just wondering if anything would happen if you were to connect the wires back together, but backwards.
    Please let me know so I can get this fixed quickly.
    Thanks,
    -Officer_Dufus

    http://www.eminent-tech.com/music/multimediatest.html
    Go to the above link and listen to the Pink Noise Monaural and the Pink Noise with Reversed Polarity tests. When a speaker is of opposite polarity, that means that the wires are connected backwards (positive to black and negative to red). If this happens, when your speaker should move out it'll move in and vice versa. This makes the sound seem to come from everywhere but the speaker. Play the tests to see what I mean.
    If you solder your wires back together and notice this problem with your speaker, just reverse the connection to the speaker itself and you should be okay.

  • SB Live! 5.1 how to connect with my HiFi-Syst

    Hey,
    I want to connect my Soundblaster Li've! 5. with my amplifier.
    The amplifier is connected to two stereo loudspeaker.
    It?s kind of an old amplifier with DIN connections, so I have to use a Cinch to DIN adapter. But the adapter has two cinch cables, one for the left signal and one for the right. (http://www.starcom-edv.de/shop/pics/AudioAdapter_DIN_St-Cinch_St_s.jpg)
    I really don?t know how to connect it to my Soundcard. (http://img55.imageshack.us/img55/4602/picture0032vf.jpg)
    Aditional I want to connect my PC with my TV, so I need one Line-Out for my TV.
    Please hep me!

    Is tha 2 x RCA to DIN intend for playback or recording input (or are those DIN connector wires equal for Input and output purposes)?
    You need another cable: 2 x RCA to 3.5mm (/8") stereo mini jack plug to get the SB and Hi-Fi connected. Connect the PC just the way, you would connect a casette deck into a stereo system (if you know how to do this) so, you get both playback and recording capability or just connect the output from PC into Hi-Fi system).
    For that TV connection, do you mean you would use the TV as output device (PC --> TV)?
    If that's what you're after, then just use the another line-out from your Li've card for that connection, otherwise (TV -> PC) connect the TV output into Li've Line-In or into some of those, possible available, internal analog inputs (AUX_IN, CD-AUDIO, etc.).
    jutapaMessage Edited by jutapa on 0-2-20072:52 PM

  • How to connect serial or usb or parallel port

    telll me how to connect our application with serial port and how to detect weather the usb is connected or not

    Without spending too much time looking at the documentation, from using previous similar transducers, the primary two connections you are concerned with are the "+output" and the "ground". The other connections are also important, but aren't directly the concern of the Data Acquisition Hardware. The "+output" produces a variable dc voltage relative to the "ground" input that is a product of the applied pressure, with the maximum output voltage, as shown on the chart, indicating maximum pressure. So you would connect those, through a suitably shielded, moderately short length, wire to the "DAQ" analog input. You will also have to supply the transducer with it's required + supply voltage at pin 2 "vs", and the power supplie's negative to the ground. There are whitepapers on this NI site about connecting to data acquistion hardware and I strongly suggest reading them a couple of times as well as anything online you can find. Writing the DAQ software, in LabVIEW or whatever, can be trivial compared to making the correct, low noise, connections to sensor hardware.
    I didn't read how the offset adjustment signal is produced or its use, figure it is designed to allow you to "tune out" some signal offset caused by differences between sensors of a type (they will not all have the same "fullscale output" as indicated in the chart).
    As you proceed, please keep asking.
    Putnam
    Certified LabVIEW Developer
    Senior Test Engineer
    Currently using LV 6.1-LabVIEW 2012, RT8.5
    LabVIEW Champion

  • How do I wire my SCXI-1520 quarter bridge completion in parallel?

    I have a quarter bridge set-up using the qtr completion in the system. In parallel with my strain gauge I'd like to have another resistor. I'm using the strain gauge as a dummy, and am interested in the current traveling in a parallel path through a film resistor as I change the film thickness. However, I'm not convinced that what I've wired has the strain gauge and other resistor in parallel. I have the strain gauge wired to: P+, QTR, S+ and the other gauge wired to QTR, S+.
    Thanks!

    Greetings,
    You will want to wire the other gauge between P+ and QTR or between P+ and S+. To make this decision, you will need to determine how the lead resistances will affect your measurement. Page 2-4 of the SCXI-1520 User Manual (link available below) contains the wiring diagram for Quarter-Bridge Configuration I.
    SCXI-1520 User Manual
    Good luck with your application.
    Spencer S.

  • Measure 16 strain gages simultaneously with SBRIO 9626 + 2 NI 9235 modules

    Hello everyone,
    I am currently working on a deformation measuring application. I need to measure 16 strain gages simultaneously, using a SBRIO 9626 + NI 9693 (two C Slots) + 2 NI 9235. I started checking out the example included for measuring strain with shunt calibration on 4 gages with one 9235 module through the FPGA, modded it to read 8 gages without any issue.
    The problem comes up when i try to use two modules for reading 16 gages. I modified the getting started example in serveral ways without any success, only can read continuously the 9235 on slot 1, and apparently, i only get an initial reading on the other 9235, and then nothing more. Below is attached the code (note that i am no expert on FPGAs, and i've been trying different configurations, and at some point, not precisely knowing what i was doing ), any advice on how to put both 9235 to work simultaneously will be appreciated.
    thank you.

    Hi s.jimenez
    First of all, thank you for your prompt answer. Regarding your suggestion, i hadn't checked the Basic Synchronizing NI 923x example, my bad . After reading and understanding it a little, i tried to adapt it to my situation, sadly with no success. Some of the mistakes that i corrected were:
    one FIFO for acquiring data from both modules
    used single I/O nodes for start/stop and acquiring data from both modules
    changed the clock settings of the modules to: slot1 module providing clock signal for the slot2 module
    the only big difference that i notice between the example is that i added the shunt calibration procedure to the sequence, do you think this is the problem?
    summing up, i havent been able to obtain continuous readings from the second module, any additional help could be greatly appreciated.
    best regards.
     