Ni PCI 6013 syncroniza​tion

I wonder if you can help me, i want to give the PCI 6013 I/O card a reset signal every 1 second. Is it possible in anyway to do this. I aim to syncronize a few of the cards i am going to use in a measurement campiagn with the help of a GPS SYSTEM. I would appreciate it tremendously if you could give me any suggestions on how i can accomplish this, with the use of your PCI 6013 I/O/ card. I am using Labview 7 to do the measurements.
Kind regards

Hey Pieter,
Another suggestion. If you are wanting to synchronize the analog inputs with the gps digital signal, an easy way to accomplish this would be to use the gps digital signal as an external clock for the analog input operations.
If you use a digital signal as an external clock, you will acquire one reading from the channel(s) in the scan list everytime the digital signal has a transition. This would create an easy way for the gps signal to set when you are actually reading data. Take a look at the LabVIEW shipping example 'Cont Acq&Graph Voltage-Ext Clk VI' that illustrates how to continuously acquire from analog input channel(s) using an external clock.
I hope this helps. Good luck with your application.
Regards,
Todd

Similar Messages

  • Clearing configuration FIFO on PCI-6013

    I am writing a driver for a PCI-6013 for a legacy DOS based system. Everything I need is working except that I cannot clear the configuration FIFO so I can change the channel setup during use. According to the E-Series RLPM writing a 1 to the Write_Strobe_0 register (windowed write) will do this. As far as I can tell this has no effect at all. Is it possible that there is something in the STC setup that is preventing clearing the config FIFO or might this be something unique to the 6013?

    John,
    You may want to consider posting this question to the Driver Development Kit (DDK) discussion forum (link available below). This forum deals with questions concerning register-level programming (RLP), alternative operating systems, etc.
    Driver Development Kit (DDK) Discussion Forum
    Good luck with your application.
    Spencer S.

  • Trouble in the configuration of a PCI-6013

    Hi:
    I´m triying to get the MAX to recognize my PCI-6013 but i haven´t been able to get the DAQ board configured. I installed the package included with the board that contains LabVIEW 6.01 evaluation software and NI-DAQ 6.9.Of7 repeatedly, and then installed the board in a pci slot for several times. WindowsXP recognizes new hardware but when i try to install it no driver is found. I tried it whith the NI-DAQ 7.0 and the board got configured, but when i run a program i developed with the same board in another pc (and it worked perfectly) y get a 10001 error, something about wrong string orther or something like that. I thouth that maybe it had to do with the wrong channel definition (i did it as virtual channels, for an analog
    input and 8 analog I/O). Well, i would like to get the sistem going with the NI-DAQ 6.9.Of7 since it worked in the past in another Pc ,or some hint in why i get that error working with the NI-DAQ 7.0. I would really apreciate any help, and thank you in advance, it´s nice to be able to rely in someone else with more knowledge and that is willing to help the others. Thank you again.

    I am not quite sure why the disk which shipped with your PCI-6013 included version 6.9.0f7. The PCI-6013 is one of our newer boards, and as such, is only supported by NI-DAQ 6.9.3 or Traditional NI-DAQ 7.0. These are the only two versions of the driver which will support this board.
    Traditional NI-DAQ 7.0 and NI-DAQ 6.9.3 are completely compatible. The VIs which you will use in LabVIEW to program your PCI-6013 are the same. Both of these drivers are available for free download from our web site.
    As for the -10001 error you are receiving, usually this has to do with an error in the channel string applied. Please see the following Knowledge Base article on our website for more information on this error.
    How Do I Handle the DAQ -10001 Error?
    http:/
    /digital.ni.com/public.nsf/websearch/42643645D554D15986256A250054ACEB?OpenDocument
    Regards,
    Justin Britten
    Applications Engineer
    National Instruments

  • Can I use Labview real time with PCI 6013?

    How much does Labview real time cost?
    Do i need to have some additional hardware to use labview real time with PCI 6013?

    You can not use the PCI-6013 board with LabVIEW Real-Time.
    We offer three platforms for LabVIEW Real-time. PXI embedded controllers, 7030 data-acquisition plug-in boards (7030/xxxx) and FieldPoint. For more information on these options go to www.ni.com/products and then choose Real-Time from the menu on the left. There are PCI versions of the 7030 boards. Each 7030 board has a data-acquistion board attached to it (6030E, 6040E or 6533). When you embed a LabVIEW Real-Time program on the 7030 you will be able to access the daughter board. You can not access other boards from the embedded program. If you want to use several daq cards in your real-time system I would recommend a PXI chassi and PXI daq boards.

  • PCI-6013 intermitant bad counts good after hardware reset

    I am using a PCI-6013 board to two edge separation counting into 10 bins over 1 second.  Most of the time the counter works great but some of the time I get garbage data eg 0,6050955,6050951,0,6050944,6050954,0,6051007,0,6050951.  The counts are coming from a PMT that has a max linear response of 1,000,000 counts per second.  Each data point represents 100ms of counting.  Powering the PMT off and then on has no effect.  After restarting my program the reponse returns to normal until the counts freak out again maybe 20 minutes later.  Does any one have any ideas why this is happening?

    What two signals are your trying to perform a two edge separation measurement on? What do you mean by bins? What language are you doing this in and can you post your code?
    Matt
    Applications Engineer
    National Instruments

  • PCI-6013 Labview SE software support?

    I am a retired engineer trying to do some experiments measuring the motion of clock pendulums. Since I am operating on a limited budget, I chose to purchase a NI PCI-6013 DAQ card after consultation about its limitations with a NI support engineer. I had hoped to do my experiments using a copy of Labview student edition since the purchase of a full version of Labview is beyond my means. Unfortunately, Labview SE does not appear to support the 6013 DAQ card and installation of a NI-DAQ that supports the 6013 is not functional with Labview SE! NI support confirmed that the 6013 was unusable with Labview SE even though there is no clue that is the case in the online support files. NI support suggested I try the Labview 7.0 Express
    Demo software until a student version of 7.0 is available but the demo software does not seem to support the PCI-6013. Does anyone have any suggestions on how to solve this problem?
    Thanks for any help someone can offer.
    Bob

    To preface my answer, here is some information that is located on our website ( FAQ for LabVIEW Student Edition). http://www.ni.com/labviewse/labviewse_faq.htm.
    Can I get technical support for LabVIEW Student Edition?
    There will NOT be any direct (voice or e-mail) technical support for the LabVIEW Student Edition. Students are asked to consult their instructors about their problems.
    Note:  The LabVIEW Student Edition is available to students, faculty, and staff for personal educational use only. It is not intended for research or institutional use. For more information about institutional or research licensing, please follow this link for information on our Academic Site Licensing.
    That being said, the reason the PCI-6013 won't work with the LabVIEW
    6i Student Edition (SE) is that only an old and "modified" version of the NI-DAQ driver works with the SE. That board was not yet released at the time of that driver release, thus not supported in the SE. Apparently, that board will work in version 7 of the LabVIEW SE, provided that the appropriate drivers are installed. This can be found through the following KnowledgeBase (KB).
    KB 2ZK6EHQJ
    Drivers for LabVIEW 7.0 Express Student Edition
    http://digital.ni.com/public.nsf/websearch/AA99FA8​704D5E22686256D6A004464ED?OpenDocument
    Keep in mind that you can program a DAQ board in other languages than LabVIEW such as C++, C, Visual Basic and .NET languages in you have any of these available to you.

  • Is it possible to use SCXI-1531 together with PCI-6024E or PCI-6013?

    I want to perform acoustic measurements as well as measurements with Hot-wire anemometers and need therefore low-pass filter and simultaneous sampling.

    Greetings,
    The PCI-6024E can be used with the SCXI-1531, but the PCI-6013 cannot. Furthermore, you will need an SCXI chassis to house the 1531 (you may have already known this). The SCXI-1000 can hold up to four modules. Below is a link to information concerning the SCXI product line:
    http://sine.ni.com/apps/we/nioc.vp?cid=1604〈=US
    Good luck with your application.
    Spencer S.

  • Syncroniza​tion

    Centro (Sprint)
    Posts: 1
    Registered: 05-02-2010
    0
    Syncroniza tion
    Options
    Edit Message
    Delete
    Mark as New
    Bookmark
    Subscribe
    Subscribe to RSS Feed
    Highlight
    Print
    Email to a Friend
    Report Inappropriate Content
    05-02-2010 11:47 AM
    I tried to hotzync my centro in February and was not able to complete the process. I was told I had to many duplications, and I could only eliminate them by deleting them manually.
    Last February I started deleting events. It took me forever and in the meanwhile I still could not hotzync.
    I was deleting on friday once again, in my spare time, and I accidentally PURGED all my diary until this last week.
    Unfortunately, the last time I hotzynced to my desktop was sometime in Dec 09!
    I cannot hotzync now because my desktop has all these duplicates. There was once a time that I could eliminate the duplicates by pushing a button. I no longer see that application on the program.
    Do you have any experience with this?
    Is there anyway to retrieve the info i purged if I did not hotzync the last 6 months?
    Can I delete duplicates on the desktop without doing it manually?
    Message 1 of 1 (0 Views)
    Add Tag...

    Are you still experiencing this issue? If so, try doing a clean uninstall of the BlackBerry Desktop Software and then reinstall it. Here is the KB article for instructions. http://bbry.lv/aQCW6K
    -FS
    Come follow your BlackBerry Technical Team on Twitter! @BlackBerryHelp
    Be sure to click Kudos! for those who have helped you.
    Click Solution? for posts that have solved your issue(s)!

  • Need device code for PCI-6013

    Help!
    Where can I get the device code of a PCI-6013 card?
    Thanks
    Mike

    Mike,
    What programming environment are you using? If LabVIEW, what version? Also, what version of the DAQ drivers are you using?
    Thanks,
    David
    David H.
    Systems Engineer
    National Instruments

  • About pci-6013 digital input

    i want to know what volume resistor i can used between 5vcc and digital input, order to keep input higt.
    i found it must less than 2k om.
    i need your propose.( i use pci-6013)
    thank you !

    Hi,
    in pci-6013 card, DIO state is pulled down by putting a resistor of 1.5 Kohm between DIO & DGND. If you connect 1.5K resistor between Vcc & DIO, you will get 2.5V approx.
    Further to increase voltage at this DIO, you have to reduce the value of this resistor. Lets say, if you connect a resistor of 150 ohm between Vcc & DIO, you will get (10/11)x5V which is near to 4.5 V which is surely TTL high signal.
    So you can connect any value between 150 ohm to 1.5 Kohm to get high TTL logic at DIO.
    Hope it works. Your feedbacks are welcome.
    Best Regards,
    Nirmal

  • Q10 contacts and calendar syncroniza​tion

    Using outlook account, syncronization of contacs and calendar by usb connection fails, with message at the end of the process "writing of the coomit on the device failed". It was not like this at the beginning. Computer and device software are updateted. All other usb options work fine, like sync images and music.Somebody knows how to solve it?

    First I did a sync only for calendar and it went through. Then I did the sync for contacts only and got the same error, so the problem in my case is in the contacts. So I generate the BB link logs enabling it in the BB link but without the windows registry modification (I'm not that skilled). The Log has been created anyhow, I went through it and found the line "outgoing payload" (there was only one) where a problem with one contact was indicated. I cancelled the contact on the phone and on the computer and did the sync again, but I got the same error. So I created again the log and found another problem with another contact. After three times I realised that with each try a new error with a different contact was created (only one at a time). So the problem is not solved, calendar sync works fine if done separately, contacts sync doesn't work properly: contacts added on the phone are copied to the computer but any change on the computer is not written on the phone. Maybe the way is to cancel all the contacts on the device and redo the sync from scratch, but I do not know how to cancel the contacts on the device (other than one by one, I have more than 400....) and the risk is not to get the contacts back on the device never again.
    Bye and thanks again,
    Maurizio

  • Blackberry Desktop Software syncroniza​tion errors

    I have not been able to find a solution to an error I am getting with a MAPi logon failure, I am using Microsoft Office 365 Pro Plus which has a 2013 version of Outlook (POP email not Mapi). I assume this error started after an automatic upgrade as I have been using this same set up for over 6 months

    Hi and Welcome to the Community!
    Interesting...I actually cannot determine at all why it ever worked for you...the Desktop Software is not compatible with Outlook 2013...refer:
    KB12268 Synchronization support for BlackBerry Desktop Software for Windows
    Unless it's Outlook that got upgraded? You were not clear about exactly what was upgraded...but, if it was Outlook, then that would indeed explain it...the upgrade took you to a version of Outlook that is not supported.
    Good luck and let us know!
    Occam's Razor nearly always applies when troubleshooting technology issues!
    If anyone has been helpful to you, please show your appreciation by clicking the button inside of their post. Please click here and read, along with the threads to which it links, for helpful information to guide you as you proceed. I always recommend that you treat your BlackBerry like any other computing device, including using a regular backup schedule...click here for an article with instructions.
    Join our BBM Channels
    BSCF General Channel
    PIN: C0001B7B4   Display/Scan Bar Code
    Knowledge Base Updates
    PIN: C0005A9AA   Display/Scan Bar Code

  • Problemi con la scheda PCI 6013

    Gentili signori,
    La nostra società ha acquistato nell'anno precedente 3 delle schede in oggetto, con le quali abbiamo creato 3 sistemi di acquisizione analogica di stazioni
    geodetiche.
    Già dall'inizio abbiamo avuto dei piccoli problemi sull'acquisizione, nel senso che il segnale acquisito per certi periodi sembrava venire amplificato da un rumore di fondo.
    Su una di queste stazioni abbiamo notato da quasi 2 mesi che ci sono degli spyke anomali e, come potete vedere dalle figure che vi ho inviato (geobox1 e geobox2), sono comparsi contemporaneamente su tutti i canali dal 26 Ottobre.
    L'acquisizione viene fatta in differenziale con campionamento a 10 secondi.
    Per fare una prova ho cortocircuitato sulla SCB-68 i Canali 3 e 4 (ovvero
    CH03+CH11 e CH04+CH12) in modo da misurare lo zero, ma i risultati di
    questa
    acquisizione sono evidenti nelle figure prova1 e prova2 che rappresentano rispettivamente il canale 3 e il 4.
    Avete qualche suggerimento da darmi?
    Grazie Mille
    Claudio Serio
    [email protected]
    SST s.r.l.
    Via F. De Lauzieres 5
    80046 - S. Giorgio a Cremano (NA)

    Ho un'altra prova che ho fatto tra ieri e oggi che Vi invio in allegato.
    In verità in tutte le comunicazioni che vi ho fatto fino ad ora mi sono dimenticato di precisare che le acquisizioni che Vi ho mandato finora sono state fatte con un software realizzato in VB.Net che si interfaccia alla scheda tramite le dll per .Net fornite con il pacchetto NiDAQ 8.
    Nell'ultima prova che ho fatto ho cortocircuitato tutti i canali tranne il n.3, dove ho acquisito un segnale variabile.
    Ho suddiviso in tre file le acquisizioni fatte nel seguente modo:
    Acquisizione software 18.01.07.XLS
    - acquisisco con il mio software tutti i canali a 1 secondo
    Acquisizione software 19.01.07.XLS
    - acquisisco con il mio software tutti i canali a 1 secondo
    Acquisizione VI 18.01.07.XLS
    - acquisisco con il VI Logger Tasks di Measurement & Automation  tutti i canali a 1 secondo
    Per semplicità ho graficato (i grafici sono all'interno del foglio excel) per ognuno di questi files, solo i primi tre canali(cortocircuitati) e per circa 600 acquisizioni.
    Come si vede c'è una finestra maggiore nei dati acquisiti dal software (quindi più rumore) rispetto a quelli acquisiti con VI.
    Il problema quindi potrebbe essere di carattere software, però non riesco ad immaginare che un software che gira e fa sempre le stesse operazioni comincia a funzionare male (senza un riavvio o un crash di sistema) nel mezzo del funzionamento. Infatti il grafico che ho mandato la prima volta fa riferimento ad un file acquisito in continuo.
    Avete dei suggerimenti da darmi o delle prove da farmi fare?
    Grazie mille.
    Allegati:
    Acquisizioni.zip ‏2769 KB

  • Simulink and labview syncroniza​tion

    I have a simulink model that has inputs controlled by Labview and outputs displayed in Labview. On my desktop computers the simulink and labview are running in sync, their times are always together. When I load the programs onto my two laptops, the simulink model is running 2-10 times faster than the Labview program, and when I change controls in Labview, they don't take effect in simulink unless I change the control MANY times. I think this lack of communication between labview and simulink on my laptops is happening because they are out of sync, but not sure, any suggestions?

    Hello,
    You mentioned that the program works fine when you are using a desktop, is simulink and LabVIEW running together on the same desktop or do you have them networked together? By default simulink communicates through port 6011 using TCP/IP. It possible that for some reason that the communication is inhibited by some process on your laptop that is also using that port and causing only some of the simulink data to get through. It is possible to change ports when starting simulink if you know that a port is blocked or being used by another process. When starting the SIT server, type NiMatlabServer ('start',xxxx) You can choose any port that you want, but many times firewalls will close ports. Each firewall is different so you will have to look into what ports that you have available for TCP/IP communication with the SIT server. Have a great day.
    Regards,
    Danny G.
    Applications Engineer
    National Instruments

  • NI PCI-6013 and Windows XPE

    We are porting our project to XPE on a diskless system.
    We were told that NI does not support that directly (is that true ?).
    We would need to create a Target Designer component for this module.
    I am looking for a plug and play (.inf stype) driver installer or else for instruction on how to do installl the driver manually.
    Using the .exe installer provided with the board is not acceptable in our environment because of the footprint. We even experienced an installation failure possibly due to the type of non volatile media (a standard CF card  that appears a removable drive).
    Please help us.
    Saluti.

    >NI does not support Windows XP Embedded. This does not necessarily mean a DAQ Board or Driver....
    It would suffice to discover how to install the driver manually and assign configuration data without using MAX.
    >Would realtime be an option for you?
    No, our project is tied to Windows.
    Saluti.