Full-Bridge Strain Vs. Custom Voltage with Ex: outputs opposite polarities of eachother (SCXI 1314)

Please see my Sig for current hardware specs.
I have a manufactured load cell that I'm plugging into an SCXI-1314 wired for full bridge. When I condition and calibrate the load cell as a Full-Bridge Strain channel (shunt cal) my signal output is POSITIVE in compression.
If I take the same load cell plugged into the same channel in the 1314 and select Custom Voltage with Excitation and give it a mV/V slope curve in the Custom Scale the signal output is NEGATIVE in compression.
Why is this? I fear I'm overlooking something very rudementary.
Other than the polarities  being opposite of eachother the signals appear just fine.
SCXI- 1000 Chassis w/ 1346 adapter
PCI 6281 DAQ card
SCXI- 1520 Bridge Board w/ 1314 Terminal Block (x2)
SCXI- 1180 Feedthrough Panel w/ 1302 Block
Signal Express 2014.
Win7 Enterprise

Here's a drawing of our wiring/cabling convention for load cell/scxi interface.
SCXI- 1000 Chassis w/ 1346 adapter
PCI 6281 DAQ card
SCXI- 1520 Bridge Board w/ 1314 Terminal Block (x2)
SCXI- 1180 Feedthrough Panel w/ 1302 Block
Signal Express 2014.
Win7 Enterprise
Attachments:
MS House to SCXI-1314 Full Bridge.jpg ‏47 KB

Similar Messages

  • What is the best way to zero a 'custom voltage with excitation' sensor from the front panel?

    What is the best way to zero a 'custom voltage with excitation' sensor from the front panel?
    Thanks for any help

    Hi Donavan,
    There are two options for zeroing with the NI-9237.  You can perform offset nulling and shunt calibration with a strain task, or if you use custom excitation you can use software compensation.
    See the Bridge Balancing, Offset Nulling section in  the Load, Force, and Torque Measurements tutorial. Take a measurement before the strain is applied and subtract this from all subsequent measurements.  See the above tutorial to consider the tradeoff in accuracy between these two options.
    Hope this helps,
    Jennifer O.
    Applications Engineer
    National Instruments

  • DAQmx Create Virtual Channel (Custom Voltage with Excitation)

    Hi,
    My problem is this. I am using the custom voltage w/ excitation for my application, together with the scxi 1520. Of course, you have to put values on the excitation value, say 5 V.
    But whenever I turn off the chassis aftr measurements (the LabVIEW program is still open) and then do the measurements all over again, I need to enter a different value at the excitation value, say 3 V, to ensure that excitation will be provided. Otherwise, I get no signal. I can then return to the previous 5 V to do the same measurements again.
    Is there a way to solve this problem. Thanks!

    Ahh, ok now I understand your question. This behavior is due to a combination of 2 reasons. First the DAQ driver is caching the values for the SCXI registers, so if you set the same values for the registers the driver doesn't write to the hardware again. Second, the driver doesn't get any notification from the hardware that the hardware has been turned off, so the driver continues on as if nothing happened.
    So, the best solution would be to do a device reset on the SCXI module before you start the acquisition. A device reset would clear the cache and cause data to be written out again. This would of course make things a tiny bit slower, but if you are turning off the chassis, the that little bit of time probably does not matter.

  • How to generate custom report with PDF output?(Payables)

    Hi All,
    we are using some custom reports in payables modual production server the reports are working fine .
    my question is i need creat same reports in test server i have done following activities.
    1. copied rdf file from prod server and pasted into test server
    2.given excutable name in system administrator modual(concurrent-progrem-excutable)
    3.define report in system administrator modual(cocurrent-program-define) output selected PDF and given parameters.
    4. switch to payables moduale run the report ,following error showing in log file plese help how to resolve it we don,t have any custom reports in test server this is first time to creat the report.
    APPLLCSP Environment Variable set to :
    XML_REPORTS_XENVIRONMENT is :
    /APPS/dkvuatora/8.0.6/guicommon6/tk60/admin/Tk2Motif_UTF8.rgb
    XENVIRONMENT is set to: /APPS/dkvuatora/8.0.6/guicommon6/tk60/admin/Tk2Motif_UTF8.rgb
    Current NLS_LANG and NLS_NUMERIC_CHARACTERS Environment Variables are :
    American_America.UTF8
    Spawned Process 2707514
    X connection to oraclefinof1:0.0 broken (explicit kill or server shutdown).
    Report Builder: Release 6.0.8.25.0 - Production on Wed Nov 18 16:06:17 2009
    (c) Copyright 1999 Oracle Corporation. All rights reserved.
    Enter Username:
    Start of log messages from FND_FILE
    End of log messages from FND_FILE
    Program exited with status 1
    Concurrent Manager encountered an error while running Oracle*Report for your concurrent request 339487.
    Review your concurrent request log and/or report output file for more detailed information.
    Executing request completion options...
    Finished executing request completion options.
    Concurrent request completed
    Current system time is 19-NOV-2009 16:24:52
    ---------------------------------------------------------------------------

    Hello.
    Have a look to Metalink Doc ID 252715.1
    You have to login to server and issue the 'xhost +' command.
    Hope this helps.
    Octavio

  • Can I measure full bridge gage configurations with SCXI1121 and 1321 terminal

    Hello
    I designed a load cell that uses a full bridge gage configuration. I have at my lab a SCXI 1121 module with a SCXI 1321 terminal block. I need to know if it is possible to use that hardware configuration in order to make de load cell and the data acquisition system fully functional.
    Thank you.
    Best regards from Bogotá, Colombia, South America

    Juan,
    The full brige configuration you are speaking of requires four wires, two for voltage and two for excitation, is this correct? If so here is a link that you might find helpful.
    Full Bridge Configuration
    This reference is for strain gauges, but the connection for a full bridge is going to be the same. The channel plus and minus and excitation plus and minus can be connected to the 1321 for the individual channel. To configure the proper gain, filtering, and excitation mode and level, please refer to the Getting Started with SCXI on pages 3-14 through 3-23 (pages 51-60 in the .pdf).
    Regards,
    Andy F.
    Applications Engineer
    National Instruments
    National Instruments

  • Issue with fixed point number output from 9219 module for full bridge measurement (cRIO/FPGA)

    Hi,
    I have a question regarding the fixed point output acquired from a 9219 module (in FPGA on a cRIO) when setup to acquire a strain measurement (full bridge).
    Software: Labview 2009
    Hardware: cRIO-9012 (NI-RIO 3.2.1), NI-9219 module
    The 9219 module is configured in the project as follow:
    Ch0: Full-Bridge +/- 7.8mV/V
    Ch1: Voltage +/- 4V
    Ch2: Voltage +/- 15V
    Ch3: Voltage +/- 60V
    The calibration mode is 'Calibrated' so the FPGA outputs fixed point numbers. 
    My issue is that the precision of the fixed point number for Ch0 (strain) is "(fixed point <+/-32,-1>[-2.5e1,2.5e1]:1.16e-10)", which indicates that the fixed point number is a value between +/- 0.25 and not the +7.8mV/V as I expected?
    The fixed point number does not change in precision whether the range for the strain measurement is set at +/- 7.8mV/V or +/-64mV/V (the two available options).
    As the fixed point number doesn't change precision I'm assuming that changing the range of the strain measurement changes the resolution of the acquired number? And as such I will need to perform additional scaling on the fixed point number to convert it to the expected range?
    There is no mention of scaling of the voltage or strain measurements mentioned in any documentation or examples, with the only scaling example provided for the thermocouple measurements.
    Any help/clarification is much appreciated.
    Regards,
    Mike

    Hello Mike,
    Hopefully I can help clarify some of the behavior you are seeing.
    My issue is that the precision of the fixed point number for Ch0 (strain) is "(fixed point <+/-32,-1>[-2.5e1,2.5e1]:1.16e-10)", which indicates that the fixed point number is a value between +/- 0.25 and not the +7.8mV/V as I expected?
    For calibrated values on the FPGA VI, the returned data is a Voltage measurement, not a directly calculated strain value. Based on the specified ranges for the 9219 in a Full bridge configuration, 250 mV will encompass all possible input values, at the module provided excitation (2-2.7V dependent on the sensor gage resistance).
    The fixed point number does not change in precision whether the range for the strain measurement is set at +/- 7.8mV/V or +/-64mV/V (the two available options).
    As you may notice from the fix-point definition, the fixed point data contains 32-bit precision which is larger than the acquired precision of 24-bits provided by the 9219. The fixed point data-type is coded to accept input encompassing both the range and precision of the instrument; such that, no additional coercion of the input data values is required based on user-defined software settings, i.e. the bridge sensitivities +/- 7.8mV/V or +/-64mV/V. 
    As the fixed point number doesn't change precision I'm assuming that changing the range of the strain measurement changes the resolution of the acquired number? And as such I will need to perform additional scaling on the fixed point number to convert it to the expected range?
    The documentation does not clearly define that by varying the discrete levels of strain input (+/- 7.8mV/V or +/-64mV/V) the range of the ADC on the module is also adjusted. I am working to follow up further on this topic, to provide a clarification on the module documentation. As for scaling, the voltage values acquired regardless of the ADC resolution will still be related strain via the bridge sensitivity. The resolution of the ADC will simply define the smallest measurable change in the strain value. 
    There is no mention of scaling of the voltage or strain measurements mentioned in any documentation or examples, with the only scaling example provided for the thermocouple measurements.
    For converting the acquired voltage values to a strain measurement, I would recommend the documentation linked here for a detailed explanation on strain calculation. Often, users will forward the acquired voltage data as fixed point values through a DMA FIFO to the RT controller on the Compact RIO; such that, they may handle conversion from voltage to strain using floating point math in real-time. 
    I hopes these responses provided a bit of clarity. I will continue to work to provide additional information on the 9219 specification information. Please post back any further questions.
    Cheers!
     Edit: Forgot to add the link.
    Message Edited by Pcorcs on 04-14-2010 04:55 PM
    Patrick Corcoran
    Application Engineering Specialist | Control
    National Instruments

  • SCXI 1520 quarter bridge setup will not output excitation voltage, but full bridge will

    My SCXI system will not output an excitation voltage to the quarter bridge(verified with voltmeter). It works for a full bridge setup. When connected to a 'quarter bridge I' configuration, I can use MAX to null and shunt calibrate. The quarter bridge circuit appears to be balanced after the calibration, but the calibration results post a software gain factor close to zero, as well as a close to zero excitation voltage (eventhough I have the voltage configured for 5 volts). I have the SCXI 1520 module, the virtual channels, and the wiring properly configured (several years experience with this device). I'm using an old E series DAQ card AI-16XE-50, with the 50 pin cable that plugs in the back of the SCXI chassis. I noticed I had two bent pins with the cable connector module that plugs into the SCXI backplane.  I was able to straighten them, so the cable connects properly, but I suspect the bent pin incident caused electrical circuit damage. Any suggestions on trouble shooting? None of the three 1520 modules I have will output an excitation voltage for a quarter bridge.

    Sorry that I don't have an answer for you tyea, but I'm interested in your thread because I have a similar issue with a 1520. My excitation votltage, as measured with a DVM, is always 1V. I've tried 0V, 1V, 5V --- it's always 1V, on Full or Quarter bridge. I'm using a similar DAQ to yours - a PCI 16XE-10 (and a PCI-6030E on antother system).
    I don't have a lot of experience with these modules. This could be a wiring or programmatic thing I'm doing wrong.
    Richard

  • 9237 with full-bridge load cell: load cell_null_off_shuntcal.vi throws error 200077

    Hi,
    I'm trying to use the example
    load_cell_null_off_shuntcal.vi  with a full-bridge load cell (Honeywell
    Model 31, unamplified). I am using LabView 8.6,  cDAQ-9172 and NI9237. The
    load cell is connected to pins 2,3,6 and 7.
    The inputs for the VI front panel
    are: internal excitation10V; mV/V 2.1492 (calibration sheet); max weight 10
    lbs; bridge resistance 350 ohms (Honeywell specs); 9237 internal shunt
    resistance 100 kohms; shunt location R4 (default setting). I have selected
    "Do offset null" and "Do shunt cal".
    This is the error I receive:
    Error -200077 occurred at DAQmx
    Perform Shunt Calibration (Bridge).vi:1 Possible reason(s):
    Measurements: Requested value is not
    a supported value for this property.
     Property:
    AI.Bridge.ShuntCal.GainAdjust
    You Have Requested: -61.980405e3
    Valid Values Begin with: 500.0e-3
    Valid Values End with: 1.500000
    If the "Do shunt cal"
    green button is not selected, there is no error. I understand that the Gain
    adjust value should be approx 1, whereas the one I get is much larger. The  subVI  DAQmx PerformShuntCalibration
    (bridge).vi contains a "Call library function node" which I don't
    know how to interrogate.  
    Has anyone else had experience
    with this error? Do you have any advice on:
    1)    
    How to "see" the calculations being
    perfomed inside the "call library function node"?
    2)    
    What the correct shunt element
    location for a full bridge load cell is? (although changing this location does
    not eliminate the error, I can't find this info).
    3)    
    Anything I may be doing wrong with
    my inputs to cause this error?
    Thanks,
    Claire.
    Solved!
    Go to Solution.

    Hi Claire,
    You have to physically connect the SC terminals to one arm of the bridge (normally R3). The terminal is not provided for connecting external resistors.
    See example 
    C:\Program Files\National Instruments\LabVIEW 8.6\examples\DAQmx\Analog In\Measure Strain.llb\Cont Acq Strain Samples (with Calibration) - NI 9237.vi
    "A VI inside a Class is worth hundreds in the bush"
    യവന്‍ പുലിയാണു കേട്ടാ!!!

  • Customer Experince with nokia N76 from may 2007 ti...

    Customer Experince with nokia !!!!
    today 20-09/2008 and still under progress!!!!!
    they wait till my warrenty end to reply to my problem...
    this just simple of emails i have the rest.....
    Till now more than 20 day's investagtion?what a careline ur just lazy and careless line but promise u will have a very big problem.bye
    @nokia.com wrote:
    > Dear Samir,
    > Thank you for emailing Nokia Careline.
    > With regarding to your inquiry kindly note that I did check the case again for you and it is still under process and once we got any new feedbacks be sure that we will come back to you, we are so sorry for being late.
    > Should you require any further assistance, please do not hesitate to contact the Nokia Careline and speak to any of our Customer Service Executives on 19876, between the hours of 9am and 8pm (local time), Saturdays to Thursdays. For online assistance, please visit http://www.nokia.com/mea/contactus.
    > Register with Nokia Connections and you will receive a bimonthly fun newsletter about the latest products and events. Click below to register.
    > http://www.nokia.com/mea/nokiaconnections
    > Kind regards,
    > Nokia Careline
    > Nokia Middle East & Africa
    >
    > [THREAD ID:2-48X8FP]
    > -----Original Message-----
    > From: @yahoo.com
    > Sent: 28/03/2008 04:46:14
    > To:
    > Subject: Nokia Careline - Samir Khalil - Nokia N76
    > Dear Mr. *****,
    > from you sent case no. 2-262950130 u told me 48hrs and
    > i will have reply till now 10 days no one from your
    > team call me or told me anything about my complaont is
    > it Nokia cant find any answer or what? u will
    > investigate forever and ur careline number in Egypt
    > doesnt work its give Error which means NO nokia at all
    > in Egypt so plz dont send ur customer service in ur
    > emails its just Fake number not working number i hope
    > if u in egypt to try it by yourself.
    > i will print all emails about my case today and all
    > paper's related to my esclate and insurance and the
    > warranty and will wait Last 48hrs am sorry i will take
    > legal action to nokia Egypt & middle east,hope to
    > contact me with final result not to wait anymore not
    > accepted by any mean.
    > ***This msg. to Mr.
    > Samir Maher
    > EGYPT
    Dear Samir,
    Thank you for emailing Nokia Careline.
    With regards to your enquiry, kindly be informed that your case is still under investigation and a resolution is expected shortly. Rest assured, your case has not gone forgotten and we will revert as soon as a resolution is achieved.
    Should you require any further assistance, please do not hesitate to contact the Nokia Careline and speak to any of our Customer Service Executives on 19876, between the hours of 9am and 8pm (local time), Saturdays to Thursdays. For online assistance, please visit ‘ASK Nokia’ at our website www.nokia.com.
    Register with Nokia Connections and you will receive a bimonthly fun newsletter about the latest products and events. Click below to register.
    http://mea.nokia.com/A4211102
    Kind regards,
    Nokia Careline
    Nokia Middle East & Africa
    [THREAD ID:2-48X8FP]
    -----Original Message-----
    From: @yahoo.com
    Sent: 20/03/2008 06:12:29
    To:
    Subject: complian is done
    Dear ,
    48hrs passed and nothing happen no one come back tome
    with any feedback frpm complain department actualy its
    72hrs my case no. 2-262950130 for nokia n76.
    that's the maximum respect you show to you customer's?
    am asked lot's to call me or your manager to call me
    samir or the resposible to this
    department call me and nothing happen.
    Dear Samir,
    Thank you for emailing Nokia Careline.
    Thanks you for your cooperation, I do appreciate that
    and I did a
    complain for you and here is the Service Requisite
    for you complain:
    2-262950130. our complain department will come back
    to you within maximum 48
    Hours.
    Should you require any further assistance, please do
    not hesitate to
    contact the Nokia Careline and speak to any of our
    Customer Service
    Executives on 19876, between the hours of 9am and 8pm
    (local time),
    Saturdays to Thursdays. For online assistance, please
    visit
    http://www.nokia.com/mea/contactus.
    Register with Nokia Connections and you will receive a
    bimonthly fun
    newsletter about the latest products and events.
    Click below to register.
    http://www.nokia.com/mea/nokiaconnections
    Kind regards,
    Nokia Careline
    Nokia Middle East & Africa
    [THREAD ID:2-48X8FP]
    -----Original Message-----
    From: @yahoo.com
    Sent: 16/03/2008 23:46:13
    To:
    Subject: Advice about your complian
    Dear ,
    this 2nd time i send this information in 3 weeks :
    Nokia N76:
    1. Job Order Number./1526584 & 1537749
    2. Authorized Service Center./Raya
    3. your full name and your phone number./Samir Maher
    Khalil-
    4. issues./1. Camera with very bad Quality &
    resultion. 2.Voice incoming calls&outgoing calls. 3.
    Material of the mobile its very very bad & not safty
    for humane being for use it.. 4. Its restart suddenly
    wizout any reason manytimes on daily basis .
    5.Software not helpful at all very completcated & have
    errors cant be able to download any new proudcts.
    6.low or NO at all performence battery there no
    battery u can say so empty every 12hrs maybe less
    maybe more deping. 7.Handfree with errors in the
    cable. 8.sms date not accurate with local time .
    9.high price with no need 3650 LE in 25/05/2007.
    my mobile after this date sent to Raya to fix it
    twice everytime not less than week with very bad
    customer service attuide which bad experince for
    customer nokia(could u plz visit any of them as
    mystery shopper)let's feel as customer,after all they
    changed for me the battery & handfree & sent my mobile
    again for the same problem's till now suffring from
    this mobile while my job at Mobinil which means
    Egyptian mobile opertator,i think now u got it how
    much my mobile important in my life & my work.please i
    need a help from Nokia i have been esclated from 3
    weeks and till now u didnt solve anything 4 me or even
    give me any helpful feedback.waiting ur feedback plz
    offer me solve for every problem from those and
    clairfy with details in ur email ok.
    5. faulty.
    6. Serial Number./
    7. Date of purchase./ 25-05-2007
    8. Date of send to repair./1st. time : 16/06/2007 &
    2nd. time: 19/06/2007
    9. Person name that you did contact with./sure can't
    remmber but i contact with Raya- Mesadak shop manager.
    ****Case no. : 2-258561699
    ***I will wait call from you
    Thanks
    Samir
    Mpderator note: personal information removed. It is unwise to publish personal contact information and unique serial numbers on the internet.
    Message Edited by michaels on 20-Sep-2008 03:25 PM

    The issue of firmware on operator-variant models is a common sore point on these boards.
    Unfortunately, Nokia's hands are tied. The operators have to approve new versions of the software before they can be applied to any phone sold to them by Nokia as their own variant.
    You will probably find that the operators' customer service staff are trained only to give advice on network services, if you want answers on issues other than that you need to locate the operator's head office and pester them. If you are a contract customer, that address should appear on your bill.

  • CS5 custom build with GTX 285 for long-form 4k projects

    I've been researching nonstop for several days, and it's time to ask for your expert opinions.  I would appreciate any help.
    I'm ordering a new system for Production Premium CS5, and mostly I will use Premiere Pro CS5 to edit long-form projects (huge timelines with thousands of clips) of Red 4k footage.  It's a feature film with only a few F/X shots, but we'd like to do as much F/X, sound mixing, and color correction in CS5 as possible.
    Here is what I'm considering:
    CUSTOM BUILD from CYBERPOWERPC (about $1950 including items not listed):
    OS:  Windows 7 Professional x64
    CPU: Intel® Core™ i7-930 2.80 GHz 8M Intel Smart Cache LGA1366
    HDD: 1TB SATA-II 3.0Gb/s 16MB Cache 7200RPM HDD (Single Hard Drive)
    MOTHERBOARD: * (3-Way SLI Support) GigaByte GA-X58A-UD5 Intel X58 Chipset SLI/CrossFireX Ultra Durable™3 Mainboard DDR3/1600 ATX Mainboard w/7.1 Dobly Audio,eSATA,Dual GbLAN, USB3.0, 2 x SATA-III RAID,IEEE1394a,4 Gen2 PCIe,2 PCIe X1 &1 PCI [+132]
    MEMORY: 12GB (2GBx6) DDR3/1600MHz Triple Channel Memory Module [+219] (Kingston HyperX)
    SOUND: Creative Labs SB X-Fi Xtreme Audio 24-BIT PCI Sound Card [+48]
    VIDEO: Nvidia GeForce GTX 285 (I will have to take out their card and install this myself for GPU acceleration)
    1.  Do you see any poor or strange decisions?
    2.  I liked that mobo because of the chipset's possibilities and because it has 1394 for old-style DV capture.  Plus it has 8 SATA 2.0's, 2 SATA 3.0's, 2. eSATA 2.0's, and 2 USB 3.0's for huge/fast storage options in the future.  Is there some better alternative if I'm keen to have at least one of each of these formats?
    3.  I'll clone their 1TB OS drive onto a smaller 250GB drive to use as the OS drive, then use their 1TB drive as part of a 2TB RAID 0 video storage.  Will using the mobo's raid controller or some type of virtual RAID (like I did in Xp64) be fine, or do I need a separate RAID controller (like I had in the late 90's).
    4.  What do you think about that processor and RAM for what I'm doing?  Any reason to prefer 2 Quad-Core's (if dual processors are even compatible with CS5)?  Please let me know if the current configuration will lag with huge timelines with thousands of 4k clips.
    5.  Maybe I don't need the Creative sound card and could just use the mobo's sound?
    6.  Will it be simple for me to remove their graphics card and install an Nvidia GeForce GTX 285 myself?
    7.  The Nvidia GeForce GTX 285 seems like a good fit for GPU acceleration for me since I'm not expecting to have 4+ layers of videos, though I will be doing multiple effects on 1 or 2 layers.  Would having a Quadro FX give any advantage over the GTX 285 for huge timelines with thousands of 4k clips?  Or any advantages for any other use?
    8.  The GTX 285 has 2 DVI's as its only connectors (plus something that looks like S-video).  Can I use one DVI for my workstation and one DVI for monitoring video in realtime on an HDMI/DVI-equipped LCD TV or professional HD video monitor.  Any problems with viewing/outputting my 4k video in HD in realtime?  What is the best setup to view your video on an external monitor?  Back in the CS2 days I used a 1394 through a camcorder into an analog professional monitor, but that's not going to be HD, plus I heard there's a sync problem these days.
    Thanks for your advice.

    Thanks for your advice, everyone! 
    I'm gaining a lot of knowledge, but I have some followups so I'll break it down into subcategories and hopefully you can chime in on one or more topics.
    To give more info about my daily use and purpose:
    --smoothly edit huge amounts of RED 4k footage/clips and view the video externally in as high a resolution as possible (2k or minimum of 1k).
    ---no reason to have more than 3 layers visible at a time.
    --the vast majority of the time only 1 video layer (with video effects) will be visible, with perhaps a handful of fades/dissolves during the entire 2-hour feature. 
    --I don't mind if it takes slightly more time rendering DPX files at the end, but I don't want any hiccups and never want to render during day-to-day editing and monitoring.
    GRAPHICS CARD and VIDEO MONITORS
    1.  I'm pretty much set on the GTX 285, but I wondered if for my purposes I'd get any significant benefit from the Quadro FX 5800 (+$2000-$2600) or the GTX 480 (+$100)?  Either in speed of playback or in output to external video monitors. The 5800 has 4GB RAM versus the GTX 285's 1GB, and the GTX 480 has 1.5 GB.   Does this RAM matter for what I'm doing or are the cores much more important?  Also, if I chose the GTX 480 then I'd have to pray that everything would be stable with a MPE/CUDA/GPU-acceleration hack, or hope Adobe comes through with full support of that card.
    2.  What is the 3rd output on the GTX 285?  There are two DVI outputs and then something that looks like S-Video.  Is it S-video out? Or an HD audio out that travels alongside the DVI output to an external video monitor?
    3. Does it slow down the system to send the video at 1k or 2k to an external video monitor since you'd be using both DVI outputs simultaneously?
    RED 4k PLAYBACK and RAM vs CPU
    1.  For day-to-day editing, I'm fine scaling to 1/4 resolution if necessary and monitoring video externally at 1k or 2k.  Would my current system already achieve that smoothly?
    2.  From what I've gathered here and elsewhere, for better 2k or 4k playback it might be better to spend extra on CPU rather than RAM.  I could spend
    --$1000 more for 24GB RAM (instead of 12GB)
    --$290 more for an i7 975 Extreme 3.33 GHz 8M Cache
    --$686 more for an i7 980x Extreme 3.33 GHz 12M cache.
    Given my needs and the current total system cost of $1950, which of these 3 would you choose?  Or should I stay with the cheaper i7 930 2.8 GHz and overclock (which I've never done before).
    ECBowen:
    It sounds like you've used a system similar to what I'm buying to playback RED 4k on a timeline in full, 1/2, and 1/4 scale.  I'd greatly appreciate if you could clarify YES/NO for which are possible to have SMOOTH PLAYBACK with your 980x/12GB RAM/GTX 285 system.  I know it's a lot to ask, but it would make me feel so much better about this big purchase:
    -full scale 4k of 1 layer with 0 video effects
    -full scale 4k of 3 layers with 0 video effects
    -full scale 4k of 1 layer with 3 video effects
    -full scale 4k of 3 layers with 3 video effects (NO, right?)
    -1/2 scale of 1 layer with 0 video effects
    -1/2 scale of 3 layers with 0 video effects
    -1/2 scale of 1 layer with 3 video effects
    -1/2 scale of 3 layers with 3 video effects
    -1/4 scale of 3 layers with 3 video effects
    RAID CONTROLLER
    1.  I won't use the virtual/software controller.  The GigaByte UD5 says it has RAID 0, 1, 5, 10.  Does that mean it has an onboard hardware controller that I can configure in Windows 7 or using some software supplied with the mobo?

  • Unknown namespace prefix Error - when using custom Aliases with RDF aliases

    I am getting unknown namespace prefix error when I use custome SEM_ALIAS with rdf SEM_ALIAS. It seems once you specify custom SEM_ALIAS, the default rdf alias is not recognized. Following are the details of queries
    I have a sem_Model "test" which has "event" as a class and "Merger" as a sub-class of event, with "acquiringorg" as a property having the domain "org". "Org" is also defined as a class with the literal attribute "hasname". I have added one instance of "merger" class with appropriate values for "acquiringorg" property. There is one instance or "Org" as well for reference in the instance above.
    the following query(return all objects having property "acquiringorg" and its .e. "hasName" attribute for the object) runs fine and returns appropriate resultset
    select x event,z acquiringorg ,l acquireeorg from table(SEM_MATCH(
    '(?x Evnt:AcquiringOrg ?y)(?k orgn:hasName ?l)',
    SEM_MODELS('test'),
    null,
    SEM_ALIASES(SEM_ALIAS('Evnt','http://www.abc.com/Event/Merger/'),
    SEM_ALIAS('orgn','http://www.abc.com/Org/')),
    null));
    However when I want to add another criteria i.e. show me events of type merger(?x rdf:type :Merger) , the query fails with the "unknown namespace prefix error" as above
    select x event,z acquiringorg ,l acquireeorg from table(SEM_MATCH(
    '(?x rdf:type :Merger)(?x Evnt:AcquiringOrg ?y)(?k orgn:hasName ?l)',
    SEM_MODELS('test'),
    null,
    SEM_ALIASES(SEM_ALIAS('Evnt','http://www.abc.com/Event/Merger/'),
    SEM_ALIAS('orgn','http://www.abc.com/Org/')),
    null));
    specifying "rdf" ALIAS explicitly also does not work
    select x event,z acquiringorg ,l acquireeorg from table(SEM_MATCH(
    '(?x rdf:type :Merger)(?x Evnt:AcquiringOrg ?y)(?k orgn:hasName ?l)',
    SEM_MODELS('test'),
    null,
    SEM_ALIASES(_SEM_ALIAS('rdf', 'http://www.w3.org/1999/02/22-rdf-syntax-ns#'),_
    SEM_ALIAS('Evnt','http://www.abc.com/Event/Merger/'),
    SEM_ALIAS('orgn','http://www.abc.com/Org/')),
    null));
    I am unable to figure out why default namespace i.e. rdf is returning this error
    Stuck at this point badly!!Any pointers would be useful!!

    Full error details are as below
    ORA-29532: Java call terminated by uncaught Java exception: oracle.spatial.rdf.server.ParseException: Unknown namespace prefix ''
    ORA-06512: at "MDSYS.RDF_MATCH_IMPL_T", line 153
    ORA-06512: at "MDSYS.RDF_MATCH_IMPL_T", line 842
    ORA-06512: at "MDSYS.RDF_MATCH_IMPL_T", line 235
    ORA-06512: at line 1
    29532. 00000 - "Java call terminated by uncaught Java exception: %s"
    *Cause:    A Java exception or error was signaled and could not be
    resolved by the Java code.
    *Action:   Modify Java code, if this behavior is not intended.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   

  • Only Bridge CS 5 will open with PS CS 6

    I installed PS CS 6 without first deleting PS CS 5 first. I was told I did not need to. Anyway now only Bridge CS 5 will open with PS CS 6, not Bridge CS 6.  Bridge CS 6 is loaded. I can access Bridge CS 6 through the Mini Bridge which doesn't work well for me. I need the full Bridge CS 6 to open, not Bridge CS 5. Can anyone help ?

    Both programs are seperate so no need to delete previous version.
    You probably have Bridge CS5 running in the background.  Only one version can be running.
    Open task manager and in processes look for Bridge.exe and close.
    You might have Bridge CS5 set to open at login so you should change this in advanced preferences.

  • 9237 + full-bridge load cell: load cell_null_off_shuntcal.vi - error 200077

    I'm trying to use 
    load_cell_null_off_shuntcal.vi  with load cell (Honeywell Model 31,
    unamplified). I am using LabView 8.6,  cDAQ-9172 and NI9237. Inputs:
    internal excitation10V; mV/V 2.1492 (calib. sheet); max weight 10 lbs; bridge
    resistance 350 ohms (Honeywell specs); 9237 internal shunt resistance 100
    kohms; shunt location R4 (default setting). Have selected "Do offset
    null" and "Do shunt cal".
    Error -200077 occurred at DAQmx
    Perform Shunt Calibration
    (Bridge).vi:1 Possible reason(s):
    Measurements: Requested value is not
    a supported value for
    this property.
     Property:
    AI.Bridge.ShuntCal.GainAdjust
    You Have Requested: -61.980405e3
    Valid Values Begin with: 500.0e-3
    Valid Values End with: 1.500000
    If "Do shunt cal" green
    button not selected, no error. Gain adjust should be approx 1. subVI 
    DAQmx PerformShuntCalibration (bridge).vi contains "Call library function
    node" which is locked (?).  
    Any ideas?
    What is the correct shunt element
    location for a full bridge load cell? Changing this location does not eliminate
    error.
    Solved!
    Go to Solution.

    Hello CFJ,
    The problem is most likely in your external connections of the NI 9237 and the load cell.  As referenced in the NI 9237 Operating Instructions and Specifications, page 9, the SC+ and SC- pins should be connected across the resistor specified in the DAQmx Perform Shunt Calibration (Bridge).vi (in the case of a full bridge it would be R3).
    Let me know if you are still having issues with your calibration.
    Message Edited by Dan_K on 01-22-2009 04:31 PM
    Regards,
    Dan King

  • Can only get the "Draft" stamp when insert customer stamp with C#+JS

    I am trying to insert customer stamp in batch files with C#+JS.
    My JS works well in Acrobat. However, it doesn't work when use C#+JS. I always get the Acrobat's "Draft" stamp other than my own ones (the AP value is confirmed correct).
    In c#, it runs as this ( with the objectype.InvokeMember())
    1. initial an annotation with addAnnot first and get the properties object with method of getProp().
    2. Set the type entry of the properties object as "Stamp"
    3. With setProp(), modify the annotation to a Stamp annotation, and then get the properties object of the Stamp annotation.
    4. Set the entries (rect, page, AP ) of the properties object. Then use setProp() again to modify the properties of the Stamp annotation.
    The program runs without any error message. But I can only get the default stamp "Draft".
    One thing I found so far, in JS console in Acrobat, if I set the AP properties AFTER set the annotation type to "Stamp", it also get the "Draft" stamp only. as this,
         annot = this.addAnnot();
         prop = annot.getProps();
         prop.type = "Stamp";
         annot.setProps(prop);
         prop = annot.getProps();
         prop.page = 0;
         prop.rect = [0, 0, 100, 100];
         prop.AP = "#xxxxxxxx";
         annot.setProps(prop);
    If I set the prop.Ap before the forth line in above para, I can get the right stamp.
         annot = this.addAnnot();
         prop = annot.getProps();
         prop.type = "Stamp";
         prop.AP = "#xxxxxxxx";
         annot.setProps(prop);
         prop = annot.getProps();
         prop.page = 0;
         prop.rect = [0, 0, 100, 100];
         annot.setProps(prop);
    Or, if I set the props in the typcial JS format like below, I also get the right one
    annot = this.addAnnot({
    type: stamp,
    page: 0,
    rect: [0,0,100,100]
    AP: #xxxxxxxx)});
    But, the problem is, in C#, the AP properties can only be set after set the annotation type as Stamp (step 4). I think this might be the reason, but I don't know how to get over this.
    Please help. Thanks!

    No, it can't – but you could do that yourself as part of the watermarking process (ie. Two watermarks or fields)
    From: santa-satan <[email protected]<mailto:[email protected]>>
    Reply-To: "[email protected]<mailto:[email protected]>" <[email protected]<mailto:[email protected]>>
    Date: Mon, 6 Feb 2012 18:35:19 -0800
    To: Leonard Rosenthol <[email protected]<mailto:[email protected]>>
    Subject: Can only get the "Draft" stamp when insert customer stamp with C#+JS
    Re: Can only get the "Draft" stamp when insert customer stamp with C#+JS
    created by santa-satan<http://forums.adobe.com/people/santa-satan> in Acrobat SDK - View the full discussion<http://forums.adobe.com/message/4190155#4190155

  • Customer Statement with opening and closing balances

    Dear Forum,
    The users want to generate the Customer Statement with opening and closing balances like the traditional one. The statement now generated gives the list of all open items as on date, but the users want the statement with opening balances as on the date of the begining of the range specified and the closing balance at the end of the period for which the statement is generated. Is there a way to generate the same from the system.
    Thanks for the help.
    Regards,

    Hi,
    SPRO> Financial Accounting (New) > Accounts Receivable and Accounts Payable > Customer Accounts > Line Items > Correspondence > Make and Check Settings for Correspondence
    You can use the program RFKORD10 with correspondance type SAP06 for your company code
    This program prints account statements and open items lists for customers and vendors in letter form. For account statements, all postings between two key dates, as well as the opening and closing balance, are listed.
    Regards,
    Gaurav

Maybe you are looking for

  • Mouse System Preferences not saved

    I´m using a MacMini with the latest updtes and only apple harware, suddenly the mouse stoped puping up the dashboard, when I go to system preferences the scroll botton is not asigned to the dashboard, a click on it and then it works fine as long as i

  • Preview doesn't recognize CR2 files

    Hello, I had recently bought an iMac but I'm unable to open cr2 files when I'm trying to do a double click to open in Yosemite. Although I can open them in Lightroom 5.7 and could not in CS6(understand I need to upgrade it). I know there should be a

  • Two important problems with GWIA

    I have two major problems in GroupWise 8 SP2 with the latest HP. First, something I arose at another facility with HP2 and that in this new server with HP3 still happens, the file is lost gwac.db every 5 days at 0 hours exactly. Something I could fix

  • Using xgrid - shared file system possible?

    I have a question about running a sequence of unix commands with xgrid. I just started fiddling with xgrid and I am already able to run a sequence of renderings in parallel. This is cool and relatively easy. The software I use for rendering is called

  • Feature Request: Apply effects to group of clip instances

    I have been trying to supply this feature request into the Adobe feedback form. But it appeared that it was too large to be posted in a single feedback form. So I had to split it into three distinct feedback forms. I don't want this feature request t