LV6.1/GPIB-PCII/Keithley 617

hi,
just found some old stuff i wanted to try out while getting to know
lv. gpib device works ok, although WinMe(card not happy with NT/2K)
did not assign DAK/IRQ, I changed this and it appears to be working
with them as well.
when i tried to connect the 617 i did find a device that did respond,
but when i dl'd the 617 drivers all i get is a unknown VISA device?
anyone got any experience with this? i could not find anything on the
subject in the lv manuals.
Thor Anders

On Fri, 11 Jan 2002 07:12:41 -0800 (PST), Dennis Knutson
wrote:
>I'm a liitle confused. Do you get the error in LabVIEW or in MAX? The
>driver is for LabVIEW to communicate to the instrument. LabVIEW uses
>the GPIB card configured in MAX to talk to the instrument via the VISA
>api or direct GPIB calls. If the card is correctly configued in MAX,
>you should be able to open Devices and Interfaces, right click on the
>GPIB card and select Scan for Instruments. Any instrument found, you
>should be able to use MAX to read and write to. Also, what version of
>MAX and NI.488.2 are you using? MAX 2.1 and NI.488.2 doesn't list the
>GPIB-PCII card as supported. If you're using an older version, you
>might be having a compata
bility problem with either VISA or GPIB
>driver and Win/ME. What did you use to successfully communicate with
>the one instrument?
ok, from the top:
I have LabView 6.0 installed, running MAX 2.0.3.6 and NI-488.2 driver
1.60 on Windows ME.
I installed the GPIB-PCII and windows installed it at a certain IO
adress. I set the jumpers for this IO and when I then scanned for the
instrument I found 'something that I could communicate with.'
I figured the GPIB should use IRQ and DMA so I reinstalled the card
until MAX were satisfied with it using DMA and IRQ.
I then created a ke617 folder under instr.lib and put the ke617.lib
and dir.mnu files there.
now, in MAX, i cannot find an instrument when I scan for it, and the
'unknown VISA resource' pops up.
I'll try to upgrade max and the NI-488-2 driver and se what happens.
Thor A

Similar Messages

  • Labview VI's for keithley 617 meter

    Respected Sir,
    Now I m trying  to interface keithley 617 electrometer using Labview and GPIB card. For that we have created code. Using this we want to measure continuous readings form 617.
    We are getting reading from 617 meter. But we want to measure current for that we have connected 1 kilo ohm resistor to the device using BNC cable(According to Keithley manual) voltage range is -1 to 1 increment  of the order of .1 means after -1 automatically labview must take -.9 next -.8 next -.7 till up to 1 after that it must stop collecting data.
    I m going to attach my code. Please try to solve according to our need.
    With regards
    Shridhar
    Attachments:
    s1Keithley 617.vi ‏82 KB

    Hi Shridhar,
    Good afternoon and I hope your well today.
    Are you using the VIs which I found here, at ni.com/ident
    Keithley ke617 Meter
    http://sine.ni.com/apps/utf8/niid_web_display.download_page?p_id_guid=E3B19B3E94A2659CE034080020E748...
    The VI titled  Keithley 617 array.VI setups/reads/close the take for 100 measurements - for one setting.
    Now I have seen for other instrument drivers they have coded a sweep - which is what you're after.
    The process to achieve this is loop through this code, but changing the voltage range and hence reading back different data. Have you got code that updates the voltage range? 
    I am sorry but I can't code in LabVIEW 5, 
    Hope this helps, 
    Kind Regards
    James Hillman
    Applications Engineer 2008 to 2009 National Instruments UK & Ireland
    Loughborough University UK - 2006 to 2011
    Remember Kudos those who help!

  • GPIB-PCII software for Windows 2000

    My system is running Windows 2000, but the NI-488.2 software does not support my GPIB-PCII board under Windows 2000. Where can I find softward that will support my board under Windows 2000?

    These links provide additional information on GPIB-PCII/IIA board. I'm facing the same problem but I don't want to part away with the boards yet so I continue to use them with Windows 95.
    http://digital.ni.com/public.nsf/3efedde4322fef19862567740067f3cc/2499c1cf1322e98e8625659500026d0b?OpenDocument
    http://www.ni.com/support/gpib/versions.htm

  • MAX for GPIB/PCIIA card

    Why MAX 3.0 can not see GPIB/PCIIA card?
    What software allow me to see it?

    Hi,
    You need the NI-488.2 driver to detect the boards. MAX is just a configuration utility, the driver detects and configures the device.
    You may have NI-488.2 installed, but this is an old GPIB interface, so it is not supported in the newest version of the driver. You can find the available driver version for at this page
    DiegoF
    National Instruments.

  • Can I install a GPIB-PCII/IIA Card into a windows XP computer and if so how?

    Can I install a GPIB-PCII/IIA ISA Card into a windows XP computer? If this is possible can you give me the stsps needed to do so.

    Hello,
    No, this is not possible. This "old" board isn't supported under Windows XP. Check this site for an overview: http://www.ni.com/support/gpib/versions.htm#pc
    Regards,
    Luca
    Regards,
    Luca

  • Driver W3.1 for GPIB - PCII/IIA

    It's not possible to download on this web site the driver W3.1 for GPIB - PCII / IIA. If someone have this "antic" version thanks.

    Hi,
    You should be able to download the driver at http://www.ni.com/support/gpib/versions.htm#pc_dos. Did you find a link to the driver elsewhere that was broken?
    Best Regards,
    Aaron K.
    Application Engineer
    National Instruments

  • GPIB-PCII/IIA Drivers for Windows 98

    Hello , Could someone can help me ? I got a trouble in GPIB-PCII/IIA card.
    I want to driver it in Windows 98 , but Windows hardware manager can not find this card.
    Could someone can tell me how to do for complete step .
    Thanks a lot !!

    大家好,我现在遇到了问题了,希望大家可以帮帮我。
        我最近在做轻集料混凝土的交流阻抗谱测试,我先叙述一下遇到的问题:
       我的电脑是windows98、仪器是PARC公司的M273A/锁相5210,所用GPIB卡为:GPIB PCIIA IEEE488.2,测试软件为M398,程序时在DOS下运行的。运行Eis.bat程序run时,出现
         potentiostat not found,
         Dev_poll 1 error caused error in gpib_ error _check
    32768 _0
         Dev_poll 3 error caused error in gpib_error_check
    32768 _0
         当运行M398.exe时,提示:
    ibfind board 1caused error in gpib_error_check 32768_7
        gpib.com improperly configured,
         gpib card not installed,
    GSS is either not loaded or improperly installed.
    ABORTED:Error Level 1,Error Code 8
         请问是驱动问题还是GPIB卡出了问题?前几次也出现过类似的问题,当时关闭电脑,关闭仪器,过一会重新开机就正常了,今天有出现了类似的问题,可是关机,关闭程序之后再开机就不再正常工作了,
          还有好多次测试快运行完在读取数据时程序突然就没有反应了,只能强制关闭程序,之后重新启动就再不能正常运行了,出现和上面一样的问题。
    期待大家的帮助!

  • Win2000 support for GPIB-PCII/IIA

    Is there a driver that allows to use the:
    GPIB-PCII/IIA (rev.5)
    with Win2000 or must I downgrade to win98 and use the drivers present in NI webpage.
    Thanks in advance.

    ThomasD wrote in message news:<[email protected]>...
    > Hello,
    >
    > the GPIB-PCII/IIA is not supported under Win 2000. You can also see
    > this on the following link:
    > http://digital.ni.com/public.nsf/websearch/0AE0DAD0C1142EF8862565230058DAE2?OpenDocument
    > So, you have to downgrade your system to Win 9x and download the
    > driver on this page:
    > http://www.ni.com/support/gpib/versions.htm
    >
    I feared the worst...back to instability and blue screen.
    Thank you!

  • Gpib pcii

    need to use under win2000. card not found. any emulation or drivers available or do I have to revert back to an old win98 machine. I have several cards and this seem a poor choice for upgradability.

    Hi,
    The GPIB-PCII is an obselete product. The newer drivers will not support this board. You can find the supported GPIB hardware vs OS and the appropriate driver version to use, in the following page:
    http://www.ni.com/support/gpib/versions.htm
    DiegoF

  • GPIB Instument- keithley 1753

    Hi!
    I want to comand my Kenthley 1753 multimeter by GPIB, but afther the GPIB initialization, when I use the "GPIB Status" VI, I receive the GPIB error #1 i.e. "Command Requires GPIB Controller to be CIC" but it also tell me that the CIC stutus is True!
    What may I do?
    thank you
    eddy

    Check this link:
    http://digital.ni.com/public.nsf/websearch/1CA1670830A3C59686256341007ADD5A?OpenDocument
    you should resolve your problem.
    Best regards
    GIUNA

  • QuickBasic GPIB PCII Help

    Hi-
    I'm new to QuickBasic and the world of GPIB but I'm not new to programming. I need help on something I'm completely stumped on. I have a feeling this is an easy and quick fix. I'll try to provide as much information as possible.
    I have a Windows 98 SE (yuck!) machine with a NI GPIB ISA/AT card in it. Connected to the card is a Sony RTD710 Digitizer. I created a QuickBasic program with the code that came in the manual and it doesn't seem to be working. The only thing it does is return the ID of the digitizer. However, when I use the Interactive Control (in the Measurement and Automation Explorer), I can successfully communicate with the machine.
    Here is my code:
    REM $INCLUDE: 'QBDECL.BAS'
    DIM READING AS STRING * 30
    CALL IBDEV(0, 7, 0, 10, 1, 0, RTD710%)
    CALL IBCLR(RTD710%)
    CALL IBWRT(RTD710%, "ID?")
    CALL IBRD(RTD710%, READING$)
    PRINT READING$
    CALL IBONL(RTD710%, 0)
    END
    This is what I'm doing in the Interactive Control:
    ibdev 0, 7, 0, 10, 1, 0
    ibclr
    ibwrt "ID?"
    ibrd 50
    ibonl 0
    The QuickBasic code compiles with no errors. The variable RTD710% returns as -1 and the variable READING$ does not return anything. In the Interactive Control, the variables are inputed one by one when prompted. When calling the ibrd function, the ID of the machine is returned "SONY TK/RTD710 ....." as it should.
    I also noticed that when I used the NI Spy program, I can see the functions being logged when using Interactive Control but I don't see the functions being logged with the compiled QuickBasic program.
    I followed the instructions here: http://sine.ni.com/apps/we/niepd_web_display.display_epd4?p_guid=B45EACE3DE7556A4E034080020E74861&p_node=DZ52353&p_source=external
    Again, I'm very new to GPIB and I tried searching the forums to see if this topic has been covered before. But I don't know what to search for. So, I apologize if this has come up before. Any help would be greatly appreciated.

    You mentioned that you are new to QuickBasic, but QuickBasic is a very old application that dates back to the days of DOS.
    You didn't happen to mention which version of NI-488.2 that you have installed on Win 98. From the problem you describe, it seems as if you have a Win 32 version installed. With a Win 32 version of the driver, by default, it does not have DOS support enabled (for applications such as QuickBasic). This type of problem is described in:
    http://digital.ni.com/public.nsf/allkb/D8AFF54234E2788486256A760070FB2E
    If this is the case, it is very easy to enable DOS support with the Win 32 driver. Please refer to this KnowledgeBase article:
    http://digital.ni.com/public.nsf/3efedde4322fef19862567740067f3cc/c293f23d6b87992c86256ab6000afbeb?O...
    Hope this information helps. If not, please report back with additional information about which NI-488.2 driver you have installed.

  • Whereis the gpib pcii/iia driver for windows 98 ?or 95

    ok

    Jo18,
    Visit the following link: http://www.ni.com/support/gpib/versions.htm
    Hope this helps,
    L Aguila
    Applications Engineer
    National Instruments

  • Control a Keithley 3706 with USB or GPIB ?

    Hello,
    I have to control a Keithley 3706 with Labview: I have to realize an automated test equipment. The Keithley is connected to the computer by USB port but on forums I can see people who use GPIB port. What are the differences ? Can I still use the USB port on the instruments to realize the automated test equipment ?
    Best regards
    Solved!
    Go to Solution.

    Hi Elohim,
    most people will not use GPIB functions, but VISA functions!
    You can use the very same device driver even when using USB as communication layer instead of GPIB (providing Keithley did a good job)…
    What's wrong with the driver you have installed already?
    Best regards,
    GerdW
    CLAD, using 2009SP1 + LV2011SP1 + LV2014SP1 on WinXP+Win7+cRIO
    Kudos are welcome

  • New and need help with keithley 2700 GPIB threw labview 2010

    Hello everyone,
    I have the drivers for the keithley 2700 dmm and its ok with labview idntifying the device.
    but when im trying to run the "Keithley 27XX Single Measurement.vi" and take a single resistance measurment
    I always get this error, (pic related)
    can anyone help please?
    btw, where can i start learn from zero how to program in labview in GPIB protocol (not visa) ?
    thanks alot.
    E.
    Attachments:
    Error.jpg ‏500 KB

    Zerez,
    Definately search the forums as I found another forum wher they tried troubleshooting a 2700 that was locking up. Unfortunately a resolution wasn't determined. 
    http://forums.ni.com/t5/Instrument-Control-GPIB-Serial/keithley-2700-7700-4-wire-Measurement-GPIB/td...
    Can you use max to request values from the device? Using visa test panels. It should reply to the identification command if it is functioning properly.
    Kyle Hartley
    RIO Product Support Engineer
    National Instruments

  • I have problems with pci-gpib card under windows me, What I need to know what to do in order to fix this problems?

    Hi, I have installed a pci-gpib card in my computer, I have a problem with
    installation, the message of troubleshooting wizard is as follow:
    "One or more missing GPIB interfaces.
    There are several reasons why you might encounter this problem:
    Reason: Your GPIB interface is incorrectly installed.
    Solution: Please verify that your GPIB interface is correctly installed
    and rerun the NI-488.2 Troubleshooting Wizard.
    Reason: You are using a non-Plug-and-Play interface (e.g., AT-GPIB/TNT,
    GPIB-PCII/IIA, GPIB-ENET) and you did not use the Add New Hardware Wizard
    to add the interface to your system properly.
    Solution: Click th
    e Add Hardware button and the Add GPIB Hardware wizard
    will guide you through the process of adding GPIB hardware.
    Reason: For the GPIB-USB and GPIB-1394, the host controller (USB or 1394)
    must be working properly in order to use the GPIB interface.
    Solution: Click on the System Details button for details about the USB and
    1394 host controllers.
    Reason: For the PCMCIA-GPIB and PCMCIA-GPIB+, the PCMCIA host controller
    must be working properly in order to use the GPIB interface.
    Solution: Use the Windows Device Manager to verify that your PCMCIA host
    controller is functioning correctly. Start the control panel by selecting
    "Start>>Settings>>Control Panel", open the System icon and select the
    Device Manager page. If you select "View devices by type", you can look at
    the PCMCIA socket properties to ensure it is correctly enabled."
    And on system details the message is the next:
    "USB Driver type: USB universal host controller"
    "This version of windows does not su
    pport IEEE 1394."
    I am using Windows Me, so i want to know what i need to do to fix this
    problems.
    Best regards.

    If you have a PCI-GPIB card, you needn't worry about the USB port or whether or not Windows ME supports 1394. All you need in NI-488.2 version 1.70 from National Instruments. Install this driver and then insert your card. Windows should detect the card and find the proper driver.
    If this did not happen in the proper sequence, look in the device manager and see if the card was installed as an "other device". If this happens, remove the card in the device manager and have windows redetect it. If this is not a solution, you may consider the ask an engineer page at www.ni.com/ask. Provide as many details as possible.

Maybe you are looking for

  • Error while Compiling

    hi guys. this is my first post on the forum. im a newbee to java and need help with an GUI app that i wrote. i cant get it to complie. public class ButtonTest extends javax.swing.JFrame // extend a frame so that it is in a window      public ButtonTe

  • Org Unit Asssignment for Activities

    Hi,     We have recently upgraded from CRM 4.0 --> 5.0 SP 11, for certain individuals who now create data in the mobile applicatio and conntrans it ,  the Organizational Unit is not being assigned to the activities that they are creating. The Sales O

  • Can anyone explain why a window pops up with my IP configuration

    every few minutes. When it does this I have to click on OK, then wait while it reconnects so that I can send or contact a website. It is driving me crazy, I am using an airport Express base station through a D_link router modem. I want to add a pc to

  • Default Foreign trade header data

    Dear Lakshmipathi, I want foreign trade header data by default while creating Export Invoice, for that I did necessary customization & maintained Default Values For Foreign Trade Header Data. I also used SAP enhancement V50EPROP to process foreign tr

  • Can anyone tell if I have an error in XML?

    Our feed has not updated for a couple weeks... validator only (wrongly) shows a need for "explicit" label, for which Dr. Edlund's Fitcast does not qualify. Is there something wrong with our XML file: http://www.doctoredlund.com/vodcast/DrMEfitcast.xm