Trans-receiver system using PXI

Hi,
I need to build a trans-receiver system using PXI. For the transmission part I can use a PXI 5652 to generate a single tone continuous waveform. 
For the receiver part, I need an antenna to capture the signal transmitted by 5652 (which will have a transmission antenna ). The received signal will then be down converted by PXI 5601 (RF OUT of 5652 is connected to LO IN of 5601). The intermediate frequency will be then digitized by PXI 5622 (IF OUT of 5601 is connected to IF IN of 5622). 
But my question is these connections are already there between the slots/chassis.(The next stage has I/Q demodulation but I want to get clear on this part first). But there is not output port on the PXI 5622, so how can I check the digitized output?
 I have attached a file for reference.
Thanks,
Sharmi
Attachments:
Ref.docx ‏689 KB

Hi Sharmi,
To address your questions:
Is the data, after it is down converted and digitized by the 5622 is automatically sent to 5450?
No, the data is not automatically sent to the 5450.  The digitized data is sent to the backplane and can be accessed on your controller/laptop through LabVIEW or the RFSA Soft Front Panel.  The 5450 is used to take digital signals generated by LabVIEW and turn them into an analog signal.  It is not used in any part of the downconversion process.  
If you want to take the data that you receive from your 5663 VSA and send it to your 5673 VSG, you will need to do so using software.  There are RFSA and RFSG examples that ship with the driver that you can use as a starting point on how to acquire a signal and send a signal using LabVIEW.  These examples can be found by going to Start » All Programs » National Instruments » NI-RFSA (or NI-RFSG for generation examples) » Examples.
5450 has “400 MS/s I/Q Signal Generator” written on it, so does it generate I and Q channel data separately?
The 5450 does generate IQ data on 2 separate channels.  LabVIEW is used to generate an IQ bit stream.  The IQ data is combined after being upconverted with the 5611.
Also 5611 has “I/Q Vector Modulator” written. So, can it demodulate a signal (after it is split into I and Q components by 5450) and then combine the resultant output?
The 5611 is an upconverter and is used to take the IQ signals generated by the 5450, combine them, upconvert them from some IF frequency to an RF frequency and transmit them.  It can be used to do some basic types of modulation, but not demodulation.  The NI RF Signal Generators Help ( Start » All Programs » National Instruments » NI-RFSG » Documentation) has a section about the 5611 that includes the hardware block diagram and a more detailed explanation about each part of the block diagram.  This is where you can find out more about what basic types of modulation the 5611 is capable of performing.
Let me know if you have more questions, I'd be happy to help.
Sarah Y
SDR Product Manager
National Instruments | Ettus Research

Similar Messages

  • Same idoc to multiple EDI receiver system using rule

    Dear Friends!
      Good day!  I am using desadv.delvry03 idoc and I want to send same idoc from same communication component to multiple receiver parties without mapping.
    I have created first scenario successfully using idoc, when I create another scenario it gives me message like "Object already exist" .  for that I have to use Service interface.  I dont want to use service interface for this purpose.  In my scenarios mapping is not even required. one idoc to and multiple receivers thats it.
    Can you please guide me how to do the above requirement ?
    your any help will be appreciated greatly
    Thanking you
    Regards
    Dipen

    Hi,
    For this... U can create 1 sender agreent for that idoc if necesseary....otherwise we dont required IDOC sender Agreement.
    Then Create 1 reciver determination in that u can mention the 2 recivers details...
    same thing for interface determination also.
    and finally create 2 reciver agreement and 2 reciver communication channels for 2 systems.
    Let me know if you require any further information.
    Thanks,
    Sreenivas

  • Synchroniz​e 2 PXI Systems through GPS using PXI 6552

    I want to synchronize two PXI systems separated by a distance of 5kms through GPS.
    PXI 6682 is a timing and synchronization card which can be used to achieve this. I also have a requirement to use PXI 6552 HSDIO in my application. Few channels in PXI 6552 are free. So i want to know if it is possible to achieve GPS Synchronization of the two PXI systems using PXI 6552.
    Regards,
    Sundar Ganesh

    Hi Ryan,
    This is a system configuration which i am doing for a requirement. The requirement is RF Generation in one PXI system and RF Acquisition in another PXI system separated by a distance of 2kms. The applications that will be running at both these PXI systems may be started at different times. So i am thinking of using a PXI card in both the chassis to synchronize them through GPS.
    Regarding this, i have already contacted the local NI FSE and together with him only we have decided to go ahead with PXI 6552 for GPS Synchronization. But i started doubting the usage of 6552 for GPS synchronization when i couldnt see the card in the list of Timing and Synchronization cards.
    I have attached a schematic of the configuration. Let me know which PXI card should be used for synchronising these two systems.
    Regards,
    Sundar
    Attachments:
    PXI Configuration.JPG ‏21 KB

  • Multiple data acquisition using PXI-5124 digitizer

    I have a system in which I do data acquisition of the present state of the physical system (using PXI-5124 digitizer),
    give the system some change (using labview program itself) and do data data acquisition again corresponding to 
    the changed state of the system.
    So I need to do two data acquisitions in one loop. How do I do this ? Labview examples have some programs (attached file) that
    does one data acquisition in one loop but I need to do two.
    I tried to make some changes to this program but it doesn't seem to work well.
    Thanks
    Attachments:
    PXI5124-data-acquisition.vi ‏46 KB

    From the explanation you have given about reconfiguring the digitizer,
    I think the example you are looking at is the one you want.
    The digitizer is reconfigured with new settings every time with in that loop,
    so the user can change the settings such as triggers and vertical range on the fly.
    The original state of the example showed how it was reconfiguring the digitizer,
    was there something else you wanted to add  to that example?

  • How to implement a dual booting PXI system using windows 7 and labview RT

    Good Day
    My name is Mariano Ocana from the University of Manitoba, and I want to implement a dual system using windows 7 and labview RT in a PXIe-1082 chassis with a PXIe-8102 embedded controlled (the chassis came with windows 7 installed). What are the steps to use both systems (windows and RT)  in the same chassis?, in the NI help comes an example implementing a dual system using windows xp and RT, it is the same procedure?.
    Thank you in advance for all your help.
    Regards
    M. Ocana

    Greetings, Mariano.
    The real issue you have to contend with is whether or not you have a LabVIEW RT license.  If the 8102 controller was purchased Windows-Only, you likely do not have a license to run LabVIEW RT on the controller - you can purchase one, I think it's about $500, and at that point you can request a LabVIEW RT Configuration CD (that's not what it's called, but the sales guy will know what I'm talking about).  The Config CD is a CD that will partition your controller correctly with both FAT32 and NTFS filesystems (in the appropriate way) to allow you to run Windows 7 and LabVIEW Real-Time on the controller correctly.  Once done, you can then switch within the BIOS between LabVIEW Real-Time and Windows in order to dual-boot.  
    Either way, you likely need to contact NI sales to either purchase an RT license (if one was not received at the time of purchase) or to request a Configuration CD (it might also be called a dual-boot install CD or similar).
    -Danny

  • Passing Changed  extended fields using ALE-IDOC to receiving system

    Hi all,
    Please tell me how passing changed  extended fields using ALE-IDOC to receiving system can be achieved. Is it possible by Chnage pointer concept.Do i need to do some coading after data enters into receiving system as it is required while creating the entries in database table.
    For ex. in matmas05 extended for classification and inspection view and then created using 'CREATION'
    related bapi.Do i also need to incorporate the changes by change related bapi or ir can be handled by change pointer concept.
    Thanks.

    Hi Sanu
    1) Please tell me how passing changed extended fields using ALE-IDOC to receiving system can be achieved. Is it possible by Chnage pointer concept.
    => Looks like you have extended standard IDOC here. so you need to find User Exit or BADI for the Push program to populate the additional fields. Yes, it is possible thru change pointers as well.
    1) Do i need to do some coading after data enters into receiving system as it is required while creating the entries in database table.
    => Yes, as suggested in 1 try finding User Exits/BADIs.
    Regards
    Shital

  • ALE using HRMDA message type - Comit not perform in receiving system

    Hi,
    I want to use ALE to transfer HR master data between two ECC instances. I have set-up all the configuration; distribution profile, filter etc... I tried to transfer the master data using standard SAP report RHALEINI.
    The iDoc is generated and transfer correctly in the receiving system. The final status of the iDoc is 53 - Application document posted. The issue we have is that the comit to tables is not happening. Infotypes remains empty even if the status of the iDoc says it was posted. We are trying to transfer infotype 0000, 0002, 0001 and 0006
    Anyone can help ?....
    Thanks
    Louis

    Hi Ted,
    Thanks for your answer,
    Autorization is the first thing that we consider. We generate the IDOC by the program RBDMIDOC .
    The job's user is the sap_all authorization. So it should not related with auth.
    And this does not always occur. For example, two days ago it occurs, but today does not occur.
    Best Regards
    Hasan.

  • How to Send One file to different Receiver Systems without using ESR Object

    Hi Gurus
    I have developed two interafces in my current project and they are both By pass scenarios. My requirement was to Send Data from MDM system to BW system. This development does not involve any ESR objects, By Pass sceanrio.
    Now there is the requirement to Send one file or files from MDM system to different BW( two BW systems for different regions)
    How can i accomplish this.
    here we are not using any ESR Objects, so to maintain any condition in Receiver Determination is not possible.
    Please let me know how to do this.
    Tahnx in advance.

    Hi Arjun,
    If i understand your situation correctly, You need to define two separate Business Systems/Components in SLD for each BW system.
    if you intend to send the file from MDM to both the BW systems, then you need to configure the following:
    1. If you want to follow the classical approach (for this by-pass scenario):
    a) Two receiver channels - One for each BW system
    b) Two receiver agreements for each channel
    c) One Receiver determination - with both the receiver systems added as receivers (w/o condition)
    d) Two interface determination - Dummy mappings as it is a by-pass scenario
    2. If yo want to go ahead with the Integrated Configuration (most preferred in this case):
    a)  Two receiver channels - One for each BW system
    b) One Integrated Configuration Scenario (ICO) with both the receivers configured in Outbound Processing tab
    Let us know if you were looking for something else (as your problem statement is not very clear)
    Hope this helps.
    Regards, Gaurav

  • Measurment through a vibration sensor interfaced with Ni-PXI is very low & less responsive. I wana know any frequency limitations is there for PXi or the system uses any clipping ckts inside ???

    Measurment through a vibration sensor interfaced with Ni-PXI is very low & less responsive. I wana know any frequency limitations is there for PXi or the system uses any clipping ckts inside ???

    It would be helpfull to know what PXI hardware and what sensor you are using.
    Up to now I only found mechanical limitations ....
    (Creating a 100kHz sinus exitation with 160nm amplitude need more than 63km/s² (>6400g) acceleration  )
    Greetings from Germany
    Henrik
    LV since v3.1
    “ground” is a convenient fantasy
    '˙˙˙˙uıɐƃɐ lɐıp puɐ °06 ǝuoɥd ɹnoʎ uɹnʇ ǝsɐǝld 'ʎɹɐuıƃɐɯı sı pǝlɐıp ǝʌɐɥ noʎ ɹǝqɯnu ǝɥʇ'

  • Using PXI 7352 Motion card and 6025E DAQ

    Dear Sir,
    I am using SCXI 1520 along with PXI 1011 & E Series DAQ (6025E) to measure the Load Value of a load cell. I am also using PXI 7352 Motion card to control the velocity of a Plunger. Now my requirement is to stop the movement of the Plunger depending on the value I read from the load cell. I am not able to achieve the stoping accuracy for high speeds (around 150mm/sec) in the software. Can it be possible with the use of Motion Control card and RTSI signals? If so, how can I do that? (I am not used any limit switches as the load value at which the Plunger should stop is user programmable and I am using DAQ card in Multiplexed mode because i am reading more number of analog inputs with only one DAQ card)
    Ramkumar. D
    Soliton Automation,
    India.

    Hello.
    Based on your post I am guessing that you are performing the control in software, i.e. starting a velocity move, and then reading the load cell until it reaches a threshold, then stopping the motion. As such you are essentially limited by the software control loop rates.
    You are using PXI so it is possible that you are using real-time. If not, one solution could be to implement a real-time application that would be more deterministic, but again this would still depend on the software.
    Depending on your system (stepper or servo), you could implement a direct analog feedback into the motion card itself, but this would require a signal which has a range much greater than your load cells. This is only a good option for servo axes. If this was somehow possible, you can tell the motion controller to move to a specified LOAD instead of a specified position/velocity. Provided you could appropriately condition the signal, this is the best choice.
    Unfortunately, you cannot implement limit switches/etc over RTSI, so I believe that this will not be a good option.
    A third, and probably most un-desireable option, is to output a voltage on the PXI-6025E proportional to the load received from the load cell. This voltage could then be used as the analog feedback for your motion device. This, however, still adds the software layer of delays.
    I hope these ideas help you out.
    Regards,
    Colin C.
    Applications Engineering
    Colin Christofferson
    Community Web Marketing
    Blog

  • How to use PXI-8532 in LabVIEW Development Environment?

    Hi,
    I want to know how to use PXI-8532 in LabVIEW Development Environment.
    I'm using
    1. Windows7 32bit
    2. LabVIEW Pro 2012 SP1 32bit
    3. NI-Industrial Communications for DeviceNet 2.3
    4. PXIe-1062Q
    5. PXIe-8135
    6. PXI-8532 *2 (Names are "dnet0" and "dnet1".)
    * Both PXI-8532 are installed in the same chassis (PXIe-1062Q).
    I want to construct the system that "dnet0" sends data, and "dnet1" receives it. The data type I want to use is an array of bools.
    I made VI (test.vi), but it does not work.
    Could you tell me the reason why my VI doesn't work correctly.
    Thank you so much for your help and time.
    Best Regards,
    volcanon2
    Attachments:
    test.vi ‏28 KB

    activeX

  • Error "codepage could not be determined for the receiver-system"

    Hi,
    we use the idoc-adapter to send message from mySAP ERP to PI.
    All worked fine.
    So, this morning I have changed the datatype of the messageId from "char" to "int" in mySAP ERP in the customizing menue (install param). So, no message received in SAP PI.
    When I start transction "we05" in mySAPERP" I get the following error:
    "codepage could not be determined for the receiver-system" status 02.
    I have changed the datatyp to "char" but the same probleme
    Can anybody help me please?
    Regards
    Stefan

    This error shouldn't occur unless there is some modification at the OS level codepages.
    Please check the entried in the RFC destination for connecting to the XI system. Unicode settings.
    Regards,
    Prateek

  • Issue:Creation of Extended iDOC data entries in receiving system.

    Hi all,
    I have extended standard IDOC MATMAS05 for classification view.I am receiving the data successfully on recewing side.And i  have done coading in user exit to store data in database  table inside
    IDOC_INPUT_MATMAS01- function module.
    When i execute the BD10 from sending sytem and view the idoc at receiving end from WE05 i can see the idoc but when i try to see the material i cant see the classification(extended) view in mm03.Rest of the views are created .When i execute the idoc from WE19 ,Classification(extended) view gets created.
    Please guide me so that the CLASSIFICATION view get automatically created in receiving system without using WE19.
    Thanks.

    Yes i used this function module-
    CALL FUNCTION 'BAPI_OBJCL_CREATE'
                  EXPORTING
                    objectkeynew            = objectkeynew1
                    objecttablenew          = 'MARA'
                    classnumnew             = wa_classnumnew1
                    classtypenew            = z1sklart-klart              "classtypenew1
                    status                  = '1'
      STANDARDCLASS           =
      CHANGENUMBER            =
                    keydate                 = sy-datum
      NO_DEFAULT_VALUES       = ' '
    IMPORTING
      CLASSIF_STATUS          =
                  TABLES
      ALLOCVALUESNUM          =
                    allocvalueschar         = allocvalueschar2[]
      ALLOCVALUESCURR         =
                    return                  = return1
    *WAIT UP TO 1 SECONDS.
                CALL FUNCTION 'BAPI_TRANSACTION_COMMIT'
                 EXPORTING
                   wait          = 'X'
    IMPORTING
      RETURN        =
    Please explain me how to maintain the Views Status at MARA general info segment level.

  • Outbound IDOC not received at the receiver system.

    Hi,
    Im sending an IDOC from sap to a non-sap system. During this process, from sap system it is executing successfully without any erros and the status is also '03' .
    The problem is that the other system is unable to receive the IDOC sent from SAP system.
    The connection is like TCP/IP and the registered server program is given as 'PROGRID' in RFC destination.
    I would like to know what would be the possible cases where this IDOC is not received at the receiver system.
    The configuration is perfect as per my knowledge. Just want to confirm one thing 'do i need to include any message type such as "SYNCH".
    Awaiting for your solutions.
    Thanks
    rohith

    Hi,
    If the IDOC is sent successfully from the sender system then the status of the IDOC will be '03' , no matter whether the receiver has received the IDOC or not.
    As per your case,
    Check the Partner Profile parameters , whether the basic and extended message types are used properly.
    The message type Synchron (Sync) is generally needed in Inbound IDOC side. Check whether you have added SYNCH or not.
    Hope this Helps.
    Regards,
    Abhinab Mishra.
    Experience is the name everyone gives to their mistakes.

  • Missing document items in Sales Order in the receiving system

    Hello.
    We're doing passing of message type ORDERS (ORDERS05) from one system to another. Idocs were posted successfully but we can't find the document items in the receiving system if we view a particular sales order in VA03. Document items are present in the sending system.
    I checked segment E1EDP01 (document items segment) for the inbound idocs and found all items attached to the sales order but we can't see them in VA03 display.
    Appreciate help on how to proceed/check this issue.
    Thanks. =)

    If you have verified that the IDOC was complete in the receiving system, I would use WE19 to copy the IDOC in the receiving system (or copy to a development instance of that system) and run the process in debug....someone will have to watch the process, I think and see what is happening to the VBAP table during sales order build.

Maybe you are looking for

  • Excel can not exit

    Dear Sir: I have the same issue. As Joe mentioned before, "Everytime when an Excel activex automation is opened, it needs to be properly closed. If your vi opened excel but without either closing it or disposing its reference, it will hang in the mem

  • What is the open purchage order

    ·     Developed a report that displays open purchase order.

  • Analog readings on PCI7332 motion card

    I am using three analog inputs with a single axis open loop strpper. channel1,3,4 Channels 1 & 3 work fine, but channel 4 doesn't. First of all it seems too drag the input from 3volts to .8voltd I have an Analog Device 5B Islation amp connecting sign

  • O.T.   Obituary

    Today we mourn the passing of a beloved friend by the name of COMMON SENSE, who has been with us for many years . No one knows for sure how old he was,since his birth records were long ago lost in bureaucratic red tape . He will be remembered as havi

  • BOM qty. V/s Actual Qty. on yield basis.

    Dear Mangalraj S., I have got standard figures from mentioned t. Codes, but here my clients requirement is slightly different. Here they are creating process order for 1000 kg. in which RM is 100 kg. as per BOM, now client produce only 800 kg, in whi