Keithley 2000 DMM AC Read Overflow

I’ve been having problems controlling a Keithley 2000 DMM during AC volts measurement using IVI. When reading AC Volts the DMM will intermittently sense an ‘overflow’ and return the value NaN. I’m using IVI and TestStand 3.0 but get the same problem with IVI and LabWindows 7.0.
The set-up is as follows:
Device: Keithley 2000 DMM, GPIB device number 6
Device interface: GPIB, DMM set-up in MAX as GIPB0::6::INSTR
IVI Driver version: 3.02, downloaded from the NI web site
IVI Compliance Package version: 2.1
NI-488.2 version: 1.70
NI-DAQ version: 7.1.0f1
MAX version: 3.0.2.3005
For test purposes I’ve connected the DMM front panel input high and low and set the inputs mode on the Keithley to front, this should mean that I get a very low Voltage reading. I’ve then run the LabWindows/CVI IVI control sample code provided with the IVI driver. At the initialisation panel I’ve entered the logical name. At the configuration panel I’ve selected Volt AC measurement but left everything else as default. The second configuration panel I’ve left as default. At the Read panel I’ve left all setting as default. Pressing the read button up to about 20 times with about a second gap in between each press will eventually lead to an ‘overflow’ reading.
Any help/explantion will be appreciated
Steve

Here are the results of my investigation into this problem, it’s two fold:
1. Keithley have informed me that their engineers in the US have just seen the problem I’ve been seeing and are working on a firmware update. Currently here is no date for the release of the updated firmware but I’ll be informed when this it’s ready.
Note: I’ve found that the read AC Volts overflow problem only occurs when the read trigger delay is set to zero seconds.
2. The TestStand DMM IVI configuration step always sends a trigger delay of zero seconds regardless of the trigger delay set.
In my TestStand sequence the DMM IVI configuration sets the read trigger delay to 3 seconds but as noted above this is sent a zero seconds thus invoking the Keithley DMM AC Volts read overflow problem.
See attached VISA call log captured by NI Spy. The first 22 call are made when the DMM IVI configuration step is called form TestStand the next 22 calls are made by the IVI DMM sample code configuration provided by National Instruments when DMM configuration is run thorough LabWindows. Note call 12 (IVI DMM TestStand) has the trigger delay set to zero seconds (should be 3 seconds) and call 34 (IVI DMM LabWindows) has a delay of three seconds.
I’ve found a workaround as follows:
After each call to the TestStand DMM IVI configuration step an additional step that calls the TestStand IVI Tools step can be added. The IVI Tools step can be set-up to specifically configure the DMM Trigger Delay to 3 seconds. I’ve proven that this individual call does in fact send the correct trigger delay command to the DMM thus alleviating the AC Volts read overflow problem.
Can NI confirm that this problem does exist with the DMM IVI step and when they plan to fix it. I’ve attached the TestStand sequence file that I’ve been using to investigate the problem.
Thanks
Steve
Attachments:
DMM Overflow.zip ‏9 KB

Similar Messages

  • Obtaining readings from Keithley 2000 DMM

    Hello,
    I have a Keithley 2000 DMM with RS-232 communication, a PC running LV 7.1. I have established communctions between the two using RS-232 and I am able to do *IDN? to generate a query of the instrument. However, I would like to read in the readings (DC voltage) from the meter at a button press. What command string would I send to the unit to get this data? Any help is greatly appreciated!
    Thanks!
    Mike

    Either read the manual on the instrument or download the drivers from
    Keithley 2000 Drivers

  • Expression Error (#:-1074001424) with Keithley 2000 DMM

    I wrote a program to that uses the Keithley DMM to check voltages on multiple channels, wait a period of time, and then check again. This cycle continues for the time specified.  I made this program on one laptop and it worked fine.  When I moved it over to a different laptop and a different Keithley DMM, then I started to see this issue.
    After stepping thru the program, It seems that the error comes from the "Configure Scan.vi".  The exact error message is:
    <ERR> ke2000 Error Query;
    instrument reports:
    -260,"Expression error"
    and the error code is: -1074001424.
    I searched for these myself and couldnt find anything helpful.  i changed a few things around and it seems that the "scan mode" parameter seems to be the problem.  It needs to be set to "internal" to read the channels on the additional scan card, but the program will return an error if it is not set to "none"... and when its set to "none" nothing happens.
    Please help me out! Thanks

    Nvm.... fixed it.  For the Configure Scan.vi, in some cases, one of the parameters, "Stop Channel", would be set to 1, when the starting channel was, by default, set to 1. Apparently this caused a conflict of some kind but I changed my program around so this doesnt happen anymore.

  • Keithley 2000 current measurement

    Hi All,
    I am using keithley 2000 DMM to measure current. I have configured the DMM under IVI Drivers in MAX. I am using the example program in CVI to configure and read the current drawn. When I read the current drawn from the DMM the current value displayed is incorrect. I have used the auto adjust to zero function also. Even on multiple readings with delay, the value keeps fluctuating and is no where near to the expected value. Can anyone suggest why I am not able to read the correct value? Where can I find the configuration parameters for this DMM in MAX? How do I know if I have configured the DMM correctly in MAX or not?
    Any pointers would be helpful.
    Thanks in Advance,
    Pradeep

    Hi Pradeep,
    Try downloading the drivers for the Keithley DMM from http://sine.ni.com/apps/utf8/niid_web_display.model_page?p_model_id=248
    Nxt question is as to how you are connecting the DMM to your PC (serial, gpib??)
    If you have connected over GPIB, verify that it is showing in MAX under devices and interfaces.
    You can then right click on the instrument and select communicate with instrument. Use simple commands that should be available in the manual to read data from 
    the DMM. Once you have verified that communication is established, try using the example programs in CVI.
    Regards,
    Denver
    AE, NI India
    Message Edited by denver_ on 11-24-2008 07:46 PM

  • Keithley 2000 starting channel

    Hello all,
    I am using LabVIEW 8.0 connected to Keithley 2000 DMM with 10 channels scanner card.
    For configuration I am using the "configure scan.vi" from the appropriate driver.
    The "starting channel" option is capable of having numbers from 1 to 10 (for internal card).
    I choose to start from channel 1, but the scan opearation keeps starting from channel 2!
    I tried setting the starting channel to be 0 but then I get an error message saying "data out of range".
    Does anyone have an idea?
    thank you,
    Noam Tal
    Ben-Gurion University
    Israel

    There is no Test.vi there.
    Anyway I tried running the scanner.vi but it does too many errors and I couldn't find a clue to my problem.
    My scan8.1.vi is attached. It is very simple.
    If you could take a look and try figure it out i would be grateful.
    Attachments:
    scan8.1.vi ‏31 KB

  • Keithley DMM IVI Read Timeout

    We are controlling a Keithley 2000 series DMM using the IVI Dmm Step in TestStand version 3.0.
    The logical name has been set-up using MAX version 3.0.2, the logical name references the ke2000 driver session. When run under TestStand the DMM configures correctly and will take a first reading correctly but subsequent reads fail with a timeout error. We’ve tried using the LabWindows example code found in directory \program files\ivi\drivers\ividmm\examples\cvi, this exhibits the same problem found use TestStand. Is this a MAX set-up problem, any help will be appreciated.
    Below is listed the software/environment we are using with the DMM:
    Software Version
    IVI Compliance Package 2.1
    NI VISA 3.0.1
    NI-488.2 1.7
    Instrume
    nt Driver (LabWindows/CVI 6.0) 3.02
    Measurement and Automation Explorer 3.0.2.3005
    TestStand 3.0
    CVI Runtime 7.0.0.393
    LabWindows/CVI 7.0
    WindowsXP Pro

    Hello Steve Penny �
    What is the exact error code you are receiving? You can find a list of common GPIB error codes and solutions here.
    Make sure you are using the correct driver for your GPIB board. Windows XP also uses NI 488.2 version 2.1. Click here for a list of current drivers by platform.
    If these suggestions do not solve your problem, please post the error code so the matter can be further investigated. Have a great day!
    Becky B.
    Applications Engineer
    National Instruments
    Becky Linton
    National Instruments
    Field Engineer
    Office: 734-464-2463
    Cell: 248-709-2822
    Email: [email protected]

  • Why does the integratio​n time setting of Keithley 2000 (number of PLC) changes to medium (1PLC) from fast (10 PLC) and slow (0.1 PLC) settings, when I tried to read from Keithley?

    Dear All,
    I am trying to change the NPLC (the integration time) of Keithley 2000 and take voltage readings. Whenever I change the settings to fast or slow and then run the VI, the NPLC changes to medium. Could you kindly tell me what is causing this? Thank you and I look forward to receiving your replies.
    Regards

    For the NPLC setting, the higher the number, the longer it takes to get a reading.  So 10 PLC is much slower than 0.1 PLC. 
    See if you can set the NPLC setting from the front panel.  Then try changing the vi to set NPLC only, then read it back to see if it matches your setting.  Sometimes changing ranges will automatically set the NPLC depending on the range chosen.  So you have to send the range command before sending the NPLC command.  This may be what is happening.
    - tbob
    Inventor of the WORM Global

  • Keithley 2000 voltage measurement

    Hi,
    I want to write a code to display the voltage measured from a keithley 2000 meter using RS232 serial port  and display it on a   graph using Labview 8.5 on Windows Vista based Laptop
    . I guess the first step is to download the driver for keithley2000 http://sine.ni.com/apps/utf8/niid_web_display.model_page?p_model_id=248. I want to go step by step. Please give me the main key steps on how to proceed on this problem and i will then work in detail based on your directions.
    How do i know that laptop  is recognizing the meter and RS232 connection is sound? What are the next key steps i need to work on? Please give the directions.
    Thanks

    For fast reading, it is important not to use the MEAS? command, but READ followed by FETCH? (if I remember correctly). Normally these DMMs operate at 1 PLC (power line cycle -> 50/60 Hz), so you need to adjust the NPLC (number of power line cycles).
    There is to options to work with the Keithley DMMs inside LabVIEW. You could use the LabVIEW drivers for the Keithley 2000, those can be found on ni.com and on the keithey homepage. Propably they only implement GPIB, the you need to make some modifications on them for RS232. The other option is to send the commands as found in the DMM manual via VISA.
    Felix
    www.aescusoft.de
    My latest community nugget on producer/consumer design
    My current blog: A journey through uml

  • Keithley 2000 and 7002 with trigger link

    Hello!
    I am writing an application for the 2000 DMM and a 7002 switch system. The instruments are set up with trigger link. What I want to do is simply to load a scan list to the 7002, scan and read back the results from the 2000. Here is what I have so far in terms of SCPI (I have looked at the Keithley LV drivers but I think they complicate stuff + there are no high level functions for trigger link setup included):
    DMM *RST
    SW *RST
    SW Open
    SW Scan (@2!1:2!40)
    SW :TRIGgerOURce TLINk
    SW :TRIGger:TCONfigureROTocol ASYNchronous
    SW :TRIGger:COUNt 40
    SW :ARM:LAYer2:COUNt 1
    DMM :TRIGger:COUNt 40
    DMM :TRIGgerOURce EXTernal
    DMM :INIT:CONT ON
    SW :ARM:LAY1OUR IMM;COUN 1
    SW :ARM:LAY2OUR IMM
    SW :INIT
    DMM :TRACEATA?
    Please advise,
    Stefan

    hi morgue!
    I am doing a program like your, but with Keitlehy 2001.
    To make the syncronization you wrote just these lines? I am doig this in labview, with the programs found in the drivers given by keithley but is not working properly. When I put the triger link, it scans and show just the first measure.
    Do you still have this programm?
    thank you 

  • Looking for examples/help on using Ethernet with Keithley 2701 DMM

    All of the example code utilizing their IVI driver is based on GPIB or RS232. I specifically bought the 2701 to utilize the Ethernet port, but cannot find CVI examples on how to do this. Can anyone point me to an ethernet example using their driver and the Ethernet port? Or perhaps post a non-Keithley ehternet driver and examples on its use?

    Hi Paul,
    What you have been told is exactly what you need to do.
    Here is a document in NI web that explains about it:
    How Can I Communicate With a Keithley 2701 DMM?
    There is also an Instrument driver that NI built for these DMM:
    LabVIEW Certified Plug and Play Instrument Driver
    To find this documents you need to go to NI.com and at the top on the search box enter 2701. That will give you all the documents related to this.
    Thank you
    Ricar
    do S.
    National Instruments.

  • Acquiring a measure every x seconds from a Keithley 2000 multimeter

    Hi everybody,
    I'm working on a program on labview which should acquire a measure every x seconds from a Keithley2000 multimeter.
    I managed to acquire a single reading using the "Single Read VI" from Keithley 2000 Instrument Drivers.
    To acquire a reading every x seconds, i have tried 2 possibilities without havinng the desired result:
    - First i have tried the "Multi Read VI" (also from Keithley Instrument Drivers), but the problem with this VI is that i could controle the desired number of readings without being able to controle their frequecy ( i can't choose the elapsed time between 2 successive readings) 
    - The second possibiltie I've tried is to use the "Single Read VI" inside a "Timed loop" for which i could choose the desired number and frequency of iterations, but the problem with this methode is that
     at the end of the program i still get only a single reading!
    Can sombody please help me with that?
    THANK YOU in advance.

    Hello,
    As dennis said, I think you made a mistake in your code.
    Could you post your code for us to see what you're doing wrong?
    Regards,
    Olivier L. | Certified LabVIEW Developer

  • Measure with Keithley 2000 over RS232

    Hey,
    ich have a problem with my Keithley 2000 by measurements over RS232.
    My programm is running, its a simple serial read and write VI. I send him an command to measure and nothing is happening. I got nothing back....
    I add the VI and the commands, maybe somebody knows whats wrong with my VI!
    Thanks a lot
    Greetings
    Ralf
    Attachments:
    RS232 Serial Write and Read.vi ‏22 KB
    Unbenannt.JPG ‏7 KB

    I have no time to look into your code, but most likley you miss the termination char. On a string constant you can switch to \-display and then use \n or \n\r depending on the settings. You can also look into the LV drivers for the keithley 2000, they are available on ni.com and/or on keithley.com
    Felix 
    www.aescusoft.de
    My latest community nugget on producer/consumer design
    My current blog: A journey through uml

  • Communication GPIB avec un keithley 2000

    Bonjour à tous,
    j'essai d'utiliser mon multimètre keithley 2000 en communication GPIB sur mon labview 8.6. J'ai télécharger un vi (pièce jointe) qui est un exemple. J'arrive à initialiser sans pb... mon pb se situe au niveau de la fonction lecture. Le sous vi atteint le time out de 25s et me renvoie l'erreur suivante :
    L'erreur -1073807339 s'est produite à : VISA Wait On Event dans Wait for RQS.VI:1->_ke2000 Wait for SRQ.vi->ke2000 Data Read Single.vi->Keithley channel test.vi
    Raisons possibles :
    VISA:  (Hex 0xBFFF0015) Le délai d'attente (timeout) a expiré avant que l'opération ne soit achevée.
    Je ne comprends pas du tout ce qui se passe, quelqu'un aurait-il une idée !!! Merci à tous
    Attachments:
    Keithley channel test.vi ‏48 KB

    duplicate post

  • Keithley 2000: Strange timeouts

    Halloechen!
    I have to read multiple values from a Keithley 2000 (DC volts data),
    and even that has to be done multiple times. So I did the
    following:
    1. ke2000 Initialize.vi
    2. ke2000 Configure DC volts.vi (no parameters, only "DC Volts"
    selected from the pull down menue)
    3. In a loop:
    a) ke2000 Configure Trigger.vi with a proper calculation of the
    timeout (taken from a Keithley example code)
    b) ke2000 Data Read Multiple.vi (with the correct number of
    readouts)
    c) ke2000 Reset.vi
    4. ke2000 Close.vi
    Unfortunately, with this programming I get timeouts regularly.
    Well, not always but every other measurement say. If I increase the
    timeout time, it simply takes longer, but the timeout is inevitable.
    When it
    happens, the Keithley seems to be unable to leave 3 b).
    Then I created a WHILE loop that repeats the measurement if there
    was a timeout. But the loop is repeated until I stop the program.
    However: If I take the initialisation vi *within* the loop, it
    works! I still get timeouts, but now it cleanly repeats the
    measurement and proceeds.
    But this is neither elegant nor fast. Apparently some delays are
    missing. Does anybody know where to put them? Where is the
    critical phase for the instrument?
    Or is it something different?
    Thank you!
    Tschoe,
    Torsten.
    Torsten Bronger, aquisgrana, europa vetus

    I find this rather interesting. I ahve some older drivers for the KE2000 and they work without problem. But when I seen your question it rasised an interest because you said DC volts from the pull down. I went and downloaded the drivers from NI and installed them wrote just a simple program from them and sure enough the problem is there just like you said it was. I also found that if you do not use the initialize vi it works fine without the timeouts. Seems that the initialize vi is putting the instrument in a funny mode. Anybody have any Idea of wht this is happening.Will do some investigating and get back to you.
    Joe
    Joe.
    "NOTHING IS EVER EASY"

  • Keithley 2000

    Hello,
    I am having a problem with connecting Keithley 2000 with Labview. I downloaded the driver from Ni , however when I run it, the error syntax occur Error +800 in the Keithley 2000. It does not do any measurement. Would you like to help me to deal with this problem? 
    Thank you very much.!

    Hi stranger1!
    I have some steps that might help you:
    First, make sure you have the 488.2 driver installed for GPIB.
    http://joule.ni.com/nidu/cds/view/p/id/3786/lang/en
    Second, make sure the device shows up in NI MAX and you are able to communicate with it there.
    http://www.ni.com/gettingstarted/setuphardware/instrumentcontrol/gpibconnect.htm
    Are you using the IVI driver or the Plug and Play driver?
    IVI: http://www.ni.com/white-paper/3377/en/
    Plug and Play: http://zone.ni.com/wv/app/doc/p/id/wv-1426
    You also might find the information in this forum post useful:
    http://forums.ni.com/t5/Digital-Multimeters-DMMs-and/How-do-I-interface-with-a-Keithley-2000-multime...
    If those documents are not able to solve your problem, we might need a little bit more information. IVI or Plug and Play? You say you try to run the driver—do you mean a particular VI or example included in the driver? The error you are getting is a SCPI error (http://forum.keithley.com/phpBB3/viewtopic.php?f=13&t=70)—are you sending a particular command to the device? 
    Regards,
    Kelsey Johnson
    Applications Engineering
    National Instruments

Maybe you are looking for

  • Need help to generate "where current of" from ODBC

    Hi I am trying to develop a small stand-alone program which uses Oracle Drivers. I have installed PERL and came up with a code (structure) like below But I dont know how to use the current cursor reference with a statement This is what I came up with

  • T500 Problem with BIOS - Win7

    I have a T500 2081CTO. I took a stab this weekend putting Windows 7 on it, but I haven't gotten very far. Aside from the fact that I am only finding BETA versions of drivers for it, my biggest problem is that the latest BIOS which purportedly adds Wi

  • ITunes won't install! Help please! :(

    Okay, so I tried to install iTunes today so I could put some music on my iPod nano. When I tried to install iTunes it said "The iTunes + Quicktime installer requires Windows XP Service pack 2 or later. To install the latest Windows XP Service Pack, u

  • Upgrade from Windows Vista 32 bit to Windows 7 32 bit - L500D-11P

    Hello, i bought since few years the notebook L500D-11P. The OS Windows Vista 32 bit was preinstalled. So i found out that i can upgrade my OS to Windows 7 32 bit. But i need a Windows 7 Upgrade Disc. I didnt find it on my Notebook package. Where can

  • Internal error message from cybertan

    When I startup lightroom 5 I get an internal error message from cybertan and it renders lr useless.