SCXI-1120D input voltage protection:

Hello All,
I have a SCXI-1120D input voltage protection question:
the data sheet specified 250Vrms. Does it mean signal plus CMV only? What if my signal itself is has a large potential difference (thus not CMV) , say 20V? Will the SCXI 1120D be damaged?
Thanks
Feilong

Hi Feilong,
You are correct. The 250 Vrms refers to your signal plus the CMV. See Figure 3-2 on page 32 of the SCXI 1120D User Manual. The paragraph above Figure 3-2 also explains this. I hope this helps!
Regards,
Ima
Applications Engineer
National Instruments
LabVIEW Introduction Course - Six Hours
Getting Started with NI-DAQmx

Similar Messages

  • Input Over-voltage protection during Power Off

    Hi,
      I am using NI-PCI-4474 & NI-PCI-4472B for my data acquisition application. The normal input voltage range is +/-10V while input over-voltage protection range is +/-42V. Now if the PC in which these devices are installed is powered off, will the input over-voltage range will remain same? Actually in our application, there may be a chance that input signal is coming while these devices are powered off. Do I need to use some extra protection mechanism in this case?
    Solved!
    Go to Solution.

    Hi aNIta B,
      Thanks for your support.  
    Basically input voltage signal will remain in the range of +/-10V (and worst case +/-25V) when device is powered off. I am just concerned since usually for an operational amplifier present at input stage, if there is no input current limiting mechanism, the input voltage can damage operational amplifier if its supply voltage is cut-off (which will obviously be the case when device is powered off). This is what typically mentioned in operational amplifier datasheets. Also if over-voltage protection (+/-42V) is based upon active components then it should also be dependent on supply voltage condition.
    I hope you understand my point now. I shall be waiting for your expert advice since you may be much more aware about the input circuitry and protection mechanism used.

  • I'm not able to set Analog Input voltages, etc

    ok,
    so i have an NI DAQ I/O card (NI PCI 6723, or something to that effect) and i'm trying to configure the channels for Analog Input signals.
    when i try to choose any kind of setting for Anolog signal by:
    >>drop DAQ Assistant vi onto the block diagram
    >>>>choose Analog Input
    >>>>>>>Voltage
    i get this message "No supported devices found"
    i thought since I have the I/O card, i should be able to configure channels for both input/output.
    i also have ADAC card (from IOTECH) installed but when i look at MAX, i only see the DAQ card.
    i'm really confused at this point. can anyone break things down for me ... what i'm doing wrong?
    -r

    stuartG,
    you were right. so what i'm using for the input is a ADAC 5501MF from IOTECH
    following their manual (http://www.iotech.com/productmanuals/adac_lvi.pdf) it says that the ADAC-LVi libraries can be located at "Functions->User Libraries" palette of labview.
    does this mean that i have to install the ADAC-LVi driver in LabView->User Libraries directory?
    if that's not the case then i don't see why the ADAC-LVi is not showing up in the LabView User Libraries since i've installed the driver correctly (in C\Program File\ADAC )
    do you have any idea why?
    thanks
    -r

  • Varying input voltage refrigerator testing application

    I am working on an application for refrigerator testing. One of the specs that we need to comply requires that we vary the input voltage, (ex. 127 VAC for a 120 VAC unit) through some sort of voltage regulator. The biggest unit we test has a 2 hp compressor with a max current draw of 18 amps. We would like to know what equipment we would need in order to accomplish this through a LabVIEW application that we are working on. We could do this through GPIB, serial, etc. Any help on this matter will be appreciated.

    You might be right.  I was think of one small 1.5 HP controller (Commander SE) we are using that is being powered by 1-phase 240 VAC supply.  So I was thinking the motor was 1-phase, but actually it does the voltage chopping to create a 3-phase supply to the motor.
    It may be possible to still use one where you just use 1 phase relative to ground, but I'm not sure, and I'm not an electrical engineer to be able to tell how to do such a thing.  It may be worthwhile to ask them if such a thing can be done.  Or look around for some other motor speed controller manufactures that make a 1-phase motor controller.
    I'm sorry, but I don't know where else to look or be of any help.

  • Input voltage on hub

    I have a Linksys EFAH16W work group hub I wish to use, however the power transformer is missing. I find no markings upon the case or circuit board stating the input voltage value or, if its AC or DC, from the transformer.
    If you own one and can spare a moment, look at the transformer and find the output listings and share them.
    Thank you very much..

    Can anyone tell me where I can purchase one of these. I have had problems locating one.

  • Error when trying to edge-trigger to a input voltage smaller 300mVpp

    The programm shown in a cut below (just a cut of it) reads multiple channels from the PXI5105. The program allows to choose between different triggers. When I choose edge-triggering and the value of the input voltage I am triggering to is smaller than 300 mVpp the program stops and sets an error. With the 'Highlight function' I found out that the error is set at the '1D WDT'-vi.
    The error message:
    ERROR
    "Possible reason(s):
    Driver Status: (Hex
    0xBFFA4009)
    A previous acquisition is still in progress. If you are
    attempting to change an attribute, note you can only change fetch attributes
    while an acquisition is still in progress.
    Status Code: -1074118647
    I guess, if the voltage gain isn`t high enough the '1D WDT'-vi waits the 'timeout' = 5s for the trigger, isn`able to get one, and stops. But what can I do to trigger to smaller voltages as 300 mVpp? There is no option to change it. The trigger level of the 'Analog Edge Ref Trigger' is set to 0.
    Attachments:
    error1.jpg ‏120 KB

    Well then that's not it.
    Not triggering at low voltages- can be caused by either scale too high
    (so that 1 bit is bigger than the trigger level, not the case here) or
    scale too low (because of saturation the acquired signal never reaches
    the trigger level).

  • Why input voltage couldn't display in chart ?

    In my vi,why input voltage couldn't display in chart ?
    please see the attachment
    Attachments:
    singal requency.vi ‏33 KB

    Well, your ai runs at 10000Hz and your ao runs at 20Hz, so start there .
    If they should be really synchronous, run them off the same clock.
    LabVIEW Champion . Do more with less code and in less time .

  • Random errors using SCXI analog input on SCXI1100

    This system is using a AT-MIO16E card and labview. I am performing single channel analog scans using the NI SCXI analog input example, modified to use the RMS and extract frequency examples. During static testing (ANy one input), the readings are fine. In one frame, the 1125 is polled 4x, and the 1100 is polled once. The scan for the 1100 module does an RMS and frequency extraction. In this frame, the result from the frequency is NAN and the RMS is 0 about 50% of the time. Individually, the modules acquire the data flawlessly (1 volt, 49.5kHz). The 1100 is is passed scan rate, buffers, scans. The signal acquired is a 50kHz square wave. I've tried scan rates from 150 to 250k. Any ideas on what could be causing this?

    Hello;
    It's hard to tell what might be happening without seeing the code, but the best way to test that out is to acquire the data without applying the RMS functionality, and double check if the data is the expected.
    If it is, then the problem is inside the RMS function. I'm sorry I couldn't give you a definite answer, but at this point the problem might be anywhere in the code, and we probably need to narrow the possibilities down.
    Regards
    Filipe A.
    Applications Engineer
    National Instruments

  • Analog input voltage fluctuatio​n in myDAQ

    -Hello !
    I have a turbine blade rotation sensing and measuring device (i.e. picoturn) connected to AI4 pin of BNC 2110 I/O board. I am using a PXI-1033 chassis with  PCI-6251 M series card in it. I have programmed an algorithm which acquires voltage from AI4 channel through DAQ Assistant and converts voltage into RPM. 
    Picoturn is connected to DC power supply (ranges 9 - 36V)
    These are the following problems:
    1. I will get error voltage readings if picoturn is connected to other channels except AI4 (i.e AI0,AI1,etc)
    2. The reading of the voltage keeps on fluctuating between 0.49 and 0.51, in both the cases of when turbine is stationary and rotating.
    3. If the power supply to picoturn is removed, even then I could see the fluctuations in the voltage reading from -0.0005 to 0.1 and above till I stopped running the program.
    Could you please help me to solve these problems?
    Attached is the VI of the program and manual of the instrument Picoturn. 
    Attachments:
    Picoturn.pdf ‏515 KB
    Analog_ RPM.vi ‏64 KB

    Hello
    What Error Voltages do you read when the picoturn is connected to other AI Inputs than AI4?
    What are you using for the ground reference of your measurement? 
    Checkout this White-paper: http://www.ni.com/white-paper/3344/en/
    Are you able to measure other analog input voltages from a different voltage source? e.g. a 1V DC signal from a power supply or a sinusoidal voltage from a frequency generator?
    Regards
    Corinne

  • Elvis II over voltage protection

    Hi,
    Does Elvis II have over voltage protection at 5V/2A output?
    Is it possible to destroy 5V power line, connecting supplies 5V and 15V together on the PROTOTYPING BOARD?
    We are using Elvis II boards in our student labs, so we must suppose all the possible cases.
    Documentation tells only about over voltage protection in the Variable Power Supply Circuitry.
    What is the difference between "Resettable circuit breaker" and "Self-resetting current limiter" (terms from NI Elvis Docs)?
    best regards
    ppawl

    Hi,
    thanks for the answer, but I'm not sure if you are right. Circuit breaker works with overload and this is not the case. Overvoltage at the output doesn't have to cause that the current exceeds 2A. Also the current flows in reverse direction - to the output. This is not a typical overload situation like e.g. a short connect to the ground.
    If we connect outputs of +15V and +5V we also know, that the power supply at +15V in ELVIS II has 0.5A current limiter, so it is less than 2A.
    The ELVIS II at 5V/2A has UCC3819DP current limiter from TI. It has overload protection and circuit breaker, but it is designed for unidirectional current flow and accepts voltages up to 6V (maximum absolute rating is 8V). This IC doesn't have overvoltage protection.
    best regards,
    ppawl

  • PC-6023 MAX INPUT VOLTAGE

    IN THE NAME OF GOD
    I have a pci-6023E card and potentiometer . The ouput of the pot. is differential .
    The (-)end is always 13.5 Volt and the (+) end is between 5to22 Volt. I connected the output of the pot. to the input of the card in dif mode and i could measure the output of pot( ie 13.5-(5to22) ) successfully.Is this a correct connection to the card?
    The maximum working voltage of each input of the card , according to manual, is 11 volt, but my sensor (i.e pot.) outputs are 13.5 and (5-22) !!
    I used BNC-2110 terminal in dif. mode for this exprement.
    Please explain me that how the card can successfully measured this input?
    Can this demage the card at ling time or not?
    thanks
    Edward. Hb.

    Thanks Mr. Alan A
    I am "434343" memeber.
    I recive your response to my question about "Max input voltage of PCI-6023E" .But I have some questions.
    According to the "Data Acquistion Specification- a Glossary by Richard House" ( Application Note 092):
    "Maximum Working Voltage (signal+common mode)= the highest voltage level that can be input to board without saturating the input."
    And according to your response:
    "measuring a signal with higher common mode voltage than that(11V) might not accurate"
    But in may experiment with a potentiometer (floating differntial source with (-)13.5V and (+) 5to22V),I could successfully measure the signal without saturating and without error (Max error is about +/- 0.02 volt)!!
    I think that the maximum working voltage of PCI-6023E is greater then 11 volt??!!
    or my measurement has a little error.
    I think that I don't need to use a signal conditioning circuit.
    Please Help me.
    Thans again.
    Edward Hb.

  • Input voltage in LAP1242AG?

    Following up the previous discussions on powering the LAP1242AG through PoE IEEE 802.3af, is there a way to monitor, using SNMP, the input voltage if (a) using IEEE 802.3af and/or (b) using external power 36-57 VDC?

    Hi AaronJudge,
    Welcome to the NI discussion forums.  I have a suggestion about what may be occurring.  You mentioned that you are using SCC-SG04 module with a SC-2345 signal conditioning box. Do you
    have the LVDT connected to the SCC-SG04? 
    If so, then you will not be able to change the range above +/- 100
    mV.  The SCC-SG04 is a full bridge strain
    gauge input module.  It has a fixed input
    range of +/- 100 mV. 
    In order to read the full range of your
    LVDT you will need a different setup. 
    According to this Knowledge Base (Is There an
    SCC Module that Can Be Used with an LVDT?) there is
    not currently an SCC module that provides both the excitation requirements for
    most LVDTs.  If you have and external
    excitation source an SCC-AIxx module could be used.  There may be other options as well that do
    not use SCC.
    Jared T.

  • WLS 9205 setting input voltage range

    Hi folks. I need some assistance in setting my input voltage range for my WLS 9205. There are three settings (it would appear) from the documents supplied with the DAQ device ranging from +/- 200mV, +/-1V and +/-10V. How can i create a dynamic menu in standard Labview to alter the settings within a .vi. I am not using field point or anything of the sort.The DAQ mx does not give me the option to select the range either.
    On the old devices, you could alter setting using dip switches but there are no dip switches on this device. Any advice would be greatly appreciated.
    Thanks,
    Daragh

    Here is the equivalent information for NI-DAQmx (since the WLS-9205 does not use LabVIEW FPGA):
    How Do I Set the Analog Input Gain using DAQmx?
    Using Min/Max or High/Low to Set NI-DAQmx Channel Gain
    Brad
    Brad Keryan
    NI R&D

  • Does the iPad mini have a setting so it can be used with either 120 or 240 electrical current? If yes, how do you go about setting the desired input voltage? I will travel to Northern Europe and want to know if I will need a voltage converter.

    Does the iPad mini have a setting so it can be used with either 120 or 240 electrical current? If yes, how do you go about setting the desired input voltage? I will travel to Northern Europe and want to know if I will need a voltage converter.

    The iPad charger is dual voltage. You need only a plug adapter to be able to plug your US charger into a European wall outlet. A basic one costs about $2.00. - $6.00. Like this...
    http://www.amazon.com/CVID-Grounded-European-Adapter-Adaptor/dp/B0098B5GN8/

  • SCXI 1100 input protection

    Can the SCXI 1100's inputs take an input of about 10 mA? (from a current source?). If a stimulator (a current source) were connected directly into one of its channels, would there be any physical damage caused to the SCXI 1100?
    It doesn't matter if the SCXI shuts off the channel temporarily to protect itself, we just want to make sure that the board/channel itself will not be damaged. And also, is there any "saturation time" of some sort before we can re-use that channel to take regular level readings again?

    Hello,
    Thank you for contacting National Instruments. Yes, the SCXI 1100 can handle a current input of 10 mA. The module is designed to be used for current inputs ranging from 0-20 mA or voltage inputs ranging from negative to positive 10 V.
    This information, as well as other specifications for the SCXI 1100 can be found in the product manual for that device, located at:
    SCXI 1100 User Manual
    http://digital.ni.com/manuals.nsf/webAdvsearch/0790D3C92CF4B16E8625665E00712865?OpenDocument&vid=niwc&node=132100_US
    The current input specification is mentioned in the "About This Manual" section.
    Let me know if you have any additional questions on this issue.
    Regards,
    Scott R.
    Applications Engineer
    National Instruments

Maybe you are looking for